The initial experience of the upper abdominal CT angiography using low-concentration contrast medium on dual energy spectral CT

Authors: Lei Xin, Xiaotang Yang, Ning Huang, Xiaosong Du, Jianxin Zhang, Yanyan Wang, Lina Hou, Jinfang Gao

Published in: Abdominal Radiology | Issue 7/2015

Login to get access

Abstract

Objective

To investigate the feasibility of using Spectral CT imaging with low contrast medium in abdominal CT angiography (CTA).

Subjects and methods

70 consecutive patients (40 men, 42.6 ± 20.4 years; 30 women, 46.7 ± 18.8 years) with suspected abdominal focal lesions were referred to CTA exam. They were randomly assigned into two groups. Group A: 35 patients underwent conventional CT scan of Tube voltage 120 kVp, automatic current modulation with a Noise Index of 12, ASIR 30%, and injected with Iohexol (350 mgI/ml). Group B: 35 patients underwent Spectral CT Imaging, with Tube current of 600 mA, injected with Iodixanol (270 mgI/ml). The optimal mono-energy keV was achieved using the optimal contrast noise ratio in abdominal aorta at the renal artery level relative to the erector spine muscle. Both groups were injected with an injection rate of 3.5 ml/s, and a contrast volume of 1.5 ml/kg body weight. The Hounsfield units (HU) and noise of the bilateral renal arteries and muscle of both groups, as well as the optimal monochromatic image set of Group B were measured. Two radiologists assessed all images with a 5-points scale. CTDIvol and DLP were recorded. Data were analyzed using student t test.

Results

The total iodine intake of Group B was 28% lower than that of Group A. The CNR of abdomen artery, celiac trunk, superior mesenteric artery, and renal artery in spectral group (at the best mono-energy of 53.0 keV) were higher than those in conventional CTA group (p < 0.001). The subjective image quality score of spectral CTA group was also rated higher than conventional CTA group (p < 0.001). CTDIvol, DLP, and effective dose of spectral group were all lower than conventional group, but there were no significant differences (p > 0.05).

