Open Access 01-08-2011 | Research

Precision of the transpulmonary thermodilution measurements

Authors: Xavier Monnet, Romain Persichini, Mariem Ktari, Mathieu Jozwiak, Christian Richard, Jean-Louis Teboul

Published in: Critical Care | Issue 4/2011

Login to get access

Abstract

Introduction

We wanted to determine the number of cold bolus injections that are necessary for achieving an acceptable level of precision for measuring cardiac index (CI), indexed global end-diastolic volume (GEDVi) and indexed extravascular lung water (EVLWi) by transpulmonary thermodilution.

Methods

We included 91 hemodynamically stable patients (age 59 (25% to 75% interquartile range: 39 to 79) years, simplified acute physiologic score (SAPS)II 59 (53 to 65), 56% under norepinephrine) who were monitored by a PiCCO2 device. We performed five successive cold saline (15 mL, 6°C) injections and recorded the measurements of CI, GEDVi and EVLWi.

Results

Considering five boluses, the coefficient of variation (CV, calculated as standard deviation divided by the mean of the five measurements) was 7 (5 to 11)%, 7 (5 to 12)% and 7 (6 to 12)% for CI, GEDVi and EVLWi, respectively. If the results of two bolus injections were averaged, the precision (2 × CV/√ number of boluses) was 10 (7 to 15)%, 10 (7 to 17)% and 8 (7 to 14)% for CI, GEDVi and EVLWi, respectively. If the results of three bolus injections were averaged, the precision dropped below 10%, that is, the cut-off that is generally considered as acceptable (8 (6 to 12)%, 8 (6 to 14)% and 8 (7 to 14)% for CI, GEDVi and EVLWi, respectively). If two injections were performed, the least significant change, that is, the minimal change in value that could be trusted to be significant, was 14 (10 to 21)%, 14 (10 to 24)% and 14 (11 to 23)% for CI, GEDVi and EVLWi, respectively. If three injections were performed, the least significant change was 12 (8 to 17)%, 12 (8 to 19)% and 12 (9 to 19)% for CI, GEDVi and EVLWi, respectively, that is, below the 15% cut-off that is usually considered as clinically relevant.

