Initial exploration of coronary stent image subtraction using dual-layer spectral CT

Authors: Le Qin, ShengJia Gu, ChiHua Chen, Huan Zhang, ZhenBin Zhu, XingBiao Chen, Qun Han, FuHua Yan, WenJie Yang

Published in: European Radiology | Issue 8/2019

Login to get access

Abstract

Objectives

This study aimed to investigate the feasibility of coronary stent image subtraction using spectral tools derived from dual-layer spectral computed tomography (CT).

Methods

Forty-three patients (65 stents) who underwent coronary CT angiography using dual-layer spectral CT were included. Conventional, 50-keV (kilo electron-volt), 100-keV, and virtual non-contrast (VNC) images were reconstructed from the same cardiac phase. Stents were subtracted on VNC images from conventional (convsub), 100-keV (100-keVsub), and 50-keV (50-keVsub) images. The in-stent lumen diameters were measured on subtraction, conventional, and 100-keV images. Subjective evaluation of reader confidence and subtractive quality was evaluated. Friedman tests were performed to compare in-stent lumen diameters and subjective evaluation among different images. Correlation between stent diameter and subjective evaluation was expressed as Spearman’s rank correlation coefficient (rs). The diagnostic accuracy was assessed according to invasive coronary angiography (ICA) performed in 11 patients (20 stents).

Results

In-stent lumen diameters were significantly larger on subtraction images than those on conventional and 100-keV images (p < 0.05). Higher reader confidence was found on 100-keV, convsub, 100-keVsub, and 50-keVsub images compared with conventional images (p < 0.05). Subtractive quality of 100-keVsub images was better than that of convsub images (p = 0.037). A moderate-to-strong correlation between stent diameter and subjective evaluation was found (rs = 0.527~0.790, p < 0.05). Higher specificity, positive predictive value, and negative predictive value of subtraction images were shown by ICA results.

Conclusions

Subtraction images derived from dual-layer spectral CT enhanced in-stent lumen visibility and could potentially improve diagnostic performance for evaluating coronary stents.