Conclusion

With 28% contrast medium reduction and reduced radiation dose, CT angiography using spectral imaging and lower concentration contrast agent provided better image quality than conventional CTA.
Literature
1.
go back to reference Li J, Tang GS, Huang Y, et al. (2009) Application of digital subtraction angiography and interventional treatment in gastrointestinal arterial hemorrhage. Chin J Gastrointest 12(3):252–256 Li J, Tang GS, Huang Y, et al. (2009) Application of digital subtraction angiography and interventional treatment in gastrointestinal arterial hemorrhage. Chin J Gastrointest 12(3):252–256
2.
go back to reference Huang L, Zheng P, Mao J, Li F, Zheng Y., Cheng Z (2014) Application value of three-dimensional surgery planning system in the preoperative evaluation of primary liver cancer resection. Chin J Hepat Surg 3(2), 79–83 Huang L, Zheng P, Mao J, Li F, Zheng Y., Cheng Z (2014) Application value of three-dimensional surgery planning system in the preoperative evaluation of primary liver cancer resection. Chin J Hepat Surg 3(2), 79–83
3.
go back to reference Mehran R, Nikolsky E (2006) Contrast-induced nephropathy: definition, epidemiology, and patients at risk. Kidney Int Suppl 100:S11–S15CrossRefPubMed Mehran R, Nikolsky E (2006) Contrast-induced nephropathy: definition, epidemiology, and patients at risk. Kidney Int Suppl 100:S11–S15CrossRefPubMed
4.
go back to reference Yuan R, Shuman WP, Earls JP, et al. (2012) Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography—a prospective randomized trial. Radiology 262(1):290–297CrossRefPubMed Yuan R, Shuman WP, Earls JP, et al. (2012) Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography—a prospective randomized trial. Radiology 262(1):290–297CrossRefPubMed
5.
go back to reference Marin D, Nelson RC, Schindera ST, et al. (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm-initial clinical experience. Radiology 254(1):145–153CrossRefPubMed Marin D, Nelson RC, Schindera ST, et al. (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm-initial clinical experience. Radiology 254(1):145–153CrossRefPubMed
6.
go back to reference Cho E-S, Yu J-S, et al. (2012) CT angiography of the renal arteries: comparison of lower-tube-voltage CTA with moderate-concentration iodinated contrast material and conventional CTA. Genitourin Imaging 199:96–102 Cho E-S, Yu J-S, et al. (2012) CT angiography of the renal arteries: comparison of lower-tube-voltage CTA with moderate-concentration iodinated contrast material and conventional CTA. Genitourin Imaging 199:96–102
7.
go back to reference Liu B, Bai M (2010) Variations of effective dose for head, thoracic, abdomen and coronary CT examinations according to new ICRP publication. Chin Med Equip J 31(03):27–31 Liu B, Bai M (2010) Variations of effective dose for head, thoracic, abdomen and coronary CT examinations according to new ICRP publication. Chin Med Equip J 31(03):27–31
8.
go back to reference Stolzmann P, Scheffel H, Rentsch K, et al. (2008) Dual-energy computed tomography for the differentiation of uric acid stones: exvivo performance evaluation. Urol Res 36(3–4):133–138CrossRefPubMed Stolzmann P, Scheffel H, Rentsch K, et al. (2008) Dual-energy computed tomography for the differentiation of uric acid stones: exvivo performance evaluation. Urol Res 36(3–4):133–138CrossRefPubMed
9.
go back to reference Zhang Z, Lu G (2007) Dual source computed tomography and its clinical application. J Med Postgrad 20(4):416–418 Zhang Z, Lu G (2007) Dual source computed tomography and its clinical application. J Med Postgrad 20(4):416–418
10.
go back to reference Ren Q, Hua Y, Li J (2011) The basic principle and clinical applications of CT spectral imaging. Int J Med Radiol 34(6):16 Ren Q, Hua Y, Li J (2011) The basic principle and clinical applications of CT spectral imaging. Int J Med Radiol 34(6):16
11.
go back to reference Matsumoto K, Jinzaki M, Tanami Y, et al. (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259(1):257–262CrossRefPubMed Matsumoto K, Jinzaki M, Tanami Y, et al. (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259(1):257–262CrossRefPubMed
12.
go back to reference Sun Y, Mao D, Li M, et al. (2013) The value of CT angiography before transchatheter in the use of chemoembolization of hepatocellular carcinoma using dual-energy CT. J Med Imaging 23(11):1718–1723 Sun Y, Mao D, Li M, et al. (2013) The value of CT angiography before transchatheter in the use of chemoembolization of hepatocellular carcinoma using dual-energy CT. J Med Imaging 23(11):1718–1723
13.
go back to reference Fenchel S, Fleeter TR, Aschoff AJ, et al. (2004) Effect of iodine concentration of contrast media on contrast enhancement in multislice CT of the pancreas. Br J Radiol 77:821–830CrossRefPubMed Fenchel S, Fleeter TR, Aschoff AJ, et al. (2004) Effect of iodine concentration of contrast media on contrast enhancement in multislice CT of the pancreas. Br J Radiol 77:821–830CrossRefPubMed
14.
go back to reference Chen X, Xie Q, Wei C, et al. (2009) Comparison of two iodine concentration of contrast media on contrast enhancement in multislice CT of kidney. Prog Mod Biomed 9(13):2528–2530 Chen X, Xie Q, Wei C, et al. (2009) Comparison of two iodine concentration of contrast media on contrast enhancement in multislice CT of kidney. Prog Mod Biomed 9(13):2528–2530
16.
go back to reference Bickham P, Golembiewski J (2010) Contrast media use in the operation room. J Perianestbesia Nurs 25:94–103CrossRef Bickham P, Golembiewski J (2010) Contrast media use in the operation room. J Perianestbesia Nurs 25:94–103CrossRef
17.
go back to reference Rihal CS, Textor SC, Grill DE, et al. (2002) Incidence and prognostic importance of acute renal failure after percutaneous coronary intervention. Circulation 105(19):2259–2264CrossRefPubMed Rihal CS, Textor SC, Grill DE, et al. (2002) Incidence and prognostic importance of acute renal failure after percutaneous coronary intervention. Circulation 105(19):2259–2264CrossRefPubMed
18.
go back to reference Abujudeh HH, Gee MS, Kaewlai (2009) In emergency situations, should serum creatinine be checked in all patients before performing second contrast CT examinations within 24 hours. J Am Coll Radiol 6(4):268–273CrossRefPubMed Abujudeh HH, Gee MS, Kaewlai (2009) In emergency situations, should serum creatinine be checked in all patients before performing second contrast CT examinations within 24 hours. J Am Coll Radiol 6(4):268–273CrossRefPubMed
19.
go back to reference Balemans CEA, Reichert LJM, van Schelven BIH, van den Brand JAJG, Wetzels JFM (2012) Epidemiology of contrast material-induced nephropathy in the era of hydration. Radiology 263(3):706–713CrossRefPubMed Balemans CEA, Reichert LJM, van Schelven BIH, van den Brand JAJG, Wetzels JFM (2012) Epidemiology of contrast material-induced nephropathy in the era of hydration. Radiology 263(3):706–713CrossRefPubMed
Metadata
Title
The initial experience of the upper abdominal CT angiography using low-concentration contrast medium on dual energy spectral CT
Authors
Lei Xin
Xiaotang Yang
Ning Huang
Xiaosong Du
Jianxin Zhang
Yanyan Wang
Lina Hou
Jinfang Gao
Publication date
01-10-2015
Publisher
Springer US
Published in
Abdominal Radiology / Issue 7/2015
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-015-0462-4

Other articles of this Issue 7/2015

Abdominal Radiology 7/2015 Go to the issue

Classics in Abdominal Imaging

The wall-echo-shadow (WES) sign

Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.