Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study

Authors: Emmanuelle Jaillette, Farid Zerimech, Julien De Jonckheere, Demosthenes Makris, Malika Balduyck, Alain Durocher, Alain Duhamel, Saad Nseir

Published in: BMC Anesthesiology | Issue 1/2013

Login to get access

Abstract

Background

The primary objective of this study was to determine the efficiency of a pneumatic device in controlling cuff pressure (Pcuff) in patients intubated with polyurethane-cuffed tracheal tubes. Secondary objectives were to determine the impact of continuous control of Pcuff, and cuff shape on microaspiration of gastric contents.

Methods

Prospective randomized controlled study. All patients requiring intubation and mechanical ventilation ≥48 h were eligible. The first 32 patients were intubated with tapered polyurethane-cuffed, and the 32 following patients were intubated with cylindrical polyurethane-cuffed tracheal tubes. Patients randomly received 24 h of continuous control of Pcuff using a pneumatic device (Nosten®), and 24 h of routine care of Pcuff using a manometer. Target Pcuff was 25 cmH2O. Pcuff was continuously recorded, and pepsin was quantitatively measured in all tracheal aspirates during these periods.

Results

The pneumatic device was efficient in controlling Pcuff (med [IQ] 26 [24, 28] vs 22 [20, 28] cmH2O, during continuous control of Pcuff and routine care, respectively; p = 0.017). In addition, percentage of patients with underinflation (31% vs 68%) or overinflation (53% vs 100%) of tracheal cuff, and percentage of time spent with underinflation (0.9 [0, 17] vs 14% [4, 30]) or overinflation (0 [0, 2] vs 32% [9, 54]) were significantly (p < 0.001) reduced during continuous control of Pcuff compared with routine care.
No significant difference was found in microaspiration of gastric content between continuous control of Pcuff compared with routine care, or between patients intubated with tapered compared with cylindrical polyurethane-cuffed tracheal tubes.

Conclusion

The pneumatic device was efficient in controlling Pcuff in critically ill patients intubated with polyurethane-cuffed tracheal tubes.

