Skip to main content
Top
Published in: Surgical Endoscopy 5/2014

01-05-2014 | New Technology

Ultrasound-based tumor movement compensation during navigated laparoscopic liver interventions

Authors: Osama Shahin, Armin Beširević, Markus Kleemann, Alexander Schlaefer

Published in: Surgical Endoscopy | Issue 5/2014

Login to get access

Abstract

Background

Image-guided navigation aims to provide better orientation and accuracy in laparoscopic interventions. However, the ability of the navigation system to reflect anatomical changes and maintain high accuracy during the procedure is crucial. This is particularly challenging in soft organs such as the liver, where surgical manipulation causes significant tumor movements. We propose a fast approach to obtain an accurate estimation of the tumor position throughout the procedure.

Methods

Initially, a three-dimensional (3D) ultrasound image is reconstructed and the tumor is segmented. During surgery, the position of the tumor is updated based on newly acquired tracked ultrasound images. The initial segmentation of the tumor is used to automatically detect the tumor and update its position in the navigation system. Two experiments were conducted. First, a controlled phantom motion using a robot was performed to validate the tracking accuracy. Second, a needle navigation scenario based on pseudotumors injected into ex vivo porcine liver was studied.

Result

In the robot-based evaluation, the approach estimated the target location with an accuracy of 0.4 ± 0.3 mm. The mean navigation error in the needle experiment was 1.2 ± 0.6 mm, and the algorithm compensated for tumor shifts up to 38 mm in an average time of 1 s.

