08-12-2022 | ORIGINAL ARTICLE

Dose-consistent dynamic SPECT

Authors: A. Krakovich, MSc, E. Gelbart, MSc, I. Moalem, A. Naimushin, MD, E. Rozen, MD, M. Scheinowitz, PhD, R. Goldkorn, MD

Published in: Journal of Nuclear Cardiology | Issue 4/2023

Login to get access

Abstract

Background

Coronary flow reserve (CFR) values measured by dynamic SPECT systems are typically consistent with other modalities (e.g., PET). However, large discrepancies are often observed for individual patients. Positioning of the region-of-interest (ROI), representing the arterial input function (AIF) could explain some of these discrepancies. We explored the possibility of positioning the ROI in a manner that evaluates its consistency with patient-based injected radiotracer doses.

Methods

Dose-consistent dynamic SPECT methodology was introduced, and its application was demonstrated in a twenty-patient clinical study. The effect of various ROI positions was investigated and comparison to myocardial perfusion imaging was performed.

Results

Mean AIF ratios were consistent with the injected dose ratios for all examined ROI positions. Good agreement (> 80%) between total perfusion deficit and CFR was found in the detection of obstructive CAD patients for all ROIs considered. However, for individual patients, significant dependence on ROI position was observed (altering CFR by typically 30%). The proposed methodology’s uncertainty was evaluated (~ 7%) and found to be smaller than the variability due to choice of ROI position.