Conclusions

These results support the injection of at least three cold boluses for obtaining an acceptable precision when transpulmonary thermodilution is used for measuring CI, GEDVi and EVLWi.
Appendix
Available only for authorised users
Literature
1.
go back to reference Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK: Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg. 2010, 110: 799-811. 10.1213/ANE.0b013e3181cc885a.PubMedCrossRef Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK: Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg. 2010, 110: 799-811. 10.1213/ANE.0b013e3181cc885a.PubMedCrossRef
2.
go back to reference Gondos T, Marjanek Z, Kisvarga Z, Halasz G: Precision of transpulmonary thermodilution: how many measurements are necessary?. Eur J Anaesthesiol. 2009, 26: 508-512. 10.1097/EJA.0b013e32832a2d23.PubMedCrossRef Gondos T, Marjanek Z, Kisvarga Z, Halasz G: Precision of transpulmonary thermodilution: how many measurements are necessary?. Eur J Anaesthesiol. 2009, 26: 508-512. 10.1097/EJA.0b013e32832a2d23.PubMedCrossRef
3.
go back to reference Sakka SG, Ruhl CC, Pfeiffer UJ, Beale R, McLuckie A, Reinhart K, Meier-Hellmann A: Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution. Intensive Care Med. 2000, 26: 180-187. 10.1007/s001340050043.PubMedCrossRef Sakka SG, Ruhl CC, Pfeiffer UJ, Beale R, McLuckie A, Reinhart K, Meier-Hellmann A: Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution. Intensive Care Med. 2000, 26: 180-187. 10.1007/s001340050043.PubMedCrossRef
4.
go back to reference Craig TR, Duffy MJ, Shyamsundar M, McDowell C, McLaughlin B, Elborn JS, McAuley DF: Extravascular lung water indexed to predicted body weight is a novel predictor of intensive care unit mortality in patients with acute lung injury. Crit Care Med. 38: 114-120. Craig TR, Duffy MJ, Shyamsundar M, McDowell C, McLaughlin B, Elborn JS, McAuley DF: Extravascular lung water indexed to predicted body weight is a novel predictor of intensive care unit mortality in patients with acute lung injury. Crit Care Med. 38: 114-120.
5.
go back to reference Martin GS, Eaton S, Mealer M, Moss M: Extravascular lung water in patients with severe sepsis: a prospective cohort study. Crit Care. 2005, 9: R74-82. 10.1186/cc3025.PubMedPubMedCentralCrossRef Martin GS, Eaton S, Mealer M, Moss M: Extravascular lung water in patients with severe sepsis: a prospective cohort study. Crit Care. 2005, 9: R74-82. 10.1186/cc3025.PubMedPubMedCentralCrossRef
6.
go back to reference Perkins GD, McAuley DF, Thickett DR, Gao F: The beta-agonist lung injury trial (BALTI): a randomized placebo-controlled clinical trial. Am J Respir Crit Care Med. 2006, 173: 281-287. 10.1164/rccm.200508-1302OC.PubMedCrossRef Perkins GD, McAuley DF, Thickett DR, Gao F: The beta-agonist lung injury trial (BALTI): a randomized placebo-controlled clinical trial. Am J Respir Crit Care Med. 2006, 173: 281-287. 10.1164/rccm.200508-1302OC.PubMedCrossRef
8.
go back to reference Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD: Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med. 2009, 35: 1801-1808. 10.1007/s00134-009-1570-9.PubMedCrossRef Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD: Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med. 2009, 35: 1801-1808. 10.1007/s00134-009-1570-9.PubMedCrossRef
9.
go back to reference Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM: Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies--with specific reference to the measurement of cardiac output. Crit Care. 2009, 13: 201-10.1186/cc7129.PubMedPubMedCentralCrossRef Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM: Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies--with specific reference to the measurement of cardiac output. Crit Care. 2009, 13: 201-10.1186/cc7129.PubMedPubMedCentralCrossRef
10.
go back to reference Della Rocca G, Costa MG, Chiarandini P, Bertossi G, Lugano M, Pompei L, Coccia C, Sainz-Barriga M, Pietropaoli P: Arterial pulse cardiac output agreement with thermodilution in patients in hyperdynamic conditions. J Cardiothorac Vasc Anesth. 2008, 22: 681-687. 10.1053/j.jvca.2008.02.021.PubMedCrossRef Della Rocca G, Costa MG, Chiarandini P, Bertossi G, Lugano M, Pompei L, Coccia C, Sainz-Barriga M, Pietropaoli P: Arterial pulse cardiac output agreement with thermodilution in patients in hyperdynamic conditions. J Cardiothorac Vasc Anesth. 2008, 22: 681-687. 10.1053/j.jvca.2008.02.021.PubMedCrossRef
11.
go back to reference Segal E, Katzenelson R, Berkenstadt H, Perel A: Transpulmonary thermodilution cardiac output measurement using the axillary artery in critically ill patients. J Clin Anesth. 2002, 14: 210-213. 10.1016/S0952-8180(02)00345-8.PubMedCrossRef Segal E, Katzenelson R, Berkenstadt H, Perel A: Transpulmonary thermodilution cardiac output measurement using the axillary artery in critically ill patients. J Clin Anesth. 2002, 14: 210-213. 10.1016/S0952-8180(02)00345-8.PubMedCrossRef
