Published in: BMC Medical Imaging 1/2020

Open Access 01-12-2020 | Research article

Correlation between spatial resolution and ball distortion rate of panoramic radiography

Authors: Han-Gyeol Yeom, Jo-Eun Kim, Kyung-Hoe Huh, Won-Jin Yi, Min-Suk Heo, Sam-Sun Lee, Soon-Chul Choi

Published in: BMC Medical Imaging | Issue 1/2020

Login to get access

Abstract

Background

The purpose of this study was to analyze the correlation between spatial resolution and ball distortion rate of panoramic radiography and to elucidate the minimum criterion for ball distortion rate, which is very relevant to clinical readability.

Methods

Horizontal and vertical spatial resolution and ball distortion rates were calculated in the same position, such as the incisor, premolar, molar, and temporomandibular joint area with various object depths corresponding to 48 mm. Three devices were evaluated. A region showing spatial resolution above the reference standard was selected, and the ball distortion rate corresponding to the same region was divided into horizontal and vertical phantom groups.
The mean and standard deviation of the obtained ball distortion rates were calculated. Student’s t-test was used to statistically analyze the mean difference in ball distortion rates between vertical and horizontal phantom groups.

Results

In all devices, the horizontal line pair phantom, but not the vertical line pair phantom, was readable in all areas measured at the line pair value of at least 1.88 lp/mm. The line pair value tended to be higher toward the center and lower toward the outside. The ball distortion rate tended to decrease closer to the center and increased further away. In addition, ball distortion rates could not be measured at some areas as they were not recognized as balls due to the high degree of distortion at the outermost and innermost sides. The number of balls satisfying the reference value using the horizontal line pair phantom was 102 (mean of ball distortion rates, 20.98; standard deviation, 15.25). The number of balls satisfying the reference value using the vertical line pair phantom was 49 (mean of ball distortion rates, 16.33; standard deviation, 14.25). However, mean ball distortion rate was not significantly different between the two groups.

