Published in: European Radiology 11/2019

Open Access 01-11-2019 | Ultrasound | Ultrasound

Intra- and inter-operator reproducibility of US point shear-wave elastography in various organs: evaluation in phantoms and healthy volunteers

Authors: Riwa Kishimoto, Katsuhiko Kikuchi, Atsuhisa Koyama, Jeff Kershaw, Tokuhiko Omatsu, Yasuhiko Tachibana, Mikio Suga, Takayuki Obata

Published in: European Radiology | Issue 11/2019

Login to get access

Abstract

Purpose

This study was conducted in order to assess the intra- and interoperator reproducibility of shear-wave speed (SWS) measurement on elasticity phantoms and healthy volunteers using ultrasound-based point shear-wave elastography.

Materials and methods

This study was approved by the institutional review board. Two operators measured the SWS of five elasticity phantoms and seven organs (thyroid, lymph node, muscle, spleen, kidney, pancreas, and liver) of 30 healthy volunteers with 1.0–4.5 MHz convex (4C1) and 4.0–9.0 MHz linear (9L4) transducers. The phantom measurements were repeated ten times, while the volunteer measurements were performed five times each. Intra- and interoperator reproducibility was assessed. Interoperator reproducibility was also evaluated with the 95% Bland–Altman limits of agreement (LOA).

Results

In phantoms, all intraclass correlation coefficients (ICCs) were above 0.90 and the 95% LOA between the two operators were less than ± 18%. In volunteers, intraoperator ICCs were > 0.75 for all regions except the pancreas. Interoperator ICC was above 0.75 for the right lobe of the liver (depth 4 cm) and the kidney, but the 95% LOA was less than ± 25% only for the liver.

Conclusion

Although excellent in phantoms, interoperator reproducibility was insufficient for all regions in the volunteers other than the right hepatic lobe at a depth of 4 cm. Clinicians should be aware of the 95% LOA when using SWS in patients.

