Open Access 04-07-2022 | Digital Volume Tomography | Original Article

Prerequisites for the clinical implementation of a markerless SGRT-only workflow for the treatment of breast cancer patients

Authors: Tim-Oliver Sauer, Oliver J. Ott, Godehard Lahmer, Rainer Fietkau, Christoph Bert

Published in: Strahlentherapie und Onkologie | Issue 1/2023

Login to get access

Abstract

Purpose

A markerless workflow for the treatment of breast cancer patients has been introduced and evaluated retrospectively. It includes surface-guided radiation therapy (SGRT)-only positioning for patients with small cone beam CT (CBCT) position corrections during the first five fractions. Prerequisites and the frequency of its clinical application were evaluated, as well as potential benefits in terms of treatment time and dose savings, the frequency of CBCT scans, and the accuracy of the positioning.

Methods

A group of 100 patients treated with the new workflow on two Versa HD linacs has been compared to a matched control group of patients treated with the former workflow, which included prepositioning with skin markings and lasers, SGRT and daily CBCT. The comparison was based on the evaluation of logfiles.

Results

Of the patients treated with the new workflow, 40% did not receive daily CBCT scans. This resulted in mean time savings of 97 s, 166 s and 239 s per fraction for the new workflow, for patients treated without daily CBCT and for SGRT-only fractions, respectively, when compared to the old workflow. Dose savings amounted to a weighted computed tomography dose index reduction of CTDIW = 2.56 cGy on average for normofractionated treatment and weekly CBCTs, while for patients not treated with daily CBCT, SGRT-based positioning accuracy was 5.2 mm for the mean translational magnitude, as evaluated by CBCT.

