Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings

Authors: S. Busch, T. R. C. Johnson, B. J. Wintersperger, N. Minaifar, A. Bhargava, C. Rist, M. F. Reiser, C. Becker, K. Nikolaou

Published in: European Radiology | Issue 3/2008

Login to get access

Abstract

Cardiac magnetic resonance imaging and echocardiography are currently regarded as standard modalities for the quantification of left ventricular volumes and ejection fraction. With the recent introduction of dual-source computedtomography (DSCT), the increased temporal resolution of 83 ms should also improve the assessment of cardiac function in CT. The aim of this study was to evaluate the accuracy of DSCT in the assessment of left ventricular functional parameters with cardiac magnetic resonance imaging (MRI) as standard of reference. Fifteen patients (two female, 13 male; mean age 50.8 ± 19.2 years) underwent CT and MRI examinations on a DSCT (Somatom Definition; Siemens Medical Solutions, Forchheim, Germany) and a 3.0-Tesla MR scanner (Magnetom Trio; Siemens Medical Solutions), respectively. Multiphase axial CT images were analysed with a semiautomatic region growing algorithms (Syngo Circulation; Siemens Medical Solutions) by two independent blinded observers. In MRI, dynamic cine loops of short axis slices were evaluated with semiautomatic contour detection software (ARGUS; Siemens Medical Solutions) independently by two readers. End-systolic volume (ESV), end-diastolic volume (EDV), ejection fraction (EF) and stroke volume (SV) were determined for both modalities, and correlation coefficient, systematic error, limits of agreement and inter-observer variability were assessed. In DSCT, EDV and ESV were 135.8 ± 41.9 ml and 54.9 ± 29.6 ml, respectively, compared with 132.1 ± 40.8 ml EDV and 57.6 ± 27.3 ml ESV in MRI. Thus, EDV was overestimated by 3.7 ml (limits of agreement −46.1/+53.6), while ESV was underestimated by 2.6 ml (−36.6/+31.4). Mean EF was 61.6 ± 12.4% in DSCT and 57.9 ± 9.0% in MRI, resulting in an overestimation of EF by 3.8% with limits of agreement at −14.7 and +22.2%. Rank correlation rho values were 0.81 for EDV (P = 0.0024), 0.79 for ESV (P = 0.0031) and 0.64 for EF (P = 0.0168). The kappa value of inter-observer variability were amounted to 0.85 for EDV, ESV and EF. DSCT offers the possibility to quantify left ventricular function from coronary CT angiography datasets with sufficient diagnostic accuracy, adding to the value of the modality in a comprehensive cardiac assessment. The observed differences in the measured values may be due to different post-processing methods and physiological reactions to contrast material injection without beta-blocker medication.
Literature
1.
go back to reference Hundley WG, Hamilton CA, Thomas MS, Herrington DM, Salido TB, Kitzman DW, Little WC, Link KM (1999) Utility of fast cine magnetic resonance imaging and display for the detection of myocardial ischemia in patients not well suited for second harmonic stress echocardiography. Circulation 100:1697–1702PubMed Hundley WG, Hamilton CA, Thomas MS, Herrington DM, Salido TB, Kitzman DW, Little WC, Link KM (1999) Utility of fast cine magnetic resonance imaging and display for the detection of myocardial ischemia in patients not well suited for second harmonic stress echocardiography. Circulation 100:1697–1702PubMed
2.
go back to reference Juergens KU, Fischbach R (2006) Left ventricular function studied with MDCT. Eur Radiol 16:342–357PubMedCrossRef Juergens KU, Fischbach R (2006) Left ventricular function studied with MDCT. Eur Radiol 16:342–357PubMedCrossRef
3.
go back to reference Wintersperger B, Nikolaou K, Dietrich O, Rieber J, Nittka M, Reiser M, Schoenberg S (2003) Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm. Eur Radiol 13:1931–1936PubMedCrossRef Wintersperger B, Nikolaou K, Dietrich O, Rieber J, Nittka M, Reiser M, Schoenberg S (2003) Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm. Eur Radiol 13:1931–1936PubMedCrossRef
4.