Conclusion

We demonstrated a navigation approach based on tracked laparoscopic ultrasound (LUS), and focused on the neighborhood of the tumor. Our experimental results indicate that this approach can be used to quickly and accurately compensate for tumor movements caused by surgical manipulation during laparoscopic interventions. The proposed approach has the advantage of being based on the routinely used LUS; however, it upgrades its functionality to estimate the tumor position in 3D. Hence, the approach is repeatable throughout surgery, and enables high navigation accuracy to be maintained.
Literature
1.
go back to reference Rao A, Rao G, Ahmed I (2012) Laparoscopic vs. open liver resection for malignant liver disease: a systematic review. Surgeon 10(4):194–201PubMedCrossRef Rao A, Rao G, Ahmed I (2012) Laparoscopic vs. open liver resection for malignant liver disease: a systematic review. Surgeon 10(4):194–201PubMedCrossRef
2.
go back to reference Bechstein W (2007) Which liver metastases are resectable? Eur J Cancer Suppl 5(5):301–306CrossRef Bechstein W (2007) Which liver metastases are resectable? Eur J Cancer Suppl 5(5):301–306CrossRef
3.
go back to reference Nguyen KT, Marsh JW, Tsung A, Steel JJ, Gamblin TC, Geller DA (2011) Comparative benefits of laparoscopic vs. open hepatic resection: a critical appraisal. Arch Surg 146:348–356PubMedCrossRef Nguyen KT, Marsh JW, Tsung A, Steel JJ, Gamblin TC, Geller DA (2011) Comparative benefits of laparoscopic vs. open hepatic resection: a critical appraisal. Arch Surg 146:348–356PubMedCrossRef
4.
go back to reference Edwin B, Nordin A, Kazaryan A et al (2011) Laparoscopic liver surgery: new frontiers. Scand J Surg 100(1):54PubMed Edwin B, Nordin A, Kazaryan A et al (2011) Laparoscopic liver surgery: new frontiers. Scand J Surg 100(1):54PubMed
5.
go back to reference Cannon RM, Brock GN, Marvin MR, Buell JF (2011) Laparoscopic liver resection: an examination of our first 300 patients. J Am Coll Surg 213:501–507PubMedCrossRef Cannon RM, Brock GN, Marvin MR, Buell JF (2011) Laparoscopic liver resection: an examination of our first 300 patients. J Am Coll Surg 213:501–507PubMedCrossRef
6.
go back to reference Westebring-van der Putten E, Goossens R, Jakimowicz J, Dankelman J (2008) Haptics in minimally invasive surgery: a review. Minim Invasive Ther Allied Technol 17(1):3–16PubMedCrossRef Westebring-van der Putten E, Goossens R, Jakimowicz J, Dankelman J (2008) Haptics in minimally invasive surgery: a review. Minim Invasive Ther Allied Technol 17(1):3–16PubMedCrossRef
7.
go back to reference Solberg OV, Langø T, Tangen GA, Mårvik R, Ystgaard B, Rethy A, Hernes TAN (2009) Navigated ultrasound in laparoscopic surgery. Minim Invasive Ther Allied Technol 18(1):36–53PubMedCrossRef Solberg OV, Langø T, Tangen GA, Mårvik R, Ystgaard B, Rethy A, Hernes TAN (2009) Navigated ultrasound in laparoscopic surgery. Minim Invasive Ther Allied Technol 18(1):36–53PubMedCrossRef
8.
go back to reference Martens V, Beširević A, Shahin O, Kleemann M (2010) LapAssistent: computer assisted laparoscopic liver surgery. In: Proceedings of biomedizinischen technik (BMT) conference Martens V, Beširević A, Shahin O, Kleemann M (2010) LapAssistent: computer assisted laparoscopic liver surgery. In: Proceedings of biomedizinischen technik (BMT) conference
9.
go back to reference Rauth TP, Bao PQ, Galloway RL, Bieszczad J, Friets EM, Knaus DA, Kynor DB, Herline AJ et al (2007) Laparoscopic surface scanning and subsurface targeting: implications for image-guided laparoscopic liver surgery. Surgery 142(2):207–214PubMedCrossRef Rauth TP, Bao PQ, Galloway RL, Bieszczad J, Friets EM, Knaus DA, Kynor DB, Herline AJ et al (2007) Laparoscopic surface scanning and subsurface targeting: implications for image-guided laparoscopic liver surgery. Surgery 142(2):207–214PubMedCrossRef
10.
go back to reference Olesch J, Beuthien B, Heldmann S, Papenberg N, Fischer B (2011) Fast intra-operative non-linear registration of 3D-CT to tracked, selected 2D-ultrasound slices. In: Proceedings of SPIE medical imaging: 79642R Olesch J, Beuthien B, Heldmann S, Papenberg N, Fischer B (2011) Fast intra-operative non-linear registration of 3D-CT to tracked, selected 2D-ultrasound slices. In: Proceedings of SPIE medical imaging: 79642R
