Abstract

Purpose

Flat-panel computed tomography (FP-CT) is increasingly available in angiographic rooms and hybrid OR’s. Considering its easy access, cerebral imaging using FP-CT is an appealing modality for intra-procedural applications. The purpose of this systematic review is to assess the diagnostic accuracy of FP-CT compared with perfusion computed tomography (CTP) and perfusion magnetic resonance (MRP) in cerebral perfusion imaging.

Methods

We performed a systematic literature search in the Cochrane Library, MEDLINE, Embase, and Web of Science up to June 2019 for studies directly comparing FP-CT with either CTP or MRP in vivo. Methodological quality was assessed using the QUADAS-2 tool. Data on diagnostic accuracy was extracted and pooled if possible.

Results

We found 11 studies comparing FP-CT with CTP and 5 studies comparing FP-CT with MRP. Most articles were pilot or feasibility studies, focusing on scanning and contrast protocols. All patients studied showed signs of cerebrovascular disease. Half of the studies were animal trials. Quality assessment showed unclear to high risks of bias and low concerns regarding applicability. Five studies reported on diagnostic accuracy; FP-CT shows good sensitivity (range 0.84–1.00) and moderate specificity (range 0.63–0.88) in detecting cerebral blood volume (CBV) lesions.

Conclusions

Even though FP-CT provides similar CBV values and reconstructed blood volume maps as CTP in cerebrovascular disease, additional studies are required in order to reliably compare its diagnostic accuracy with cerebral perfusion imaging.
Appendix
Available only for authorised users
Literature
1.
go back to reference Doerfler A, Gölitz P, Engelhorn T, Kloska S, Struffert T (2015) Flat-panel computed tomography (DYNA-CT) in neuroradiology. From High-Resolution Imaging of Implants to One-Stop-Shopping for Acute Stroke. Clin Neuroradiol 25:291–297CrossRef Doerfler A, Gölitz P, Engelhorn T, Kloska S, Struffert T (2015) Flat-panel computed tomography (DYNA-CT) in neuroradiology. From High-Resolution Imaging of Implants to One-Stop-Shopping for Acute Stroke. Clin Neuroradiol 25:291–297CrossRef
2.
go back to reference Frolich AM, Buhk JH, Fiehler J, Kemmling A (2016) Voxel-based sensitivity of flat-panel ct for the detection of intracranial hemorrhage: comparison to multi-detector CT. PLoS ONE 11(11):1–10CrossRef Frolich AM, Buhk JH, Fiehler J, Kemmling A (2016) Voxel-based sensitivity of flat-panel ct for the detection of intracranial hemorrhage: comparison to multi-detector CT. PLoS ONE 11(11):1–10CrossRef
3.
go back to reference Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender WA et al (2009) Visualisation of intracerebral hemorrhage with flat-detector CT compared to multislice CT: Results in 44 cases. Eur Radiol 19(3):619–625CrossRef Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender WA et al (2009) Visualisation of intracerebral hemorrhage with flat-detector CT compared to multislice CT: Results in 44 cases. Eur Radiol 19(3):619–625CrossRef
4.
go back to reference Lindner T, Ahmeti H, Lübbing I, Helle M, Jansen O, Synowitz M et al (2017) Intraoperative resection control using arterial spin labeling - proof of concept, reproducibility of data and initial results. NeuroImage Clin 15(April):136–142CrossRef Lindner T, Ahmeti H, Lübbing I, Helle M, Jansen O, Synowitz M et al (2017) Intraoperative resection control using arterial spin labeling - proof of concept, reproducibility of data and initial results. NeuroImage Clin 15(April):136–142CrossRef
5.
go back to reference Wintermark M, Reichhart M, Cuisenaire O, Maeder P, Thiran JP, Schnyder P et al (2002) Comparison of admission perfusion computed tomography and qualitative diffusion- and perfusion-weighted magnetic resonance imaging in acute stroke patients. Stroke. 33(8):2025–2031CrossRef Wintermark M, Reichhart M, Cuisenaire O, Maeder P, Thiran JP, Schnyder P et al (2002) Comparison of admission perfusion computed tomography and qualitative diffusion- and perfusion-weighted magnetic resonance imaging in acute stroke patients. Stroke. 33(8):2025–2031CrossRef
6.
