Acetabular positioning is more consistent with the use of a novel miniature computer-assisted device

Authors: Ivan Jacob, Jessica Benson, Kate Shanaghan, Alejandro Gonzalez Della Valle

Published in: International Orthopaedics | Issue 3/2020

Login to get access

Abstract

Introduction

Computer-assisted surgery (CAS) relying on registration of the anterior pelvic plane (APP) allows precise acetabular component placement. We determined the variability of cup placement in patients who underwent THA with and without the use of CAS that does not rely on the registration of APP.

Methods

Fifty-one patients who underwent staged-bilateral THAs, one without CAS (control), and a subsequent one with CAS (study group), were included. Acetabular inclination and anteversion were measured on standardized post-operative radiographs. Variance in cup position and Lewinnek’s zone proportionality were compared between the groups. Multiple regressions were performed to identify factors affecting variability in acetabular component placement.

Results

The mean inclination for the control and study group was 42.7° (SD 4.5) and 42.5° (SD 2.9), respectively. The inclination variance was 20.5° and 8.2° respectively (p = < 0.001). Cup inclination was more consistent in the study than in the control group (deviation from the mean: 2.3° vs. 3.8°, p < 0.001). The mean anteversion for the control and the study group was 25.5° (SD 7.4) and 26.8° (SD 4.3), respectively. The anteversion variance was 54.2° and 18.2° respectively (p = <0.001). Consistency in cup anteversion was significantly improved with CAS (deviation from the mean: 3.4° vs. 5.8°; p = 0.002). Lewinnek’s zone proportionality was not affected by the use of  CAS. In the linear regression analysis, CAS significantly increased consistency in cup inclination (p = 0.01). Patient’s factors including BMI and laterality affected consistency of cup placement.