12.
go back to reference Friesecke S, Heinrich A, Abel P, Felix SB: Comparison of pulmonary artery and aortic transpulmonary thermodilution for monitoring of cardiac output in patients with severe heart failure: validation of a novel method. Crit Care Med. 2009, 37: 119-123. 10.1097/CCM.0b013e31819290d5.PubMedCrossRef Friesecke S, Heinrich A, Abel P, Felix SB: Comparison of pulmonary artery and aortic transpulmonary thermodilution for monitoring of cardiac output in patients with severe heart failure: validation of a novel method. Crit Care Med. 2009, 37: 119-123. 10.1097/CCM.0b013e31819290d5.PubMedCrossRef
13.
go back to reference Zollner C, Polasek J, Kilger E, Pichler B, Jaenicke U, Briegel J, Vetter HO, Haller M: Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: a prospective criterion standard study. Crit Care Med. 1999, 27: 293-298. 10.1097/00003246-199902000-00033.PubMedCrossRef Zollner C, Polasek J, Kilger E, Pichler B, Jaenicke U, Briegel J, Vetter HO, Haller M: Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: a prospective criterion standard study. Crit Care Med. 1999, 27: 293-298. 10.1097/00003246-199902000-00033.PubMedCrossRef
14.
go back to reference Wiesenack C, Prasser C, Keyl C, Rodig G: Assessment of intrathoracic blood volume as an indicator of cardiac preload: single transpulmonary thermodilution technique versus assessment of pressure preload parameters derived from a pulmonary artery catheter. J Cardiothorac Vasc Anesth. 2001, 15: 584-588. 10.1053/jcan.2001.26536.PubMedCrossRef Wiesenack C, Prasser C, Keyl C, Rodig G: Assessment of intrathoracic blood volume as an indicator of cardiac preload: single transpulmonary thermodilution technique versus assessment of pressure preload parameters derived from a pulmonary artery catheter. J Cardiothorac Vasc Anesth. 2001, 15: 584-588. 10.1053/jcan.2001.26536.PubMedCrossRef
15.
go back to reference McLuckie A, Murdoch IA, Marsh MJ, Anderson D: A comparison of pulmonary and femoral artery thermodilution cardiac indices in paediatric intensive care patients. Acta Paediatr. 1996, 85: 336-338. 10.1111/j.1651-2227.1996.tb14027.x.PubMedCrossRef McLuckie A, Murdoch IA, Marsh MJ, Anderson D: A comparison of pulmonary and femoral artery thermodilution cardiac indices in paediatric intensive care patients. Acta Paediatr. 1996, 85: 336-338. 10.1111/j.1651-2227.1996.tb14027.x.PubMedCrossRef
16.
go back to reference Holm C, Melcer B, Horbrand F, Henckel von Donnersmarck G, Muhlbauer W: Arterial thermodilution: an alternative to pulmonary artery catheter for cardiac output assessment in burn patients. Burns. 2001, 27: 161-166. 10.1016/S0305-4179(00)00088-7.PubMedCrossRef Holm C, Melcer B, Horbrand F, Henckel von Donnersmarck G, Muhlbauer W: Arterial thermodilution: an alternative to pulmonary artery catheter for cardiac output assessment in burn patients. Burns. 2001, 27: 161-166. 10.1016/S0305-4179(00)00088-7.PubMedCrossRef
17.
go back to reference Pauli C, Fakler U, Genz T, Hennig M, Lorenz HP, Hess J: Cardiac output determination in children: equivalence of the transpulmonary thermodilution method to the direct Fick principle. Intensive Care Med. 2002, 28: 947-952. 10.1007/s00134-002-1334-2.PubMedCrossRef Pauli C, Fakler U, Genz T, Hennig M, Lorenz HP, Hess J: Cardiac output determination in children: equivalence of the transpulmonary thermodilution method to the direct Fick principle. Intensive Care Med. 2002, 28: 947-952. 10.1007/s00134-002-1334-2.PubMedCrossRef
18.
go back to reference Tibby SM, Hatherill M, Marsh MJ, Morrison G, Anderson D, Murdoch IA: Clinical validation of cardiac output measurements using femoral artery thermodilution with direct Fick in ventilated children and infants. Intensive Care Med. 1997, 23: 987-991. 10.1007/s001340050443.PubMedCrossRef Tibby SM, Hatherill M, Marsh MJ, Morrison G, Anderson D, Murdoch IA: Clinical validation of cardiac output measurements using femoral artery thermodilution with direct Fick in ventilated children and infants. Intensive Care Med. 1997, 23: 987-991. 10.1007/s001340050443.PubMedCrossRef
19.
go back to reference Critchley LA, Lee A, Ho AM: A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010, 111: 1180-1192. 10.1213/ANE.0b013e3181f08a5b.PubMedCrossRef Critchley LA, Lee A, Ho AM: A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010, 111: 1180-1192. 10.1213/ANE.0b013e3181f08a5b.PubMedCrossRef
20.
go back to reference Godje O, Thiel C, Lamm P, Reichenspurner H, Schmitz C, Schutz A, Reichart B: Less invasive, continuous hemodynamic monitoring during minimally invasive coronary surgery. Ann Thorac Surg. 1999, 68: 1532-1536. 10.1016/S0003-4975(99)00956-X.PubMedCrossRef Godje O, Thiel C, Lamm P, Reichenspurner H, Schmitz C, Schutz A, Reichart B: Less invasive, continuous hemodynamic monitoring during minimally invasive coronary surgery. Ann Thorac Surg. 1999, 68: 1532-1536. 10.1016/S0003-4975(99)00956-X.PubMedCrossRef