Key Points

• Our phantom study indicated a high reproducibility for shear-wave speed (SWS) measurements with point shear-wave elastography (pSWE).
• In volunteers, intraoperator reproducibility was generally high, but the interoperator reproducibility was not high enough except for the right hepatic lobe at 4 cm depth.
• To evaluate interoperator reproducibility, the 95% limits of agreement (LOA) between operators should be considered in addition to the intraclass correlation coefficient (ICC).
Appendix
Available only for authorised users
Literature
1.
go back to reference Ferraioli G, Filice C, Castera L et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 3: liver. Ultrasound Med Biol 41(5):1161–1179PubMedCrossRef Ferraioli G, Filice C, Castera L et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 3: liver. Ultrasound Med Biol 41(5):1161–1179PubMedCrossRef
2.
go back to reference Barr RG, Nakashima K, Amy D et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 2: breast. Ultrasound Med Biol 41(5):1148–1160PubMedCrossRef Barr RG, Nakashima K, Amy D et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 2: breast. Ultrasound Med Biol 41(5):1148–1160PubMedCrossRef
3.
go back to reference Cosgrove D, Barr R, Bojunga J et al (2017) WFUMB guidelines and recommendations on the clinical use of ultrasound elastography: part 4. Thyroid. Ultrasound Med Biol 43(1):4–26PubMedCrossRef Cosgrove D, Barr R, Bojunga J et al (2017) WFUMB guidelines and recommendations on the clinical use of ultrasound elastography: part 4. Thyroid. Ultrasound Med Biol 43(1):4–26PubMedCrossRef
4.
go back to reference Barr RG, Cosgrove D, Brock M et al (2017) WFUMB guidelines and recommendations on the clinical use of ultrasound elastography: part 5. Prostate. Ultrasound Med Biol 43(1):27–48PubMedCrossRef Barr RG, Cosgrove D, Brock M et al (2017) WFUMB guidelines and recommendations on the clinical use of ultrasound elastography: part 5. Prostate. Ultrasound Med Biol 43(1):27–48PubMedCrossRef
5.
go back to reference Shiina T, Nightingale KR, Palmeri ML et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 1: basic principles and terminology. Ultrasound Med Biol 41(5):1126–1147PubMedCrossRef Shiina T, Nightingale KR, Palmeri ML et al (2015) WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 1: basic principles and terminology. Ultrasound Med Biol 41(5):1126–1147PubMedCrossRef
7.
go back to reference Sandrin L, Fourquet B, Hasquenoph JM et al (2003) Transient elastography: a new noninvasive method for assessment of hepatic fibrosis. Ultrasound Med Biol 29(12):1705–1713PubMedCrossRef Sandrin L, Fourquet B, Hasquenoph JM et al (2003) Transient elastography: a new noninvasive method for assessment of hepatic fibrosis. Ultrasound Med Biol 29(12):1705–1713PubMedCrossRef
8.
go back to reference Barr RG, Ferraioli G, Palmeri ML et al (2015) Elastography assessment of liver fibrosis: Society of Radiologists in ultrasound consensus conference statement. Radiology 276(3):845–861PubMedCrossRef Barr RG, Ferraioli G, Palmeri ML et al (2015) Elastography assessment of liver fibrosis: Society of Radiologists in ultrasound consensus conference statement. Radiology 276(3):845–861PubMedCrossRef
9.
go back to reference Tang A, Cloutier G, Szeverenyi NM, Sirlin CB (2015) Ultrasound elastography and MR elastography for assessing liver fibrosis: part 1, principles and techniques. AJR Am J Roentgenol 205(1):22–32PubMedPubMedCentralCrossRef Tang A, Cloutier G, Szeverenyi NM, Sirlin CB (2015) Ultrasound elastography and MR elastography for assessing liver fibrosis: part 1, principles and techniques. AJR Am J Roentgenol 205(1):22–32PubMedPubMedCentralCrossRef
10.
go back to reference Bercoff J, Tanter M, Muller M, Fink M (2004) The role of viscosity in the impulse diffraction field of elastic waves induced by the acoustic radiation force. IEEE Trans Ultrason Ferroelectr Freq Control 51(11):1523–1536PubMedCrossRef Bercoff J, Tanter M, Muller M, Fink M (2004) The role of viscosity in the impulse diffraction field of elastic waves induced by the acoustic radiation force. IEEE Trans Ultrason Ferroelectr Freq Control 51(11):1523–1536PubMedCrossRef
