Abstract

Objectives

To update the national diagnostic reference levels (DRLs) for adult CT in Switzerland using dose management software and to compare them to the previous Swiss DRLs from 2010.

Methods

CT dose data from 14 radiological institutes with a total of 50 CT scanners were collected with locally installed dose management software between 2014 and 2017. Data were assigned to 15 defined master protocols. Data cleaning steps were developed and adjusted individually for each participating institute and protocol. The DRLs for each master protocol were calculated as the 75th percentile of the distribution of the median volume computed tomography dose index (CTDIvol) and dose-length product (DLP) values per CT scanner.

Results

In total, 220,269 CT exams were available after data cleaning. Updated DRLs showed a clear trend towards lower doses compared with previous DRLs. The average relative change in the DRLs for CTDIvol was − 30% (0 to − 47%) and − 22% for DLP (+ 20 to − 40%). The largest relative decrease in the DRL for DLP was observed for the cervical spine protocol (− 40%), the two chest protocols (chest, − 37%; and exclusion of pulmonary embolism, − 33%), and the two neck protocols (neck, − 32%; and carotid angiography, − 28%). The DRLs for other protocols, for example the head and the abdomen-pelvis protocol, showed smaller relative changes (− 11% and − 17%).

Conclusions

The updated national DRLs are substantially lower than the previous values from 2010, demonstrating technological progress and the efforts of the radiological community to lower CT radiation exposure.

