Published in: Clinical Orthopaedics and Related Research® 1/2013

01-01-2013 | Symposium: Papers Presented at the Annual Meetings of the Knee Society

The 2012 Mark Coventry Award: A Retrieval Analysis of High Flexion versus Posterior-stabilized Tibial Inserts

Authors: Nicholas R. Paterson, BScH, Matthew G. Teeter, BScH, PhD, Steven J. MacDonald, MD, FRCSC, Richard W. McCalden, MD, MPhil(Edin), FRCSC, Douglas D. R. Naudie, MD, FRCSC

Published in: Clinical Orthopaedics and Related Research® | Issue 1/2013

Login to get access

Abstract

Background

High flexion (HF) implants were introduced to increase ROM and patient satisfaction, but design changes to the implant potentially have deleterious effects on polyethylene wear. It is unclear whether the HF implants affect wear.

Questions/purposes

We therefore examined whether the design changes between HF and posterior-stabilized (PS) tibial inserts would affect overall damage or damage on their articular surface, backside, and tibial post and whether flexion angle achieved related to damage.

Methods

We matched 20 retrieved HF inserts to 20 retrieved PS inserts from the same implant system on the basis of duration of implantation, body mass index, and age. Inserts were divided into 16 zones and a microscopic analysis of surface damage was carried out. Five inserts were scanned using micro-CT to further quantify instances of severe post notching. We determined overall damage with a scoring system.

Results

We found greater backside and post damage in the HF group but no difference in the articular surface or overall damage scores. Backside and post damage scores correlated to flexion angle in the HF group. There was no flexion/damage correlation in the PS group. Notch depths around the post in both groups ranged from 0.6 to 1.9 mm.

Conclusions

HF inserts are more susceptible to post damage, possibly as a result of higher contact stresses from greater flexion. The increased backside damage was unexpected because the two groups have the same tibial component, locking mechanism, and sterilization method.

