Published in: Journal of Cardiovascular Magnetic Resonance 1/2014

Open Access 01-12-2014 | Research

Calibration of myocardial T2 and T1 against iron concentration

Authors: John-Paul Carpenter, Taigang He, Paul Kirk, Michael Roughton, Lisa J Anderson, Sofia V de Noronha, A John Baksi, Mary N Sheppard, John B Porter, J Malcolm Walker, John C Wood, Gianluca Forni, Gualtiero Catani, Gildo Matta, Suthat Fucharoen, Adam Fleming, Mike House, Greg Black, David N Firmin, Timothy G St. Pierre, Dudley J Pennell

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2014

Login to get access

Abstract

Background

The assessment of myocardial iron using T2* cardiovascular magnetic resonance (CMR) has been validated and calibrated, and is in clinical use. However, there is very limited data assessing the relaxation parameters T1 and T2 for measurement of human myocardial iron.

Methods

Twelve hearts were examined from transfusion-dependent patients: 11 with end-stage heart failure, either following death (n = 7) or cardiac transplantation (n = 4), and 1 heart from a patient who died from a stroke with no cardiac iron loading. Ex-vivo R1 and R2 measurements (R1 = 1/T1 and R2 = 1/T2) at 1.5 Tesla were compared with myocardial iron concentration measured using inductively coupled plasma atomic emission spectroscopy.

Results

From a single myocardial slice in formalin which was repeatedly examined, a modest decrease in T2 was observed with time, from mean (±SD) 23.7 ± 0.93 ms at baseline (13 days after death and formalin fixation) to 18.5 ± 1.41 ms at day 566 (p < 0.001). Raw T2 values were therefore adjusted to correct for this fall over time. Myocardial R2 was correlated with iron concentration [Fe] (R2 0.566, p < 0.001), but the correlation was stronger between LnR2 and Ln[Fe] (R2 0.790, p < 0.001). The relation was [Fe] = 5081•(T2)-2.22 between T2 (ms) and myocardial iron (mg/g dry weight). Analysis of T1 proved challenging with a dichotomous distribution of T1, with very short T1 (mean 72.3 ± 25.8 ms) that was independent of iron concentration in all hearts stored in formalin for greater than 12 months. In the remaining hearts stored for <10 weeks prior to scanning, LnR1 and iron concentration were correlated but with marked scatter (R2 0.517, p < 0.001). A linear relationship was present between T1 and T2 in the hearts stored for a short period (R2 0.657, p < 0.001).

