01-02-2011 | Physiological and Technical Notes

Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study

Authors: Alberto Zanella, Vittorio Scaravilli, Stefano Isgrò, Manuela Milan, Massimo Cressoni, Nicolò Patroniti, Roberto Fumagalli, Antonio Pesenti

Published in: Intensive Care Medicine | Issue 2/2011

Login to get access

Abstract

Purpose

Standard polyvinylchloride (PVC) endotracheal tube (ETT) cuffs do not protect from aspiration across the cuff, a leading cause of ventilator-associated pneumonia (VAP). In a long-lasting in vitro study we compared the effect of different cuff materials (PVC, polyurethane, and guayule latex), shapes (cylindrical, conical), and positive end expiratory pressures (PEEP) in reducing fluid leakage across the cuff.

Methods

We compared fluid leakage across a cylindrical double-layer guayule latex prototype cuff, three cylindrical PVC cuffs (Mallinckrodt Hi-Lo, Mallinckrodt HighContour, Portex Ivory), one conical PVC cuff (Mallinckrodt TaperGuard), and two polyurethane cuffs (Mallinckrodt SealGuard, conical; Microcuff, cylindrical). Ten centimeters of dyed water was poured above the cuffs inflated (pressure 30 cmH2O) in a vertical cylinder (diameter 20 mm). A respiratory circuit connected the bottom of the cylinder to a breathing bag inflated at four pressures (PEEP = 0, 5, 10, 15 cmH2O). Pictures were taken every 60 s for 24 h to measure leakage as a reduction in the water column above the cuff. Five new ETTs of each type were tested.

Results

The guayule latex cuffs showed no leakage at all the PEEP levels. Both the cylindrical and conical polyurethane cuffs showed limited leakage (2.1 ± 1.8 cm of water) only for PEEP zero. The PVC cuffs showed reduced leakage with increasing PEEP: 8.4 ± 1.5, 7.8 ± 2.2, 2.2 ± 1.0, and 0 cm of water at 0, 5, 10, and 15 cmH2O, respectively. Among all the PVC cuffs, the conical shape ensured higher sealing properties.

Conclusions

The guayule latex cuffs always prevented fluid leakage; the polyurethane and PVC cuffs required incremental levels of PEEP to prevent fluid leakage ever-present at zero PEEP.
Literature
1.
go back to reference Cooper JD, Grillo HC (1969) The evolution of tracheal injury due to ventilatory assistance through cuffed tubes: a pathologic study. Ann Surg 169:334–348CrossRefPubMed Cooper JD, Grillo HC (1969) The evolution of tracheal injury due to ventilatory assistance through cuffed tubes: a pathologic study. Ann Surg 169:334–348CrossRefPubMed
2.
go back to reference Seegobin RD, van Hasselt GL (1986) Aspiration beyond endotracheal cuffs. Can Anaesth Soc J 33:273–279CrossRefPubMed Seegobin RD, van Hasselt GL (1986) Aspiration beyond endotracheal cuffs. Can Anaesth Soc J 33:273–279CrossRefPubMed
3.
4.
go back to reference Janson BA, Poulton TJ (1986) Does PEEP reduce the incidence of aspiration around endotracheal tubes? Can Anaesth Soc J 33:157–161CrossRefPubMed Janson BA, Poulton TJ (1986) Does PEEP reduce the incidence of aspiration around endotracheal tubes? Can Anaesth Soc J 33:157–161CrossRefPubMed
5.
go back to reference Young PJ, Rollinson M, Downward G, Henderson S (1997) Leakage of fluid past the tracheal tube cuff in a benchtop model. Br J Anaesth 78:557–562PubMed Young PJ, Rollinson M, Downward G, Henderson S (1997) Leakage of fluid past the tracheal tube cuff in a benchtop model. Br J Anaesth 78:557–562PubMed
6.
go back to reference Lucangelo U, Zin WA, Antonaglia V, Petrucci L, Viviani M, Buscema G, Borelli M, Berlot G (2008) Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit. Crit Care Med 36:409–413CrossRefPubMed Lucangelo U, Zin WA, Antonaglia V, Petrucci L, Viviani M, Buscema G, Borelli M, Berlot G (2008) Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit. Crit Care Med 36:409–413CrossRefPubMed
7.
go back to reference Manzano F, Fernandez-Mondejar E, Colmenero M, Poyatos ME, Rivera R, Machado J, Catalan I, Artigas A (2008) Positive-end expiratory pressure reduces incidence of ventilator-associated pneumonia in nonhypoxemic patients. Crit Care Med 36:2225–2231CrossRefPubMed Manzano F, Fernandez-Mondejar E, Colmenero M, Poyatos ME, Rivera R, Machado J, Catalan I, Artigas A (2008) Positive-end expiratory pressure reduces incidence of ventilator-associated pneumonia in nonhypoxemic patients. Crit Care Med 36:2225–2231CrossRefPubMed
8.
go back to reference Zanella A, Cressoni M, Epp M, Stylianou M, Kolobow T (2008) A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study. Intensive Care Med 34:1145–1149CrossRefPubMed Zanella A, Cressoni M, Epp M, Stylianou M, Kolobow T (2008) A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study. Intensive Care Med 34:1145–1149CrossRefPubMed
9.
go back to reference Dave MH, Koepfer N, Madjdpour C, Frotzler A, Weiss M (2010) Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication. J Anesth 24:247–252CrossRefPubMed Dave MH, Koepfer N, Madjdpour C, Frotzler A, Weiss M (2010) Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication. J Anesth 24:247–252CrossRefPubMed
Metadata
Title
Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study
Authors
Alberto Zanella
Vittorio Scaravilli
Stefano Isgrò
Manuela Milan
Massimo Cressoni
Nicolò Patroniti
Roberto Fumagalli
Antonio Pesenti
Publication date
01-02-2011
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 2/2011
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-010-2106-z

Other articles of this Issue 2/2011

Intensive Care Medicine 2/2011 Go to the issue