01-09-2020 | KNEE

Varus or valgus positioning of the tibial component of a unicompartmental fixed-bearing knee arthroplasty does not increase wear

Authors: Matthias Woiczinski, Christian Schröder, Alexander Paulus, Manuel Kistler, Volkmar Jansson, Peter E. Müller, Patrick Weber

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 9/2020

Login to get access

Abstract

Purpose

Higher revision rates were shown in varus- or valgus-positioned tibias in unicompartmental knee arthroplasty (UKA), but more than 15% of UKA prostheses are implanted with more than 5° of varus or valgus. This study aimed to analyze the wear rate in UKA when implanting the tibial component in either varus or valgus position versus a neutral placement at 90° to the tibial anatomical axis. The study hypothesized that a 5° varus or valgus positioning of the tibial plateau will generate less wear compared to a neutral alignment.

Methods

Wear was experimentally analyzed on a medial anatomical fixed-bearing unicompartmental knee prosthesis (Univation, Aesculap, Germany) in vitro with a customized, four-station, servohydraulic knee wear simulator, reproducing the walking cycle. The forces, loading and range of motion were applied as specified in the ISO 14243–1:2002, 5 million cycles were analyzed. The tibial components of the medial prostheses were inserted in a neutral position, with 5° varus, and 5° valgus (n = 3, each group).

Results

The wear rate decreased significantly with a 5° varus positioning (6.30 ± 1.38 mg/million cycles) and a 5° valgus positioning (4.96 ± 2.47 mg/million cycles) compared to the neutral position (12.16 ± 1.26 mg/million cycles) (p < 0.01 for the varus and the valgus position). The wear area on the inlay was slightly reduced in the varus and valgus group.

Conclusion

A varus or valgus “malpositioning” up to 5° will not lead to an increased wear. Wear was even less because of the reduced articulating contact area between the inlay and the femur. A slight varus positioning of the tibial component (parallel to the anatomical joint line) positioning can be advocated from a point of wear.