Conclusion

Myocardial T2 correlates well with myocardial iron concentration, which raises the possibility that T2 may provide additive information to T2* for patients with myocardial siderosis. However, ex-vivo T1 measurements are less reliable due to the severe chemical effects of formalin on T1 shortening, and therefore T1 calibration may only be practical from in-vivo human studies.
Appendix
Available only for authorised users
Literature
1.
go back to reference Anderson LJ, Westwood MA, Holden S, Davis B, Prescott E, Wonke B, Porter JB, Walker JM, Pennell DJ: Myocardial iron clearance during reversal of siderotic cardiomyopathy with intravenous desferrioxamine: a prospective study using T2* cardiovascular magnetic resonance. Br J Haematol. 2004, 127: 348-55. 10.1111/j.1365-2141.2004.05202.x.CrossRefPubMed Anderson LJ, Westwood MA, Holden S, Davis B, Prescott E, Wonke B, Porter JB, Walker JM, Pennell DJ: Myocardial iron clearance during reversal of siderotic cardiomyopathy with intravenous desferrioxamine: a prospective study using T2* cardiovascular magnetic resonance. Br J Haematol. 2004, 127: 348-55. 10.1111/j.1365-2141.2004.05202.x.CrossRefPubMed
2.
go back to reference Miskin H, Yaniv I, Berant M, Hershko C, Tamary H: Reversal of cardiac complications in thalassemia major by long-term intermittent daily intensive iron chelation. Eur J Haematol. 2003, 70: 398-403. 10.1034/j.1600-0609.2003.00075.x.CrossRefPubMed Miskin H, Yaniv I, Berant M, Hershko C, Tamary H: Reversal of cardiac complications in thalassemia major by long-term intermittent daily intensive iron chelation. Eur J Haematol. 2003, 70: 398-403. 10.1034/j.1600-0609.2003.00075.x.CrossRefPubMed
3.
go back to reference Anderson LJ, Holden S, Davis B, Prescott E, Charrier CC, Bunce NH, Firmin DN, Wonke B, Porter J, Walker JM, Pennell DJ: Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J. 2001, 23: 2171-9. 10.1053/euhj.2001.2822.CrossRef Anderson LJ, Holden S, Davis B, Prescott E, Charrier CC, Bunce NH, Firmin DN, Wonke B, Porter J, Walker JM, Pennell DJ: Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J. 2001, 23: 2171-9. 10.1053/euhj.2001.2822.CrossRef
4.
go back to reference Nienhuis AW, Griffith P, Strawczynski H, Henry W, Borer J, Leon M, Anderson WF: Evaluation of cardiac function in patients with thalassemia major. Ann N Y Acad Sci. 1980, 344: 384-96. 10.1111/j.1749-6632.1980.tb33677.x.CrossRefPubMed Nienhuis AW, Griffith P, Strawczynski H, Henry W, Borer J, Leon M, Anderson WF: Evaluation of cardiac function in patients with thalassemia major. Ann N Y Acad Sci. 1980, 344: 384-96. 10.1111/j.1749-6632.1980.tb33677.x.CrossRefPubMed
5.
go back to reference Westwood MA, Anderson LJ, Firmin DN, Gatehouse PD, Lorenz CH, Wonke B, Pennell DJ: Interscanner reproducibility of cardiovascular magnetic resonance T2* measurements of tissue iron in thalassemia. J Magn Reson Imaging. 2003, 18: 616-20. 10.1002/jmri.10396.CrossRefPubMed Westwood MA, Anderson LJ, Firmin DN, Gatehouse PD, Lorenz CH, Wonke B, Pennell DJ: Interscanner reproducibility of cardiovascular magnetic resonance T2* measurements of tissue iron in thalassemia. J Magn Reson Imaging. 2003, 18: 616-20. 10.1002/jmri.10396.CrossRefPubMed
6.
go back to reference Westwood M, Anderson LJ, Firmin DN, Gatehouse PD, Charrier CC, Wonke B, Pennell DJ: A single breath-hold multiecho T2* cardiovascular magnetic resonance technique for diagnosis of myocardial iron overload. J Magn Reson Imaging. 2003, 18: 33-9. 10.1002/jmri.10332.CrossRefPubMed Westwood M, Anderson LJ, Firmin DN, Gatehouse PD, Charrier CC, Wonke B, Pennell DJ: A single breath-hold multiecho T2* cardiovascular magnetic resonance technique for diagnosis of myocardial iron overload. J Magn Reson Imaging. 2003, 18: 33-9. 10.1002/jmri.10332.CrossRefPubMed