11.
go back to reference Friedrich-Rust M, Wunder K, Kriener S et al (2009) Liver fibrosis in viral hepatitis: noninvasive assessment with acoustic radiation force impulse imaging versus transient elastography. Radiology 252(2):595–604PubMedCrossRef Friedrich-Rust M, Wunder K, Kriener S et al (2009) Liver fibrosis in viral hepatitis: noninvasive assessment with acoustic radiation force impulse imaging versus transient elastography. Radiology 252(2):595–604PubMedCrossRef
12.
go back to reference Woo H, Lee JY, Yoon JH, Kim W, Cho B, Choi BI (2015) Comparison of the reliability of acoustic radiation force impulse imaging and supersonic shear imaging in measurement of liver stiffness. Radiology 277(3):881–886PubMedCrossRef Woo H, Lee JY, Yoon JH, Kim W, Cho B, Choi BI (2015) Comparison of the reliability of acoustic radiation force impulse imaging and supersonic shear imaging in measurement of liver stiffness. Radiology 277(3):881–886PubMedCrossRef
13.
go back to reference Van Beers BE, Daire JL, Garteiser P (2015) New imaging techniques for liver diseases. J Hepatol 62(3):690–700PubMedCrossRef Van Beers BE, Daire JL, Garteiser P (2015) New imaging techniques for liver diseases. J Hepatol 62(3):690–700PubMedCrossRef
14.
go back to reference Tozaki M, Saito M, Benson J, Fan L, Isobe S (2013) Shear wave velocity measurements for differential diagnosis of solid breast masses: a comparison between virtual touch quantification and virtual touch IQ. Ultrasound Med Biol 39(12):2233–2245PubMedCrossRef Tozaki M, Saito M, Benson J, Fan L, Isobe S (2013) Shear wave velocity measurements for differential diagnosis of solid breast masses: a comparison between virtual touch quantification and virtual touch IQ. Ultrasound Med Biol 39(12):2233–2245PubMedCrossRef
15.
go back to reference Samir AE, Dhyani M, Anvari A et al (2015) Shear-wave elastography for the preoperative risk stratification of follicular-patterned lesions of the thyroid: diagnostic accuracy and optimal measurement plane. Radiology 277(2):565–573PubMedCrossRef Samir AE, Dhyani M, Anvari A et al (2015) Shear-wave elastography for the preoperative risk stratification of follicular-patterned lesions of the thyroid: diagnostic accuracy and optimal measurement plane. Radiology 277(2):565–573PubMedCrossRef
16.
go back to reference Kawada N, Tanaka S, Uehara H et al (2014) Potential use of point shear wave elastography for the pancreas: a single center prospective study. Eur J Radiol 83(4):620–624PubMedCrossRef Kawada N, Tanaka S, Uehara H et al (2014) Potential use of point shear wave elastography for the pancreas: a single center prospective study. Eur J Radiol 83(4):620–624PubMedCrossRef
17.
go back to reference Guo LH, Xu HX, Fu HJ, Peng A, Zhang YF, Liu LN (2013) Acoustic radiation force impulse imaging for noninvasive evaluation of renal parenchyma elasticity: preliminary findings. PLoS One 8(7):e68925PubMedPubMedCentralCrossRef Guo LH, Xu HX, Fu HJ, Peng A, Zhang YF, Liu LN (2013) Acoustic radiation force impulse imaging for noninvasive evaluation of renal parenchyma elasticity: preliminary findings. PLoS One 8(7):e68925PubMedPubMedCentralCrossRef
18.
go back to reference Lu Q, Wen JX, Huang BJ, Xue LY, Wang WP (2015) Virtual touch quantification using acoustic radiation force impulse (ARFI) technology for the evaluation of focal solid renal lesions: preliminary findings. Clin Radiol 70(12):1376–1381PubMedCrossRef Lu Q, Wen JX, Huang BJ, Xue LY, Wang WP (2015) Virtual touch quantification using acoustic radiation force impulse (ARFI) technology for the evaluation of focal solid renal lesions: preliminary findings. Clin Radiol 70(12):1376–1381PubMedCrossRef
19.