Conclusion

Dose-consistent dynamic SPECT may contribute to evaluating uncertainty of CFR measurements and may potentially decrease uncertainty by allowing improved ROI positioning for individual patients.
Appendix
Available only for authorised users
Literature
1.
go back to reference Murthy VL, Bateman TM, Beanlands RS, Berman DS, Borges-Neto S, Chareonthaitawee P. Clinical quantification of myocardial blood flow using PET: Joint position paper of the SNMMI cardiovascular council and the ASNC. J Nucl Med 2018;25:269‐97. Murthy VL, Bateman TM, Beanlands RS, Berman DS, Borges-Neto S, Chareonthaitawee P. Clinical quantification of myocardial blood flow using PET: Joint position paper of the SNMMI cardiovascular council and the ASNC. J Nucl Med 2018;25:269‐97.
2.
go back to reference Herzog BA, Husmann L, Valenta I, Gaemperli O, Siegrist PT, Tay FM, et al. Long-term prognostic value of 13N-ammonia myocardial perfusion positron emission tomography: Added value of coronary flow reserve. J Am Coll Cardiol 2009;54:150‐6.CrossRefPubMed Herzog BA, Husmann L, Valenta I, Gaemperli O, Siegrist PT, Tay FM, et al. Long-term prognostic value of 13N-ammonia myocardial perfusion positron emission tomography: Added value of coronary flow reserve. J Am Coll Cardiol 2009;54:150‐6.CrossRefPubMed
3.
go back to reference Holly TA, Abbott BG, Al-Mallah M, Calnon DA, Cohen MC, DiFilippo FP, et al. Single photon-emission computed tomography. J Nucl Cardiol 2010;17:941‐73.CrossRefPubMed Holly TA, Abbott BG, Al-Mallah M, Calnon DA, Cohen MC, DiFilippo FP, et al. Single photon-emission computed tomography. J Nucl Cardiol 2010;17:941‐73.CrossRefPubMed
4.
go back to reference Lima RS, Watson DD, Goode AR, Siadaty MS, Ragosta M, Beller GA, et al. Incremental value of combined perfusion and function over perfusion alone by gated SPECT myocardial perfusion imaging for detection of severe three-vessel coronary artery disease. J Am Coll Cardiol 2003;42:64‐70.CrossRefPubMed Lima RS, Watson DD, Goode AR, Siadaty MS, Ragosta M, Beller GA, et al. Incremental value of combined perfusion and function over perfusion alone by gated SPECT myocardial perfusion imaging for detection of severe three-vessel coronary artery disease. J Am Coll Cardiol 2003;42:64‐70.CrossRefPubMed
8.
go back to reference Bocher M, Blevis IM, Tsukerman L, Shrem Y, Kovalski G, Volokh L. A fast cardiac gamma camera with dynamic SPECT capabilities: Design, system validation and future potential. Eur J Nucl Med Mol Imaging 2010;37:1887‐902.CrossRefPubMedPubMedCentral Bocher M, Blevis IM, Tsukerman L, Shrem Y, Kovalski G, Volokh L. A fast cardiac gamma camera with dynamic SPECT capabilities: Design, system validation and future potential. Eur J Nucl Med Mol Imaging 2010;37:1887‐902.CrossRefPubMedPubMedCentral
9.
go back to reference Gabrani-Juma H, Clarkin OJ, Pourmoghaddas A, Driscoll B, Wells RG, deKemp RA, et al. Validation of a multimodality flow phantom and its application for assessment of dynamic SPECT and PET technologies. IEEE Trans Med Imaging 2017;36:132‐41.CrossRefPubMed Gabrani-Juma H, Clarkin OJ, Pourmoghaddas A, Driscoll B, Wells RG, deKemp RA, et al. Validation of a multimodality flow phantom and its application for assessment of dynamic SPECT and PET technologies. IEEE Trans Med Imaging 2017;36:132‐41.CrossRefPubMed
10.
go back to reference Krakovich A, Zaretsky U, Moalem I, Naimushin A, Rozen E, Scheinowitz M, et al. A new cardiac phantom for dynamic SPECT. J Nucl Cardiol 2021;28:2299‐309.CrossRefPubMed Krakovich A, Zaretsky U, Moalem I, Naimushin A, Rozen E, Scheinowitz M, et al. A new cardiac phantom for dynamic SPECT. J Nucl Cardiol 2021;28:2299‐309.CrossRefPubMed
12.
go back to reference Ben-Haim S, Murthy VL, Breault C, Allie R, Sitek A, Roth N, et al. Quantification of myocardial perfusion reserve using dynamic SPECT imaging in humans: A feasibility study. J Nucl Med 2013;54:873‐9.CrossRefPubMed Ben-Haim S, Murthy VL, Breault C, Allie R, Sitek A, Roth N, et al. Quantification of myocardial perfusion reserve using dynamic SPECT imaging in humans: A feasibility study. J Nucl Med 2013;54:873‐9.CrossRefPubMed
13.
go back to reference Wells RG, Timmins R, Klein R, Lockwood J, Marvin B, deKemp RA, et al. Dynamic SPECT measurement of absolute myocardial blood flow in a porcine model. J Nucl Med 2014;55:1685‐91.CrossRefPubMed Wells RG, Timmins R, Klein R, Lockwood J, Marvin B, deKemp RA, et al. Dynamic SPECT measurement of absolute myocardial blood flow in a porcine model. J Nucl Med 2014;55:1685‐91.CrossRefPubMed
14.