go back to reference Connor SEJ, Flis C (2005) The contribution of high-resolution multiplanar reformats of the skull base to the detection of skull-base fractures. Clin Radiol 60(8):878–885CrossRef Connor SEJ, Flis C (2005) The contribution of high-resolution multiplanar reformats of the skull base to the detection of skull-base fractures. Clin Radiol 60(8):878–885CrossRef
7.
go back to reference Grigoryan M, Tung CE, Albers GW (2011) Role of diffusion and perfusion MRI in selecting patients for reperfusion therapies. Neuroimaging Clin N Am 21:247–257CrossRef Grigoryan M, Tung CE, Albers GW (2011) Role of diffusion and perfusion MRI in selecting patients for reperfusion therapies. Neuroimaging Clin N Am 21:247–257CrossRef
8.
go back to reference Fouke SJ, Benzinger T, Gibson D, Ryken TC, Kalkanis SN, Olson JJ (2015) The role of imaging in the management of adults with diffuse low grade glioma: a systematic review and evidence-based clinical practice guideline. J Neuro-Oncol 125:457–479CrossRef Fouke SJ, Benzinger T, Gibson D, Ryken TC, Kalkanis SN, Olson JJ (2015) The role of imaging in the management of adults with diffuse low grade glioma: a systematic review and evidence-based clinical practice guideline. J Neuro-Oncol 125:457–479CrossRef
9.
go back to reference Provenzale JM, Shah K, Patel U, McCrory DC (2008) Systematic review of CT and MR perfusion imaging for assessment of acute cerebrovascular disease. AJNR Am J Neuroradiol 29(8):1476–1482CrossRef Provenzale JM, Shah K, Patel U, McCrory DC (2008) Systematic review of CT and MR perfusion imaging for assessment of acute cerebrovascular disease. AJNR Am J Neuroradiol 29(8):1476–1482CrossRef
10.
go back to reference Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche PC, Ioannidis JP, Clarke M, Devereaux PJ, Kleijnen J, Moher D (2009) The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. J Clin Epidemiol 62(10):e1–34CrossRef Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche PC, Ioannidis JP, Clarke M, Devereaux PJ, Kleijnen J, Moher D (2009) The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. J Clin Epidemiol 62(10):e1–34CrossRef
11.
go back to reference Whiting PF, Rutjes AWS, Westwood ME, Mallett S, Deeks JJ, Reitsma JB et al (2011) Quadas-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med 155:529–536CrossRef Whiting PF, Rutjes AWS, Westwood ME, Mallett S, Deeks JJ, Reitsma JB et al (2011) Quadas-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med 155:529–536CrossRef
12.
go back to reference Struffert T, Deuerling-Zheng Y, Engelhorn T, Kloska S, Golitz P, Kohrmann M et al (2012) Feasibility of cerebral blood volume mapping by flat-panel detector CT in the angiography suite: first experience in patients with acute middle cerebral artery occlusions. AJNR Am J Neuroradiol 33(4 PG-618-25):618–625CrossRef Struffert T, Deuerling-Zheng Y, Engelhorn T, Kloska S, Golitz P, Kohrmann M et al (2012) Feasibility of cerebral blood volume mapping by flat-panel detector CT in the angiography suite: first experience in patients with acute middle cerebral artery occlusions. AJNR Am J Neuroradiol 33(4 PG-618-25):618–625CrossRef
13.
go back to reference Zhang Q, Wang B, Han J, Zhang Z, Sun Q, Beilner J et al (2013) Feasibility study of perfusion imaging using flat detector CT with an intra-arterial injection protocol compared to conventional multi-slice perfusion CT with an intravenous injection protocol. Interv Neuroradiol 19(4 PG-409-15):409–415CrossRef Zhang Q, Wang B, Han J, Zhang Z, Sun Q, Beilner J et al (2013) Feasibility study of perfusion imaging using flat detector CT with an intra-arterial injection protocol compared to conventional multi-slice perfusion CT with an intravenous injection protocol. Interv Neuroradiol 19(4 PG-409-15):409–415CrossRef
14.
go back to reference Zhang YW, Hong B, Huang QH, Xu Y, Zhao WY, Yang ZG et al (2012) The feasibility and clinical application of flat-panel detector computer tomography in evaluating cerebral blood volume: An initial prospective study. Vol. 21. J Interv Radiol 21(4):265–269 Zhang YW, Hong B, Huang QH, Xu Y, Zhao WY, Yang ZG et al (2012) The feasibility and clinical application of flat-panel detector computer tomography in evaluating cerebral blood volume: An initial prospective study. Vol. 21. J Interv Radiol 21(4):265–269