Conclusion

CAS without referencing the APP allows a more consistent orientation of the acetabular component when compared to freehand placement.
Literature
1.
go back to reference Gonzalez Della Valle A, Padgett DE, Salvati EA (2005) Preoperative planning for primary total hip arthroplasty. J Am Acad Orthop Surg 13(7):455–462CrossRef Gonzalez Della Valle A, Padgett DE, Salvati EA (2005) Preoperative planning for primary total hip arthroplasty. J Am Acad Orthop Surg 13(7):455–462CrossRef
2.
go back to reference Murphy WS, Yun HH, Hayden B, Kowal JH, Murphy SB (2018) The safe zone range for cup anteversion is narrower than for inclination in THA. Clin Orthop Relat Res 476(2):325–335CrossRef Murphy WS, Yun HH, Hayden B, Kowal JH, Murphy SB (2018) The safe zone range for cup anteversion is narrower than for inclination in THA. Clin Orthop Relat Res 476(2):325–335CrossRef
3.
go back to reference Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ (1998) Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplast 13(5):530–534CrossRef Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ (1998) Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplast 13(5):530–534CrossRef
4.
go back to reference Hassani H, Cherix S, Ek ET, Rüdiger HA (2014) Comparisons of preoperative three-dimensional planning and surgical reconstruction in primary cementless total hip arthroplasty. J Arthroplast 29(6):1273–1277CrossRef Hassani H, Cherix S, Ek ET, Rüdiger HA (2014) Comparisons of preoperative three-dimensional planning and surgical reconstruction in primary cementless total hip arthroplasty. J Arthroplast 29(6):1273–1277CrossRef
5.
go back to reference Lee YK, Biau DJ, Yoon BH, Kim TY, Ha YC, Koo KH (2014) Learning curve of acetabular cup positioning in total hip arthroplasty using a cumulative summation test for learning curve (LC-CUSUM). J Arthroplast 29(3):586–589CrossRef Lee YK, Biau DJ, Yoon BH, Kim TY, Ha YC, Koo KH (2014) Learning curve of acetabular cup positioning in total hip arthroplasty using a cumulative summation test for learning curve (LC-CUSUM). J Arthroplast 29(3):586–589CrossRef
6.
go back to reference Heinert G, Hendricks J, Loeffler MD (2009) Digital templating in hip replacement with and without radiological markers. J Bone Joint Surg Br 91(4):459–462CrossRef Heinert G, Hendricks J, Loeffler MD (2009) Digital templating in hip replacement with and without radiological markers. J Bone Joint Surg Br 91(4):459–462CrossRef
7.
go back to reference Beckmann J, Lüring C, Tingart M, Anders S, Grifka J, Köck FX (2009) Cup positioning in THA: current status and pitfalls. A systematic evaluation of the literature. Arch Orthop and Trauma Surg 129(7):863–872CrossRef Beckmann J, Lüring C, Tingart M, Anders S, Grifka J, Köck FX (2009) Cup positioning in THA: current status and pitfalls. A systematic evaluation of the literature. Arch Orthop and Trauma Surg 129(7):863–872CrossRef
8.
go back to reference Schloemann DT, Edelstein AI, Barrack RL (2019) Changes in acetabular orientation during total hip arthroplasty. Bone Joint J 101(6 Suppl B):45–50CrossRef Schloemann DT, Edelstein AI, Barrack RL (2019) Changes in acetabular orientation during total hip arthroplasty. Bone Joint J 101(6 Suppl B):45–50CrossRef
9.
go back to reference Gonzalez Della Valle A, Shanaghan K, Benson JR, Carroll K, Cross M, McLawhorn A, Sculco PK (2019) Pelvic pitch and roll during total hip arthroplasty performed through a posterolateral approach. A potential source of error in free-hand cup positioning. Int Orthop 43(8):1823–1829CrossRef Gonzalez Della Valle A, Shanaghan K, Benson JR, Carroll K, Cross M, McLawhorn A, Sculco PK (2019) Pelvic pitch and roll during total hip arthroplasty performed through a posterolateral approach. A potential source of error in free-hand cup positioning. Int Orthop 43(8):1823–1829CrossRef
10.
go back to reference Rojas J, Bautista M, Bonilla G, Amado O, Huerfano E, Monsalvo D, Llinas A, Navas J (2018) A retrospective study on the relationship between altered native acetabular angle and vertical implant malpositioning. Int Orthop 42(4):769–775CrossRef Rojas J, Bautista M, Bonilla G, Amado O, Huerfano E, Monsalvo D, Llinas A, Navas J (2018) A retrospective study on the relationship between altered native acetabular angle and vertical implant malpositioning. Int Orthop 42(4):769–775CrossRef
11.
go back to reference Amirouche F, Solitro GF, Walia A, Gonzalez M, Bobko A (2017) Segmental acetabular rim defects, bone loss, oversizing, and press fit cup in total hip arthroplasty evaluated with a probabilistic finite element analysis. Int Orthop 41(8):1527–1533CrossRef Amirouche F, Solitro GF, Walia A, Gonzalez M, Bobko A (2017) Segmental acetabular rim defects, bone loss, oversizing, and press fit cup in total hip arthroplasty evaluated with a probabilistic finite element analysis. Int Orthop 41(8):1527–1533CrossRef
12.
go back to reference McArthur BA, Vulcano E, Cross M, Nguyen J, Gonzalez Della Valle A, Salvati E (2014) Acetabular component orientation in total hip arthroplasty: the impact of obesity. Hip Int 24(3):263–269CrossRef McArthur BA, Vulcano E, Cross M, Nguyen J, Gonzalez Della Valle A, Salvati E (2014) Acetabular component orientation in total hip arthroplasty: the impact of obesity. Hip Int 24(3):263–269CrossRef
13.
