01-08-2017 | Brief Report

Limitations of Rb-82 weight-adjusted dosing accuracy at low doses

Authors: Jonathan B. Moody, PhD, Kerri M. Hiller, BS, CNMT, RT(N), Benjamin C. Lee, PhD, James R. Corbett, MD, Edward P. Ficaro, PhD, Venkatesh L. Murthy, MD, PhD

Published in: Journal of Nuclear Cardiology | Issue 4/2017

Login to get access

Abstract

Introduction

Weight-adjusted dosing is important to maintain accurate quantification for dynamic cardiac three-dimensional positron emission tomography (PET). However, the manufacturer of the only Food and Drug Administration approved rubidium-82 (Rb-82) generator (CardioGen-82, Bracco Diagnostics, Inc.) recommends recalibration after each change in dose which is inefficient in a busy clinical PET lab. The objective of this study was to evaluate the accuracy of Rb-82 dosing without this recalibration.

Methods

After daily calibration to either 30 mCi (1110 MBq) or 12 mCi (444 MBq), Rb-82 doses between 6 and 45 mCi (222-1665 MBq) were eluted and measured on an external dose calibrator. This was repeated for four generators at weeks 1, 2, or 4 of the generator cycle. The measurements were compared with values reported by the infusion system.

Results

For requested doses less than 30 mCi, the measured dose was consistently lower than the requested dose, ranging from −3.2 mCi at 25 mCi to −5.5 mCi at 10 mCi. The error exceeded 10% for doses less than or equal to 26 mCi for the 30 mCi calibration. Residual activity in the infusion system dead volume accounted for the discrepancy between requested and delivered doses for calibration to 30 mCi but not 12 mCi.