21.
go back to reference Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL: Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010, 14: R109-10.1186/cc9058.PubMedPubMedCentralCrossRef Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL: Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010, 14: R109-10.1186/cc9058.PubMedPubMedCentralCrossRef
22.
go back to reference Cecconi M, Dawson D, Grounds RM, Rhodes A: Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Intensive Care Med. 2009, 35: 498-504. 10.1007/s00134-008-1292-4.PubMedCrossRef Cecconi M, Dawson D, Grounds RM, Rhodes A: Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Intensive Care Med. 2009, 35: 498-504. 10.1007/s00134-008-1292-4.PubMedCrossRef
23.
go back to reference Stetz CW, Miller RG, Kelly GE, Raffin TA: Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis. 1982, 126: 1001-1004.PubMed Stetz CW, Miller RG, Kelly GE, Raffin TA: Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis. 1982, 126: 1001-1004.PubMed
24.
go back to reference Nilsson LB, Nilsson JC, Skovgaard LT, Berthelsen PG: Thermodilution cardiac output--are three injections enough?. Acta Anaesthesiol Scand. 2004, 48: 1322-1327. 10.1111/j.1399-6576.2004.00514.x.PubMedCrossRef Nilsson LB, Nilsson JC, Skovgaard LT, Berthelsen PG: Thermodilution cardiac output--are three injections enough?. Acta Anaesthesiol Scand. 2004, 48: 1322-1327. 10.1111/j.1399-6576.2004.00514.x.PubMedCrossRef
25.
go back to reference Phillips CR, Chesnutt MS, Smith SM: Extravascular lung water in sepsis-associated acute respiratory distress syndrome: indexing with predicted body weight improves correlation with severity of illness and survival. Crit Care Med. 2008, 36: 69-73. 10.1097/01.CCM.0000295314.01232.BE.PubMedCrossRef Phillips CR, Chesnutt MS, Smith SM: Extravascular lung water in sepsis-associated acute respiratory distress syndrome: indexing with predicted body weight improves correlation with severity of illness and survival. Crit Care Med. 2008, 36: 69-73. 10.1097/01.CCM.0000295314.01232.BE.PubMedCrossRef
26.
go back to reference Groeneveld AB, Verheij J: Extravascular lung water to blood volume ratios as measures of permeability in sepsis-induced ALI/ARDS. Intensive Care Med. 2006, 32: 1315-1321. 10.1007/s00134-006-0212-8.PubMedCrossRef Groeneveld AB, Verheij J: Extravascular lung water to blood volume ratios as measures of permeability in sepsis-induced ALI/ARDS. Intensive Care Med. 2006, 32: 1315-1321. 10.1007/s00134-006-0212-8.PubMedCrossRef
27.
go back to reference Squara P: Bioreactance: A new method for non-invasive Cardiac Output monitoring. Yearbook of Intensive Care and Emergency Medicine. Edited by: Vincent JL. 2008, Berlin, Heidelberg: Springer, 619-630.CrossRef Squara P: Bioreactance: A new method for non-invasive Cardiac Output monitoring. Yearbook of Intensive Care and Emergency Medicine. Edited by: Vincent JL. 2008, Berlin, Heidelberg: Springer, 619-630.CrossRef
28.
go back to reference Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL: Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med. 2008, 36: 434-440. 10.1097/01.CCM.OB013E318161FEC4.PubMedCrossRef Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL: Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med. 2008, 36: 434-440. 10.1097/01.CCM.OB013E318161FEC4.PubMedCrossRef
29.
go back to reference Jabot J, Teboul JL, Richard C, Monnet X: Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009, 35: 85-90. 10.1007/s00134-008-1293-3.PubMedCrossRef Jabot J, Teboul JL, Richard C, Monnet X: Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009, 35: 85-90. 10.1007/s00134-008-1293-3.PubMedCrossRef
30.
go back to reference Monnet X, Osman D, Ridel C, Lamia B, Richard C, Teboul JL: Predicting volume responsiveness by using the end-expiratory occlusion in mechanically ventilated intensive care unit patients. Crit Care Med. 2009, 37: 951-956. 10.1097/CCM.0b013e3181968fe1.PubMedCrossRef Monnet X, Osman D, Ridel C, Lamia B, Richard C, Teboul JL: Predicting volume responsiveness by using the end-expiratory occlusion in mechanically ventilated intensive care unit patients. Crit Care Med. 2009, 37: 951-956. 10.1097/CCM.0b013e3181968fe1.PubMedCrossRef
Metadata
Title
Precision of the transpulmonary thermodilution measurements
Authors
Xavier Monnet
Romain Persichini
Mariem Ktari
Mathieu Jozwiak
Christian Richard
Jean-Louis Teboul
Publication date
01-08-2011
Publisher
BioMed Central
Published in
Critical Care / Issue 4/2011
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc10421

Other articles of this Issue 4/2011

Critical Care 4/2011 Go to the issue