go back to reference Hassan DM, Johnston GH, Dust WN, Watson G, Dolovich AT (1998) Accuracy of intraoperative assessment of acetabular prosthesis placement. J Arthroplast 13(1):80–84CrossRef Hassan DM, Johnston GH, Dust WN, Watson G, Dolovich AT (1998) Accuracy of intraoperative assessment of acetabular prosthesis placement. J Arthroplast 13(1):80–84CrossRef
14.
go back to reference Minoda Y, Kadowaki T, Kim M (2006) Acetabular component orientation in 834 total hip arthroplasties using a manual technique. Clin Orthop Relat Res 445:186–191PubMed Minoda Y, Kadowaki T, Kim M (2006) Acetabular component orientation in 834 total hip arthroplasties using a manual technique. Clin Orthop Relat Res 445:186–191PubMed
15.
go back to reference González Della Valle A, Sharrock N, Barlow M, Caceres L, Go G, Salvati A (2016) The modern, hybrid total hip arthroplasty for primary osteoarthritis at the Hospital for Special Surgery. Bone Joint J 98(1_Supple_A):54–59CrossRef González Della Valle A, Sharrock N, Barlow M, Caceres L, Go G, Salvati A (2016) The modern, hybrid total hip arthroplasty for primary osteoarthritis at the Hospital for Special Surgery. Bone Joint J 98(1_Supple_A):54–59CrossRef
16.
go back to reference Paprosky WG, Muir JM (2016) Intellijoint HIP®: a 3D mini-optical navigation tool for improving intraoperative accuracy during total hip arthroplasty. Med Devices (Auckland, NZ) 9:401–408 Paprosky WG, Muir JM (2016) Intellijoint HIP®: a 3D mini-optical navigation tool for improving intraoperative accuracy during total hip arthroplasty. Med Devices (Auckland, NZ) 9:401–408
17.
go back to reference Parvizi J, Benson JR, Muir JM (2018) A new mini-navigation tool allows accurate component placement during anterior total hip arthroplasty. Med Devices (Auckland, NZ) 11:95–104 Parvizi J, Benson JR, Muir JM (2018) A new mini-navigation tool allows accurate component placement during anterior total hip arthroplasty. Med Devices (Auckland, NZ) 11:95–104
18.
go back to reference Bachhal V, Jindal N, Saini G, Sament R, Kumar V, Chouhan D, Dhillon M (2012) A new method of measuring acetabular cup anteversion on simulated radiographs. Int Orthop 36(9):1813–1818CrossRef Bachhal V, Jindal N, Saini G, Sament R, Kumar V, Chouhan D, Dhillon M (2012) A new method of measuring acetabular cup anteversion on simulated radiographs. Int Orthop 36(9):1813–1818CrossRef
19.
go back to reference Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220CrossRef Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220CrossRef
20.
go back to reference Jolles BM, Genoud P, Hoffmeyer P (2004) Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement. Clin Orthop Relat Res 426:174–179CrossRef Jolles BM, Genoud P, Hoffmeyer P (2004) Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement. Clin Orthop Relat Res 426:174–179CrossRef
21.
go back to reference Malik AT, Jain N, Scharschmidt TJ, Li M, Glassman AH, Khan SN (2018) Does surgeon volume affect outcomes following primary total hip arthroplasty? A systematic review. J Arthroplast 33(10):3329–3342CrossRef Malik AT, Jain N, Scharschmidt TJ, Li M, Glassman AH, Khan SN (2018) Does surgeon volume affect outcomes following primary total hip arthroplasty? A systematic review. J Arthroplast 33(10):3329–3342CrossRef
22.
go back to reference Najarian BC, Kilgore JE, Markel DC (2009) Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplast 24(1):15–21CrossRef Najarian BC, Kilgore JE, Markel DC (2009) Evaluation of component positioning in primary total hip arthroplasty using an imageless navigation device compared with traditional methods. J Arthroplast 24(1):15–21CrossRef
23.
go back to reference Ryan JA, Jamali AA, Bargar WL (2010) Accuracy of computer navigation for acetabular component placement in THA. Clin Orthop Relat Res 468(1):169–177CrossRef Ryan JA, Jamali AA, Bargar WL (2010) Accuracy of computer navigation for acetabular component placement in THA. Clin Orthop Relat Res 468(1):169–177CrossRef
24.
go back to reference Sendtner E, Schuster T, Wörner M, Kalteis T, Grifka J, Renkawitz T (2011) Accuracy of acetabular cup placement in computer-assisted, minimally-invasive THR in a lateral decubitus position. Int Orthop 35(6):809–815CrossRef Sendtner E, Schuster T, Wörner M, Kalteis T, Grifka J, Renkawitz T (2011) Accuracy of acetabular cup placement in computer-assisted, minimally-invasive THR in a lateral decubitus position. Int Orthop 35(6):809–815CrossRef
25.
go back to reference Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2011) Limitations of imageless computer-assisted navigation for total hip arthroplasty. J Arthroplast 26(4):596–605CrossRef Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2011) Limitations of imageless computer-assisted navigation for total hip arthroplasty. J Arthroplast 26(4):596–605CrossRef
26.
go back to reference Hohmann E, Bryant A, Tetsworth K (2011) A comparison between imageless navigated and manual freehand technique acetabular cup placement in total hip arthroplasty. J Arthroplast 26(7):1078–1082CrossRef Hohmann E, Bryant A, Tetsworth K (2011) A comparison between imageless navigated and manual freehand technique acetabular cup placement in total hip arthroplasty. J Arthroplast 26(7):1078–1082CrossRef
27.
go back to reference Lass R, Kubista B, Olischar B, Frantal S, Windhager R, Giurea A (2014) Total hip arthroplasty using imageless computer-assisted hip navigation: a prospective randomized study. J Arthroplast 29(4):786–791CrossRef Lass R, Kubista B, Olischar B, Frantal S, Windhager R, Giurea A (2014) Total hip arthroplasty using imageless computer-assisted hip navigation: a prospective randomized study. J Arthroplast 29(4):786–791CrossRef
