Published in: Clinical Research in Cardiology 1/2016

01-01-2016 | Original Paper

Two-year outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon: lessons from the bonus freeze protocol

Authors: Andreas Metzner, Christian-Hendrik Heeger, Peter Wohlmuth, Bruno Reißmann, Andreas Rillig, Roland Richard Tilz, Shibu Mathew, Christine Lemes, Sebastian Deiß, Tilman Maurer, Ardan Saguner, Feifan Ouyang, Karl-Heinz Kuck, Erik Wißner

Published in: Clinical Research in Cardiology | Issue 1/2016

Login to get access

Abstract

Background

Pulmonary vein isolation (PVI) using the second-generation cryoballoon (CB2) in patients with paroxysmal and persistent atrial fibrillation (AF) has demonstrated encouraging acute and mid-term results. Follow-up data on outcome beyond 1-year is lacking. The purpose of this analysis was to investigate the 2-year clinical outcome after CB2-based PVI.

Methods

Sixty patients (age 62 ± 11 years) with paroxysmal [45/60 (75 %) patients] or short-standing persistent AF [15/60 (25 %) patients] underwent 28-mm CB2-based PVI. Freeze-cycle duration was 240 s. After successful PVI, a bonus freeze-cycle of the same duration was applied. Follow-up was based on outpatient clinic visits at 3, 6, 12, 18, and 24 months including 24 h Holter-ECGs and telephone interviews. Recurrence was defined as any symptomatic and/or documented atrial tachyarrhythmia (ATA) episode >30 s following a 3-month blanking period.

Results

A total of 231 pulmonary veins (PV) were identified and 230/231 (99.6 %) PVs were successfully isolated. Phrenic nerve palsy occurred in 2/60 (3.3 %) patients. No other periprocedural complications occurred. Follow-up was available for 59/60 (98 %) patients with a mean duration of 838 ± 67 days. A total of 43/59 (73 %) patients remained in stable sinus rhythm. In 10/16 (63 %) patients with ATA recurrence, a repeat procedure was performed using radiofrequency energy. The overall success rate after a maximum of two ablation procedures and a follow-up period of 838 ± 67 days was 88 % (52/59) patients.

