Transpulmonary thermodilution measurements are not affected by continuous veno-venous hemofiltration at high blood pump flow

Authors: Nicolas Dufour, Marianne Delville, Jean-Louis Teboul, Laurent Camous, Aude Favier du Noyer, Christian Richard, Xavier Monnet

Published in: Intensive Care Medicine | Issue 7/2012

Login to get access

Abstract

Purpose

To assess whether continuous veno-venous hemofiltration (CVVH) with high blood pump flow alters the measurements of cardiac index (CI), global end-diastolic volume indexed (GEDVI), and extravascular lung water indexed (EVLWI) performed by transpulmonary thermodilution.

Methods

Sixty-nine patients were included if they were monitored by a PiCCO2 device and received CVVH through a femoral (n = 62) or an internal jugular (n = 7) dialysis catheter. The blood pump flow was set at 250 mL/min (n = 31) or 350 mL/min (n = 38) and the filtration flow at 6,000 mL/h. A first set of data was collected with a first transpulmonary thermodilution (TDon). The blood pump was stopped and the continuous CI derived from pulse contour analysis was recorded (PCoff). A second data set (TDoff) was collected before and a last one (TDon-last) after restarting the blood pump.

Results

\( {\text{CI}}_{{{\text{TD}}_{\text{on}} }} \), \( {\text{CI}}_{{{\text{PC}}_{\text{off}} }} \), \( {\text{CI}}_{{{\text{TD}}_{\text{off}} }} \) , and \( {\text{CI}}_{{{\text{TD}}_{{{\text{on}} - {\text{last}}}} }} \) were not significantly different in patients with a femoral dialysis catheter (3.49 ± 0.96, 3.51 ± 0.96, 3.51 ± 0.99, and 3.44 ± 1.00 L min−1 m−2, respectively). This was observed with a blood pump flow at 350 mL/min and at 250 mL/min. In these patients with a femoral dialysis catheter, GEDVI did not significantly change when the blood pump was stopped. EVLWI significantly decreased when the blood pump was stopped but to a non-clinically relevant extent (−0.3 ± 0.8 mL/kg). No significant changes in CI, GEDVI, and EVLWI were observed in patients with an internal jugular dialysis catheter over the study period.