Key Points

• Dual-layer spectral CT enabled good subtractive quality of coronary stents without misregistration artifacts.
• Subtraction images could improve in-stent lumen visibility.
• Reader confidence and diagnostic performance were enhanced with subtraction images.
Literature
1.
go back to reference André F, Korosoglou G, Hosch W et al (2013) Performance of dual source versus 256-slice multi-slice CT in the evaluation of 16 coronary artery stents. Eur J Radiol 82:601–607CrossRefPubMed André F, Korosoglou G, Hosch W et al (2013) Performance of dual source versus 256-slice multi-slice CT in the evaluation of 16 coronary artery stents. Eur J Radiol 82:601–607CrossRefPubMed
2.
go back to reference Maintz D, Burg MC, Seifarth H et al (2008) Update on multidetector coronary CT angiography of coronary stents: in vitro evaluation of 29 different stent types with dual-source CT. Eur Radiol 19:42–49CrossRefPubMed Maintz D, Burg MC, Seifarth H et al (2008) Update on multidetector coronary CT angiography of coronary stents: in vitro evaluation of 29 different stent types with dual-source CT. Eur Radiol 19:42–49CrossRefPubMed
3.
go back to reference Ebersberger U, Tricarico F, Schoepf UJ et al (2013) CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction. Eur Radiol 23:125–132CrossRefPubMed Ebersberger U, Tricarico F, Schoepf UJ et al (2013) CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction. Eur Radiol 23:125–132CrossRefPubMed
4.
go back to reference Stehli J, Fuchs TA, Singer A et al (2015) First experience with single-source, dual-energy CCTA for monochromatic stent imaging. Eur Heart J Cardiovasc Imaging 16:507–512CrossRefPubMed Stehli J, Fuchs TA, Singer A et al (2015) First experience with single-source, dual-energy CCTA for monochromatic stent imaging. Eur Heart J Cardiovasc Imaging 16:507–512CrossRefPubMed
5.
go back to reference Alani A, Nakanishi R, Budoff MJ (2014) Recent improvement in coronary computed tomography angiography diagnostic accuracy. Clin Cardiol 37:428–433CrossRefPubMed Alani A, Nakanishi R, Budoff MJ (2014) Recent improvement in coronary computed tomography angiography diagnostic accuracy. Clin Cardiol 37:428–433CrossRefPubMed
6.
go back to reference Kalisz K, Halliburton S, Abbara S et al (2017) Update on cardiovascular applications of multienergy CT. Radiographics 37:1955–1974CrossRef Kalisz K, Halliburton S, Abbara S et al (2017) Update on cardiovascular applications of multienergy CT. Radiographics 37:1955–1974CrossRef
7.
go back to reference Rajiah P, Rong R, Martinez-Rios C, Rassouli N, Landeras L (2017) Benefit and clinical significance of retrospectively obtained spectral data with a novel detector-based spectral computed tomography - initial experiences and results. Clin Imaging 49:65–72CrossRefPubMed Rajiah P, Rong R, Martinez-Rios C, Rassouli N, Landeras L (2017) Benefit and clinical significance of retrospectively obtained spectral data with a novel detector-based spectral computed tomography - initial experiences and results. Clin Imaging 49:65–72CrossRefPubMed
8.
go back to reference Fuchs A, Kühl JT, Chen MY et al (2015) Feasibility of coronary calcium and stent image subtraction using 320-detector row CT angiography. J Cardiovasc Comput Tomogr 9:393–398CrossRefPubMedPubMedCentral Fuchs A, Kühl JT, Chen MY et al (2015) Feasibility of coronary calcium and stent image subtraction using 320-detector row CT angiography. J Cardiovasc Comput Tomogr 9:393–398CrossRefPubMedPubMedCentral
9.
go back to reference Yamaguchi T, Ichikawa K, Takahashi D, Sugaya T, Furuya J, Igarashi K (2017) A new contrast enhancement protocol for subtraction coronary computed tomography requiring a short breath-holding time. Acad Radiol 24:38–44 Yamaguchi T, Ichikawa K, Takahashi D, Sugaya T, Furuya J, Igarashi K (2017) A new contrast enhancement protocol for subtraction coronary computed tomography requiring a short breath-holding time. Acad Radiol 24:38–44
10.
go back to reference Ananthakrishnan L, Rajiah P, Ahn R et al (2017) Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT. Abdom Radiol (NY) 42:702–709CrossRef Ananthakrishnan L, Rajiah P, Ahn R et al (2017) Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT. Abdom Radiol (NY) 42:702–709CrossRef
11.
go back to reference Hickethier T, Baeßler B, Kroeger JR et al (2017) Monoenergetic reconstructions for imaging of coronary artery stents using spectral detector CT: in-vitro experience and comparison to conventional images. J Cardiovasc Comput Tomogr 11:33–39CrossRefPubMed Hickethier T, Baeßler B, Kroeger JR et al (2017) Monoenergetic reconstructions for imaging of coronary artery stents using spectral detector CT: in-vitro experience and comparison to conventional images. J Cardiovasc Comput Tomogr 11:33–39CrossRefPubMed
12.
go back to reference American Association of Physicists in Medicine (2008) The measurement, reporting, and management of radiation dose in CT: report of AAPM task group 23 of the diagnostic imaging council CT committee. Technical report, College Park, MD: AAPM, 2008: AAPM report 96 American Association of Physicists in Medicine (2008) The measurement, reporting, and management of radiation dose in CT: report of AAPM task group 23 of the diagnostic imaging council CT committee. Technical report, College Park, MD: AAPM, 2008: AAPM report 96
13.
go back to reference Mangold S, Cannaó PM, Schoepf UJ et al (2016) Impact of an advanced image-based monoenergetic reconstruction algorithm on coronary stent visualization using third generation dual-source dual-energy CT: a phantom study. Eur Radiol 26:1871–1878CrossRefPubMed Mangold S, Cannaó PM, Schoepf UJ et al (2016) Impact of an advanced image-based monoenergetic reconstruction algorithm on coronary stent visualization using third generation dual-source dual-energy CT: a phantom study. Eur Radiol 26:1871–1878CrossRefPubMed
14.