Level of evidence

Experimental study.
Appendix
Available only for authorised users
Literature
1.
go back to reference Arnout N, Vanlommel L, Vanlommel J, Luyckx JP, Labey L, Innocenti B et al (2015) Post-cam mechanics and tibiofemoral kinematics: a dynamic in vitro analysis of eight posterior-stabilized total knee designs. Knee Surg Sports Traumatol Arthrosc 23:3343–3353CrossRef Arnout N, Vanlommel L, Vanlommel J, Luyckx JP, Labey L, Innocenti B et al (2015) Post-cam mechanics and tibiofemoral kinematics: a dynamic in vitro analysis of eight posterior-stabilized total knee designs. Knee Surg Sports Traumatol Arthrosc 23:3343–3353CrossRef
2.
go back to reference Barbadoro P, Ensini A, Leardini A, d'Amato M, Feliciangeli A, Timoncini A et al (2014) Tibial component alignment and risk of loosening in unicompartmental knee arthroplasty: a radiographic and radiostereometric study. Knee Surg Sports Traumatol Arthrosc 22:3157–3162CrossRef Barbadoro P, Ensini A, Leardini A, d'Amato M, Feliciangeli A, Timoncini A et al (2014) Tibial component alignment and risk of loosening in unicompartmental knee arthroplasty: a radiographic and radiostereometric study. Knee Surg Sports Traumatol Arthrosc 22:3157–3162CrossRef
4.
go back to reference Cartier P, Sanouiller JL, Grelsamer RP (1996) Unicompartmental knee arthroplasty surgery. 10-year minimum follow-up period. J Arthroplasty 11:782–788CrossRef Cartier P, Sanouiller JL, Grelsamer RP (1996) Unicompartmental knee arthroplasty surgery. 10-year minimum follow-up period. J Arthroplasty 11:782–788CrossRef
5.
go back to reference Chatellard R, Sauleau V, Colmar M, Robert H, Raynaud G, Brilhault J et al (2013) Medial unicompartmental knee arthroplasty: does tibial component position influence clinical outcomes and arthroplasty survival? J Orthopaedics 99:S219–S225 Chatellard R, Sauleau V, Colmar M, Robert H, Raynaud G, Brilhault J et al (2013) Medial unicompartmental knee arthroplasty: does tibial component position influence clinical outcomes and arthroplasty survival? J Orthopaedics 99:S219–S225
6.
go back to reference Diezi C, Wirth S, Meyer DC, Koch PP (2010) Effect of femoral to tibial varus mismatch on the contact area of unicondylar knee prostheses. Knee 17:350–355CrossRef Diezi C, Wirth S, Meyer DC, Koch PP (2010) Effect of femoral to tibial varus mismatch on the contact area of unicondylar knee prostheses. Knee 17:350–355CrossRef
7.
go back to reference Grimberg A, Jansson V, Liebs T, Melsheimer O, Steinbrück A (2018) Endoprothesen register Deutschland. Springer, Berlin Grimberg A, Jansson V, Liebs T, Melsheimer O, Steinbrück A (2018) Endoprothesen register Deutschland. Springer, Berlin
8.
go back to reference Grochowsky JC, Alaways LW, Siskey R, Most E, Kurtz SM (2006) Digital photogrammetry for quantitative wear analysis of retrieved TKA components. J Biomed Mater Res B Appl Biomater 79:263–267CrossRef Grochowsky JC, Alaways LW, Siskey R, Most E, Kurtz SM (2006) Digital photogrammetry for quantitative wear analysis of retrieved TKA components. J Biomed Mater Res B Appl Biomater 79:263–267CrossRef
9.
go back to reference Grupp TM, Utzschneider S, Schroder C, Schwiesau J, Fritz B, Maas A et al (2010) Biotribology of alternative bearing materials for unicompartmental knee arthroplasty. Acta Biomater 6:3601–3610CrossRef Grupp TM, Utzschneider S, Schroder C, Schwiesau J, Fritz B, Maas A et al (2010) Biotribology of alternative bearing materials for unicompartmental knee arthroplasty. Acta Biomater 6:3601–3610CrossRef
10.
go back to reference Gulati A, Pandit H, Jenkins C, Chau R, Dodd CA, Murray DW (2009) The effect of leg alignment on the outcome of unicompartmental knee replacement. J Bone Jt Surg Br 91:469–474CrossRef Gulati A, Pandit H, Jenkins C, Chau R, Dodd CA, Murray DW (2009) The effect of leg alignment on the outcome of unicompartmental knee replacement. J Bone Jt Surg Br 91:469–474CrossRef
11.
go back to reference Innocenti B, Pianigiani S, Ramundo G, Thienpont E (2016) Biomechanical effects of different varus and valgus alignments in medial unicompartmental knee arthroplasty. J Arthroplasty 31:2685–2691CrossRef Innocenti B, Pianigiani S, Ramundo G, Thienpont E (2016) Biomechanical effects of different varus and valgus alignments in medial unicompartmental knee arthroplasty. J Arthroplasty 31:2685–2691CrossRef
12.
go back to reference Kang KT, Son J, Kwon SK, Kwon OR, Koh YG (2018) Preservation of femoral and tibial coronal alignment to improve biomechanical effects of medial unicompartment knee arthroplasty: computational study. Biomed Mater Eng 29:651–664PubMed Kang KT, Son J, Kwon SK, Kwon OR, Koh YG (2018) Preservation of femoral and tibial coronal alignment to improve biomechanical effects of medial unicompartment knee arthroplasty: computational study. Biomed Mater Eng 29:651–664PubMed
13.