Trial registration

The Australian New Zealand Clinical Trials Registry (NCT01351259)
Appendix
Available only for authorised users
Literature
1.
go back to reference Nseir S, Zerimech F, Jaillette E, Artru F, Balduyck M: Microaspiration in intubated critically ill patients: diagnosis and prevention. Infect Disor Drug Target. 2011, 11: 413-423. 10.2174/187152611796504827.CrossRef Nseir S, Zerimech F, Jaillette E, Artru F, Balduyck M: Microaspiration in intubated critically ill patients: diagnosis and prevention. Infect Disor Drug Target. 2011, 11: 413-423. 10.2174/187152611796504827.CrossRef
2.
go back to reference Craven DE, Chroneou A, Zias N, Hjalmarson KI: Ventilator-associated tracheobronchitis: the impact of targeted antibiotic therapy on patient outcomes. Chest. 2009, 135: 521-528. 10.1378/chest.08-1617.CrossRefPubMed Craven DE, Chroneou A, Zias N, Hjalmarson KI: Ventilator-associated tracheobronchitis: the impact of targeted antibiotic therapy on patient outcomes. Chest. 2009, 135: 521-528. 10.1378/chest.08-1617.CrossRefPubMed
3.
go back to reference Bouadma L, Wolff M, Lucet J-C: Ventilator-associated pneumonia and its prevention. Curr Opin Infect Dis. 2012, 25: 395-404. 10.1097/QCO.0b013e328355a835.CrossRefPubMed Bouadma L, Wolff M, Lucet J-C: Ventilator-associated pneumonia and its prevention. Curr Opin Infect Dis. 2012, 25: 395-404. 10.1097/QCO.0b013e328355a835.CrossRefPubMed
5.
go back to reference Lucangelo U, Zin WA, Antonaglia V, Petrucci L, Viviani M, Buscema G, Borelli M, Berlot G: Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit. Crit Care Med. 2008, 36: 409-413. 10.1097/01.CCM.0000297888.82492.31.CrossRefPubMed Lucangelo U, Zin WA, Antonaglia V, Petrucci L, Viviani M, Buscema G, Borelli M, Berlot G: Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit. Crit Care Med. 2008, 36: 409-413. 10.1097/01.CCM.0000297888.82492.31.CrossRefPubMed
6.
go back to reference Nseir S, Zerimech F, De Jonckheere J, Alves I, Balduyck M, Durocher A: Impact of polyurethane on variations in tracheal cuff pressure in critically ill patients: a prospective observational study. Intensive Care Med. 2010, 36: 1156-1163. 10.1007/s00134-010-1892-7.CrossRefPubMed Nseir S, Zerimech F, De Jonckheere J, Alves I, Balduyck M, Durocher A: Impact of polyurethane on variations in tracheal cuff pressure in critically ill patients: a prospective observational study. Intensive Care Med. 2010, 36: 1156-1163. 10.1007/s00134-010-1892-7.CrossRefPubMed
7.
go back to reference Dullenkopf A, Gerber A, Weiss M: Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Med. 2003, 29: 1849-1853. 10.1007/s00134-003-1933-6.CrossRefPubMed Dullenkopf A, Gerber A, Weiss M: Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Med. 2003, 29: 1849-1853. 10.1007/s00134-003-1933-6.CrossRefPubMed
8.
go back to reference Lorente L, Lecuona M, Jiménez A, Mora ML, Sierra A: Influence of an endotracheal tube with polyurethane cuff and subglottic secretion drainage on pneumonia. Am J Respir Crit Care Med. 2007, 176: 1079-1083. 10.1164/rccm.200705-761OC.CrossRefPubMed Lorente L, Lecuona M, Jiménez A, Mora ML, Sierra A: Influence of an endotracheal tube with polyurethane cuff and subglottic secretion drainage on pneumonia. Am J Respir Crit Care Med. 2007, 176: 1079-1083. 10.1164/rccm.200705-761OC.CrossRefPubMed
9.
go back to reference Poelaert J, Depuydt P, De Wolf A, Van de Velde S, Herck I, Blot S: Polyurethane cuffed endotracheal tubes to prevent early postoperative pneumonia after cardiac surgery: a pilot study. J Thor Cardiovascular Surg. 2008, 135: 771-776. 10.1016/j.jtcvs.2007.08.052.CrossRef Poelaert J, Depuydt P, De Wolf A, Van de Velde S, Herck I, Blot S: Polyurethane cuffed endotracheal tubes to prevent early postoperative pneumonia after cardiac surgery: a pilot study. J Thor Cardiovascular Surg. 2008, 135: 771-776. 10.1016/j.jtcvs.2007.08.052.CrossRef
10.