Conclusions

The CardioGen-82 infusion system is capable of accurate weight-adjusted doses without recalibration for Rb-82 doses as low as 26 mCi when calibrated to 30 mCi. For doses less than 26 mCi, the generator-reported residual activity within the infusion system can be used to correct the delivered dose.
Appendix
Available only for authorised users
Literature
1.
go back to reference deKemp RA, Yoshinaga K, Beanlands RSB. Will 3-dimensional PET-CT enable the routine quantification of myocardial blood flow? J Nucl Cardiol. 2007;14:380–97.CrossRef deKemp RA, Yoshinaga K, Beanlands RSB. Will 3-dimensional PET-CT enable the routine quantification of myocardial blood flow? J Nucl Cardiol. 2007;14:380–97.CrossRef
2.
go back to reference Cerqueira MD, Allman KC, Ficaro EP, et al. Recommendations for reducing radiation exposure in myocardial perfusion imaging. J Nucl Cardiol. 2010;17:709–18.CrossRef Cerqueira MD, Allman KC, Ficaro EP, et al. Recommendations for reducing radiation exposure in myocardial perfusion imaging. J Nucl Cardiol. 2010;17:709–18.CrossRef
3.
go back to reference Dorbala S, Blankstein R, Skali H, et al. Approaches to reducing radiation dose from radionuclide myocardial perfusion imaging. J Nucl Med. 2015;56:592–9.CrossRef Dorbala S, Blankstein R, Skali H, et al. Approaches to reducing radiation dose from radionuclide myocardial perfusion imaging. J Nucl Med. 2015;56:592–9.CrossRef
4.
go back to reference Gerber TC, Carr JJ, Arai AE, et al. Ionizing radiation in cardiac imaging. A science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention. Circulation. 2009;119:1056–65.CrossRef Gerber TC, Carr JJ, Arai AE, et al. Ionizing radiation in cardiac imaging. A science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention. Circulation. 2009;119:1056–65.CrossRef
5.
go back to reference Fihn SD, Gardin JM, Abrams J, et al. 2012 ACCF/AHA/ACP/AATS/PCNA/SCAI/STS Guideline for the Diagnosis and Management of Patients With Stable Ischemic Heart Disease. A Report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, and the American College of Physicians, American Association for Thoracic Surgery, Preventive Cardiovascular Nurses Association, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons. J Am Coll Cardiol. 2012;60:e44–164.CrossRef Fihn SD, Gardin JM, Abrams J, et al. 2012 ACCF/AHA/ACP/AATS/PCNA/SCAI/STS Guideline for the Diagnosis and Management of Patients With Stable Ischemic Heart Disease. A Report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, and the American College of Physicians, American Association for Thoracic Surgery, Preventive Cardiovascular Nurses Association, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons. J Am Coll Cardiol. 2012;60:e44–164.CrossRef
6.
go back to reference Fazel R, Gerber TC, Balter S, et al. Approaches to enhancing radiation safety in cardiovascular imaging a scientific statement from the American Heart Association. Circulation. 2014;130:1730–48.CrossRef Fazel R, Gerber TC, Balter S, et al. Approaches to enhancing radiation safety in cardiovascular imaging a scientific statement from the American Heart Association. Circulation. 2014;130:1730–48.CrossRef
7.
go back to reference Bracco Diagnostics. CardioGen-82® (Rubidium Rb 82 Generator) Infusion System User Guide. 2013. Bracco Diagnostics. CardioGen-82® (Rubidium Rb 82 Generator) Infusion System User Guide. 2013.
8.
go back to reference Tout D, Tonge C, Muthu S, Arumugam P. Assessment of a protocol for routine simultaneous myocardial blood flow measurement and standard myocardial perfusion imaging with rubidium-82 on a high count rate positron emission tomography system. Nucl Med Commun. 2012;33:1202–11.CrossRef Tout D, Tonge C, Muthu S, Arumugam P. Assessment of a protocol for routine simultaneous myocardial blood flow measurement and standard myocardial perfusion imaging with rubidium-82 on a high count rate positron emission tomography system. Nucl Med Commun. 2012;33:1202–11.CrossRef
9.
go back to reference Saha GB, Go RT, Macintyre WJ, et al. Use of the 82Sr/82Rb generator in clinical PET studies. Int J Rad Appl Instrum B. 1990;17:763–8.CrossRef Saha GB, Go RT, Macintyre WJ, et al. Use of the 82Sr/82Rb generator in clinical PET studies. Int J Rad Appl Instrum B. 1990;17:763–8.CrossRef
10.
go back to reference Gennaro GP, Bergner BC, Haney PS, Kramer RH, Loberg MD. Radioanalysis of 82Rb generator eluates. Int J Rad Appl Instrum. 1987;38:219–25.CrossRef Gennaro GP, Bergner BC, Haney PS, Kramer RH, Loberg MD. Radioanalysis of 82Rb generator eluates. Int J Rad Appl Instrum. 1987;38:219–25.CrossRef
11.
go back to reference Renaud J, Huettges J, Gelbach J, Bush E, Port S, deKemp R. Dynamic range is acceptable for myocardial blood flow quantification with the Scintron LSO 3D-PET system. J Nucl Med. 2015;56:1851. Renaud J, Huettges J, Gelbach J, Bush E, Port S, deKemp R. Dynamic range is acceptable for myocardial blood flow quantification with the Scintron LSO 3D-PET system. J Nucl Med. 2015;56:1851.
12.
go back to reference Klein R, Adler A, Beanlands RS, deKemp RA. Precision-controlled elution of a 82Sr/82Rb generator for cardiac perfusion imaging with positron emission tomography. Phys Med Biol. 2007;52:659–73.CrossRef Klein R, Adler A, Beanlands RS, deKemp RA. Precision-controlled elution of a 82Sr/82Rb generator for cardiac perfusion imaging with positron emission tomography. Phys Med Biol. 2007;52:659–73.CrossRef
Metadata
Title
Limitations of Rb-82 weight-adjusted dosing accuracy at low doses
Authors
Jonathan B. Moody, PhD
Kerri M. Hiller, BS, CNMT, RT(N)
Benjamin C. Lee, PhD
James R. Corbett, MD
Edward P. Ficaro, PhD
Venkatesh L. Murthy, MD, PhD
Publication date
01-08-2017
Publisher
Springer International Publishing
Published in
Journal of Nuclear Cardiology / Issue 4/2017
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-016-0531-2

Other articles of this Issue 4/2017

Journal of Nuclear Cardiology 4/2017 Go to the issue