go back to reference Hsu B, Hu LH, Yang BH, Chen LC, Chen YK, Ting CH, et al. SPECT myocardial blood flow quantitation toward clinical use: A comparative study with 13N-ammonia PET myocardial blood flow quantitation. Eur J Nucl Med Mol Imaging 2017;44:117‐28.CrossRefPubMed Hsu B, Hu LH, Yang BH, Chen LC, Chen YK, Ting CH, et al. SPECT myocardial blood flow quantitation toward clinical use: A comparative study with 13N-ammonia PET myocardial blood flow quantitation. Eur J Nucl Med Mol Imaging 2017;44:117‐28.CrossRefPubMed
15.
go back to reference Nkoulou R, Fuchs TA, Pazhenkottil AP, Kuest SM, Ghadri JR, Stehli J, et al. Absolute myocardial blood flow and flow reserve assessed by gated SPECT with cadmium–zinc–telluride detectors using 99mTc-tetrofosmin: Head-to-head comparison with 13N-ammonia PET. J Nucl Med 2016;57:1887‐92.CrossRefPubMed Nkoulou R, Fuchs TA, Pazhenkottil AP, Kuest SM, Ghadri JR, Stehli J, et al. Absolute myocardial blood flow and flow reserve assessed by gated SPECT with cadmium–zinc–telluride detectors using 99mTc-tetrofosmin: Head-to-head comparison with 13N-ammonia PET. J Nucl Med 2016;57:1887‐92.CrossRefPubMed
16.
go back to reference Wells RG, Marvin B, Poirier M, Renaud J, deKemp RA, Ruddy TD. Optimization of SPECT measurement of myocardial blood flow with corrections for attenuation, motion, and blood binding compared with PET. J Nucl Med 2017;58:2013‐9.CrossRefPubMed Wells RG, Marvin B, Poirier M, Renaud J, deKemp RA, Ruddy TD. Optimization of SPECT measurement of myocardial blood flow with corrections for attenuation, motion, and blood binding compared with PET. J Nucl Med 2017;58:2013‐9.CrossRefPubMed
17.
go back to reference Agostini D, Roule V, Nganoa C, Roth N, Baavour R, Parienti JJ, et al. First validation of myocardial flow reserve assessed by dynamic 99m Tc-sestamibi CZT-SPECT camera: Head to head comparison with 15 O-water PET and fractional flow reserve in patients with suspected coronary artery disease. The WATERDAY study. Eur J Nucl Med Mol Imaging 2018;45:1079‐90.CrossRefPubMedPubMedCentral Agostini D, Roule V, Nganoa C, Roth N, Baavour R, Parienti JJ, et al. First validation of myocardial flow reserve assessed by dynamic 99m Tc-sestamibi CZT-SPECT camera: Head to head comparison with 15 O-water PET and fractional flow reserve in patients with suspected coronary artery disease. The WATERDAY study. Eur J Nucl Med Mol Imaging 2018;45:1079‐90.CrossRefPubMedPubMedCentral
18.
go back to reference Acampa W, Assante R, Mannarino T, Zampella E, D’Antonio A, Buongiorno P, et al. Low-dose dynamic myocardial perfusion imaging by CZT-SPECT in the identification of obstructive coronary artery disease. Eur J Nucl Med Mol Imaging 2020;47:1705‐12.CrossRefPubMed Acampa W, Assante R, Mannarino T, Zampella E, D’Antonio A, Buongiorno P, et al. Low-dose dynamic myocardial perfusion imaging by CZT-SPECT in the identification of obstructive coronary artery disease. Eur J Nucl Med Mol Imaging 2020;47:1705‐12.CrossRefPubMed
19.
go back to reference Wells RG, Marvin B. Measuring SPECT myocardial blood flow at the University of Ottawa Heart Institute. J Nucl Cardiol 2021;28:1298‐303.CrossRefPubMed Wells RG, Marvin B. Measuring SPECT myocardial blood flow at the University of Ottawa Heart Institute. J Nucl Cardiol 2021;28:1298‐303.CrossRefPubMed
20.
go back to reference Renaud JM, Poitrasson-Riviere A, Hagio T, Moody JB, Arida-Moody L, Ficaro EP. Myocardial flow reserve estimation with contemporary CZT-SPECT and 99mTc-tracers lacks precision for routine clinical application. J Nucl Cardiol 2021;29:2078‐89.CrossRefPubMed Renaud JM, Poitrasson-Riviere A, Hagio T, Moody JB, Arida-Moody L, Ficaro EP. Myocardial flow reserve estimation with contemporary CZT-SPECT and 99mTc-tracers lacks precision for routine clinical application. J Nucl Cardiol 2021;29:2078‐89.CrossRefPubMed
22.
go back to reference Valentinuzzi ME, Geddes LA, Baker LE. A simple mathematical derivation of the Stewart-Hamilton formula for the determination of cardiac output. Med Biol Eng 1969;7:277‐82.CrossRefPubMed Valentinuzzi ME, Geddes LA, Baker LE. A simple mathematical derivation of the Stewart-Hamilton formula for the determination of cardiac output. Med Biol Eng 1969;7:277‐82.CrossRefPubMed
23.
go back to reference Sugihara H, Yonekura Y, Miyazaki Y, Taniguchi Y. Estimation of cardiac output by first-pass transit of radiotracers. Ann Nucl Med 1999;13:299‐302.CrossRefPubMed Sugihara H, Yonekura Y, Miyazaki Y, Taniguchi Y. Estimation of cardiac output by first-pass transit of radiotracers. Ann Nucl Med 1999;13:299‐302.CrossRefPubMed
24.
go back to reference Brown TLY, Merrill J, Volokh L, Bengel FM. Determinants of the response of left ventricular ejection fraction to vasodilator stress in electrocardiographically gated 82rubidium myocardial perfusion PET. Eur J Nucl Med Mol Imaging 2008;35:336‐42.CrossRefPubMed Brown TLY, Merrill J, Volokh L, Bengel FM. Determinants of the response of left ventricular ejection fraction to vasodilator stress in electrocardiographically gated 82rubidium myocardial perfusion PET. Eur J Nucl Med Mol Imaging 2008;35:336‐42.CrossRefPubMed