go back to reference Miller MA, Arndt JL, Konkle MA, Chenoweth CE, Iwashyna TJ, Flaherty KR, Hyzy RC: A polyurethane cuffed endotracheal tube is associated with decreased rates of ventilator-associated pneumonia. J Crit Care. 2011, 26: 280-286. 10.1016/j.jcrc.2010.05.035.CrossRefPubMed Miller MA, Arndt JL, Konkle MA, Chenoweth CE, Iwashyna TJ, Flaherty KR, Hyzy RC: A polyurethane cuffed endotracheal tube is associated with decreased rates of ventilator-associated pneumonia. J Crit Care. 2011, 26: 280-286. 10.1016/j.jcrc.2010.05.035.CrossRefPubMed
11.
go back to reference Madjdpour C, Mauch J, Dave MH, Spielmann N, Weiss M: Comparison of air-sealing characteristics of tapered- vs. cylindrical-shaped high-volume, low-pressure tube cuffs. Acta Anaesthesiol Scand. 2012, 56: 230-235. 10.1111/j.1399-6576.2011.02542.x.CrossRefPubMed Madjdpour C, Mauch J, Dave MH, Spielmann N, Weiss M: Comparison of air-sealing characteristics of tapered- vs. cylindrical-shaped high-volume, low-pressure tube cuffs. Acta Anaesthesiol Scand. 2012, 56: 230-235. 10.1111/j.1399-6576.2011.02542.x.CrossRefPubMed
12.
go back to reference Dave MH, Frotzler A, Spielmann N, Madjdpour C, Weiss M: Effect of tracheal tube cuff shape on fluid leakage across the cuff: an in vitro study. Brit J Anaesthesia. 2010, 105: 538-543. 10.1093/bja/aeq202.CrossRef Dave MH, Frotzler A, Spielmann N, Madjdpour C, Weiss M: Effect of tracheal tube cuff shape on fluid leakage across the cuff: an in vitro study. Brit J Anaesthesia. 2010, 105: 538-543. 10.1093/bja/aeq202.CrossRef
13.
go back to reference Zanella A, Scaravilli V, Isgrò S, Milan M, Cressoni M, Patroniti N, Fumagalli R, Pesenti A: Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study. Intensive Care Med. 2011, 37: 343-347. 10.1007/s00134-010-2106-z.CrossRefPubMed Zanella A, Scaravilli V, Isgrò S, Milan M, Cressoni M, Patroniti N, Fumagalli R, Pesenti A: Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study. Intensive Care Med. 2011, 37: 343-347. 10.1007/s00134-010-2106-z.CrossRefPubMed
14.
go back to reference Young PJ, Rollinson M, Downward G, Henderson S: Leakage of fluid past the tracheal tube cuff in a benchtop model. Brit J Anaesthesia. 1997, 78: 557-562. 10.1093/bja/78.5.557.CrossRef Young PJ, Rollinson M, Downward G, Henderson S: Leakage of fluid past the tracheal tube cuff in a benchtop model. Brit J Anaesthesia. 1997, 78: 557-562. 10.1093/bja/78.5.557.CrossRef
15.
go back to reference Rello J, Soñora R, Jubert P, Artigas A, Rué M, Vallés J: Pneumonia in intubated patients: role of respiratory airway care. Am J Respir Crit Care Med. 1996, 154: 111-115. 10.1164/ajrccm.154.1.8680665.CrossRefPubMed Rello J, Soñora R, Jubert P, Artigas A, Rué M, Vallés J: Pneumonia in intubated patients: role of respiratory airway care. Am J Respir Crit Care Med. 1996, 154: 111-115. 10.1164/ajrccm.154.1.8680665.CrossRefPubMed
16.
go back to reference Seegobin RD, van Hasselt GL: Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. Brit Med J. 1984, 288: 965-968. 10.1136/bmj.288.6422.965.CrossRef Seegobin RD, van Hasselt GL: Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. Brit Med J. 1984, 288: 965-968. 10.1136/bmj.288.6422.965.CrossRef
17.
go back to reference Nseir S, Brisson H, Marquette C-H, Chaud P, Di Pompeo C, Diarra M, Durocher A: Variations in endotracheal cuff pressure in intubated critically ill patients: prevalence and risk factors. Eur J Anaesthsiol. 2009, 26: 229-234. 10.1097/EJA.0b013e3283222b6e.CrossRef Nseir S, Brisson H, Marquette C-H, Chaud P, Di Pompeo C, Diarra M, Durocher A: Variations in endotracheal cuff pressure in intubated critically ill patients: prevalence and risk factors. Eur J Anaesthsiol. 2009, 26: 229-234. 10.1097/EJA.0b013e3283222b6e.CrossRef
18.
go back to reference Nseir S, Zerimech F, Fournier C, Lubret R, Ramon P, Durocher A, Balduyck M: Continuous control of tracheal cuff pressure and microaspiration of gastric contents in critically ill patients. Am J Respir Crit Care Med. 2011, 184: 1041-1047. 10.1164/rccm.201104-0630OC.CrossRefPubMed Nseir S, Zerimech F, Fournier C, Lubret R, Ramon P, Durocher A, Balduyck M: Continuous control of tracheal cuff pressure and microaspiration of gastric contents in critically ill patients. Am J Respir Crit Care Med. 2011, 184: 1041-1047. 10.1164/rccm.201104-0630OC.CrossRefPubMed