11.
go back to reference Herickhoff CD, Light ED, Bing KF, Mukundan S, Grant GA, Wolf PD, Smith SW (2009) Dual-mode intracranial catheter integrating 3D ultrasound imaging and hyperthermia for neuro-oncology: feasibility study. Ultrason Imaging 31:81–100PubMedCentralPubMedCrossRef Herickhoff CD, Light ED, Bing KF, Mukundan S, Grant GA, Wolf PD, Smith SW (2009) Dual-mode intracranial catheter integrating 3D ultrasound imaging and hyperthermia for neuro-oncology: feasibility study. Ultrason Imaging 31:81–100PubMedCentralPubMedCrossRef
12.
go back to reference Light ED, Idriss SF, Sullivan KF, Wolf PD, Smith SW (2005) Real-time 3D laparoscopic ultrasonography. Ultrason Imaging 27:129–144PubMedCrossRef Light ED, Idriss SF, Sullivan KF, Wolf PD, Smith SW (2005) Real-time 3D laparoscopic ultrasonography. Ultrason Imaging 27:129–144PubMedCrossRef
13.
go back to reference Solberg OV, Lindseth F, Torp H, Blake RE, Nagelhus Hernes TA (2007) Freehand 3D ultrasound reconstruction algorithms: a review. Ultrasound Med Biol 33:991–1009PubMedCrossRef Solberg OV, Lindseth F, Torp H, Blake RE, Nagelhus Hernes TA (2007) Freehand 3D ultrasound reconstruction algorithms: a review. Ultrasound Med Biol 33:991–1009PubMedCrossRef
14.
go back to reference Bao P, Sinha TK, Chen CC, Warmath JR, Galloway RL, Herline AJ (2007) A prototype ultrasound-guided laparoscopic radiofrequency ablation system. Surg Endosc 21:74–79PubMedCrossRef Bao P, Sinha TK, Chen CC, Warmath JR, Galloway RL, Herline AJ (2007) A prototype ultrasound-guided laparoscopic radiofrequency ablation system. Surg Endosc 21:74–79PubMedCrossRef
15.
go back to reference Boctor E, Fischer G, Choti M, Fichtinger G, Taylor R (2004) A dual-armed robotic system for intraoperative ultrasound guided hepatic ablative therapy: a prospective study. In: Proceedings of 2004 IEEE international conference on robotics and automation, vol 3. pp 2517–2522 Boctor E, Fischer G, Choti M, Fichtinger G, Taylor R (2004) A dual-armed robotic system for intraoperative ultrasound guided hepatic ablative therapy: a prospective study. In: Proceedings of 2004 IEEE international conference on robotics and automation, vol 3. pp 2517–2522
16.
go back to reference Sjølie E, Langø T, Ystgaard B, Tangen GA, Nagelhus Hernes TA, Mårvik R (2003) 3D ultrasound-based navigation for radiofrequency thermal ablation in the treatment of liver malignancies. Surg Endosc 17:933–938PubMedCrossRef Sjølie E, Langø T, Ystgaard B, Tangen GA, Nagelhus Hernes TA, Mårvik R (2003) 3D ultrasound-based navigation for radiofrequency thermal ablation in the treatment of liver malignancies. Surg Endosc 17:933–938PubMedCrossRef
17.
go back to reference Harms J, Feussner H, Baumgartner M, Schneider A, Donhauser M, Wessels G (2001) Three-dimensional navigated laparoscopic ultrasonography. Surg Endosc 15:1459–1462PubMedCrossRef Harms J, Feussner H, Baumgartner M, Schneider A, Donhauser M, Wessels G (2001) Three-dimensional navigated laparoscopic ultrasonography. Surg Endosc 15:1459–1462PubMedCrossRef
18.
go back to reference Shahin O, Martens V, Beširević A, Kleemann M, Schlaefer A (2012) Localization of liver tumors in freehand 3D laparoscopic ultrasound. In: Proceedings of SPIE medical imaging 83162C–83162C–7 Shahin O, Martens V, Beširević A, Kleemann M, Schlaefer A (2012) Localization of liver tumors in freehand 3D laparoscopic ultrasound. In: Proceedings of SPIE medical imaging 83162C–83162C–7
19.
go back to reference Scott DJMD, Young WN, Watumull LM, Lindberg G, Fleming JB, Rege RV, Brown RJ, Jones DB (2000) Development of an in vivo tumor-mimic model for learning radiofrequency ablation. J Gastrointest Surg 4(6):620–625CrossRef Scott DJMD, Young WN, Watumull LM, Lindberg G, Fleming JB, Rege RV, Brown RJ, Jones DB (2000) Development of an in vivo tumor-mimic model for learning radiofrequency ablation. J Gastrointest Surg 4(6):620–625CrossRef