Conclusions

Patients with paroxysmal or short-persistent AF undergoing PVI using the 28-mm CB2 demonstrated a 73 % 2-year single-procedure clinical success rate.
Literature
1.
go back to reference Su W, Kowal R, Kowalski M, Metzner A, Svinarich JT, Wheelan K, Wang P (2015) Best Practice Guide for Cryoballoon Ablation in Atrial Fibrillation: the compilation experience of over 3000 Procedures. Heart Rhythm. 12:1658–1666PubMedCrossRef Su W, Kowal R, Kowalski M, Metzner A, Svinarich JT, Wheelan K, Wang P (2015) Best Practice Guide for Cryoballoon Ablation in Atrial Fibrillation: the compilation experience of over 3000 Procedures. Heart Rhythm. 12:1658–1666PubMedCrossRef
2.
go back to reference Metzner A, Reissmann B, Rausch P, Mathew S, Wohlmuth P, Tilz R, Rillig A, Lemes C, Deiss S, Heeger C et al (2014) One-year clinical outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon. Circul Arrhyth Electrophysiol 7:288–292CrossRef Metzner A, Reissmann B, Rausch P, Mathew S, Wohlmuth P, Tilz R, Rillig A, Lemes C, Deiss S, Heeger C et al (2014) One-year clinical outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon. Circul Arrhyth Electrophysiol 7:288–292CrossRef
3.
go back to reference Wissner E, Heeger CH, Grahn H, Reissmann B, Wohlmuth P, Lemes C, Rausch P, Mathew S, Rillig A, Deiss S et al (2015) One-year clinical success of a ‘no-bonus’ freeze protocol using the second-generation 28 mm cryoballoon for pulmonary vein isolation. Europace (Epub ahead of print) Wissner E, Heeger CH, Grahn H, Reissmann B, Wohlmuth P, Lemes C, Rausch P, Mathew S, Rillig A, Deiss S et al (2015) One-year clinical success of a ‘no-bonus’ freeze protocol using the second-generation 28 mm cryoballoon for pulmonary vein isolation. Europace (Epub ahead of print)
4.
go back to reference Furnkranz A, Bordignon S, Dugo D, Perotta L, Gunawardene M, Schulte-Hahn B, Nowak B, Schmidt B, Chun JK (2014) Improved 1-year clinical success rate of pulmonary vein isolation with the second-generation cryoballoon in patients with paroxysmal atrial fibrillation. J Cardiovasc Electrophysiol 25:840–844PubMedCrossRef Furnkranz A, Bordignon S, Dugo D, Perotta L, Gunawardene M, Schulte-Hahn B, Nowak B, Schmidt B, Chun JK (2014) Improved 1-year clinical success rate of pulmonary vein isolation with the second-generation cryoballoon in patients with paroxysmal atrial fibrillation. J Cardiovasc Electrophysiol 25:840–844PubMedCrossRef
5.
go back to reference Chierchia GB, Di Giovanni G, Ciconte G, de Asmundis C, Conte G, Sieira-Moret J, Rodriguez-Manero M, Casado R, Baltogiannis G, Namdar M et al (2014) Second-generation cryoballoon ablation for paroxysmal atrial fibrillation: 1-year follow-up. Europace. 16:639–644PubMedCrossRef Chierchia GB, Di Giovanni G, Ciconte G, de Asmundis C, Conte G, Sieira-Moret J, Rodriguez-Manero M, Casado R, Baltogiannis G, Namdar M et al (2014) Second-generation cryoballoon ablation for paroxysmal atrial fibrillation: 1-year follow-up. Europace. 16:639–644PubMedCrossRef
6.
go back to reference Ciconte G, de Asmundis C, Sieira J, Conte G, Di Giovanni G, Mugnai G, Saitoh Y, Baltogiannis G, Irfan G, Coutino-Moreno HE et al (2015) Single 3-minute freeze for second-generation cryoballoon ablation: one-year follow-up after pulmonary vein isolation. Heart Rhythm. 12:673–680PubMedCrossRef Ciconte G, de Asmundis C, Sieira J, Conte G, Di Giovanni G, Mugnai G, Saitoh Y, Baltogiannis G, Irfan G, Coutino-Moreno HE et al (2015) Single 3-minute freeze for second-generation cryoballoon ablation: one-year follow-up after pulmonary vein isolation. Heart Rhythm. 12:673–680PubMedCrossRef
7.