go back to reference Ehn S, Sellerer T, Muenzel D et al (2018) Assessment of quantification accuracy and image quality of a full-body dual-layer spectral CT system. J Appl Clin Med Phys 19:204–217CrossRefPubMed Ehn S, Sellerer T, Muenzel D et al (2018) Assessment of quantification accuracy and image quality of a full-body dual-layer spectral CT system. J Appl Clin Med Phys 19:204–217CrossRefPubMed
15.
go back to reference de Vet HC, Terwee CB, Knol DL, Bouter LM (2006) When to use agreement versus reliability measures. J Clin Epidemiol 59:1033–1039CrossRef de Vet HC, Terwee CB, Knol DL, Bouter LM (2006) When to use agreement versus reliability measures. J Clin Epidemiol 59:1033–1039CrossRef
16.
go back to reference Kottner J, Audigé L, Brorson S et al (2011) Guidelines for reporting reliability and agreement studies (GRRAS) were proposed. J Clin Epidemiol 64:96–106CrossRef Kottner J, Audigé L, Brorson S et al (2011) Guidelines for reporting reliability and agreement studies (GRRAS) were proposed. J Clin Epidemiol 64:96–106CrossRef
17.
go back to reference Amanuma M, Kondo T, Sano T et al (2016) Assessment of coronary in-stent restenosis: value of subtraction coronary computed tomography angiography. Int J Cardiovasc Imaging 32:661–670CrossRefPubMed Amanuma M, Kondo T, Sano T et al (2016) Assessment of coronary in-stent restenosis: value of subtraction coronary computed tomography angiography. Int J Cardiovasc Imaging 32:661–670CrossRefPubMed
18.
go back to reference Kidoh M, Utsunomiya D, Oda S et al (2015) Optimized subtraction coronary CT angiography protocol for clinical use with short breath-holding time-initial experience. Acad Radiol 22:117–120CrossRefPubMed Kidoh M, Utsunomiya D, Oda S et al (2015) Optimized subtraction coronary CT angiography protocol for clinical use with short breath-holding time-initial experience. Acad Radiol 22:117–120CrossRefPubMed
19.
go back to reference Halpern EJ, Halpern DJ, Yanof JH et al (2009) Is coronary stent assessment improved with spectral analysis of dual energy CT? Acad Radiol 16:1241–1250CrossRefPubMed Halpern EJ, Halpern DJ, Yanof JH et al (2009) Is coronary stent assessment improved with spectral analysis of dual energy CT? Acad Radiol 16:1241–1250CrossRefPubMed
20.
go back to reference Ozguner O, Dhanantwari A, Halliburton S, Wen G, Utrup S, Jordan D (2018) Objective image characterization of a spectral CT scanner with dual-layer detector. Phys Med Biol 63:025027 Ozguner O, Dhanantwari A, Halliburton S, Wen G, Utrup S, Jordan D (2018) Objective image characterization of a spectral CT scanner with dual-layer detector. Phys Med Biol 63:025027
21.
go back to reference Moon JW, Park BK, Kim CK, Park SY (2012) Evaluation of virtual unenhanced CT obtained from dual-energy CT urography for detecting urinary stones. Br J Radiol 85:e176–e181CrossRefPubMedPubMedCentral Moon JW, Park BK, Kim CK, Park SY (2012) Evaluation of virtual unenhanced CT obtained from dual-energy CT urography for detecting urinary stones. Br J Radiol 85:e176–e181CrossRefPubMedPubMedCentral
22.
go back to reference Patino M, Prochowski A, Agrawal MD et al (2016) Material separation using dual-energy CT: current and emerging applications. Radiographics 36:1087–1105CrossRefPubMed Patino M, Prochowski A, Agrawal MD et al (2016) Material separation using dual-energy CT: current and emerging applications. Radiographics 36:1087–1105CrossRefPubMed
23.
go back to reference Mangold S, Thomas C, Fenchel M et al (2012) Virtual nonenhanced dual-energy CT urography with tin-filter technology determinants of detection of urinary calculi in the renal collecting system. Radiology 264:119–125CrossRef Mangold S, Thomas C, Fenchel M et al (2012) Virtual nonenhanced dual-energy CT urography with tin-filter technology determinants of detection of urinary calculi in the renal collecting system. Radiology 264:119–125CrossRef
24.
go back to reference Esposito A, Colantoni C, De Cobelli F et al (2013) Multidetector computed tomography for coronary stents imaging high-voltage (140-kVp) prospective ECG-triggered versus standard-voltage (120-kVp) retrospective ECG-gated helical scanning. J Comput Assist Tomogr 37:395–401CrossRefPubMed Esposito A, Colantoni C, De Cobelli F et al (2013) Multidetector computed tomography for coronary stents imaging high-voltage (140-kVp) prospective ECG-triggered versus standard-voltage (120-kVp) retrospective ECG-gated helical scanning. J Comput Assist Tomogr 37:395–401CrossRefPubMed
25.
go back to reference Menke J, Unterberg-Buchwald C, Staab W, Sohns JM, Seif Amir Hosseini A, Schwarz A (2013) Head-to-head comparison of prospectively triggered vs retrospectively gated coronary computed tomography angiography: meta-analysis of diagnostic accuracy, image quality, and radiation dose. Am Heart J 165:154–163 e153 Menke J, Unterberg-Buchwald C, Staab W, Sohns JM, Seif Amir Hosseini A, Schwarz A (2013) Head-to-head comparison of prospectively triggered vs retrospectively gated coronary computed tomography angiography: meta-analysis of diagnostic accuracy, image quality, and radiation dose. Am Heart J 165:154–163 e153
26.
go back to reference Andreini D, Pontone G, Bartorelli AL et al (2011) High diagnostic accuracy of prospective ECG-gating 64-slice computed tomography coronary angiography for the detection of in-stent restenosis: in-stent restenosis assessment by low-dose MDCT. Eur Radiol 21:1430–1438CrossRefPubMed Andreini D, Pontone G, Bartorelli AL et al (2011) High diagnostic accuracy of prospective ECG-gating 64-slice computed tomography coronary angiography for the detection of in-stent restenosis: in-stent restenosis assessment by low-dose MDCT. Eur Radiol 21:1430–1438CrossRefPubMed
Metadata
Title
Initial exploration of coronary stent image subtraction using dual-layer spectral CT
Authors
Le Qin
ShengJia Gu
ChiHua Chen
Huan Zhang
ZhenBin Zhu
XingBiao Chen
Qun Han
FuHua Yan
WenJie Yang
Publication date
01-08-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 8/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5990-1

Other articles of this Issue 8/2019

European Radiology 8/2019 Go to the issue