go back to reference Alan B, Utangac M, Goya C, Daggulli M (2016) Role of acoustic radiation force impulse (ARFI) elastography in determination of severity of benign prostate hyperplasia. Med Sci Monit 22:4523–4528PubMedPubMedCentralCrossRef Alan B, Utangac M, Goya C, Daggulli M (2016) Role of acoustic radiation force impulse (ARFI) elastography in determination of severity of benign prostate hyperplasia. Med Sci Monit 22:4523–4528PubMedPubMedCentralCrossRef
20.
go back to reference Fujiwara T, Tomokuni J, Iwanaga K, Ooba S, Haji T (2013) Acoustic radiation force impulse imaging for reactive and malignant/metastatic cervical lymph nodes. Ultrasound Med Biol 39(7):1178–1183PubMedCrossRef Fujiwara T, Tomokuni J, Iwanaga K, Ooba S, Haji T (2013) Acoustic radiation force impulse imaging for reactive and malignant/metastatic cervical lymph nodes. Ultrasound Med Biol 39(7):1178–1183PubMedCrossRef
21.
go back to reference Rosskopf AB, Ehrmann C, Buck FM, Gerber C, Fluck M, Pfirrmann CW (2016) Quantitative shear-wave US elastography of the supraspinatus muscle: reliability of the method and relation to tendon integrity and muscle quality. Radiology 278(2):465–474PubMedCrossRef Rosskopf AB, Ehrmann C, Buck FM, Gerber C, Fluck M, Pfirrmann CW (2016) Quantitative shear-wave US elastography of the supraspinatus muscle: reliability of the method and relation to tendon integrity and muscle quality. Radiology 278(2):465–474PubMedCrossRef
22.
go back to reference Suda T, Okawa O, Masaoka R et al (2017) Shear wave elastography in hepatitis C patients before and after antiviral therapy. World J Hepatol 9(1):64–68PubMedPubMedCentralCrossRef Suda T, Okawa O, Masaoka R et al (2017) Shear wave elastography in hepatitis C patients before and after antiviral therapy. World J Hepatol 9(1):64–68PubMedPubMedCentralCrossRef
23.
go back to reference Ma Y, Zhang S, Li J, Li J, Kang Y, Ren W (2017) Comparison of strain and shear-wave ultrasounic elastography in predicting the pathological response to neoadjuvant chemotherapy in breast cancers. Eur Radiol 27(6):2282–2291PubMedCrossRef Ma Y, Zhang S, Li J, Li J, Kang Y, Ren W (2017) Comparison of strain and shear-wave ultrasounic elastography in predicting the pathological response to neoadjuvant chemotherapy in breast cancers. Eur Radiol 27(6):2282–2291PubMedCrossRef
24.
go back to reference Ahn SJ, Lee JM, Chang W et al (2017) Prospective validation of intra- and interobserver reproducibility of a new point shear wave elastographic technique for assessing liver stiffness in patients with chronic liver disease. Korean J Radiol 18(6):926–935PubMedPubMedCentralCrossRef Ahn SJ, Lee JM, Chang W et al (2017) Prospective validation of intra- and interobserver reproducibility of a new point shear wave elastographic technique for assessing liver stiffness in patients with chronic liver disease. Korean J Radiol 18(6):926–935PubMedPubMedCentralCrossRef
25.
go back to reference Cosgrove DO, Berg WA, Doré CJ et al (2012) BE1 Study Group. Shear wave elastography for breast masses is highly reproducible. Eur Radiol 22(5):1023–1032PubMedCrossRef Cosgrove DO, Berg WA, Doré CJ et al (2012) BE1 Study Group. Shear wave elastography for breast masses is highly reproducible. Eur Radiol 22(5):1023–1032PubMedCrossRef
26.
go back to reference Anvari A, Dhyani M, Stephen AE, Samir AE (2016) Reliability of shear-wave elastography estimates of the Young modulus of tissue in follicular thyroid neoplasms. Am J Roentgenol 206(3):609–1627CrossRef Anvari A, Dhyani M, Stephen AE, Samir AE (2016) Reliability of shear-wave elastography estimates of the Young modulus of tissue in follicular thyroid neoplasms. Am J Roentgenol 206(3):609–1627CrossRef
27.
go back to reference Kishimoto R, Suga M, Koyama A, et al (2017) Measuring shear-wave speed with point shear-wave elastography and MR elastography: a phantom study. BMJ Open 7(1):e013925 Kishimoto R, Suga M, Koyama A, et al (2017) Measuring shear-wave speed with point shear-wave elastography and MR elastography: a phantom study. BMJ Open 7(1):e013925
28.