7.
go back to reference Tanner MA, He T, Westwood MA, Firmin DN, Pennell DJ: Thalassemia International Federation Heart T2* Investigators. Multi-center validation of the transferability of the magnetic resonance T2* technique for the quantification of tissue iron. Haematologica. 2006, 91: 1388-91.PubMed Tanner MA, He T, Westwood MA, Firmin DN, Pennell DJ: Thalassemia International Federation Heart T2* Investigators. Multi-center validation of the transferability of the magnetic resonance T2* technique for the quantification of tissue iron. Haematologica. 2006, 91: 1388-91.PubMed
8.
go back to reference Carpenter JP, He T, Kirk P, Roughton M, Anderson LJ, de Noronha SV, Sheppard MN, Porter JB, Walker JM, Wood JC, Galanello R, Forni G, Catani G, Matta G, Fucharoen S, Fleming A, House MJ, Black G, Firmin DN, St Pierre TG, Pennell DJ: On T2* magnetic resonance and cardiac iron. Circulation. 2011, 123: 1519-28. 10.1161/CIRCULATIONAHA.110.007641.PubMedCentralCrossRefPubMed Carpenter JP, He T, Kirk P, Roughton M, Anderson LJ, de Noronha SV, Sheppard MN, Porter JB, Walker JM, Wood JC, Galanello R, Forni G, Catani G, Matta G, Fucharoen S, Fleming A, House MJ, Black G, Firmin DN, St Pierre TG, Pennell DJ: On T2* magnetic resonance and cardiac iron. Circulation. 2011, 123: 1519-28. 10.1161/CIRCULATIONAHA.110.007641.PubMedCentralCrossRefPubMed
9.
go back to reference Kirk P, Roughton M, Porter JB, Walker JM, Tanner MA, Patel J, Wu D, Taylor J, Westwood MA, Anderson LJ, Pennell DJ: Cardiac T2* magnetic resonance for prediction of cardiac complications in thalassemia major. Circulation. 2009, 120: 1961-8. 10.1161/CIRCULATIONAHA.109.874487.PubMedCentralCrossRefPubMed Kirk P, Roughton M, Porter JB, Walker JM, Tanner MA, Patel J, Wu D, Taylor J, Westwood MA, Anderson LJ, Pennell DJ: Cardiac T2* magnetic resonance for prediction of cardiac complications in thalassemia major. Circulation. 2009, 120: 1961-8. 10.1161/CIRCULATIONAHA.109.874487.PubMedCentralCrossRefPubMed
10.
go back to reference Telfer PT, Prestcott E, Holden S, Walker M, Hoffbrand AV, Wonke B: Hepatic iron concentration combined with long-term monitoring of serum ferritin to predict complications of iron overload in thalassaemia major. Br J Haematol. 2000, 110: 971-7. 10.1046/j.1365-2141.2000.02298.x.CrossRefPubMed Telfer PT, Prestcott E, Holden S, Walker M, Hoffbrand AV, Wonke B: Hepatic iron concentration combined with long-term monitoring of serum ferritin to predict complications of iron overload in thalassaemia major. Br J Haematol. 2000, 110: 971-7. 10.1046/j.1365-2141.2000.02298.x.CrossRefPubMed
11.
go back to reference Leonardi B, Margossian R, Colan SD, Powell AJ: Relationship of magnetic resonance imaging estimation of myocardial iron to left ventricular systolic and diastolic function in thalassemia. JACC Cardiovasc Imaging. 2008, 1: 572-8. 10.1016/j.jcmg.2008.04.005.CrossRefPubMed Leonardi B, Margossian R, Colan SD, Powell AJ: Relationship of magnetic resonance imaging estimation of myocardial iron to left ventricular systolic and diastolic function in thalassemia. JACC Cardiovasc Imaging. 2008, 1: 572-8. 10.1016/j.jcmg.2008.04.005.CrossRefPubMed
12.
go back to reference Fitchett DH, Coltart DJ, Littler WA, Leyland MJ, Trueman T, Gozzard DI, Peters TJ: Cardiac involvement in secondary haemochromatosis: a catheter biopsy study and analysis of myocardium. Cardiovasc Res. 1980, 14: 719-24. 10.1093/cvr/14.12.719.CrossRefPubMed Fitchett DH, Coltart DJ, Littler WA, Leyland MJ, Trueman T, Gozzard DI, Peters TJ: Cardiac involvement in secondary haemochromatosis: a catheter biopsy study and analysis of myocardium. Cardiovasc Res. 1980, 14: 719-24. 10.1093/cvr/14.12.719.CrossRefPubMed
13.