go back to reference Ciconte G, Ottaviano L, de Asmundis C, Baltogiannis G, Conte G, Sieira J, Di Giovanni G, Saitoh Y, Irfan G, Mugnai G et al (2015) Pulmonary vein isolation as index procedure for persistent atrial fibrillation: one-year clinical outcome after ablation using the second-generation cryoballoon. Heart Rhythm. 12:60–66PubMedCrossRef Ciconte G, Ottaviano L, de Asmundis C, Baltogiannis G, Conte G, Sieira J, Di Giovanni G, Saitoh Y, Irfan G, Mugnai G et al (2015) Pulmonary vein isolation as index procedure for persistent atrial fibrillation: one-year clinical outcome after ablation using the second-generation cryoballoon. Heart Rhythm. 12:60–66PubMedCrossRef
8.
go back to reference Ciconte G, Baltogiannis G, de Asmundis C, Sieira J, Conte G, Di Giovanni G, Saitoh Y, Irfan G, Mugnai G, Hunuk B et al (2015) Circumferential pulmonary vein isolation as index procedure for persistent atrial fibrillation: a comparison between radiofrequency catheter ablation and second-generation cryoballoon ablation. Europace. Europace. 17:559–565PubMedCrossRef Ciconte G, Baltogiannis G, de Asmundis C, Sieira J, Conte G, Di Giovanni G, Saitoh Y, Irfan G, Mugnai G, Hunuk B et al (2015) Circumferential pulmonary vein isolation as index procedure for persistent atrial fibrillation: a comparison between radiofrequency catheter ablation and second-generation cryoballoon ablation. Europace. Europace. 17:559–565PubMedCrossRef
9.
go back to reference Lemes C, Wissner E, Lin T, Mathew S, Deiss S, Rillig A, Heeger C, Wohlmuth P, Reissmann B, Tilz R et al (2015) One-year clinical outcome after pulmonary vein isolation in persistent atrial fibrillation using the second-generation 28 mm cryoballoon: a retrospective analysis. Europace (Epub ahead of print) Lemes C, Wissner E, Lin T, Mathew S, Deiss S, Rillig A, Heeger C, Wohlmuth P, Reissmann B, Tilz R et al (2015) One-year clinical outcome after pulmonary vein isolation in persistent atrial fibrillation using the second-generation 28 mm cryoballoon: a retrospective analysis. Europace (Epub ahead of print)
10.
go back to reference Reddy VY, Sediva L, Petru J, Skoda J, Chovanec M, Chitovova Z, P DIS, Rubin E, Dukkipati S, Neuzil P (2015) Durability of pulmonary vein isolation with cryoballoon ablation: results from the SUstained PV isolation with ARctic Front Advance (SUPIR) Study. J Cardiovasc Electrophysiol 26:493–500 Reddy VY, Sediva L, Petru J, Skoda J, Chovanec M, Chitovova Z, P DIS, Rubin E, Dukkipati S, Neuzil P (2015) Durability of pulmonary vein isolation with cryoballoon ablation: results from the SUstained PV isolation with ARctic Front Advance (SUPIR) Study. J Cardiovasc Electrophysiol 26:493–500
11.
go back to reference Bordignon S, Furnkranz A, Perrotta L, Dugo D, Konstantinou A, Nowak B, Schulte-Hahn B, Schmidt B, Chun KR (2015) High rate of durable pulmonary vein isolation after second-generation cryoballoon ablation: analysis of repeat procedures. Europace Europace 17:725–731PubMedCrossRef Bordignon S, Furnkranz A, Perrotta L, Dugo D, Konstantinou A, Nowak B, Schulte-Hahn B, Schmidt B, Chun KR (2015) High rate of durable pulmonary vein isolation after second-generation cryoballoon ablation: analysis of repeat procedures. Europace Europace 17:725–731PubMedCrossRef
12.
go back to reference Metzner A, Burchard A, Wohlmuth P, Rausch P, Bardyszewski A, Gienapp C, Tilz RR, Rillig A, Mathew S, Deiss S et al (2013) Increased incidence of esophageal thermal lesions using the second-generation 28-mm cryoballoon. Circul Arrhyth Electrophysiol 6:769–775CrossRef Metzner A, Burchard A, Wohlmuth P, Rausch P, Bardyszewski A, Gienapp C, Tilz RR, Rillig A, Mathew S, Deiss S et al (2013) Increased incidence of esophageal thermal lesions using the second-generation 28-mm cryoballoon. Circul Arrhyth Electrophysiol 6:769–775CrossRef
13.
