The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria

Authors: Sandra Incardona, Magoma Mwancha-Kwasa, Roxanne R. Rees-Channer, Audrey Albertini, Joshua Havumaki, Peter Chiodini, Wellington Oyibo, Iveth J. Gonzalez

Published in: Malaria Journal | Issue 1/2018

Login to get access

Abstract

Background

Malaria rapid diagnostic tests (RDTs) are becoming widely adopted for case management at community level. However, reports and anecdotal observations indicate that the blood transfer step poses a significant challenge to many users. This study sought to evaluate the inverted cup device in the hands of health workers in everyday clinical practice, in comparison with the plastic pipette, and to determine the volume accuracy of the device made of a lower-cost plastic.

Methods

The volume accuracy of inverted cup devices made of two plastics, PMMA and SBC, was compared by transferring blood 150 times onto filter paper and comparing the blood spot areas with those produced by 20 reference transfers with a calibrated micropipette. The ease of use, safety and acceptability of the inverted cup device and the pipette were evaluated by 50 health workers in Nigeria. Observations were recorded on pre-designed questionnaires, by the health workers themselves and by trained observers. Focus group discussions were also conducted.

Results

The volume accuracy assessment showed that the device made from the low-cost material (SBC) delivered a more accurate volume (mean 5.4 μL, SD 0.48 μL, range 4.5–7.0 μL) than the PMMA device (mean 5.9 μL, SD 0.48 μL, range 4.9–7.2 μL). The observational evaluation demonstrated that the inverted cup device performed better than the pipette in all aspects, e.g. higher proportions of health workers achieved successful blood collection (96%, vs. 66%), transfer of the required blood volume (90%, vs. 58%), and blood deposit without any loss (95%, vs. 50%). Majority of health workers also considered it’ very easy’ to use (81%),’very appropriate’ for everyday use (78%), and 50% of them reported that it was their preferred BTD.