19.
go back to reference Chastre J, Bedock B, Clair B, Gehanno P, Lacaze T, Lesieur O, Picart-Jacq JY, Plaisance P, Ravussin P, Samain E, et al: Quel abord trachéal pour la ventilation mécanique des malades de réanimation? (à l’exclusion du nouveau né). Reanimation. 1998, 7: 438-442. Chastre J, Bedock B, Clair B, Gehanno P, Lacaze T, Lesieur O, Picart-Jacq JY, Plaisance P, Ravussin P, Samain E, et al: Quel abord trachéal pour la ventilation mécanique des malades de réanimation? (à l’exclusion du nouveau né). Reanimation. 1998, 7: 438-442.
20.
go back to reference Duguet A, D’Amico L, Biondi G, Prodanovic H, Gonzalez-Bermejo J, Similowski T: Control of tracheal cuff pressure: a pilot study using a pneumatic device. Intensive Care Med. 2007, 33: 128-132. 10.1007/s00134-006-0417-x.CrossRefPubMed Duguet A, D’Amico L, Biondi G, Prodanovic H, Gonzalez-Bermejo J, Similowski T: Control of tracheal cuff pressure: a pilot study using a pneumatic device. Intensive Care Med. 2007, 33: 128-132. 10.1007/s00134-006-0417-x.CrossRefPubMed
21.
go back to reference Nseir S, Duguet A, Copin M-C, De Jonckheere J, Zhang M, Similowski T, Marquette C-H: Continuous control of endotracheal cuff pressure and tracheal wall damage: a randomized controlled animal study. Crit Care. 2007, 11: R109-10.1186/cc6142.CrossRefPubMedPubMedCentral Nseir S, Duguet A, Copin M-C, De Jonckheere J, Zhang M, Similowski T, Marquette C-H: Continuous control of endotracheal cuff pressure and tracheal wall damage: a randomized controlled animal study. Crit Care. 2007, 11: R109-10.1186/cc6142.CrossRefPubMedPubMedCentral
22.
go back to reference Guyton DC, Barlow MR, Besselievre TR: Influence of airway pressure on minimum occlusive endotracheal tube cuff pressure. Crit Care Med. 1997, 25: 91-94. 10.1097/00003246-199701000-00018.CrossRefPubMed Guyton DC, Barlow MR, Besselievre TR: Influence of airway pressure on minimum occlusive endotracheal tube cuff pressure. Crit Care Med. 1997, 25: 91-94. 10.1097/00003246-199701000-00018.CrossRefPubMed
23.
go back to reference Blanch PB: Laboratory evaluation of 4 brands of endotracheal tube cuff inflator. Respir Care. 2004, 49: 166-173.PubMed Blanch PB: Laboratory evaluation of 4 brands of endotracheal tube cuff inflator. Respir Care. 2004, 49: 166-173.PubMed
24.
go back to reference Valta P, Corbeil C, Chassé M, Braidy J, Milic-Emili J: Mean airway pressure as an index of mean alveolar pressure. Am J Respir Crit Care Med. 1996, 153: 1825-1830. 10.1164/ajrccm.153.6.8665041.CrossRefPubMed Valta P, Corbeil C, Chassé M, Braidy J, Milic-Emili J: Mean airway pressure as an index of mean alveolar pressure. Am J Respir Crit Care Med. 1996, 153: 1825-1830. 10.1164/ajrccm.153.6.8665041.CrossRefPubMed
25.
go back to reference Sole ML, Su X, Talbert S, Penoyer DA, Kalita S, Jimenez E, Ludy JE, Bennett M: Evaluation of an intervention to maintain endotracheal tube cuff pressure within therapeutic range. Am J Crit Care. 2011, 20: 109-117. 10.4037/ajcc2011661.CrossRefPubMedPubMedCentral Sole ML, Su X, Talbert S, Penoyer DA, Kalita S, Jimenez E, Ludy JE, Bennett M: Evaluation of an intervention to maintain endotracheal tube cuff pressure within therapeutic range. Am J Crit Care. 2011, 20: 109-117. 10.4037/ajcc2011661.CrossRefPubMedPubMedCentral
Metadata
Title
Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study
Authors
Emmanuelle Jaillette
Farid Zerimech
Julien De Jonckheere
Demosthenes Makris
Malika Balduyck
Alain Durocher
Alain Duhamel
Saad Nseir
Publication date
01-12-2013
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2013
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/1471-2253-13-50

Other articles of this Issue 1/2013

BMC Anesthesiology 1/2013 Go to the issue