Conclusions

CVVH with a high blood flow pump does not alter the transpulmonary thermodilution measurements of CI, GEDVI, and EVLWI.
Literature
1.
go back to reference Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK (2010) Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 110:799–811PubMedCrossRef Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK (2010) Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg 110:799–811PubMedCrossRef
2.
go back to reference Monnet X, Anguel N, Osman D, Hamzaoui O, Richard C, Teboul J-L (2007) Assessing pulmonary permeability by transpulmonary thermodilution allows differentiation of hydrostatic pulmonary edema from ALI/ARDS. Intensive Care Med 33:448–453PubMedCrossRef Monnet X, Anguel N, Osman D, Hamzaoui O, Richard C, Teboul J-L (2007) Assessing pulmonary permeability by transpulmonary thermodilution allows differentiation of hydrostatic pulmonary edema from ALI/ARDS. Intensive Care Med 33:448–453PubMedCrossRef
3.
go back to reference Sakka SG, Hanusch T, Thuemer O, Wegscheider K (2007) The influence of venovenous renal replacement therapy on measurements by the transpulmonary thermodilution technique. Anesth Analg 105:1079–1082PubMedCrossRef Sakka SG, Hanusch T, Thuemer O, Wegscheider K (2007) The influence of venovenous renal replacement therapy on measurements by the transpulmonary thermodilution technique. Anesth Analg 105:1079–1082PubMedCrossRef
4.
go back to reference Heise D, Faulstich M, Morer O, Brauer A, Quintel M (2012) Influence of continuous renal replacement therapy on cardiac output measurement using thermodilution techniques. Minerva Anestesiol 78:315–321 Heise D, Faulstich M, Morer O, Brauer A, Quintel M (2012) Influence of continuous renal replacement therapy on cardiac output measurement using thermodilution techniques. Minerva Anestesiol 78:315–321
5.
go back to reference Saugel B, Umgelter A, Schuster T, Phillip V, Schmid RM, Huber W (2010) Transpulmonary thermodilution using femoral indicator injection: a prospective trial in patients with a femoral and a jugular central venous catheter. Crit Care 14:R95PubMedCrossRef Saugel B, Umgelter A, Schuster T, Phillip V, Schmid RM, Huber W (2010) Transpulmonary thermodilution using femoral indicator injection: a prospective trial in patients with a femoral and a jugular central venous catheter. Crit Care 14:R95PubMedCrossRef
6.
go back to reference Monnet X, Persichini R, Ktari M, Jozwiak M, Richard C, Teboul JL (2011) Precision of the transpulmonary thermodilution measurements. Crit Care 15:R204PubMedCrossRef Monnet X, Persichini R, Ktari M, Jozwiak M, Richard C, Teboul JL (2011) Precision of the transpulmonary thermodilution measurements. Crit Care 15:R204PubMedCrossRef
7.
go back to reference Stewart GN (1921) The pulmonary circulation time, the quantity of blood in the lungs and the output of the heart. Am J Physiol 58:20–44 Stewart GN (1921) The pulmonary circulation time, the quantity of blood in the lungs and the output of the heart. Am J Physiol 58:20–44
8.
go back to reference Böck JC, Lewis FR (1990) Clinical relevance of lung water measurement with the thermal-dye dilution technique. J Surg Res 48:254–265PubMedCrossRef Böck JC, Lewis FR (1990) Clinical relevance of lung water measurement with the thermal-dye dilution technique. J Surg Res 48:254–265PubMedCrossRef
9.
go back to reference Bazaral MG, Petre J, Novoa R (1992) Errors in thermodilution cardiac output measurements caused by rapid pulmonary artery temperature decreases after cardiopulmonary bypass. Anesthesiology 77:31–37PubMedCrossRef Bazaral MG, Petre J, Novoa R (1992) Errors in thermodilution cardiac output measurements caused by rapid pulmonary artery temperature decreases after cardiopulmonary bypass. Anesthesiology 77:31–37PubMedCrossRef
10.
go back to reference Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ (1971) A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396PubMedCrossRef Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ (1971) A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396PubMedCrossRef
11.
go back to reference Griffin K, Benjamin E, DelGiudice R, Schechter C, Iberti TJ (1997) Thermodilution cardiac output measurement during simultaneous volume infusion through the venous infusion port of the pulmonary artery catheter. J Cardiothorac Vasc Anesth 11:437–439PubMedCrossRef Griffin K, Benjamin E, DelGiudice R, Schechter C, Iberti TJ (1997) Thermodilution cardiac output measurement during simultaneous volume infusion through the venous infusion port of the pulmonary artery catheter. J Cardiothorac Vasc Anesth 11:437–439PubMedCrossRef