20.
go back to reference Chen T K, Thurston AD, Moghari MH, Ellis RE, Abolmaesumi P (2008) A real-time ultrasound calibration system with automatic accuracy control and incorporation of ultrasound section thickness. In: Proceedings of SPIE medical imaging 69182A–69182A–11 Chen T K, Thurston AD, Moghari MH, Ellis RE, Abolmaesumi P (2008) A real-time ultrasound calibration system with automatic accuracy control and incorporation of ultrasound section thickness. In: Proceedings of SPIE medical imaging 69182A–69182A–11
21.
go back to reference Pagoulatos N, Haynor DR, Kim Y (2001) A fast calibration method for 3-D tracking of ultrasound images using a spatial localizer. Ultrasound Med Biol 27:1219–1229PubMedCrossRef Pagoulatos N, Haynor DR, Kim Y (2001) A fast calibration method for 3-D tracking of ultrasound images using a spatial localizer. Ultrasound Med Biol 27:1219–1229PubMedCrossRef
22.
go back to reference Gobbi DG, Peters TM (2002) Interactive intra-operative 3D ultrasound reconstruction and visualization. In: Medical image computing and computer-assisted intervention MICCAI. Springer, Berlin, pp 156–163 Gobbi DG, Peters TM (2002) Interactive intra-operative 3D ultrasound reconstruction and visualization. In: Medical image computing and computer-assisted intervention MICCAI. Springer, Berlin, pp 156–163
23.
go back to reference Noble JA, Boukerroui D (2006) Ultrasound image segmentation: a survey. IEEE Trans Med Imaging 25:987–1010PubMedCrossRef Noble JA, Boukerroui D (2006) Ultrasound image segmentation: a survey. IEEE Trans Med Imaging 25:987–1010PubMedCrossRef
24.
go back to reference Caselles V, Kimmel R, Sapiro G (1995) Geodesic active contours. Int J Computer Vis 22:61–79CrossRef Caselles V, Kimmel R, Sapiro G (1995) Geodesic active contours. Int J Computer Vis 22:61–79CrossRef
25.
go back to reference Ibanez L, Schroeder W, Ng L, Cates J (2003) The ITK software guide, 1st edn. Kitware Inc., Prentice-Hall. ISBN 1-930934-10-6 Ibanez L, Schroeder W, Ng L, Cates J (2003) The ITK software guide, 1st edn. Kitware Inc., Prentice-Hall. ISBN 1-930934-10-6
26.
go back to reference Mattes D, Haynor DR, Vesselle H, Lewellyn TK, Eubank W (2001) Nonrigid multimodality image registration. In: Proceedings of SPIE medical imaging. pp 1609–1620 Mattes D, Haynor DR, Vesselle H, Lewellyn TK, Eubank W (2001) Nonrigid multimodality image registration. In: Proceedings of SPIE medical imaging. pp 1609–1620
27.
go back to reference Yoo TS (ed) (2004) Insight into images: principles and practice for segmentation, registration, and image analysis. A K Peters/CRC Press, Natick Yoo TS (ed) (2004) Insight into images: principles and practice for segmentation, registration, and image analysis. A K Peters/CRC Press, Natick
28.
go back to reference Schroeder WJ, Martin K, Lorensen W (2003) The visualization toolkit: an object-oriented approach to 3D graphics, 3rd edn. Kitware Inc., Prentice-Hall Schroeder WJ, Martin K, Lorensen W (2003) The visualization toolkit: an object-oriented approach to 3D graphics, 3rd edn. Kitware Inc., Prentice-Hall
29.
go back to reference Maurer CR, Fitzpatrick JM, Wang MY, Member S, Galloway RL, Maciunas RJ, Allen GS (1997) Registration of head volume images using implantable fiducial markers. IEEE Trans Med Imaging 16:447–462PubMedCrossRef Maurer CR, Fitzpatrick JM, Wang MY, Member S, Galloway RL, Maciunas RJ, Allen GS (1997) Registration of head volume images using implantable fiducial markers. IEEE Trans Med Imaging 16:447–462PubMedCrossRef
30.
go back to reference Powell MJD (1964) An efficient method for finding the minimum of a function of several variables without calculating derivatives. Comput J 7:155–162CrossRef Powell MJD (1964) An efficient method for finding the minimum of a function of several variables without calculating derivatives. Comput J 7:155–162CrossRef
31.
go back to reference Wein W, Cheng J-Z, Khamene A (2008) Ultrasound based respiratory motion compensation in the abdomen. In: MICCAI, workshop on image guidance and computer assistance for soft-tissue interventions. Springer, New York, p 294 Wein W, Cheng J-Z, Khamene A (2008) Ultrasound based respiratory motion compensation in the abdomen. In: MICCAI, workshop on image guidance and computer assistance for soft-tissue interventions. Springer, New York, p 294