Conclusions

The good volume accuracy and high acceptability of the inverted cup device shown in this study, along with observed ease of use and safety in hands of health workers, further strengthens prior findings which demonstrated its higher accuracy as compared with other BTDs in a laboratory setting. Altogether, these studies suggest that the inverted cup device should replace other types of devices for use in day-to-day malaria diagnosis with RDTs.
Literature
2.
go back to reference Premji Z, Minjas JN, Shiff CJ. Laboratory diagnosis of malaria by village health workers using the rapid manual ParaSight-F test. Trans R Soc Trop Med Hyg. 1994;88:418.CrossRefPubMed Premji Z, Minjas JN, Shiff CJ. Laboratory diagnosis of malaria by village health workers using the rapid manual ParaSight-F test. Trans R Soc Trop Med Hyg. 1994;88:418.CrossRefPubMed
3.
go back to reference Bojang KA. The diagnosis of Plasmodium falciparum infection in Gambian children, by field staff using the rapid, manual, ParaSight-F test. Ann Trop Med Parasitol. 1999;93:685–7.PubMed Bojang KA. The diagnosis of Plasmodium falciparum infection in Gambian children, by field staff using the rapid, manual, ParaSight-F test. Ann Trop Med Parasitol. 1999;93:685–7.PubMed
4.
go back to reference Mayxay M, Newton PN, Yeung S, Pongvongsa T, Phompida S, Phetsouvanh R, et al. An assessment of the use of malaria rapid tests by village health volunteers in rural Laos. Trop Med Int Health. 2004;9:325–9.CrossRefPubMed Mayxay M, Newton PN, Yeung S, Pongvongsa T, Phompida S, Phetsouvanh R, et al. An assessment of the use of malaria rapid tests by village health volunteers in rural Laos. Trop Med Int Health. 2004;9:325–9.CrossRefPubMed
5.
go back to reference Luchavez J, Lintag ME, Coll-Black M, Baik F, Bell D. An assessment of various blood collection and transfer methods used for malaria rapid diagnostic tests. Malar J. 2007;6:149.CrossRefPubMedPubMedCentral Luchavez J, Lintag ME, Coll-Black M, Baik F, Bell D. An assessment of various blood collection and transfer methods used for malaria rapid diagnostic tests. Malar J. 2007;6:149.CrossRefPubMedPubMedCentral
6.
go back to reference Fogg C, Twesigye R, Batwala V, Piola P, Nabasumba C, Kiguli J, et al. Assessment of three new parasite lactate dehydrogenase (pan-pLDH) tests for diagnosis of uncomplicated malaria. Trans R Soc Trop Med Hyg. 2008;102:25–31.CrossRefPubMed Fogg C, Twesigye R, Batwala V, Piola P, Nabasumba C, Kiguli J, et al. Assessment of three new parasite lactate dehydrogenase (pan-pLDH) tests for diagnosis of uncomplicated malaria. Trans R Soc Trop Med Hyg. 2008;102:25–31.CrossRefPubMed
7.
go back to reference Harvey SA, Jennings L, Chinyama M, Masaninga F, Mulholland K, Bell DR. Improving community health worker use of malaria rapid diagnostic tests in Zambia: package instructions, job aid and job aid-plus-training. Malar J. 2008;7:160.CrossRefPubMedPubMedCentral Harvey SA, Jennings L, Chinyama M, Masaninga F, Mulholland K, Bell DR. Improving community health worker use of malaria rapid diagnostic tests in Zambia: package instructions, job aid and job aid-plus-training. Malar J. 2008;7:160.CrossRefPubMedPubMedCentral
8.
go back to reference Roukens AH, Berg J, Barbey A, Visser LG. Performance of self-diagnosis and standby treatment of malaria in international oilfield service employees in the field. Malar J. 2008;7:128.CrossRefPubMedPubMedCentral Roukens AH, Berg J, Barbey A, Visser LG. Performance of self-diagnosis and standby treatment of malaria in international oilfield service employees in the field. Malar J. 2008;7:128.CrossRefPubMedPubMedCentral
9.
go back to reference Hopkins H, Oyibo W, Luchavez J, Mationg ML, Asiimwe C, Albertini A, et al. Blood transfer devices for malaria rapid diagnostic tests: evaluation of accuracy, safety and ease of use. Malar J. 2011;10:30.CrossRefPubMedPubMedCentral Hopkins H, Oyibo W, Luchavez J, Mationg ML, Asiimwe C, Albertini A, et al. Blood transfer devices for malaria rapid diagnostic tests: evaluation of accuracy, safety and ease of use. Malar J. 2011;10:30.CrossRefPubMedPubMedCentral
10.
go back to reference Dunnett CW. Pairwise multiple comparisons in the unequal variance case. J Am Stat Assoc. 1980;75:796–800.CrossRef Dunnett CW. Pairwise multiple comparisons in the unequal variance case. J Am Stat Assoc. 1980;75:796–800.CrossRef
11.
go back to reference Manyando C, Njunju EM, Chileshe J, Siziya S, Shiff C. Rapid diagnostic tests for malaria and health workers’ adherence to test results at health facilities in Zambia. Malar J. 2014;13:166.CrossRefPubMedPubMedCentral Manyando C, Njunju EM, Chileshe J, Siziya S, Shiff C. Rapid diagnostic tests for malaria and health workers’ adherence to test results at health facilities in Zambia. Malar J. 2014;13:166.CrossRefPubMedPubMedCentral
12.
go back to reference Uzochukwu BS, Onwujekwe E, Ezuma NN, Ezeoke OP, Ajuba MO, Sibeudu FT. Improving rational treatment of malaria: perceptions and influence of RDTs on prescribing behaviour of health workers in southeast Nigeria. PLoS ONE. 2011;6:e14627.CrossRefPubMedPubMedCentral Uzochukwu BS, Onwujekwe E, Ezuma NN, Ezeoke OP, Ajuba MO, Sibeudu FT. Improving rational treatment of malaria: perceptions and influence of RDTs on prescribing behaviour of health workers in southeast Nigeria. PLoS ONE. 2011;6:e14627.CrossRefPubMedPubMedCentral
13.
go back to reference Ezeoke OP, Ezumah NN, Chandler CC, Mangham-Jefferies LJ, Onwujekwe OE, Wiseman V, et al. Exploring health providers’ and community perceptions and experiences with malaria tests in South-East Nigeria: a critical step towards appropriate treatment. Malar J. 2012;11:368.CrossRefPubMedPubMedCentral Ezeoke OP, Ezumah NN, Chandler CC, Mangham-Jefferies LJ, Onwujekwe OE, Wiseman V, et al. Exploring health providers’ and community perceptions and experiences with malaria tests in South-East Nigeria: a critical step towards appropriate treatment. Malar J. 2012;11:368.CrossRefPubMedPubMedCentral
Metadata
Title
The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria
Authors
Sandra Incardona
Magoma Mwancha-Kwasa
Roxanne R. Rees-Channer
Audrey Albertini
Joshua Havumaki
Peter Chiodini
Wellington Oyibo
Iveth J. Gonzalez
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Malaria Journal / Issue 1/2018
Electronic ISSN: 1475-2875
DOI
https://doi.org/10.1186/s12936-018-2173-0

Other articles of this Issue 1/2018

Malaria Journal 1/2018 Go to the issue