Conclusions

Image layer of panoramic radiography could be evaluated by the spatial resolution using horizontal and vertical line pair phantoms and by assessing ball distortion rates through a ball-type panorama phantom. A ball distortion rate of 20% could be used as a threshold to evaluate the image layer of panoramic radiography.
Literature
1.
go back to reference Moll MA, Seuthe M, von See C, Zapf A, Hornecker E, Mausberg RF, et al. Comparison of clinical and dental panoramic findings: a practice-based crossover study. BMC Oral Health. 2013;13:48.CrossRef Moll MA, Seuthe M, von See C, Zapf A, Hornecker E, Mausberg RF, et al. Comparison of clinical and dental panoramic findings: a practice-based crossover study. BMC Oral Health. 2013;13:48.CrossRef
2.
go back to reference Horner K. Review article: radiation protection in dental radiology. Br J Radiol. 1994;67:1041–9.CrossRef Horner K. Review article: radiation protection in dental radiology. Br J Radiol. 1994;67:1041–9.CrossRef
3.
go back to reference Welander U, McDavid WD, Tronje G, Morris CR. An analysis of different planes within the image layer in rotational panoramic radiography. Dentomaxillofac Radiol. 1987;16:79–84.CrossRef Welander U, McDavid WD, Tronje G, Morris CR. An analysis of different planes within the image layer in rotational panoramic radiography. Dentomaxillofac Radiol. 1987;16:79–84.CrossRef
4.
go back to reference Patel JR. The focal trough of the autopan panoramic dental x-ray machine. Oral Surg Oral Med Oral Pathol. 1989;68:115–8.CrossRef Patel JR. The focal trough of the autopan panoramic dental x-ray machine. Oral Surg Oral Med Oral Pathol. 1989;68:115–8.CrossRef
5.
go back to reference McDavid WD, Morris CR, Tronje G, Welander U. Resolution of several screen-film combinations in rotational panoramic radiography. Oral Surg Oral Med Oral Pathol. 1986;61:629–34.CrossRef McDavid WD, Morris CR, Tronje G, Welander U. Resolution of several screen-film combinations in rotational panoramic radiography. Oral Surg Oral Med Oral Pathol. 1986;61:629–34.CrossRef
6.
go back to reference Hassen SM, Manson-Hing LR. A study of the zone of sharpness of three panoramic x-ray machines and the effect of screen speed on the sharpness zone. Oral Surg Oral Med Oral Pathol. 1982;54:242–9.CrossRef Hassen SM, Manson-Hing LR. A study of the zone of sharpness of three panoramic x-ray machines and the effect of screen speed on the sharpness zone. Oral Surg Oral Med Oral Pathol. 1982;54:242–9.CrossRef
7.
go back to reference Lund TM, Manson-Hing LR. A study of the focal troughs of three panoramic dental x-ray machines: part I. The area of sharpness. Oral Surg Oral Med Oral Pathol. 1975;39:318–28.CrossRef Lund TM, Manson-Hing LR. A study of the focal troughs of three panoramic dental x-ray machines: part I. The area of sharpness. Oral Surg Oral Med Oral Pathol. 1975;39:318–28.CrossRef
8.
go back to reference Manson-Hing LR, Lund TM. A study of the focal troughs of three panoramic dental x-ray machines: part II. Image dimensions. Oral Surg Oral Med Oral Pathol. 1975;39:647–53.CrossRef Manson-Hing LR, Lund TM. A study of the focal troughs of three panoramic dental x-ray machines: part II. Image dimensions. Oral Surg Oral Med Oral Pathol. 1975;39:647–53.CrossRef
9.
go back to reference Yeo DKL, Freer TJ, Brockhurst PJ. Distortions in panoramic radiographs. Aust Orthod J. 2002;18:92.PubMed Yeo DKL, Freer TJ, Brockhurst PJ. Distortions in panoramic radiographs. Aust Orthod J. 2002;18:92.PubMed
10.
go back to reference Devlin H, Yuan J. Object position and image magnification in dental panoramic radiography: a theoretical analysis. Dentomaxillofac Radiol. 2013;42:29951683.CrossRef Devlin H, Yuan J. Object position and image magnification in dental panoramic radiography: a theoretical analysis. Dentomaxillofac Radiol. 2013;42:29951683.CrossRef
11.
go back to reference Shiojima M, Bäckström A, Welander U, McDavid WD, Tronje G, Naitoh M. Layer thickness in panoramic radiography as defined by different noise-equivalent passbands. Oral Surg Oral Med Oral Pathol. 1993;76:244–50.CrossRef Shiojima M, Bäckström A, Welander U, McDavid WD, Tronje G, Naitoh M. Layer thickness in panoramic radiography as defined by different noise-equivalent passbands. Oral Surg Oral Med Oral Pathol. 1993;76:244–50.CrossRef
12.