go back to reference Olson LJ, Edwards WD, McCall JT, Ilstrup DM, Gersh BJ: Cardiac iron deposition in idiopathic hemochromatosis: histologic and analytic assessment of 14 hearts from autopsy. J Am Coll Cardiol. 1987, 10: 1239-43. 10.1016/S0735-1097(87)80124-9.CrossRefPubMed Olson LJ, Edwards WD, McCall JT, Ilstrup DM, Gersh BJ: Cardiac iron deposition in idiopathic hemochromatosis: histologic and analytic assessment of 14 hearts from autopsy. J Am Coll Cardiol. 1987, 10: 1239-43. 10.1016/S0735-1097(87)80124-9.CrossRefPubMed
14.
go back to reference Olson LJ, Edwards WD, Holmes DR, Miller FA, Nordstrom LA, Baldus WP: Endomyocardial biopsy in hemochromatosis: clinicopathologic correlates in six cases. J Am Coll Cardiol. 1989, 13: 116-20. 10.1016/0735-1097(89)90558-5.CrossRefPubMed Olson LJ, Edwards WD, Holmes DR, Miller FA, Nordstrom LA, Baldus WP: Endomyocardial biopsy in hemochromatosis: clinicopathologic correlates in six cases. J Am Coll Cardiol. 1989, 13: 116-20. 10.1016/0735-1097(89)90558-5.CrossRefPubMed
15.
go back to reference Modell B, Khan M, Darlison M: Survival in beta thalassaemia major in the UK: data from the UK Thalassaemia Register. Lancet. 2000, 355: 2051-2. 10.1016/S0140-6736(00)02357-6.CrossRefPubMed Modell B, Khan M, Darlison M: Survival in beta thalassaemia major in the UK: data from the UK Thalassaemia Register. Lancet. 2000, 355: 2051-2. 10.1016/S0140-6736(00)02357-6.CrossRefPubMed
16.
go back to reference Modell B, Khan M, Darlison M, Westwood MA, Ingram D, Pennell DJ: Improved survival of thalassaemia major in the UK and relation to T2* cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2008, 10: 42-10.1186/1532-429X-10-42.PubMedCentralCrossRefPubMed Modell B, Khan M, Darlison M, Westwood MA, Ingram D, Pennell DJ: Improved survival of thalassaemia major in the UK and relation to T2* cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2008, 10: 42-10.1186/1532-429X-10-42.PubMedCentralCrossRefPubMed
17.
go back to reference Wood JC, Otto-Duessel M, Aguilar M, Nick H, Nelson MD, Coates TD, Pollack H, Moats R: Cardiac iron determines cardiac T2*, T2, and T1 in the gerbil model of iron cardiomyopathy. Circulation. 2005, 112: 535-43. 10.1161/CIRCULATIONAHA.104.504415.PubMedCentralCrossRefPubMed Wood JC, Otto-Duessel M, Aguilar M, Nick H, Nelson MD, Coates TD, Pollack H, Moats R: Cardiac iron determines cardiac T2*, T2, and T1 in the gerbil model of iron cardiomyopathy. Circulation. 2005, 112: 535-43. 10.1161/CIRCULATIONAHA.104.504415.PubMedCentralCrossRefPubMed
18.
go back to reference St Pierre TG, Clark PR, Chua-anusorn W, Fleming AJ, Jeffrey GP, Olynyk JK, Pootrakul P, Robins E, Lindeman R: Noninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance. Blood. 2005, 105: 855-61. 10.1182/blood-2004-01-0177.CrossRefPubMed St Pierre TG, Clark PR, Chua-anusorn W, Fleming AJ, Jeffrey GP, Olynyk JK, Pootrakul P, Robins E, Lindeman R: Noninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance. Blood. 2005, 105: 855-61. 10.1182/blood-2004-01-0177.CrossRefPubMed
19.
go back to reference Wood JC, Enriquez C, Ghugre N, Tyzka JM, Carson S, Nelson MD, Coates TD: MRI R2 and R2* mapping accurately estimates hepatic iron concentration in transfusion-dependent thalassemia and sickle cell disease patients. Blood. 2005, 106: 1460-5. 10.1182/blood-2004-10-3982.PubMedCentralCrossRefPubMed Wood JC, Enriquez C, Ghugre N, Tyzka JM, Carson S, Nelson MD, Coates TD: MRI R2 and R2* mapping accurately estimates hepatic iron concentration in transfusion-dependent thalassemia and sickle cell disease patients. Blood. 2005, 106: 1460-5. 10.1182/blood-2004-10-3982.PubMedCentralCrossRefPubMed
20.