12.
go back to reference Latson TW, Whitten CW, O’Flaherty D (1993) Ventilation, thermal noise, and errors in cardiac output measurements after cardiopulmonary bypass. Anesthesiology 79:1233–1243PubMedCrossRef Latson TW, Whitten CW, O’Flaherty D (1993) Ventilation, thermal noise, and errors in cardiac output measurements after cardiopulmonary bypass. Anesthesiology 79:1233–1243PubMedCrossRef
13.
go back to reference Martinez-Simon A, Monedero P, Cacho-Asenjo E (2006) Erroneous measurement of haemodynamic parameters by PiCCO monitor in a critically ill patient with renal replacement therapy: a case report. Crit Care 10:410PubMedCrossRef Martinez-Simon A, Monedero P, Cacho-Asenjo E (2006) Erroneous measurement of haemodynamic parameters by PiCCO monitor in a critically ill patient with renal replacement therapy: a case report. Crit Care 10:410PubMedCrossRef
14.
go back to reference Sakka SG, Hanusch T, Thuemer O, Wegscheider K (2007) The influence of venovenous renal replacement therapy on measurements by the transpulmonary thermodilution technique. Anesth Analg 105:1079–1082PubMedCrossRef Sakka SG, Hanusch T, Thuemer O, Wegscheider K (2007) The influence of venovenous renal replacement therapy on measurements by the transpulmonary thermodilution technique. Anesth Analg 105:1079–1082PubMedCrossRef
15.
go back to reference Manns M, Maurer E, Steinbach B, Evering HG (1998) Thermal energy balance during in vitro continuous veno-venous hemofiltration. ASAIO 44:M601–M605CrossRef Manns M, Maurer E, Steinbach B, Evering HG (1998) Thermal energy balance during in vitro continuous veno-venous hemofiltration. ASAIO 44:M601–M605CrossRef
16.
go back to reference Santoro A, Mancini E, Canova C, Mambelli E (2003) Thermal balance in convective therapies. Nephrol Dial Transplant 18(Suppl 7):vii41–vii57PubMed Santoro A, Mancini E, Canova C, Mambelli E (2003) Thermal balance in convective therapies. Nephrol Dial Transplant 18(Suppl 7):vii41–vii57PubMed
17.
go back to reference Sakka SG (2010) Influence of an extracorporeal lung assist system on transpulmonary thermodilution-derived variables. Br J Anaesth 104:664–665PubMedCrossRef Sakka SG (2010) Influence of an extracorporeal lung assist system on transpulmonary thermodilution-derived variables. Br J Anaesth 104:664–665PubMedCrossRef
18.
go back to reference Haller M, Zollner C, Manert W, Briegel J, Kilger E, Polasek J, Hummel T, Forst H, Peter K (1995) Thermodilution cardiac output may be incorrect in patients on venovenous extracorporeal lung assist. Am J Respir Crit Care Med 152:1812–1817PubMed Haller M, Zollner C, Manert W, Briegel J, Kilger E, Polasek J, Hummel T, Forst H, Peter K (1995) Thermodilution cardiac output may be incorrect in patients on venovenous extracorporeal lung assist. Am J Respir Crit Care Med 152:1812–1817PubMed
19.
go back to reference Monnet X, Perel A (2011) Extravascular lung water. In: Vincent J, Hall J (eds) Encyclopedia of intensive care medicine. Springer-Verlag, Berlin Heidelberg Monnet X, Perel A (2011) Extravascular lung water. In: Vincent J, Hall J (eds) Encyclopedia of intensive care medicine. Springer-Verlag, Berlin Heidelberg
20.
go back to reference Van Craenenbroeck A, Van Ingelgem A, Palmers P-J, Cordemans C, Van Hootegem A, Van Regenmortel N, De Laet I, Schoonheydt K, Dits H, Malbrain M (2010) Influence of continuous venovenous hemofiltration (CVVH) and catheter position on transpulmonary thermodilution derived parameters with PiCCO. Intensive Care Med 36(Suppl 2):S294 Van Craenenbroeck A, Van Ingelgem A, Palmers P-J, Cordemans C, Van Hootegem A, Van Regenmortel N, De Laet I, Schoonheydt K, Dits H, Malbrain M (2010) Influence of continuous venovenous hemofiltration (CVVH) and catheter position on transpulmonary thermodilution derived parameters with PiCCO. Intensive Care Med 36(Suppl 2):S294
Metadata
Title
Transpulmonary thermodilution measurements are not affected by continuous veno-venous hemofiltration at high blood pump flow
Authors
Nicolas Dufour
Marianne Delville
Jean-Louis Teboul
Laurent Camous
Aude Favier du Noyer
Christian Richard
Xavier Monnet
Publication date
01-07-2012
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 7/2012
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-012-2573-5

Other articles of this Issue 7/2012

Intensive Care Medicine 7/2012 Go to the issue