go back to reference Metzner A, Rausch P, Lemes C, Reissmann B, Bardyszewski A, Tilz R, Rillig A, Mathew S, Deiss S, Kamioka M et al (2014) The incidence of phrenic nerve injury during pulmonary vein isolation using the second-generation 28 mm cryoballoon. J Cardiovasc Electrophysiol 25:466–470PubMedCrossRef Metzner A, Rausch P, Lemes C, Reissmann B, Bardyszewski A, Tilz R, Rillig A, Mathew S, Deiss S, Kamioka M et al (2014) The incidence of phrenic nerve injury during pulmonary vein isolation using the second-generation 28 mm cryoballoon. J Cardiovasc Electrophysiol 25:466–470PubMedCrossRef
14.
go back to reference Casado-Arroyo R, Chierchia GB, Conte G, Levinstein M, Sieira J, Rodriguez-Manero M, di Giovanni G, Baltogiannis Y, Wauters K, de Asmundis C et al (2013) Phrenic nerve paralysis during cryoballoon ablation for atrial fibrillation: a comparison between the first- and second-generation balloon. Heart Rhythm. 10:1318–1324PubMedCrossRef Casado-Arroyo R, Chierchia GB, Conte G, Levinstein M, Sieira J, Rodriguez-Manero M, di Giovanni G, Baltogiannis Y, Wauters K, de Asmundis C et al (2013) Phrenic nerve paralysis during cryoballoon ablation for atrial fibrillation: a comparison between the first- and second-generation balloon. Heart Rhythm. 10:1318–1324PubMedCrossRef
15.
go back to reference Ouyang F, Tilz R, Chun J, Schmidt B, Wissner E, Zerm T, Neven K, Kokturk B, Konstantinidou M, Metzner A et al (2010) Long-term results of catheter ablation in paroxysmal atrial fibrillation: lessons from a 5-year follow-up. Circulation 122:2368–2377PubMedCrossRef Ouyang F, Tilz R, Chun J, Schmidt B, Wissner E, Zerm T, Neven K, Kokturk B, Konstantinidou M, Metzner A et al (2010) Long-term results of catheter ablation in paroxysmal atrial fibrillation: lessons from a 5-year follow-up. Circulation 122:2368–2377PubMedCrossRef
16.
go back to reference Tilz RR, Chun KR, Schmidt B, Fuernkranz A, Wissner E, Koester I, Baensch D, Boczor S, Koektuerk B, Metzner A et al (2010) Catheter ablation of long-standing persistent atrial fibrillation: a lesson from circumferential pulmonary vein isolation. J Cardiovasc Electrophysiol 21:1085–1093PubMedCrossRef Tilz RR, Chun KR, Schmidt B, Fuernkranz A, Wissner E, Koester I, Baensch D, Boczor S, Koektuerk B, Metzner A et al (2010) Catheter ablation of long-standing persistent atrial fibrillation: a lesson from circumferential pulmonary vein isolation. J Cardiovasc Electrophysiol 21:1085–1093PubMedCrossRef
17.
go back to reference Tilz RR, Rillig A, Thum AM, Arya A, Wohlmuth P, Metzner A, Mathew S, Yoshiga Y, Wissner E, Kuck KH et al (2012) Catheter ablation of long-standing persistent atrial fibrillation: 5-year outcomes of the Hamburg Sequential Ablation Strategy. J Am Coll Cardiol 60:1921–1929PubMedCrossRef Tilz RR, Rillig A, Thum AM, Arya A, Wohlmuth P, Metzner A, Mathew S, Yoshiga Y, Wissner E, Kuck KH et al (2012) Catheter ablation of long-standing persistent atrial fibrillation: 5-year outcomes of the Hamburg Sequential Ablation Strategy. J Am Coll Cardiol 60:1921–1929PubMedCrossRef
Metadata
Title
Two-year outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon: lessons from the bonus freeze protocol
Authors
Andreas Metzner
Christian-Hendrik Heeger
Peter Wohlmuth
Bruno Reißmann
Andreas Rillig
Roland Richard Tilz
Shibu Mathew
Christine Lemes
Sebastian Deiß
Tilman Maurer
Ardan Saguner
Feifan Ouyang
Karl-Heinz Kuck
Erik Wißner
Publication date
01-01-2016
Publisher
Springer Berlin Heidelberg
Published in
Clinical Research in Cardiology / Issue 1/2016
Print ISSN: 1861-0684
Electronic ISSN: 1861-0692
DOI
https://doi.org/10.1007/s00392-015-0890-8

Other articles of this Issue 1/2016

Clinical Research in Cardiology 1/2016 Go to the issue