Conclusion

For 40% of the patients, after five fractions with small CBCT corrections, the workflow could be changed to SGRT-only positioning with weekly CBCT. This leads to imaging dose and time savings and thus also reduced intrafraction motion, potentially increased patient throughput and patient comfort, while assuring appropriate positioning accuracy.
Literature
2.
go back to reference Wang H, Xu Z, Grantham K, Zhou Y, Cui T, Zhang Y, Liu B, Wang X, Vergalasova I, Reyhan M, Weiner J, Danish SF, Yue N, Nie K (2021) Performance assessment of two motion management systems for frameless stereotactic radiosurgery. Strahlenther Onkol 197(2):150–157. https://doi.org/10.1007/s00066-020-01688-8CrossRef Wang H, Xu Z, Grantham K, Zhou Y, Cui T, Zhang Y, Liu B, Wang X, Vergalasova I, Reyhan M, Weiner J, Danish SF, Yue N, Nie K (2021) Performance assessment of two motion management systems for frameless stereotactic radiosurgery. Strahlenther Onkol 197(2):150–157. https://​doi.​org/​10.​1007/​s00066-020-01688-8CrossRef
3.
go back to reference Stanley DN, McConnell KA, Kirby N, Gutiérrez AN, Papanikolaou N, Rasmussen K (2017) Comparison of initial patient setup accuracy between surface imaging and three point localization: a retrospective analysis. J Appl Clin Med Phys 18(6):58–61. https://doi.org/10.1002/acm2.12183CrossRef Stanley DN, McConnell KA, Kirby N, Gutiérrez AN, Papanikolaou N, Rasmussen K (2017) Comparison of initial patient setup accuracy between surface imaging and three point localization: a retrospective analysis. J Appl Clin Med Phys 18(6):58–61. https://​doi.​org/​10.​1002/​acm2.​12183CrossRef
4.
go back to reference Kügele M, Mannerberg A, Norring Bekke S, Alkner S, Berg L, Mahmood F, Thornberg C, Edvardsson A, Back SAJ, Behrens CF, Ceberg S (2019) Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy. J Appl Clin Med Phys 20(9):61–68. https://doi.org/10.1002/acm2.12700CrossRef Kügele M, Mannerberg A, Norring Bekke S, Alkner S, Berg L, Mahmood F, Thornberg C, Edvardsson A, Back SAJ, Behrens CF, Ceberg S (2019) Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy. J Appl Clin Med Phys 20(9):61–68. https://​doi.​org/​10.​1002/​acm2.​12700CrossRef
5.
go back to reference Jimenez RB, Batin E, Giantsoudi D, Hazeltine W, Bertolino K, Ho AY, MacDonald SM, Taghian AG, Gierga DP (2019) Tattoo free setup for partial breast irradiation: a feasibility study. J Appl Clin Med Phys 20(4):45–50. https://doi.org/10.1002/acm2.12557CrossRef Jimenez RB, Batin E, Giantsoudi D, Hazeltine W, Bertolino K, Ho AY, MacDonald SM, Taghian AG, Gierga DP (2019) Tattoo free setup for partial breast irradiation: a feasibility study. J Appl Clin Med Phys 20(4):45–50. https://​doi.​org/​10.​1002/​acm2.​12557CrossRef
7.
go back to reference Laaksomaa M, Sarudis S, Rossi M, Lehtonen T, Pehkonen J, Remes J, Luukkanen H, Skyttä T, Kapanen M (2019) AlignRT((R)) and Catalyst in whole-breast radiotherapy with DIBH: Is IGRT still needed? J Appl Clin Med Phys 20(3):97–104. https://doi.org/10.1002/acm2.12553CrossRef Laaksomaa M, Sarudis S, Rossi M, Lehtonen T, Pehkonen J, Remes J, Luukkanen H, Skyttä T, Kapanen M (2019) AlignRT((R)) and Catalyst in whole-breast radiotherapy with DIBH: Is IGRT still needed? J Appl Clin Med Phys 20(3):97–104. https://​doi.​org/​10.​1002/​acm2.​12553CrossRef
8.
go back to reference Crop F, Pasquier D, Baczkiewic A, Doré J, Bequet L, Steux E, Gadroy A, Bouillon J, Florence C, Muszynski L, Lacour M, Lartigau E (2016) Surface imaging, laser positioning or volumetric imaging for breast cancer with nodal involvement treated by helical TomoTherapy. J Appl Clin Med Phys 17(5):200–211. https://doi.org/10.1120/jacmp.v17i5.6041CrossRef Crop F, Pasquier D, Baczkiewic A, Doré J, Bequet L, Steux E, Gadroy A, Bouillon J, Florence C, Muszynski L, Lacour M, Lartigau E (2016) Surface imaging, laser positioning or volumetric imaging for breast cancer with nodal involvement treated by helical TomoTherapy. J Appl Clin Med Phys 17(5):200–211. https://​doi.​org/​10.​1120/​jacmp.​v17i5.​6041CrossRef
9.
go back to reference Borm KJ, Kessel K, Devecka M, Muench S, Straube C, Schiller K, Schüttrumpf L, Dapper H, Wöller B, Pigorsch S, Combs SE (2020) Variability in lymph node irradiation in patients with breast cancer—results from a multi-center survey in German-speaking countries. Strahlenther Onkol 196(1):15–22. https://doi.org/10.1007/s00066-019-01537-3CrossRef Borm KJ, Kessel K, Devecka M, Muench S, Straube C, Schiller K, Schüttrumpf L, Dapper H, Wöller B, Pigorsch S, Combs SE (2020) Variability in lymph node irradiation in patients with breast cancer—results from a multi-center survey in German-speaking countries. Strahlenther Onkol 196(1):15–22. https://​doi.​org/​10.​1007/​s00066-019-01537-3CrossRef
11.
go back to reference Pazos M, Walter F, Reitz D, Schönecker S, Konnerth D, Schäfer A, Rottler M, Alongi F, Freislederer P, Niyazi M, Belka C, Corradini S (2019) Impact of surface-guided positioning on the use of portal imaging and initial set-up duration in breast cancer patients. Strahlenther Onkol 195(11):964–971. https://doi.org/10.1007/s00066-019-01494-xCrossRef Pazos M, Walter F, Reitz D, Schönecker S, Konnerth D, Schäfer A, Rottler M, Alongi F, Freislederer P, Niyazi M, Belka C, Corradini S (2019) Impact of surface-guided positioning on the use of portal imaging and initial set-up duration in breast cancer patients. Strahlenther Onkol 195(11):964–971. https://​doi.​org/​10.​1007/​s00066-019-01494-xCrossRef
14.
go back to reference Loutfi-Krauss B, Köhn J, Blümer N, Freundl K, Koch T, Kara E, Scherf C, Rödel C, Ramm U, Licher J (2015) Effect of dose reduction on image registration and image quality for cone-beam CT in radiotherapy. Strahlenther Onkol 191(2):192–200. https://doi.org/10.1007/s00066-014-0750-xCrossRef Loutfi-Krauss B, Köhn J, Blümer N, Freundl K, Koch T, Kara E, Scherf C, Rödel C, Ramm U, Licher J (2015) Effect of dose reduction on image registration and image quality for cone-beam CT in radiotherapy. Strahlenther Onkol 191(2):192–200. https://​doi.​org/​10.​1007/​s00066-014-0750-xCrossRef
16.
go back to reference Reitz D, Carl G, Schönecker S, Pazos M, Freislederer P, Niyazi M, Ganswindt U, Alongi F, Reiner M, Belka C, Corradini S (2018) Real-time intra-fraction motion management in breast cancer radiotherapy: analysis of 2028 treatment sessions. Radiat Oncol 13(1):1–9. https://doi.org/10.1186/s13014-018-1072-4CrossRef Reitz D, Carl G, Schönecker S, Pazos M, Freislederer P, Niyazi M, Ganswindt U, Alongi F, Reiner M, Belka C, Corradini S (2018) Real-time intra-fraction motion management in breast cancer radiotherapy: analysis of 2028 treatment sessions. Radiat Oncol 13(1):1–9. https://​doi.​org/​10.​1186/​s13014-018-1072-4CrossRef
Metadata
Title
Prerequisites for the clinical implementation of a markerless SGRT-only workflow for the treatment of breast cancer patients
Authors
Tim-Oliver Sauer
Oliver J. Ott
Godehard Lahmer
Rainer Fietkau
Christoph Bert
Publication date
04-07-2022
Publisher
Springer Berlin Heidelberg
Published in
Strahlentherapie und Onkologie / Issue 1/2023
Print ISSN: 0179-7158
Electronic ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-022-01966-7

Other articles of this Issue 1/2023

Strahlentherapie und Onkologie 1/2023 Go to the issue