26.
go back to reference Chander A, Brenner M, Lautamäki R, Voicu C, Merrill J, Bengel FM. Comparison of measures of left ventricular function from electrocardiographically gated 82Rb PET with contrast-enhanced CT ventriculography: A hybrid PET/CT analysis. J Nucl Med 2008;49:1643‐50.CrossRefPubMed Chander A, Brenner M, Lautamäki R, Voicu C, Merrill J, Bengel FM. Comparison of measures of left ventricular function from electrocardiographically gated 82Rb PET with contrast-enhanced CT ventriculography: A hybrid PET/CT analysis. J Nucl Med 2008;49:1643‐50.CrossRefPubMed
27.
go back to reference Barletta G, Bene MRD. Effects of dipyridamole on cardiac and systemic haemodynamics: Real-time three-dimensional stress echo beyond regional wall motion. J Cardiovasc Med 2011;12:455‐9.CrossRef Barletta G, Bene MRD. Effects of dipyridamole on cardiac and systemic haemodynamics: Real-time three-dimensional stress echo beyond regional wall motion. J Cardiovasc Med 2011;12:455‐9.CrossRef
29.
go back to reference Iaizzo PA, editor. Handbook of cardiac anatomy, physiology, and devices. Cham: Springer Science & Business Media; 2010. Iaizzo PA, editor. Handbook of cardiac anatomy, physiology, and devices. Cham: Springer Science & Business Media; 2010.
30.
go back to reference Lortie M, Beanlands RS, Yoshinaga K, Klein R, DaSilva JN, DeKemp RA. Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imaging 2007;34:1765‐74.CrossRefPubMed Lortie M, Beanlands RS, Yoshinaga K, Klein R, DaSilva JN, DeKemp RA. Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imaging 2007;34:1765‐74.CrossRefPubMed
31.
go back to reference Boarini DJ, et al. The effect of intravenous dipyridamole on the cerebral and systemic circulations of the dog. Stroke 1982;13:842‐7.CrossRefPubMed Boarini DJ, et al. The effect of intravenous dipyridamole on the cerebral and systemic circulations of the dog. Stroke 1982;13:842‐7.CrossRefPubMed
32.
go back to reference Melin JA, Becker LC. Quantitative relationship between global left ventricular thallium uptake and blood flow: Effects of propranolol, quabain, dipyridamole, and coronary artery occlusion. J Nucl Med 1986;27:641‐52.PubMed Melin JA, Becker LC. Quantitative relationship between global left ventricular thallium uptake and blood flow: Effects of propranolol, quabain, dipyridamole, and coronary artery occlusion. J Nucl Med 1986;27:641‐52.PubMed
33.
go back to reference Slomka PJ, Nishina H, Berman DS, Akincioglu C, Abidov A, Friedman JD, et al. Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005;12:66‐77.CrossRefPubMed Slomka PJ, Nishina H, Berman DS, Akincioglu C, Abidov A, Friedman JD, et al. Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005;12:66‐77.CrossRefPubMed
34.
go back to reference Sharir T, Slomka PJ, Hayes SW, DiCarli MF, Ziffer JA, Martin WH, et al. Multicenter trial of high-speed versus conventional single-photon emission computed tomography imaging: Quantitative results of myocardial perfusion and left ventricular function. J Am Coll Cardiol 2010;55:1965‐74.CrossRefPubMed Sharir T, Slomka PJ, Hayes SW, DiCarli MF, Ziffer JA, Martin WH, et al. Multicenter trial of high-speed versus conventional single-photon emission computed tomography imaging: Quantitative results of myocardial perfusion and left ventricular function. J Am Coll Cardiol 2010;55:1965‐74.CrossRefPubMed
35.
go back to reference Aarsvold JN, Galt JR, Nye JA, Grammens FD, Glass ZA. The quality field of view of a discovery 530c. In: 2012 IEEE nuclear science symposium and medical imaging conference record (NSS/MIC). IEEE; 2012. Aarsvold JN, Galt JR, Nye JA, Grammens FD, Glass ZA. The quality field of view of a discovery 530c. In: 2012 IEEE nuclear science symposium and medical imaging conference record (NSS/MIC). IEEE; 2012.
36.
go back to reference Savi A, Gerundini P, Zoli P, Maffioli L, Compierchio A, Colombo F, et al. Biodistribution of Tc-99m methoxy-isobutyl-isonitrile (MIBI) in humans. Eur J Nucl Med 1989;15:597‐600.CrossRefPubMed Savi A, Gerundini P, Zoli P, Maffioli L, Compierchio A, Colombo F, et al. Biodistribution of Tc-99m methoxy-isobutyl-isonitrile (MIBI) in humans. Eur J Nucl Med 1989;15:597‐600.CrossRefPubMed
Metadata
Title
Dose-consistent dynamic SPECT
Authors
A. Krakovich, MSc
E. Gelbart, MSc
I. Moalem
A. Naimushin, MD
E. Rozen, MD
M. Scheinowitz, PhD
R. Goldkorn, MD
Publication date
08-12-2022
Publisher
Springer International Publishing
Published in
Journal of Nuclear Cardiology / Issue 4/2023
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-022-03160-9

Other articles of this Issue 4/2023

Journal of Nuclear Cardiology 4/2023 Go to the issue