go back to reference Rist C, Johnson TR, Becker A, Leber AW, Huber A, Busch S, Becker CR, Reiser MF, Nikolaou K (2007) Dual-source cardiac CT imaging with improved temporal resolution: impact on image quality and analysis of left ventricular function. Radiologe 47:287–294, AprPubMedCrossRef Rist C, Johnson TR, Becker A, Leber AW, Huber A, Busch S, Becker CR, Reiser MF, Nikolaou K (2007) Dual-source cardiac CT imaging with improved temporal resolution: impact on image quality and analysis of left ventricular function. Radiologe 47:287–294, AprPubMedCrossRef
5.
go back to reference Tanabe K, Belohlavek M, Jakrapanichakul D et al (1998) Three dimensional echocardiography: precision and accuracy of left ventricular volume measurement using rotational geometry with variable numbers of slice resolution. Echocardiography 15:575–580PubMedCrossRef Tanabe K, Belohlavek M, Jakrapanichakul D et al (1998) Three dimensional echocardiography: precision and accuracy of left ventricular volume measurement using rotational geometry with variable numbers of slice resolution. Echocardiography 15:575–580PubMedCrossRef
6.
go back to reference Johnson T, Nikolaou K, Wintersperger BJ, Leber A, von Ziegler F, Rist C, Buhmann S, Knez A, Reiser M, Becker C (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef Johnson T, Nikolaou K, Wintersperger BJ, Leber A, von Ziegler F, Rist C, Buhmann S, Knez A, Reiser M, Becker C (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef
7.
go back to reference Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef
8.
go back to reference Becker CR, Knez A, Leber A, Treede H, Ohnesorge B, Schoepf UJ, Reiser MF (2002) Detection of coronary artery stenoses with multislice helical CT angiography. J Comput Assist Tomogr 26:750–755PubMedCrossRef Becker CR, Knez A, Leber A, Treede H, Ohnesorge B, Schoepf UJ, Reiser MF (2002) Detection of coronary artery stenoses with multislice helical CT angiography. J Comput Assist Tomogr 26:750–755PubMedCrossRef
9.
go back to reference Juergens KU, Grude M, Maintz D, Fallenberg EM, Wichter T et al (2004) Multi-detector row CT of left ventricular function with dedicated analysis software versus MR imaging: initial experiences. Radiology 230:403–410PubMedCrossRef Juergens KU, Grude M, Maintz D, Fallenberg EM, Wichter T et al (2004) Multi-detector row CT of left ventricular function with dedicated analysis software versus MR imaging: initial experiences. Radiology 230:403–410PubMedCrossRef
10.
go back to reference Mahnken AH, Koos R, Katoh M, Spuentrup E, Busch P, Wildberger JE, Kuhl HP, Gunther RW (2005) Sixteen-slice spiral CT versus MR imaging for the assessment of left ventricular function in acute myocardial infarction. Eur Radiol 15:714–720PubMedCrossRef Mahnken AH, Koos R, Katoh M, Spuentrup E, Busch P, Wildberger JE, Kuhl HP, Gunther RW (2005) Sixteen-slice spiral CT versus MR imaging for the assessment of left ventricular function in acute myocardial infarction. Eur Radiol 15:714–720PubMedCrossRef
11.
go back to reference Sugeng L, Mor-Avi V, Weinert L, Niel J, Steringer-Mascherbauer R, Schmidt F, Galuschky C, Schummers G, Lang RM, Nesser HJ (2006) Quantitative assessment of left ventricular size and function: side-by-side comparison of real-time three-dimensional echocardiography and computed tomography with magnetic resonance reference. Circulation 114:654–661PubMedCrossRef Sugeng L, Mor-Avi V, Weinert L, Niel J, Steringer-Mascherbauer R, Schmidt F, Galuschky C, Schummers G, Lang RM, Nesser HJ (2006) Quantitative assessment of left ventricular size and function: side-by-side comparison of real-time three-dimensional echocardiography and computed tomography with magnetic resonance reference. Circulation 114:654–661PubMedCrossRef
12.