go back to reference Vetter NC, Steding J, Jurk S, Ripke S, Mennigen E, Smolka MN (2017) Reliability in adolescent fMRI within two years—a comparison of three tasks. Sci Rep 7(1):2287PubMedPubMedCentralCrossRef Vetter NC, Steding J, Jurk S, Ripke S, Mennigen E, Smolka MN (2017) Reliability in adolescent fMRI within two years—a comparison of three tasks. Sci Rep 7(1):2287PubMedPubMedCentralCrossRef
29.
go back to reference Park HS, Kim YJ, Yu MH, Jung SI, Jeon HJ (2015) Shear wave elastography of focal liver lesion: intraobserver reproducibility and elasticity characterization. Ultrasound Q 31(4):262–271PubMedCrossRef Park HS, Kim YJ, Yu MH, Jung SI, Jeon HJ (2015) Shear wave elastography of focal liver lesion: intraobserver reproducibility and elasticity characterization. Ultrasound Q 31(4):262–271PubMedCrossRef
30.
go back to reference Marticorena Garcia SR, Grossmann M, Lang ST et al (2018) Full-field-of-view time-harmonic elastography of the native kidney. Ultrasound Med Biol 44(5):949–954PubMedCrossRef Marticorena Garcia SR, Grossmann M, Lang ST et al (2018) Full-field-of-view time-harmonic elastography of the native kidney. Ultrasound Med Biol 44(5):949–954PubMedCrossRef
31.
go back to reference Bartlett JW, Frost C (2008) Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables. Ultrasound Obstet Gynecol 31(4):466–475PubMedCrossRef Bartlett JW, Frost C (2008) Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables. Ultrasound Obstet Gynecol 31(4):466–475PubMedCrossRef
32.
go back to reference Goertz RS, Amann K, Heide R, Bernatik T, Neurath MF, Strobel D (2011) An abdominal and thyroid status with acoustic radiation force impulse elastometry—a feasibility study: acoustic radiation force impulse elastometry of human organs. Eur J Radiol 80(3):e226–e230PubMedCrossRef Goertz RS, Amann K, Heide R, Bernatik T, Neurath MF, Strobel D (2011) An abdominal and thyroid status with acoustic radiation force impulse elastometry—a feasibility study: acoustic radiation force impulse elastometry of human organs. Eur J Radiol 80(3):e226–e230PubMedCrossRef
33.
go back to reference Bob F, Bota S, Sporea I, Sirli R, Petrica L, Schiller A (2014) Kidney shear wave speed values in subjects with and without renal pathology and inter-operator reproducibility of acoustic radiation force impulse elastography (ARFI)—preliminary results. PLoS One 9(11):e113761PubMedPubMedCentralCrossRef Bob F, Bota S, Sporea I, Sirli R, Petrica L, Schiller A (2014) Kidney shear wave speed values in subjects with and without renal pathology and inter-operator reproducibility of acoustic radiation force impulse elastography (ARFI)—preliminary results. PLoS One 9(11):e113761PubMedPubMedCentralCrossRef
34.
go back to reference Cortez CD, Hermitte L, Ramain A, Mesmann C, Lefort T, Pialat JB (2016) Ultrasound shear wave velocity in skeletal muscle: a reproducibility study. Diagn Interv Imaging 97(1):71–79PubMedCrossRef Cortez CD, Hermitte L, Ramain A, Mesmann C, Lefort T, Pialat JB (2016) Ultrasound shear wave velocity in skeletal muscle: a reproducibility study. Diagn Interv Imaging 97(1):71–79PubMedCrossRef
35.
go back to reference Balakrishnan M, Souza F, Munoz C et al (2016) Liver and spleen stiffness measurements by point shear wave elastography via acoustic radiation force impulse: intraobserver and interobserver variability and predictors of variability in a US population. J Ultrasound Med 35(11):2373–2380PubMedCrossRef Balakrishnan M, Souza F, Munoz C et al (2016) Liver and spleen stiffness measurements by point shear wave elastography via acoustic radiation force impulse: intraobserver and interobserver variability and predictors of variability in a US population. J Ultrasound Med 35(11):2373–2380PubMedCrossRef
36.
go back to reference Bota S, Sporea I, Sirli R, Popescu A, Danila M, Costachescu D (2012) Intra- and interoperator reproducibility of acoustic radiation force impulse (ARFI) elastography—preliminary results. Ultrasound Med Biol 38(7):1103–1108PubMedCrossRef Bota S, Sporea I, Sirli R, Popescu A, Danila M, Costachescu D (2012) Intra- and interoperator reproducibility of acoustic radiation force impulse (ARFI) elastography—preliminary results. Ultrasound Med Biol 38(7):1103–1108PubMedCrossRef