15.
go back to reference Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Boese J, Zellerhoff M et al (2011) Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT. Eur Radiol 21(4 PG-882-9):882–889CrossRef Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Boese J, Zellerhoff M et al (2011) Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT. Eur Radiol 21(4 PG-882-9):882–889CrossRef
16.
go back to reference Fiorella D, Turk A, Chaudry I, Turner R, Dunkin J, Roque C et al (2014) A prospective, multicenter pilot study investigating the utility of flat detector derived parenchymal blood volume maps to estimate cerebral blood volume in stroke patients. J NeuroInterventional Surg 6(6):451–456CrossRef Fiorella D, Turk A, Chaudry I, Turner R, Dunkin J, Roque C et al (2014) A prospective, multicenter pilot study investigating the utility of flat detector derived parenchymal blood volume maps to estimate cerebral blood volume in stroke patients. J NeuroInterventional Surg 6(6):451–456CrossRef
17.
go back to reference Bley T, Strother CM, Pulfer K, Royalty K, Zellerhoff M, Deuerling-Zheng Y et al (2010) C-arm CT measurement of cerebral blood volume in ischemic stroke: An experimental study in canines. Am J Neuroradiol 31(3):536–540CrossRef Bley T, Strother CM, Pulfer K, Royalty K, Zellerhoff M, Deuerling-Zheng Y et al (2010) C-arm CT measurement of cerebral blood volume in ischemic stroke: An experimental study in canines. Am J Neuroradiol 31(3):536–540CrossRef
18.
go back to reference Ahmed AS, Zellerhoff M, Strother CM, Pulfer KA, Redel T, Deuerling-Zheng Y et al (2009) C-arm CT measurement of cerebral blood volume: an experimental study in canines. AJNR Am J Neuroradiol 30(5 PG-917-22):917–922CrossRef Ahmed AS, Zellerhoff M, Strother CM, Pulfer KA, Redel T, Deuerling-Zheng Y et al (2009) C-arm CT measurement of cerebral blood volume: an experimental study in canines. AJNR Am J Neuroradiol 30(5 PG-917-22):917–922CrossRef
19.
go back to reference Yasuda R, Royalty K, Pulfer K, Consigny D, Strother CM (2012) C-arm CT measurement of cerebral blood volume using intra-arterial injection of contrast medium: an experimental study in canines. AJNR Am J Neuroradiol 33(9 PG-1696-701):1696–1701CrossRef Yasuda R, Royalty K, Pulfer K, Consigny D, Strother CM (2012) C-arm CT measurement of cerebral blood volume using intra-arterial injection of contrast medium: an experimental study in canines. AJNR Am J Neuroradiol 33(9 PG-1696-701):1696–1701CrossRef
20.
go back to reference Royalty K, Manhart M, Pulfer K, Deuerling-Zheng Y, Strother C, Fieselmann A et al (2013) C-arm CT measurement of cerebral blood volume and cerebral blood flow using a novel high-speed acquisition and a single intravenous contrast injection. AJNR Am J Neuroradiol 34(11 PG-2131-8):2131–2138CrossRef Royalty K, Manhart M, Pulfer K, Deuerling-Zheng Y, Strother C, Fieselmann A et al (2013) C-arm CT measurement of cerebral blood volume and cerebral blood flow using a novel high-speed acquisition and a single intravenous contrast injection. AJNR Am J Neuroradiol 34(11 PG-2131-8):2131–2138CrossRef
21.
go back to reference Beuing O, Boese A, Kyriakou Y, Deuerling-Zengh Y, Jollenbeck B, Scherlach C et al (2014) A novel technique for the measurement of CBF and CBV with robot-arm-mounted flat panel CT in a large-animal model. AJNR Am J Neuroradiol 35(9 PG-1740-5):1740–1745CrossRef Beuing O, Boese A, Kyriakou Y, Deuerling-Zengh Y, Jollenbeck B, Scherlach C et al (2014) A novel technique for the measurement of CBF and CBV with robot-arm-mounted flat panel CT in a large-animal model. AJNR Am J Neuroradiol 35(9 PG-1740-5):1740–1745CrossRef
22.
go back to reference Ganguly A, Fieselmann A, Marks M, Rosenberg J, Boese J, Deuerling-Zheng Y et al (2011) Cerebral CT perfusion using an interventional C-arm imaging system: cerebral blood flow measurements. AJNR Am J Neuroradiol 32(8 PG-1525-31):1525–1531CrossRef Ganguly A, Fieselmann A, Marks M, Rosenberg J, Boese J, Deuerling-Zheng Y et al (2011) Cerebral CT perfusion using an interventional C-arm imaging system: cerebral blood flow measurements. AJNR Am J Neuroradiol 32(8 PG-1525-31):1525–1531CrossRef