28.
go back to reference Buller LT, McLawhorn AS, Romero JA, Sculco PK, Mayman DJ (2019) Accuracy and precision of acetabular component placement with imageless navigation in obese patients. J Arthroplast 34(4):693–699CrossRef Buller LT, McLawhorn AS, Romero JA, Sculco PK, Mayman DJ (2019) Accuracy and precision of acetabular component placement with imageless navigation in obese patients. J Arthroplast 34(4):693–699CrossRef
29.
go back to reference Parratte S, Argenson JNA (2007) Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty: a prospective, randomized, controlled study. J Bone Joint Surg Am 89(3):494–499CrossRef Parratte S, Argenson JNA (2007) Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty: a prospective, randomized, controlled study. J Bone Joint Surg Am 89(3):494–499CrossRef
30.
go back to reference Kalteis T, Handel M, Bäthis H, Perlick L, Tingart M, Grifka J (2006) Imageless navigation for insertion of the acetabular component in total hip arthroplasty: is it as accurate as CT-based navigation? J Bone Joint Surg Br 88(2):163–167CrossRef Kalteis T, Handel M, Bäthis H, Perlick L, Tingart M, Grifka J (2006) Imageless navigation for insertion of the acetabular component in total hip arthroplasty: is it as accurate as CT-based navigation? J Bone Joint Surg Br 88(2):163–167CrossRef
31.
go back to reference Gurgel HM, Croci AT, Cabrita HA, Vicente JRN, Leonhardt MC, Rodrigues JC (2014) Acetabular component positioning in total hip arthroplasty with and without a computer-assisted system: a prospective, randomized and controlled study. J Arthroplast 29(1):167–171CrossRef Gurgel HM, Croci AT, Cabrita HA, Vicente JRN, Leonhardt MC, Rodrigues JC (2014) Acetabular component positioning in total hip arthroplasty with and without a computer-assisted system: a prospective, randomized and controlled study. J Arthroplast 29(1):167–171CrossRef
32.
go back to reference Boylan M, Suchman K, Vigdorchik J, Slover J, Bosco J (2018) Technology-assisted hip and knee arthroplasties: an analysis of utilization trends. J Arthroplast 33(4):1019–1023CrossRef Boylan M, Suchman K, Vigdorchik J, Slover J, Bosco J (2018) Technology-assisted hip and knee arthroplasties: an analysis of utilization trends. J Arthroplast 33(4):1019–1023CrossRef
33.
go back to reference Davis ET, Schubert M, Wegner M, Haimerl M (2015) A new method of registration in navigated hip arthroplasty without the need to register the anterior pelvic plane. J Arthroplast 30(1):55–60CrossRef Davis ET, Schubert M, Wegner M, Haimerl M (2015) A new method of registration in navigated hip arthroplasty without the need to register the anterior pelvic plane. J Arthroplast 30(1):55–60CrossRef
34.
go back to reference González Della Valle A, Beksac B, Peterson MGE, Salvati EA (2006) A new technique to improve cup inclination accuracy in primary total hip arthroplasty. Hip Int 16(4):250–252CrossRef González Della Valle A, Beksac B, Peterson MGE, Salvati EA (2006) A new technique to improve cup inclination accuracy in primary total hip arthroplasty. Hip Int 16(4):250–252CrossRef
35.
go back to reference Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H (2011) The John Charnley award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res 469(2):319–329CrossRef Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H (2011) The John Charnley award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res 469(2):319–329CrossRef
36.
go back to reference Song X, Ni M, Li H, Li X, Li X, Fu J, Chen J (2018) Is the cup orientation different in bilateral total hip arthroplasty with right-handed surgeons using posterolateral approach? J Orthop Surg Res 13(1):123CrossRef Song X, Ni M, Li H, Li X, Li X, Fu J, Chen J (2018) Is the cup orientation different in bilateral total hip arthroplasty with right-handed surgeons using posterolateral approach? J Orthop Surg Res 13(1):123CrossRef
37.
go back to reference Kamara E, Robinson J, Bas MA, Rodriguez JA, Hepinstall MS (2017) Adoption of robotic vs fluoroscopic guidance in total hip arthroplasty: is acetabular positioning improved in the learning curve? J Arthroplast 32(1):125–130CrossRef Kamara E, Robinson J, Bas MA, Rodriguez JA, Hepinstall MS (2017) Adoption of robotic vs fluoroscopic guidance in total hip arthroplasty: is acetabular positioning improved in the learning curve? J Arthroplast 32(1):125–130CrossRef
38.
go back to reference Kayani B, Konan S, Ayuob A, Ayyad S, Haddad FS (2019) The current role of robotics in total hip arthroplasty. EFORT Open Rev 4(11):618–625CrossRef Kayani B, Konan S, Ayuob A, Ayyad S, Haddad FS (2019) The current role of robotics in total hip arthroplasty. EFORT Open Rev 4(11):618–625CrossRef
Metadata
Title
Acetabular positioning is more consistent with the use of a novel miniature computer-assisted device
Authors
Ivan Jacob
Jessica Benson
Kate Shanaghan
Alejandro Gonzalez Della Valle
Publication date
01-03-2020
Publisher
Springer Berlin Heidelberg
Keywords
Hip-TEP
Hip-TEP
Published in
International Orthopaedics / Issue 3/2020
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-020-04484-2

Other articles of this Issue 3/2020

International Orthopaedics 3/2020 Go to the issue