The invention of microwave surgical scissors for seamless coagulation and cutting

Authors: Tohru Tani, Shigeyuki Naka, Soichiro Tani, Hisanori Shiomi, Koichiro Murakami, Atsushi Yamada, Dang Tran Khiem

Published in: Surgery Today | Issue 9/2018

Login to get access

Abstract

Purpose

We developed a microwave energy-based scissors device (MWCX) that is capable of performing cutting and coagulation using 2.45 GHz microwave energy. This paper aims to present the concept of the device and assess the basic functions including the hemostasis, cutting, and sealing abilities.

Methods

Seven beagle dogs were used in our experiments. In six dogs, we measured the coagulation time (CT), lateral thermal injury (LTI), bursting pressure (BP). The dogs were then subjected to re-laparotomy 1 week later to allow us to investigate the results. In one dog, the same factors and the quantities of smoke and mist emitted were compared to those observed when using a Harmonic Focus (HF) device.

Results

At 60 W, the MWCX could cut and seal small (5 s, diameter 1–2 mm) and medium-sized (10 s, 3–4 mm) vessels with complete hemostasis. The liver (length 2 cm) was cut for 30 s. Harvested vessels were sealed for 10 s (artery, 17 times; vein, six times). The mean BP was 887. 8 ± 41.5 mmHg in the medium arteries and 457.2 ± 118.0 mmHg in veins, with a mean diameter of 4.5 ± 1.3 mm. In a comparative study, the MWCX showed similar results to the HF with regard to the CT, BP and LTI, and emitted less smoke and mist.