go back to reference Piechnik SK, Ferreira VM, Dall’Armellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD: Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010, 12: 69-10.1186/1532-429X-12-69.PubMedCentralCrossRefPubMed Piechnik SK, Ferreira VM, Dall’Armellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD: Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010, 12: 69-10.1186/1532-429X-12-69.PubMedCentralCrossRefPubMed
21.
go back to reference Verhaert D, Thavendiranathan P, Giri S, Mihai G, Rajagopalan S, Simonetti OP, Raman SV: Direct T2 quantification of myocardial edema in acute ischemic injury. JACC Cardiovasc Imaging. 2011, 4: 269-78. 10.1016/j.jcmg.2010.09.023.PubMedCentralCrossRefPubMed Verhaert D, Thavendiranathan P, Giri S, Mihai G, Rajagopalan S, Simonetti OP, Raman SV: Direct T2 quantification of myocardial edema in acute ischemic injury. JACC Cardiovasc Imaging. 2011, 4: 269-78. 10.1016/j.jcmg.2010.09.023.PubMedCentralCrossRefPubMed
22.
go back to reference Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS: Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002, 105: 539-42. 10.1161/hc0402.102975.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS: Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002, 105: 539-42. 10.1161/hc0402.102975.CrossRefPubMed
23.
go back to reference He T, Gatehouse PD, Kirk P, Mohiaddin RH, Pennell DJ, Firmin DN: Myocardial T2* measurement in iron-overloaded thalassemia: an ex vivo study to investigate optimal methods of quantification. Magn Reson Med. 2008, 60: 350-6. 10.1002/mrm.21625.CrossRefPubMed He T, Gatehouse PD, Kirk P, Mohiaddin RH, Pennell DJ, Firmin DN: Myocardial T2* measurement in iron-overloaded thalassemia: an ex vivo study to investigate optimal methods of quantification. Magn Reson Med. 2008, 60: 350-6. 10.1002/mrm.21625.CrossRefPubMed
24.
go back to reference Collins W, Taylor WH: Determination of iron in cardiac and liver tissues by plasma emission spectroscopy. Ann Clin Biochem. 1987, 24: 482-7. 10.1177/000456328702400510.CrossRef Collins W, Taylor WH: Determination of iron in cardiac and liver tissues by plasma emission spectroscopy. Ann Clin Biochem. 1987, 24: 482-7. 10.1177/000456328702400510.CrossRef
25.
go back to reference Abbruzzese G, Cossu G, Balocco M, Marchese R, Murgia D, Melis M, Galanello R, Barella S, Matta G, Ruffinengo U, Bonuccelli U, Forni GL: A pilot trial of deferiprone for neurodegeneration with brain iron accumulation. Haematologica. 2011, 96: 1708-11. 10.3324/haematol.2011.043018.PubMedCentralCrossRefPubMed Abbruzzese G, Cossu G, Balocco M, Marchese R, Murgia D, Melis M, Galanello R, Barella S, Matta G, Ruffinengo U, Bonuccelli U, Forni GL: A pilot trial of deferiprone for neurodegeneration with brain iron accumulation. Haematologica. 2011, 96: 1708-11. 10.3324/haematol.2011.043018.PubMedCentralCrossRefPubMed
26.
go back to reference Ghugre NR, Enriquez CM, Gonzalez I, Nelson MD, Coates TD, Wood JC: MRI detects myocardial iron in the human heart. Magn Reson Med. 2006, 56: 681-6. 10.1002/mrm.20981.PubMedCentralCrossRefPubMed Ghugre NR, Enriquez CM, Gonzalez I, Nelson MD, Coates TD, Wood JC: MRI detects myocardial iron in the human heart. Magn Reson Med. 2006, 56: 681-6. 10.1002/mrm.20981.PubMedCentralCrossRefPubMed
27.
go back to reference Mavrogeni SI, Markussis V, Kaklamanis L, Tsiapras D, Paraskevaidis I, Karavolias G, Karagiorga M, Douskou M, Cokkinos DV, Kremastinos DT: A comparison of magnetic resonance imaging and cardiac biopsy in the evaluation of heart iron overload in patients with beta-thalassemia major. Eur J Haematol. 2005, 75: 241-7. 10.1111/j.1600-0609.2005.00474.x.CrossRefPubMed Mavrogeni SI, Markussis V, Kaklamanis L, Tsiapras D, Paraskevaidis I, Karavolias G, Karagiorga M, Douskou M, Cokkinos DV, Kremastinos DT: A comparison of magnetic resonance imaging and cardiac biopsy in the evaluation of heart iron overload in patients with beta-thalassemia major. Eur J Haematol. 2005, 75: 241-7. 10.1111/j.1600-0609.2005.00474.x.CrossRefPubMed