go back to reference Tronje G, Eliasson S, Julin P, Welander U. Image distortion in rotational panoramic radiography: II. Vertical distances. Acta Radiol Diagn (Stockh). 1981;22:449–55.CrossRef Tronje G, Eliasson S, Julin P, Welander U. Image distortion in rotational panoramic radiography: II. Vertical distances. Acta Radiol Diagn (Stockh). 1981;22:449–55.CrossRef
13.
go back to reference Glass BJ, McDavid WD, Welander U, Morris CR. The central plane of the image layer determined experimentally in various rotational panoramic x-ray machines. Oral Surg Oral Med Oral Pathol. 1985;60:104–12.CrossRef Glass BJ, McDavid WD, Welander U, Morris CR. The central plane of the image layer determined experimentally in various rotational panoramic x-ray machines. Oral Surg Oral Med Oral Pathol. 1985;60:104–12.CrossRef
14.
go back to reference Yeom HG, Kim JE, Huh KH, Yi WJ, Heo MS, Lee SS, et al. Development of a new ball-type phantom for evaluation of the image layer of panoramic radiography. Imaging Sci Dent. 2018;48:255–9.CrossRef Yeom HG, Kim JE, Huh KH, Yi WJ, Heo MS, Lee SS, et al. Development of a new ball-type phantom for evaluation of the image layer of panoramic radiography. Imaging Sci Dent. 2018;48:255–9.CrossRef
15.
go back to reference Paiboon C, Manson-Hing LR. Effect of border sharpness on the size and position of the focal trough of panoramic x-ray machines. Oral Surg Oral Med Oral Pathol. 1985;60:670–6.CrossRef Paiboon C, Manson-Hing LR. Effect of border sharpness on the size and position of the focal trough of panoramic x-ray machines. Oral Surg Oral Med Oral Pathol. 1985;60:670–6.CrossRef
16.
go back to reference Martinez-Cruz S, Manson-Hing LR. Comparison of focal trough dimensions and form by resolution measurements in panoramic radiography. J Am Dent Assoc. 1987;114:639–42.CrossRef Martinez-Cruz S, Manson-Hing LR. Comparison of focal trough dimensions and form by resolution measurements in panoramic radiography. J Am Dent Assoc. 1987;114:639–42.CrossRef
17.
go back to reference Choi DH, Choi BR, Choi JW, Huh KH, Yi WJ, Heo MS, et al. Reference line-pair values of panoramic radiographs using an arch-form phantom stand to assess clinical image quality. Imaging Sci Dent. 2013;43:7–15.CrossRef Choi DH, Choi BR, Choi JW, Huh KH, Yi WJ, Heo MS, et al. Reference line-pair values of panoramic radiographs using an arch-form phantom stand to assess clinical image quality. Imaging Sci Dent. 2013;43:7–15.CrossRef
18.
go back to reference International Electrotechnical Commission. Evaluation and routine testing in medical imaging departments-part 3–4: acceptance tests-imaging performance of dental X-ray equipment. Geneva: IEC; 2000. International Electrotechnical Commission. Evaluation and routine testing in medical imaging departments-part 3–4: acceptance tests-imaging performance of dental X-ray equipment. Geneva: IEC; 2000.
19.
go back to reference International Electrotechnical Commission. Evaluation and routine testing in medical imaging departments part 2–6: Constancy tests imaging performance of computed tomography X-ray equipment. Geneva: IEC; 2006. International Electrotechnical Commission. Evaluation and routine testing in medical imaging departments part 2–6: Constancy tests imaging performance of computed tomography X-ray equipment. Geneva: IEC; 2006.
20.
go back to reference Kaeppler G, Axmann-Krcmar D, Reuter I, Meyle J, Gómez-Román G. A clinical evaluation of some factors affecting image quality in panoramic radiography. Dentomaxillofac Radiol. 2000;29:81–4.CrossRef Kaeppler G, Axmann-Krcmar D, Reuter I, Meyle J, Gómez-Román G. A clinical evaluation of some factors affecting image quality in panoramic radiography. Dentomaxillofac Radiol. 2000;29:81–4.CrossRef
21.
go back to reference Brown CE Jr, Christen AC, Jerman AC. Dimensions of the focal trough in panoramic radiography. J Am Dent Assoc. 1972;84:843–7.CrossRef Brown CE Jr, Christen AC, Jerman AC. Dimensions of the focal trough in panoramic radiography. J Am Dent Assoc. 1972;84:843–7.CrossRef
Metadata
Title
Correlation between spatial resolution and ball distortion rate of panoramic radiography
Authors
Han-Gyeol Yeom
Jo-Eun Kim
Kyung-Hoe Huh
Won-Jin Yi
Min-Suk Heo
Sam-Sun Lee
Soon-Chul Choi
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2020
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-020-00472-5

Other articles of this Issue 1/2020

BMC Medical Imaging 1/2020 Go to the issue