Conclusion

The MWCX showed similar levels of functionality and safety to HF, as well as the advantages offered by the use of microwave energy. Microwave devices might be used in the majority of applications for which traditional energy devices are used.
Literature
3.
go back to reference Swan RZ, Tsirline V, Sindram D, Martinie JB, Iannitti DA. Fundamentals of microwave physics: application to hepatic ablation. J Microwave Surg. 2012;30:1–16.CrossRef Swan RZ, Tsirline V, Sindram D, Martinie JB, Iannitti DA. Fundamentals of microwave physics: application to hepatic ablation. J Microwave Surg. 2012;30:1–16.CrossRef
4.
go back to reference Zimmer RP, Ecker HA, Popovic VP. Selective electromagnetic heating of tumor in animals in deep Hypothermia. IEEE Trans on Microwave Theory Techniques. 1971;16 – 2:238–45.CrossRef Zimmer RP, Ecker HA, Popovic VP. Selective electromagnetic heating of tumor in animals in deep Hypothermia. IEEE Trans on Microwave Theory Techniques. 1971;16 – 2:238–45.CrossRef
5.
go back to reference Tabuse K. A new operative procedure of hepatic surgery using a microwave tissue coagulator. Arch Jpn Chir. 1979;48:160–72. Tabuse K. A new operative procedure of hepatic surgery using a microwave tissue coagulator. Arch Jpn Chir. 1979;48:160–72.
6.
go back to reference Tabuse K, Katsumi M, Kobayashi Y, Noguchi H, Egawa H, Aoyama O, et al. Microwave surgery: Hepatectomy using microwave tissue coagulation. World J Surgery. 1985;9:136–43.CrossRef Tabuse K, Katsumi M, Kobayashi Y, Noguchi H, Egawa H, Aoyama O, et al. Microwave surgery: Hepatectomy using microwave tissue coagulation. World J Surgery. 1985;9:136–43.CrossRef
7.
go back to reference Kurumi Y, Tani T, Naka S, Shiomi H, Shimizu T, Abe H, et al. MR-guided microwave ablation for malignancies. Int J Clin Oncol. 2007; 1285–93. Kurumi Y, Tani T, Naka S, Shiomi H, Shimizu T, Abe H, et al. MR-guided microwave ablation for malignancies. Int J Clin Oncol. 2007; 1285–93.
8.
go back to reference Nguyen Quoc V, Tani T, Naka S, Tani S. Comparison of tissue changes caused by microwave and radio frequency energy in an experimental liver ablation. Int J Clin Exp Pathol. 2016;9(2):1294–302. Nguyen Quoc V, Tani T, Naka S, Tani S. Comparison of tissue changes caused by microwave and radio frequency energy in an experimental liver ablation. Int J Clin Exp Pathol. 2016;9(2):1294–302.
9.
go back to reference Takebayashi K, Shiomi H, Naka S, Murayama H, Murakami K, Akabori H, et al. Utility of microwave surgical instrument in sealing lymphatic vessels. Am J Surg. 2013;206:229–33.CrossRefPubMed Takebayashi K, Shiomi H, Naka S, Murayama H, Murakami K, Akabori H, et al. Utility of microwave surgical instrument in sealing lymphatic vessels. Am J Surg. 2013;206:229–33.CrossRefPubMed
10.
go back to reference Tani T, Naka S, Murakami K, Tani S, Akabori K, Shiomi N, et al. Comparative study between newly developed microwave surgical devices and commercialized energy devices in animal model (in Japanese). J Biliary Tract Pancreas. 2016;37:581–8. Tani T, Naka S, Murakami K, Tani S, Akabori K, Shiomi N, et al. Comparative study between newly developed microwave surgical devices and commercialized energy devices in animal model (in Japanese). J Biliary Tract Pancreas. 2016;37:581–8.
11.
go back to reference Courtney KP, Gregory WH, Evren D, Daniel Sl, Peter AH, Mahesh MM, et al. Tissue response to surgical energy devices. Urology. 2008;71:744–8.CrossRef Courtney KP, Gregory WH, Evren D, Daniel Sl, Peter AH, Mahesh MM, et al. Tissue response to surgical energy devices. Urology. 2008;71:744–8.CrossRef
12.
go back to reference William LN, William WH, Thomas MS, Jessica JHH, James N, Amy EL, et al. Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices. Surg Endosc. 2009;23:90–6.CrossRef William LN, William WH, Thomas MS, Jessica JHH, James N, Amy EL, et al. Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices. Surg Endosc. 2009;23:90–6.CrossRef
13.
go back to reference Mukaisho K, Kurumi Y, Sugihara H, Naka S, Kamitani S, Tsubosa Y, et al. Enzyme histochemistry is useful to assess viability of tumor tissue after microwave coagulation therapy (MCT). Dig Dis Sci. 2002;47(11):2441–5.CrossRefPubMed Mukaisho K, Kurumi Y, Sugihara H, Naka S, Kamitani S, Tsubosa Y, et al. Enzyme histochemistry is useful to assess viability of tumor tissue after microwave coagulation therapy (MCT). Dig Dis Sci. 2002;47(11):2441–5.CrossRefPubMed
15.
go back to reference Milsom J, Trencheva K, Monette S, Pavoor R, Shukla P, Ma J, et al. Evaluation of the safety, efficacy, and versatility of a new surgical energy device (Thunderbeat) in comparison with harmonic ACE, LigaSure V, and EnSeal devices in a porcine model. J Laparoendosc Adv Surg Tech. 2012;22(4):378–86.CrossRef Milsom J, Trencheva K, Monette S, Pavoor R, Shukla P, Ma J, et al. Evaluation of the safety, efficacy, and versatility of a new surgical energy device (Thunderbeat) in comparison with harmonic ACE, LigaSure V, and EnSeal devices in a porcine model. J Laparoendosc Adv Surg Tech. 2012;22(4):378–86.CrossRef
Metadata
Title
The invention of microwave surgical scissors for seamless coagulation and cutting
Authors
Tohru Tani
Shigeyuki Naka
Soichiro Tani
Hisanori Shiomi
Koichiro Murakami
Atsushi Yamada
Dang Tran Khiem
Publication date
01-09-2018
Publisher
Springer Singapore
Published in
Surgery Today / Issue 9/2018
Print ISSN: 0941-1291
Electronic ISSN: 1436-2813
DOI
https://doi.org/10.1007/s00595-018-1662-7

Other articles of this Issue 9/2018

Surgery Today 9/2018 Go to the issue