28.
go back to reference Garbowski MW, Carpenter JP, Smith G, Roughton M, Alam MH, He T, Pennell DJ, Porter JB: Biopsy-based calibration of T2* magnetic resonance for estimation of liver iron concentration and comparison with R2 Ferriscan. J Cardiovasc Magn Reson. 2014, 16: 40-10.1186/1532-429X-16-40.PubMedCentralCrossRefPubMed Garbowski MW, Carpenter JP, Smith G, Roughton M, Alam MH, He T, Pennell DJ, Porter JB: Biopsy-based calibration of T2* magnetic resonance for estimation of liver iron concentration and comparison with R2 Ferriscan. J Cardiovasc Magn Reson. 2014, 16: 40-10.1186/1532-429X-16-40.PubMedCentralCrossRefPubMed
29.
go back to reference He T, Gatehouse PD, Anderson LJ, Tanner M, Keegan J, Pennell DJ, Firmin DN: Development of a novel optimized breathhold technique for myocardial T2 measurement in thalassemia. J Magn Reson Imaging. 2006, 24: 580-5. 10.1002/jmri.20681.CrossRefPubMed He T, Gatehouse PD, Anderson LJ, Tanner M, Keegan J, Pennell DJ, Firmin DN: Development of a novel optimized breathhold technique for myocardial T2 measurement in thalassemia. J Magn Reson Imaging. 2006, 24: 580-5. 10.1002/jmri.20681.CrossRefPubMed
30.
go back to reference He T, Kirk P, Firmin DN, Lam WM, Chu WC, Au WY, Chan GC, Tan RS, Ng I, Biceroglu S, Aydinok Y, Fogel MA, Cohen AR, Pennell DJ: Multi-center transferability of a breath-hold T2 technique for myocardial iron assessment. J Cardiovasc Magn Reson. 2008, 10: 11-10.1186/1532-429X-10-11.PubMedCentralCrossRefPubMed He T, Kirk P, Firmin DN, Lam WM, Chu WC, Au WY, Chan GC, Tan RS, Ng I, Biceroglu S, Aydinok Y, Fogel MA, Cohen AR, Pennell DJ: Multi-center transferability of a breath-hold T2 technique for myocardial iron assessment. J Cardiovasc Magn Reson. 2008, 10: 11-10.1186/1532-429X-10-11.PubMedCentralCrossRefPubMed
31.
go back to reference Jensen PD, Jensen FT, Christensen T, Heickendorff L, Jensen LG, Ellegaard J: Indirect evidence for the potential ability of magnetic resonance imaging to evaluate the myocardial iron content in patients with transfusional iron overload. Magma. 2001, 12: 153-66. 10.1007/BF02668097.CrossRefPubMed Jensen PD, Jensen FT, Christensen T, Heickendorff L, Jensen LG, Ellegaard J: Indirect evidence for the potential ability of magnetic resonance imaging to evaluate the myocardial iron content in patients with transfusional iron overload. Magma. 2001, 12: 153-66. 10.1007/BF02668097.CrossRefPubMed
32.
go back to reference Messroghli DR, Plein S, Higgins DM, Walters K, Jones TR, Ridgway JP, Sivananthan MU: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution–reproducibility study. Radiology. 2006, 238: 1004-12. 10.1148/radiol.2382041903.CrossRefPubMed Messroghli DR, Plein S, Higgins DM, Walters K, Jones TR, Ridgway JP, Sivananthan MU: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution–reproducibility study. Radiology. 2006, 238: 1004-12. 10.1148/radiol.2382041903.CrossRefPubMed
33.
go back to reference Feng Y, He T, Carpenter JP, Jabbour A, Alam MH, Gatehouse PD, Greiser A, Messroghli D, Firmin DN, Pennell DJ: In vivo comparison of myocardial T1 with T2 and T2* in thalassaemia major. J Magn Reson Imaging. 2013, 38: 588-93. 10.1002/jmri.24010.CrossRefPubMed Feng Y, He T, Carpenter JP, Jabbour A, Alam MH, Gatehouse PD, Greiser A, Messroghli D, Firmin DN, Pennell DJ: In vivo comparison of myocardial T1 with T2 and T2* in thalassaemia major. J Magn Reson Imaging. 2013, 38: 588-93. 10.1002/jmri.24010.CrossRefPubMed
34.