23.
go back to reference Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Lang S, Mennecke A et al (2015) Dynamic angiography and perfusion imaging using flat detector ct in the angiography suite: a pilot study in patients with acute middle cerebral artery occlusions. Am J Neuroradiol 36(10):1964–1970CrossRef Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Lang S, Mennecke A et al (2015) Dynamic angiography and perfusion imaging using flat detector ct in the angiography suite: a pilot study in patients with acute middle cerebral artery occlusions. Am J Neuroradiol 36(10):1964–1970CrossRef
24.
go back to reference Kamran M, Byrne JV (2015) C-arm flat detector computed tomography parenchymal blood volume imaging: the nature of parenchymal blood volume parameter and the feasibility of parenchymal blood volume imaging in aneurysmal subarachnoid haemorrhage patients. Vol. 57. Neuroradiology. 57(9):937–949CrossRef Kamran M, Byrne JV (2015) C-arm flat detector computed tomography parenchymal blood volume imaging: the nature of parenchymal blood volume parameter and the feasibility of parenchymal blood volume imaging in aneurysmal subarachnoid haemorrhage patients. Vol. 57. Neuroradiology. 57(9):937–949CrossRef
25.
go back to reference Chen LW, Lin CJ, Guo WY, Hung SC, Lee HJ, Chen KK et al (2018) Evaluating cerebral hemodynamics using quantitative digital subtraction angiography and flat-detector computed tomography perfusion imaging: a comparative study in patients with carotid stenosis. J Chin Med Assoc 81(11):977–984CrossRef Chen LW, Lin CJ, Guo WY, Hung SC, Lee HJ, Chen KK et al (2018) Evaluating cerebral hemodynamics using quantitative digital subtraction angiography and flat-detector computed tomography perfusion imaging: a comparative study in patients with carotid stenosis. J Chin Med Assoc 81(11):977–984CrossRef
26.
go back to reference Garcia M, Okell TW, Gloor M, Chappell MA, Jezzard P, Bieri O et al (2017) Feasibility of flat panel detector CT in perfusion assessment of brain arteriovenous malformations: initial clinical experience. Am J Neuroradiol 38(4):735–739CrossRef Garcia M, Okell TW, Gloor M, Chappell MA, Jezzard P, Bieri O et al (2017) Feasibility of flat panel detector CT in perfusion assessment of brain arteriovenous malformations: initial clinical experience. Am J Neuroradiol 38(4):735–739CrossRef
27.
go back to reference Kuriyama T, Sakai N, Beppu M, Sakai C, Imamura H, Kojima I et al (2018) Optimal dilution of contrast medium for quantitating parenchymal blood volume using a flat-panel detector. J Int Med Res 46(1):464–474CrossRef Kuriyama T, Sakai N, Beppu M, Sakai C, Imamura H, Kojima I et al (2018) Optimal dilution of contrast medium for quantitating parenchymal blood volume using a flat-panel detector. J Int Med Res 46(1):464–474CrossRef
28.
go back to reference Lin C-J, Guo W-Y, Chang F-C, Hung S-C, Chen K-K, Yu D-Z et al (2016) Using flat-panel perfusion imaging to measure cerebral hemodynamics. Medicine (Baltimore) 95(20):e3529CrossRef Lin C-J, Guo W-Y, Chang F-C, Hung S-C, Chen K-K, Yu D-Z et al (2016) Using flat-panel perfusion imaging to measure cerebral hemodynamics. Medicine (Baltimore) 95(20):e3529CrossRef
29.
go back to reference Barber PA, Demchuk AM, Zhang J, Buchan AM (2000) Validity and reliability of a quantitative computed tomography score in predicting outcome of hyperacute stroke before thrombolytic therapy. ASPECTS Study Group. Alberta Stroke Programme Early CT Score. Lancet. 355(9216):1670–1674CrossRef Barber PA, Demchuk AM, Zhang J, Buchan AM (2000) Validity and reliability of a quantitative computed tomography score in predicting outcome of hyperacute stroke before thrombolytic therapy. ASPECTS Study Group. Alberta Stroke Programme Early CT Score. Lancet. 355(9216):1670–1674CrossRef
30.