Key Points

• Dose management software allows the establishment of DRLs based on big data.
• Updated Swiss DRLs for adult CT are substantially lower compared with those from 2010.
• Swiss DRLs are low compared with other national DRLs.
Appendix
Available only for authorised users
Literature
1.
go back to reference Aroua A, Burnand B, Decka I, Vader JP, Valley JF (2002) Nation-wide survey on radiation doses in diagnostic and interventional radiology in Switzerland in 1998. Health Phys 83:46–55CrossRef Aroua A, Burnand B, Decka I, Vader JP, Valley JF (2002) Nation-wide survey on radiation doses in diagnostic and interventional radiology in Switzerland in 1998. Health Phys 83:46–55CrossRef
2.
go back to reference Aroua A, Trueb P, Vader JP, Valley JF, Verdun FR (2007) Exposure of the Swiss population by radiodiagnostics: 2003 review. Health Phys 92:442–448CrossRef Aroua A, Trueb P, Vader JP, Valley JF, Verdun FR (2007) Exposure of the Swiss population by radiodiagnostics: 2003 review. Health Phys 92:442–448CrossRef
3.
go back to reference Samara ET, Aroua A, Bochud FO et al (2012) Exposure of the Swiss population by medical X-rays: 2008 review. Health Phys 102:263–270CrossRef Samara ET, Aroua A, Bochud FO et al (2012) Exposure of the Swiss population by medical X-rays: 2008 review. Health Phys 102:263–270CrossRef
4.
go back to reference Le Coultre R, Bize J, Champendal M et al (2016) Exposure of the Swiss population by radiodiagnostics: 2013 review. Radiat Prot Dosimetry 169:221–224CrossRef Le Coultre R, Bize J, Champendal M et al (2016) Exposure of the Swiss population by radiodiagnostics: 2013 review. Radiat Prot Dosimetry 169:221–224CrossRef
6.
go back to reference Vañó E, Miller DL, Martin CJ et al (2017) ICRP Publication 135: Diagnostic reference levels in medical imaging. Ann ICRP 46(1):1–144 Vañó E, Miller DL, Martin CJ et al (2017) ICRP Publication 135: Diagnostic reference levels in medical imaging. Ann ICRP 46(1):1–144
7.
go back to reference (1996) Radiological protection and safety in medicine. A report of the International Commission on Radiological Protection. Ann ICRP 26(2):1–31 (1996) Radiological protection and safety in medicine. A report of the International Commission on Radiological Protection. Ann ICRP 26(2):1–31
8.
go back to reference Järvinen H, Vassileva J, Samei E, Wallace A, Vano E, Rehani M (2017) Patient dose monitoring and the use of diagnostic reference levels for the optimization of protection in medical imaging: current status and challenges worldwide. J Med Imaging (Bellingham) 4:031214 Järvinen H, Vassileva J, Samei E, Wallace A, Vano E, Rehani M (2017) Patient dose monitoring and the use of diagnostic reference levels for the optimization of protection in medical imaging: current status and challenges worldwide. J Med Imaging (Bellingham) 4:031214
12.
go back to reference Treier R, Aroua A, Verdun FR, Samara E, Stuessi A, Trueb PR (2010) Patient doses in CT examinations in Switzerland: implementation of national diagnostic reference levels. Radiat Prot Dosimetry 142:244–254CrossRef Treier R, Aroua A, Verdun FR, Samara E, Stuessi A, Trueb PR (2010) Patient doses in CT examinations in Switzerland: implementation of national diagnostic reference levels. Radiat Prot Dosimetry 142:244–254CrossRef
13.
go back to reference Heilmaier C, Treier R, Merkle EM, Alkadhi H, Weishaupt D, Schindera S (2018) National survey on dose data analysis in computed tomography. Eur Radiol 28:5044–5050CrossRef Heilmaier C, Treier R, Merkle EM, Alkadhi H, Weishaupt D, Schindera S (2018) National survey on dose data analysis in computed tomography. Eur Radiol 28:5044–5050CrossRef
16.
go back to reference Roch P, Célier D, Dessaud C, Etard C (2018) Using diagnostic reference levels to evaluate the improvement of patient dose optimisation and the influence of recent technologies in radiography and computed tomography. Eur J Radiol 98:68–74CrossRef Roch P, Célier D, Dessaud C, Etard C (2018) Using diagnostic reference levels to evaluate the improvement of patient dose optimisation and the influence of recent technologies in radiography and computed tomography. Eur J Radiol 98:68–74CrossRef
17.
go back to reference Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL (2017) U.S. Diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology 284:120–133CrossRef Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL (2017) U.S. Diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology 284:120–133CrossRef
19.
go back to reference Pansini M, Morsbach F, Schubert T et al (2016) Feasibility of dose optimization in a second-generation dual-source CT scanner for a manufacturer-recommended urolithiasis protocol for imaging renal stones. AJR Am J Roentgenol 206:348–354CrossRef Pansini M, Morsbach F, Schubert T et al (2016) Feasibility of dose optimization in a second-generation dual-source CT scanner for a manufacturer-recommended urolithiasis protocol for imaging renal stones. AJR Am J Roentgenol 206:348–354CrossRef
20.
go back to reference Tack D, Widelec J, De Maertelaer V, Bailly JM, Delcour C, Gevenois PA (2003) Comparison between low-dose and standard-dose multidetector CT in patients with suspected chronic sinusitis. AJR Am J Roentgenol 181:939–944CrossRef Tack D, Widelec J, De Maertelaer V, Bailly JM, Delcour C, Gevenois PA (2003) Comparison between low-dose and standard-dose multidetector CT in patients with suspected chronic sinusitis. AJR Am J Roentgenol 181:939–944CrossRef
21.
24.
go back to reference National Radiological Protection Board (1999) Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures. Documents of the NRPB 10(1) National Radiological Protection Board (1999) Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures. Documents of the NRPB 10(1)
25.
go back to reference National Council on Radiation Protection and Measurements (2012) Reference levels and achievable doses in medical and dental imaging: recommendations for the United States. NCRP Report No 172 National Council on Radiation Protection and Measurements (2012) Reference levels and achievable doses in medical and dental imaging: recommendations for the United States. NCRP Report No 172
28.
go back to reference Smith-Bindman R, Wang Y, Chu P et al (2019) International variation in radiation dose for computed tomography examinations: prospective cohort study. BMJ 364:k4931CrossRef Smith-Bindman R, Wang Y, Chu P et al (2019) International variation in radiation dose for computed tomography examinations: prospective cohort study. BMJ 364:k4931CrossRef
30.
go back to reference Vinci L, Krieger JP, Braun J et al (2019) Clustering of sociodemographic and lifestyle factors among adults with excess weight in a multilingual country. Nutrition 62:177–185CrossRef Vinci L, Krieger JP, Braun J et al (2019) Clustering of sociodemographic and lifestyle factors among adults with excess weight in a multilingual country. Nutrition 62:177–185CrossRef
32.
go back to reference Boos J, Lanzman RS, Heusch P et al (2016) Does body mass index outperform body weight as a surrogate parameter in the calculation of size-specific dose estimates in adult body CT? Br J Radiol 89:20150734CrossRef Boos J, Lanzman RS, Heusch P et al (2016) Does body mass index outperform body weight as a surrogate parameter in the calculation of size-specific dose estimates in adult body CT? Br J Radiol 89:20150734CrossRef
Metadata
Title
Update of national diagnostic reference levels for adult CT in Switzerland and assessment of radiation dose reduction since 2010
Authors
Christoph Aberle
Nick Ryckx
Reto Treier
Sebastian Schindera
Publication date
01-03-2020
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 3/2020
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06485-1

Other articles of this Issue 3/2020

European Radiology 3/2020 Go to the issue