go back to reference Kamman RL, Go KG, Brouwer W, Berendsen HJ: Nuclear magnetic resonance relaxation in experimental brain edema: effects of water concentration, protein concentration, and temperature. Magn Reson Med. 1988, 6: 265-74. 10.1002/mrm.1910060304.CrossRefPubMed Kamman RL, Go KG, Brouwer W, Berendsen HJ: Nuclear magnetic resonance relaxation in experimental brain edema: effects of water concentration, protein concentration, and temperature. Magn Reson Med. 1988, 6: 265-74. 10.1002/mrm.1910060304.CrossRefPubMed
35.
go back to reference He T, Smith G, Carpenter JP, Mohiaddin R, Pennell DJ, Firmin D: A phantom study of temperature-dependent MRI T2* measurement. J Cardiovasc Magn Reson. 2009, 11 (Suppl 1): 147-10.1186/1532-429X-11-S1-P147.CrossRef He T, Smith G, Carpenter JP, Mohiaddin R, Pennell DJ, Firmin D: A phantom study of temperature-dependent MRI T2* measurement. J Cardiovasc Magn Reson. 2009, 11 (Suppl 1): 147-10.1186/1532-429X-11-S1-P147.CrossRef
36.
go back to reference Wacker CM, Bock M, Hartlep AW, Beck G, van Kaick G, Ertl G, Bauer WR, Schad LR: Changes in myocardial oxygenation and perfusion under pharmacological stress with dipyridamole: assessment using T2* and T1 measurements. Magn Reson Med. 1999, 41: 686-95. 10.1002/(SICI)1522-2594(199904)41:4<686::AID-MRM6>3.0.CO;2-9.CrossRefPubMed Wacker CM, Bock M, Hartlep AW, Beck G, van Kaick G, Ertl G, Bauer WR, Schad LR: Changes in myocardial oxygenation and perfusion under pharmacological stress with dipyridamole: assessment using T2* and T1 measurements. Magn Reson Med. 1999, 41: 686-95. 10.1002/(SICI)1522-2594(199904)41:4<686::AID-MRM6>3.0.CO;2-9.CrossRefPubMed
37.
go back to reference Krombach GA, Hahn C, Tomars M, Buecker A, Grawe A, Günther RW, Kühl HP: Cardiac amyloidosis: MR imaging findings and T1 quantification, comparison with control subjects. J Magn Reson Imaging. 2007, 25: 1283-7. 10.1002/jmri.20917.CrossRefPubMed Krombach GA, Hahn C, Tomars M, Buecker A, Grawe A, Günther RW, Kühl HP: Cardiac amyloidosis: MR imaging findings and T1 quantification, comparison with control subjects. J Magn Reson Imaging. 2007, 25: 1283-7. 10.1002/jmri.20917.CrossRefPubMed
38.
go back to reference Hsu JC, Johnson GA, Smith WM, Reimer KA, Ideker RE: Magnetic resonance imaging of chronic myocardial infarcts in formalin-fixed human autopsy hearts. Circulation. 1994, 89: 2133-40. 10.1161/01.CIR.89.5.2133.CrossRefPubMed Hsu JC, Johnson GA, Smith WM, Reimer KA, Ideker RE: Magnetic resonance imaging of chronic myocardial infarcts in formalin-fixed human autopsy hearts. Circulation. 1994, 89: 2133-40. 10.1161/01.CIR.89.5.2133.CrossRefPubMed
40.
go back to reference Koizumi N, Hatayama F, Sumino K: Problems in the analysis of cadmium in autopsied tissues. Environ Res. 1994, 64: 192-8. 10.1006/enrs.1994.1015.CrossRefPubMed Koizumi N, Hatayama F, Sumino K: Problems in the analysis of cadmium in autopsied tissues. Environ Res. 1994, 64: 192-8. 10.1006/enrs.1994.1015.CrossRefPubMed
41.
go back to reference Olynyk JK, O’Neill R, Britton RS, Bacon BR: Determination of hepatic iron concentration in fresh and paraffin-embedded tissue: diagnostic implications. Gastroenterology. 1994, 106: 674-7.PubMed Olynyk JK, O’Neill R, Britton RS, Bacon BR: Determination of hepatic iron concentration in fresh and paraffin-embedded tissue: diagnostic implications. Gastroenterology. 1994, 106: 674-7.PubMed
42.
go back to reference Chua-anusorn W, Webb J, Macey DJ, Pootrakul P, St Pierre TG: The effect of histological processing on the form of iron in iron-loaded human tissues. Biochim Biophys Acta. 1997, 1360: 255-61. 10.1016/S0925-4439(97)00009-4.CrossRefPubMed Chua-anusorn W, Webb J, Macey DJ, Pootrakul P, St Pierre TG: The effect of histological processing on the form of iron in iron-loaded human tissues. Biochim Biophys Acta. 1997, 1360: 255-61. 10.1016/S0925-4439(97)00009-4.CrossRefPubMed
43.