32.
go back to reference Nakamoto M, Hirayama H, Sato Y, Konishi K, Kakeji Y, Hashizume M, Tamura S (2007) Recovery of respiratory motion and deformation of the liver using laparoscopic freehand 3D ultrasound system. Med Image Anal 11(5):429–442PubMedCrossRef Nakamoto M, Hirayama H, Sato Y, Konishi K, Kakeji Y, Hashizume M, Tamura S (2007) Recovery of respiratory motion and deformation of the liver using laparoscopic freehand 3D ultrasound system. Med Image Anal 11(5):429–442PubMedCrossRef
33.
go back to reference Zijlmans M, Langø T, Hofstad EF, Van Swol CF, Rethy A (2012) Navigated laparoscopy-liver shift and deformation due to pneumoperitoneum in an animal model. Minim Invasive Ther Allied Technol 21(3):241–248PubMedCrossRef Zijlmans M, Langø T, Hofstad EF, Van Swol CF, Rethy A (2012) Navigated laparoscopy-liver shift and deformation due to pneumoperitoneum in an animal model. Minim Invasive Ther Allied Technol 21(3):241–248PubMedCrossRef
34.
go back to reference Beller S, Hünerbein M, Eulenstein S, Lange T, Schlag PM (2007) Feasibility of navigated resection of liver tumors using multiplanar visualization of intraoperative 3-dimensional ultrasound data. Ann Surg 246:288–294PubMedCentralPubMedCrossRef Beller S, Hünerbein M, Eulenstein S, Lange T, Schlag PM (2007) Feasibility of navigated resection of liver tumors using multiplanar visualization of intraoperative 3-dimensional ultrasound data. Ann Surg 246:288–294PubMedCentralPubMedCrossRef
35.
go back to reference Maier-Hein L, Pianka F, Seitel A, Müller S, Tekbas A, Seitel M, Wolf I, Schmied B, Meinzer H-P (2007) Precision targeting of liver lesions with a needle-based soft tissue navigation system. In: Medical image computing and computer-assisted intervention–MICCAI 2007. Springer, New York, pp 42–49CrossRef Maier-Hein L, Pianka F, Seitel A, Müller S, Tekbas A, Seitel M, Wolf I, Schmied B, Meinzer H-P (2007) Precision targeting of liver lesions with a needle-based soft tissue navigation system. In: Medical image computing and computer-assisted intervention–MICCAI 2007. Springer, New York, pp 42–49CrossRef
36.
go back to reference Zhang H, Banovac F, Lin R, Glossop N, Wood BJ, Lindisch D, Levy E, Cleary K (2006) Electromagnetic tracking for abdominal interventions in computer aided surgery. Comput Aided Surg 11(3):127–136PubMed Zhang H, Banovac F, Lin R, Glossop N, Wood BJ, Lindisch D, Levy E, Cleary K (2006) Electromagnetic tracking for abdominal interventions in computer aided surgery. Comput Aided Surg 11(3):127–136PubMed
37.
go back to reference Khan MF, Dogan S, Maataoui A, Wesarg S, Gurung J, Ackermann H, Schiemann M, Wimmer-Greinecker G, Vogl TJ (2006) Navigation-based needle puncture of a cadaver using a hybrid tracking navigational system. Invest Radiol 41(10):713–720PubMedCrossRef Khan MF, Dogan S, Maataoui A, Wesarg S, Gurung J, Ackermann H, Schiemann M, Wimmer-Greinecker G, Vogl TJ (2006) Navigation-based needle puncture of a cadaver using a hybrid tracking navigational system. Invest Radiol 41(10):713–720PubMedCrossRef
38.
go back to reference Fichtinger G, Deguet A, Fischer G, Iordachita I, Balogh E, Masamune K, Taylor RH, Fayad LM, De Oliveira M, Zinreich SJ (2005) Image overlay for CT-guided needle insertions. Comput Aided Surg 10(4):241–255PubMed Fichtinger G, Deguet A, Fischer G, Iordachita I, Balogh E, Masamune K, Taylor RH, Fayad LM, De Oliveira M, Zinreich SJ (2005) Image overlay for CT-guided needle insertions. Comput Aided Surg 10(4):241–255PubMed
Metadata
Title
Ultrasound-based tumor movement compensation during navigated laparoscopic liver interventions
Authors
Osama Shahin
Armin Beširević
Markus Kleemann
Alexander Schlaefer
Publication date
01-05-2014
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 5/2014
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-013-3374-9

Other articles of this Issue 5/2014