Published in:
17-05-2022 | Ultrasound | Musculoskeletal
Incidence of infectious complications following ultrasound-guided percutaneous musculoskeletal interventions with the use of an uncovered transducer footprint
Authors:
Natalia Gorelik, Yousef Darwish, William R. Walter, Christopher J. Burke, Dost Sarpel, Jaron Chong, Ronald S. Adler
Published in:
European Radiology
|
Issue 10/2022
Login to get access
Abstract
Objectives
To determine the incidence of infectious complications following ultrasound-guided musculoskeletal interventions performed with a disinfected uncovered ultrasound transducer footprint.
Methods
Electronic medical records of all patients who underwent an ultrasound-guided musculoskeletal procedure (including injection, calcific lavage, or ganglion cyst aspiration) performed by any of the 14 interventional musculoskeletal radiologists at our institution between January 2013 and December 2018 were retrospectively reviewed to identify procedure site infections. Biopsies and joint aspirations were excluded. The procedures were performed using a disinfected uncovered transducer footprint. First, an automated chart review identified cases with (1) positive answers to the nurse’s post-procedure call, (2) an International Classification of Diseases (ICD) diagnostic code related to a musculoskeletal infection, or (3) an antibiotic prescription within 30 days post-procedure. Then, these cases were manually reviewed for evidence of procedure site infection.
Results
In total, 6511 procedures were included. The automated chart review identified 3 procedures (2 patients) in which post-procedural fever was reported during the nurse’s post-procedure call, 33 procedures (28 patients) with an ICD code for a musculoskeletal infection, and 220 procedures (216 patients) with an antibiotic prescription within 30 post-procedural days. The manual chart review of these patients revealed no cases of confirmed infection and 1 case (0.015%) of possible site infection.
Conclusions
The incidence of infectious complications after an ultrasound-guided musculoskeletal procedure performed with an uncovered transducer footprint is extremely low. This information allows radiologists to counsel their patients more precisely when obtaining informed consent.
Key Points
• Infectious complications after ultrasound-guided musculoskeletal procedures performed with a disinfected uncovered transducer footprint are extremely rare.