go back to reference Hosch W, Bock M, Libicher M, Ley S, Hegenbart U, Dengler TJ, Katus HA, Kauczor HU, Kauffmann GW, Kristen AV: MR-relaxometry of myocardial tissue: significant elevation of T1 and T2 relaxation times in cardiac amyloidosis. Invest Radiol. 2007, 42: 636-42. 10.1097/RLI.0b013e318059e021.CrossRefPubMed Hosch W, Bock M, Libicher M, Ley S, Hegenbart U, Dengler TJ, Katus HA, Kauczor HU, Kauffmann GW, Kristen AV: MR-relaxometry of myocardial tissue: significant elevation of T1 and T2 relaxation times in cardiac amyloidosis. Invest Radiol. 2007, 42: 636-42. 10.1097/RLI.0b013e318059e021.CrossRefPubMed
44.
go back to reference Saryan LA, Hollis DP, Economou JS, Eggleston JC: Nuclear magnetic resonance studies of cancer. IV. Correlation of water content with tissue relaxation times. J Natl Cancer Inst. 1974, 52: 599-602.PubMed Saryan LA, Hollis DP, Economou JS, Eggleston JC: Nuclear magnetic resonance studies of cancer. IV. Correlation of water content with tissue relaxation times. J Natl Cancer Inst. 1974, 52: 599-602.PubMed
45.
go back to reference Kamman RL, Go KG, Stomp GP, Hulstaert CE, Berendsen HJ: Changes of relaxation times T1 and T2 in rat tissues after biopsy and fixation. Magn Reson Imaging. 1985, 3: 245-50. 10.1016/0730-725X(85)90353-4.CrossRefPubMed Kamman RL, Go KG, Stomp GP, Hulstaert CE, Berendsen HJ: Changes of relaxation times T1 and T2 in rat tissues after biopsy and fixation. Magn Reson Imaging. 1985, 3: 245-50. 10.1016/0730-725X(85)90353-4.CrossRefPubMed
46.
go back to reference Tovi M, Ericsson A: Measurements of T1 and T2 over time in formalin-fixed human whole-brain specimens. Acta Radiol. 1992, 33: 400-4. 10.3109/02841859209172021.CrossRefPubMed Tovi M, Ericsson A: Measurements of T1 and T2 over time in formalin-fixed human whole-brain specimens. Acta Radiol. 1992, 33: 400-4. 10.3109/02841859209172021.CrossRefPubMed
47.
go back to reference Thickman DI, Kundel HL, Wolf G: Nuclear magnetic resonance characteristics of fresh and fixed tissue: the effect of elapsed time. Radiology. 1983, 148: 183-5. 10.1148/radiology.148.1.6856832.CrossRefPubMed Thickman DI, Kundel HL, Wolf G: Nuclear magnetic resonance characteristics of fresh and fixed tissue: the effect of elapsed time. Radiology. 1983, 148: 183-5. 10.1148/radiology.148.1.6856832.CrossRefPubMed
48.
go back to reference Yong-Hing CJ, Obenaus A, Stryker R, Tong K, Sarty GE: Magnetic resonance imaging and mathematical modeling of progressive formalin fixation of the human brain. Magn Reson Med. 2005, 54: 324-32. 10.1002/mrm.20578.CrossRefPubMed Yong-Hing CJ, Obenaus A, Stryker R, Tong K, Sarty GE: Magnetic resonance imaging and mathematical modeling of progressive formalin fixation of the human brain. Magn Reson Med. 2005, 54: 324-32. 10.1002/mrm.20578.CrossRefPubMed
49.
go back to reference Jensen JH, Chandra R: NMR relaxation in tissues with weak magnetic inhomogeneities. Magn Reson Med. 2000, 44: 144-56. 10.1002/1522-2594(200007)44:1<144::AID-MRM21>3.0.CO;2-O.CrossRefPubMed Jensen JH, Chandra R: NMR relaxation in tissues with weak magnetic inhomogeneities. Magn Reson Med. 2000, 44: 144-56. 10.1002/1522-2594(200007)44:1<144::AID-MRM21>3.0.CO;2-O.CrossRefPubMed
Metadata
Title
Calibration of myocardial T2 and T1 against iron concentration
Authors
John-Paul Carpenter
Taigang He
Paul Kirk
Michael Roughton
Lisa J Anderson
Sofia V de Noronha
A John Baksi
Mary N Sheppard
John B Porter
J Malcolm Walker
John C Wood
Gianluca Forni
Gualtiero Catani
Gildo Matta
Suthat Fucharoen
Adam Fleming
Mike House
Greg Black
David N Firmin
Timothy G St. Pierre
Dudley J Pennell
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2014
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-014-0062-4

Other articles of this Issue 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Go to the issue