go back to reference Schlosser T, Mohrs OK, Magedanz A, Voigtlander T, Schmermund A, Barkhausen J (2007) Assessment of left ventricular function and mass in patients undergoing computed tomography (CT) coronary angiography using 64-detector-row CT: comparison to magnetic resonance imaging. Acta Radiol 48:30–35PubMedCrossRef Schlosser T, Mohrs OK, Magedanz A, Voigtlander T, Schmermund A, Barkhausen J (2007) Assessment of left ventricular function and mass in patients undergoing computed tomography (CT) coronary angiography using 64-detector-row CT: comparison to magnetic resonance imaging. Acta Radiol 48:30–35PubMedCrossRef
13.
go back to reference Muhlenbruch G, Das M, Hohl C, Wildberger JE, Rinck D, Flohr TG, Koos R, Knackstedt C, Gunther RW, Mahnken AH (2006) Global left ventricular function in cardiac CT. Evaluation of an automated 3D region-growing segmentation algorithm. Eur Radiol 16:1117–1123PubMedCrossRef Muhlenbruch G, Das M, Hohl C, Wildberger JE, Rinck D, Flohr TG, Koos R, Knackstedt C, Gunther RW, Mahnken AH (2006) Global left ventricular function in cardiac CT. Evaluation of an automated 3D region-growing segmentation algorithm. Eur Radiol 16:1117–1123PubMedCrossRef
14.
go back to reference Mahnken AH, Muhlenbruch G, Koos R, Stanzel S, Busch PS, Niethammer M, Gunther RW, Wildberger JE (2006) Automated vs. manual assessment of left ventricular function in cardiac multidetector row computed tomography: comparison with magnetic resonance imaging. Eur Radiol 16:1416–1423PubMedCrossRef Mahnken AH, Muhlenbruch G, Koos R, Stanzel S, Busch PS, Niethammer M, Gunther RW, Wildberger JE (2006) Automated vs. manual assessment of left ventricular function in cardiac multidetector row computed tomography: comparison with magnetic resonance imaging. Eur Radiol 16:1416–1423PubMedCrossRef
15.
go back to reference Vogel-Claussen J, Finn JP, Gomes AS, Hundley GW, Jerosch-Herold M, Pearson G, Sinha S, Lima JA, Bluemke DA (2006) Left ventricular papillary muscle mass: relationship to left ventricular mass and volumes by magnetic resonance imaging. J Comput Assist Tomogr 30:426–432PubMedCrossRef Vogel-Claussen J, Finn JP, Gomes AS, Hundley GW, Jerosch-Herold M, Pearson G, Sinha S, Lima JA, Bluemke DA (2006) Left ventricular papillary muscle mass: relationship to left ventricular mass and volumes by magnetic resonance imaging. J Comput Assist Tomogr 30:426–432PubMedCrossRef
16.
go back to reference Orakzai SH, Orakzai RH, Nasir K, Budoff MJ (2006) Assessment of cardiac function using multidetector row computed tomography. J Comput Assist Tomogr 30:555–563PubMedCrossRef Orakzai SH, Orakzai RH, Nasir K, Budoff MJ (2006) Assessment of cardiac function using multidetector row computed tomography. J Comput Assist Tomogr 30:555–563PubMedCrossRef
17.
go back to reference Lasalarie JC, Serfaty JM, Carre C, Messika-Zeitoun D, Jeannot C, Schouman-Claeys E, Laissy JP (2007) Accuracy of contrast-enhanced cine-MR sequences in the assessment of left ventricular function: comparison with precontrast cine-MR sequences. Results of a bicentric study. Eur Radiol, DOI 10.1007/s00330-007-0647-5 Lasalarie JC, Serfaty JM, Carre C, Messika-Zeitoun D, Jeannot C, Schouman-Claeys E, Laissy JP (2007) Accuracy of contrast-enhanced cine-MR sequences in the assessment of left ventricular function: comparison with precontrast cine-MR sequences. Results of a bicentric study. Eur Radiol, DOI 10.​1007/​s00330-007-0647-5
18.
go back to reference Wintersperger BJ, Sincleair S, Runge VM, Dietrich O, Huber A, Reiser MF, Schoenberg SO (2007) Dual breath-hold magnetic resonance cine evaluation of global and regional cardiac function. Eur Radiol 17:73–80PubMedCrossRef Wintersperger BJ, Sincleair S, Runge VM, Dietrich O, Huber A, Reiser MF, Schoenberg SO (2007) Dual breath-hold magnetic resonance cine evaluation of global and regional cardiac function. Eur Radiol 17:73–80PubMedCrossRef
Metadata
Title
Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings
Authors
S. Busch
T. R. C. Johnson
B. J. Wintersperger
N. Minaifar
A. Bhargava
C. Rist
M. F. Reiser
C. Becker
K. Nikolaou
Publication date
01-03-2008
Publisher
Springer-Verlag
Published in
European Radiology / Issue 3/2008
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-007-0767-y

Other articles of this Issue 3/2008

European Radiology 3/2008 Go to the issue