37.
go back to reference Choi SH, Jeong WK, Kim Y et al (2016) How many times should we repeat measuring liver stiffness using shear wave elastography?: 5-repetition versus 10-repetition protocols. Ultrasonics 72:158–164PubMedCrossRef Choi SH, Jeong WK, Kim Y et al (2016) How many times should we repeat measuring liver stiffness using shear wave elastography?: 5-repetition versus 10-repetition protocols. Ultrasonics 72:158–164PubMedCrossRef
38.
go back to reference Ferraioli G, Maiocchi L, Lissandrin R, Tinelli C, De Silvestri A, Filice C (2016) Accuracy of the Elast PQ technique for the assessment of liver fibrosis in patients with chronic hepatitis C: a “real life” single center study. J Gastrointestin Liver Dis 25(3):331–335PubMed Ferraioli G, Maiocchi L, Lissandrin R, Tinelli C, De Silvestri A, Filice C (2016) Accuracy of the Elast PQ technique for the assessment of liver fibrosis in patients with chronic hepatitis C: a “real life” single center study. J Gastrointestin Liver Dis 25(3):331–335PubMed
39.
go back to reference Yoon JH, Lee JM, Han JK, Choi BI (2014) Shear wave elastography for liver stiffness measurement in clinical sonographic examinations: evaluation of intraobserver reproducibility, technical failure, and unreliable stiffness measurements. J Ultrasound Med 33(3):437–447PubMedCrossRef Yoon JH, Lee JM, Han JK, Choi BI (2014) Shear wave elastography for liver stiffness measurement in clinical sonographic examinations: evaluation of intraobserver reproducibility, technical failure, and unreliable stiffness measurements. J Ultrasound Med 33(3):437–447PubMedCrossRef
40.
go back to reference Fang C, Jaffer OS, Yusuf GT et al (2018) Reducing the number of measurements in liver point shear-wave elastography: factors that influence the number and reliability of measurements in assessment of liver fibrosis in clinical practice. Radiology 287(3):844–852PubMedCrossRef Fang C, Jaffer OS, Yusuf GT et al (2018) Reducing the number of measurements in liver point shear-wave elastography: factors that influence the number and reliability of measurements in assessment of liver fibrosis in clinical practice. Radiology 287(3):844–852PubMedCrossRef
41.
go back to reference Trout AT, Dillman JR, Xanthakos S et al (2016) Prospective assessment of correlation between US acoustic radiation force impulse and MR elastography in a pediatric population: dispersion of US shear-wave speed measurement matters. Radiology 281(2):544–552PubMedCrossRef Trout AT, Dillman JR, Xanthakos S et al (2016) Prospective assessment of correlation between US acoustic radiation force impulse and MR elastography in a pediatric population: dispersion of US shear-wave speed measurement matters. Radiology 281(2):544–552PubMedCrossRef
42.
go back to reference Dillman JR, Chen S, Davenport MS et al (2015) Superficial ultrasound shear wave speed measurements in soft and hard elasticity phantoms: repeatability and reproducibility using two ultrasound systems. Pediatr Radiol 45(3):376–385PubMedCrossRef Dillman JR, Chen S, Davenport MS et al (2015) Superficial ultrasound shear wave speed measurements in soft and hard elasticity phantoms: repeatability and reproducibility using two ultrasound systems. Pediatr Radiol 45(3):376–385PubMedCrossRef
Metadata
Title
Intra- and inter-operator reproducibility of US point shear-wave elastography in various organs: evaluation in phantoms and healthy volunteers
Authors
Riwa Kishimoto
Katsuhiko Kikuchi
Atsuhisa Koyama
Jeff Kershaw
Tokuhiko Omatsu
Yasuhiko Tachibana
Mikio Suga
Takayuki Obata
Publication date
01-11-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 11/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06195-8

Other articles of this Issue 11/2019

European Radiology 11/2019 Go to the issue