go back to reference Fieselmann A, Ganguly A, Deuerling-Zheng Y, Zellerhoff M, Rohkohl C, Boese J et al (2012) Interventional 4-D C-arm CT perfusion imaging using interleaved scanning and partial reconstruction interpolation. IEEE Trans Med Imaging 31(4 PG-892-906):892–906CrossRef Fieselmann A, Ganguly A, Deuerling-Zheng Y, Zellerhoff M, Rohkohl C, Boese J et al (2012) Interventional 4-D C-arm CT perfusion imaging using interleaved scanning and partial reconstruction interpolation. IEEE Trans Med Imaging 31(4 PG-892-906):892–906CrossRef
31.
go back to reference Hooijmans CR, Rovers MM, Vries RBMD, Leenaars M, Ritskes-hoitinga M, Langendam MW (2014) SYRCLE’ s risk of bias tool for animal studies. BMC Med Res Methodol 14(1):1–9CrossRef Hooijmans CR, Rovers MM, Vries RBMD, Leenaars M, Ritskes-hoitinga M, Langendam MW (2014) SYRCLE’ s risk of bias tool for animal studies. BMC Med Res Methodol 14(1):1–9CrossRef
32.
go back to reference Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender WA et al (2010) Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol 31(8 PG-1462-9):1462–1469CrossRef Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender WA et al (2010) Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol 31(8 PG-1462-9):1462–1469CrossRef
33.
go back to reference Gupta R, Cheung AC, Bartling SH, Lisauskas J, Grasruck M, Leidecker C et al (2008) Flat-panel volume CT: fundamental principles, technology, and applications. Radiogr A Rev Publ Radiol Soc North Am Inc 28(7):2009–2022 Gupta R, Cheung AC, Bartling SH, Lisauskas J, Grasruck M, Leidecker C et al (2008) Flat-panel volume CT: fundamental principles, technology, and applications. Radiogr A Rev Publ Radiol Soc North Am Inc 28(7):2009–2022
34.
go back to reference Abels B, Klotz E, Tomandl BF, Villablanca JP, Kloska SP, Lell MM (2011) CT perfusion in acute ischemic stroke: a comparison of 2-second and 1-second temporal resolution. Am J Neuroradiol 32(9):1632–1639CrossRef Abels B, Klotz E, Tomandl BF, Villablanca JP, Kloska SP, Lell MM (2011) CT perfusion in acute ischemic stroke: a comparison of 2-second and 1-second temporal resolution. Am J Neuroradiol 32(9):1632–1639CrossRef
35.
go back to reference Kämena A, Streitparth F, Grieser C, Lehmkuhl L, Jamil B, Wojtal K et al (2007) Dynamic perfusion CT: optimizing the temporal resolution for the calculation of perfusion CT parameters in stroke patients. Eur J Radiol 64(1):111–118CrossRef Kämena A, Streitparth F, Grieser C, Lehmkuhl L, Jamil B, Wojtal K et al (2007) Dynamic perfusion CT: optimizing the temporal resolution for the calculation of perfusion CT parameters in stroke patients. Eur J Radiol 64(1):111–118CrossRef
36.
go back to reference Wiesmann M, Berg S, Bohner G, Klingebiel R, Schöpf V, Stoeckelhuber BM et al (2008) Dose reduction in dynamic perfusion CT of the brain: effects of the scan frequency on measurements of cerebral blood flow, cerebral blood volume, and mean transit time. Eur Radiol 18(12):2967–2974CrossRef Wiesmann M, Berg S, Bohner G, Klingebiel R, Schöpf V, Stoeckelhuber BM et al (2008) Dose reduction in dynamic perfusion CT of the brain: effects of the scan frequency on measurements of cerebral blood flow, cerebral blood volume, and mean transit time. Eur Radiol 18(12):2967–2974CrossRef
37.
go back to reference Wintermark M, Smith WS, Ko NU, Quist M, Schnyder P, Dillon WP (2004) Dynamic perfusion CT: optimizing the temporal resolution and contrast volume for calculation of perfusion CT parameters in stroke patients. Am J Neuroradiol 25(5):720–729PubMed Wintermark M, Smith WS, Ko NU, Quist M, Schnyder P, Dillon WP (2004) Dynamic perfusion CT: optimizing the temporal resolution and contrast volume for calculation of perfusion CT parameters in stroke patients. Am J Neuroradiol 25(5):720–729PubMed
Metadata
Title
Diagnostic accuracy of flat-panel computed tomography in assessing cerebral perfusion in comparison with perfusion computed tomography and perfusion magnetic resonance: a systematic review
Authors
Ernst L. Stille
Ilaria Viozzi
Mark ter Laan
Frederick J.A. Meijer
Jurgen J. Futterer
Maroeska M. Rovers
Publication date
01-12-2019
Publisher
Springer Berlin Heidelberg
Published in
Neuroradiology / Issue 12/2019
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-019-02285-y

Other articles of this Issue 12/2019

Neuroradiology 12/2019 Go to the issue