go back to reference Kwon OR, Kang KT, Son J, Suh DS, Baek C, Koh YG (2017) Importance of joint line preservation in unicompartmental knee arthroplasty: finite element analysis. J Orthop Res 35:347–352CrossRef Kwon OR, Kang KT, Son J, Suh DS, Baek C, Koh YG (2017) Importance of joint line preservation in unicompartmental knee arthroplasty: finite element analysis. J Orthop Res 35:347–352CrossRef
14.
go back to reference Laurent JT, Yao J, Blanchard C, Crowninshield R (2003) In vitro lateral versus medial wear of a knee prosthesis. Wear 255:1101–1106CrossRef Laurent JT, Yao J, Blanchard C, Crowninshield R (2003) In vitro lateral versus medial wear of a knee prosthesis. Wear 255:1101–1106CrossRef
15.
go back to reference Mazzucco D, Spector M (2003) Effects of contact area and stress on the volumetric wear of ultrahigh molecular weight polyethylene. J Wear 254:514–522CrossRef Mazzucco D, Spector M (2003) Effects of contact area and stress on the volumetric wear of ultrahigh molecular weight polyethylene. J Wear 254:514–522CrossRef
16.
go back to reference Puente Reyna AL, Fritz B, Schwiesau J, Schilling C, Summer B, Thomas P et al (2018) Metal ion release barrier function and biotribological evaluation of a zirconium nitride multilayer coated knee implant under highly demanding activities wear simulation. J Biomech 79:88–96CrossRef Puente Reyna AL, Fritz B, Schwiesau J, Schilling C, Summer B, Thomas P et al (2018) Metal ion release barrier function and biotribological evaluation of a zirconium nitride multilayer coated knee implant under highly demanding activities wear simulation. J Biomech 79:88–96CrossRef
17.
go back to reference Saikko V (2017) Effect of contact area on the wear of ultrahigh molecular weight polyethylene in noncyclic pin-on-disk tests. J Tribol Int 114:84–87CrossRef Saikko V (2017) Effect of contact area on the wear of ultrahigh molecular weight polyethylene in noncyclic pin-on-disk tests. J Tribol Int 114:84–87CrossRef
18.
go back to reference Schwiesau J, Schilling C, Utzschneider S, Jansson V, Fritz B, Blomer W et al (2013) Knee wear simulation under conditions of highly demanding daily activities–influence on an unicompartmental fixed bearing knee design. Med Eng Phys 35:1204–1211CrossRef Schwiesau J, Schilling C, Utzschneider S, Jansson V, Fritz B, Blomer W et al (2013) Knee wear simulation under conditions of highly demanding daily activities–influence on an unicompartmental fixed bearing knee design. Med Eng Phys 35:1204–1211CrossRef
19.
go back to reference The-Swedish-Knee-Arthroplasty-Register (2010) Annual Report 2010. Lund 10/7/2010 The-Swedish-Knee-Arthroplasty-Register (2010) Annual Report 2010. Lund 10/7/2010
20.
go back to reference Weber P, Crispin A, Schmidutz F, Utzschneider S, Pietschmann MF, Jansson V et al (2013) Improved accuracy in computer-assisted unicondylar knee arthroplasty: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 21:2453–2461CrossRef Weber P, Crispin A, Schmidutz F, Utzschneider S, Pietschmann MF, Jansson V et al (2013) Improved accuracy in computer-assisted unicondylar knee arthroplasty: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 21:2453–2461CrossRef
21.
go back to reference Weber P, Schroder C, Schmidutz F, Kraxenberger M, Utzschneider S, Jansson V et al (2013) Increase of tibial slope reduces backside wear in medial mobile bearing unicompartmental knee arthroplasty. Clin Biomech (Bristol Avon) 28:904–909CrossRef Weber P, Schroder C, Schmidutz F, Kraxenberger M, Utzschneider S, Jansson V et al (2013) Increase of tibial slope reduces backside wear in medial mobile bearing unicompartmental knee arthroplasty. Clin Biomech (Bristol Avon) 28:904–909CrossRef
24.
go back to reference Wimmer MA, Nechtow W, Schwenke T, Moisio KC (2015) Knee flexion and daily activities in patients following total knee replacement: a comparison with ISO standard 14243. Biomed Res Int 2015:157541CrossRef Wimmer MA, Nechtow W, Schwenke T, Moisio KC (2015) Knee flexion and daily activities in patients following total knee replacement: a comparison with ISO standard 14243. Biomed Res Int 2015:157541CrossRef
25.
go back to reference Zietz C, Reinders J, Schwiesau J, Paulus A, Kretzer JP, Grupp T et al (2015) Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions. J Mater Sci Mater Med 26:134CrossRef Zietz C, Reinders J, Schwiesau J, Paulus A, Kretzer JP, Grupp T et al (2015) Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions. J Mater Sci Mater Med 26:134CrossRef
Metadata
Title
Varus or valgus positioning of the tibial component of a unicompartmental fixed-bearing knee arthroplasty does not increase wear
Authors
Matthias Woiczinski
Christian Schröder
Alexander Paulus
Manuel Kistler
Volkmar Jansson
Peter E. Müller
Patrick Weber
Publication date
01-09-2020
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 9/2020
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-019-05761-3

Other articles of this Issue 9/2020

Knee Surgery, Sports Traumatology, Arthroscopy 9/2020 Go to the issue