Clinical Relevance

The introduction of a highly crosslinked HF insert will require close scrutiny as a result of the potential for post damage demonstrated in this series.
Literature
1.
go back to reference Angelo FD, Marcolli D, Bulgheroni P, Murena L, Congiu T, Cherubino P. Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report. J Med Case Reports. 2010;4:2–9.CrossRef Angelo FD, Marcolli D, Bulgheroni P, Murena L, Congiu T, Cherubino P. Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report. J Med Case Reports. 2010;4:2–9.CrossRef
2.
go back to reference Bal BS, Greenberg D, Li S, Mauerhan DR, Schultz L, Cherry K. Tibial post failures in a condylar posterior cruciate substituting total knee arthroplasty. J Arthroplasty. 2008;23:650–655.PubMedCrossRef Bal BS, Greenberg D, Li S, Mauerhan DR, Schultz L, Cherry K. Tibial post failures in a condylar posterior cruciate substituting total knee arthroplasty. J Arthroplasty. 2008;23:650–655.PubMedCrossRef
3.
go back to reference Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468:57–63.PubMedCrossRef Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468:57–63.PubMedCrossRef
4.
go back to reference Chiu Y, Chen W, Huang C. Case report fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis. J Bone Joint Surg Am. 2004;19:1045–1049. Chiu Y, Chen W, Huang C. Case report fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis. J Bone Joint Surg Am. 2004;19:1045–1049.
5.
go back to reference Crowninshield RD, Wimmer MA, Jacobs JJ, Rosenberg AG. Clinical performance of contemporary tibial polyethylene components. J Arthroplasty. 2006;21:754–761.PubMedCrossRef Crowninshield RD, Wimmer MA, Jacobs JJ, Rosenberg AG. Clinical performance of contemporary tibial polyethylene components. J Arthroplasty. 2006;21:754–761.PubMedCrossRef
6.
go back to reference Dolan MM, Kelly NH, Nguyen JT, Wright TM, Haas SB. Implant design influences tibial post wear damage in posterior-stabilized knees. Clin Orthop Relat Res. 2011;469:160–167.PubMedCrossRef Dolan MM, Kelly NH, Nguyen JT, Wright TM, Haas SB. Implant design influences tibial post wear damage in posterior-stabilized knees. Clin Orthop Relat Res. 2011;469:160–167.PubMedCrossRef
7.
go back to reference Engh GA, Zimmerman RL, Parks NL, Engh CA. Analysis of wear in retrieved mobile and fixed bearing knee inserts. J Arthroplasty. 2009;24:28–32.PubMedCrossRef Engh GA, Zimmerman RL, Parks NL, Engh CA. Analysis of wear in retrieved mobile and fixed bearing knee inserts. J Arthroplasty. 2009;24:28–32.PubMedCrossRef
8.
go back to reference Furman BD, Lipman J, Kligman M, Wright TM, Haas SB. Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res. 2008;466:2650–2655.PubMedCrossRef Furman BD, Lipman J, Kligman M, Wright TM, Haas SB. Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res. 2008;466:2650–2655.PubMedCrossRef
9.
go back to reference Hamai S, Miura H, Matsuda S, Shimoto T, Higaki H, Iwamoto Y. Contact stress at the anterior aspect of the tibial post in posterior-stabilized total knee replacement. J Bone Joint Surg Am. 2011;92:1765–1773.CrossRef Hamai S, Miura H, Matsuda S, Shimoto T, Higaki H, Iwamoto Y. Contact stress at the anterior aspect of the tibial post in posterior-stabilized total knee replacement. J Bone Joint Surg Am. 2011;92:1765–1773.CrossRef
10.
go back to reference Hanson GR, Suggs JF, Kwon Y, Freiberg AA, Li G. In vivo anterior tibial post contact after posterior stabilizing total knee arthroplasty. J Orthop Res. 2007;25:1447–1453.PubMedCrossRef Hanson GR, Suggs JF, Kwon Y, Freiberg AA, Li G. In vivo anterior tibial post contact after posterior stabilizing total knee arthroplasty. J Orthop Res. 2007;25:1447–1453.PubMedCrossRef
11.
go back to reference Heyse TJ, Chen DX, Kelly N, Boettner F, Wright TM, Haas SB. The knee matched-pair total knee arthroplasty retrieval analysis: oxidized zirconium vs CoCrMo. Knee. 2011;18:448–452.PubMedCrossRef Heyse TJ, Chen DX, Kelly N, Boettner F, Wright TM, Haas SB. The knee matched-pair total knee arthroplasty retrieval analysis: oxidized zirconium vs CoCrMo. Knee. 2011;18:448–452.PubMedCrossRef
12.
go back to reference Heyse TJ, Davis J, Haas SB, Chen DX, Wright TM, Laskin RS. Retrieval analysis of femoral zirconium components in total knee arthroplasty preliminary results. J Arthroplasty. 2011;26:445–450.PubMedCrossRef Heyse TJ, Davis J, Haas SB, Chen DX, Wright TM, Laskin RS. Retrieval analysis of femoral zirconium components in total knee arthroplasty preliminary results. J Arthroplasty. 2011;26:445–450.PubMedCrossRef
13.
go back to reference Hood R, Wright T, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef Hood R, Wright T, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef
14.
go back to reference Implant Research Center Writing Committee, Medel FJ, Kurtz SM, Sharkey PF, Austin MS, Klein GR, Cohen AR, Patel H, Goldberg VM, Kraay MJ, Rimnac CM. Post damage in contemporary posterior-stabilized tibial inserts influence of implant design and clinical relevance. J Arthroplasty. 2011;26:606–614.PubMedCrossRef Implant Research Center Writing Committee, Medel FJ, Kurtz SM, Sharkey PF, Austin MS, Klein GR, Cohen AR, Patel H, Goldberg VM, Kraay MJ, Rimnac CM. Post damage in contemporary posterior-stabilized tibial inserts influence of implant design and clinical relevance. J Arthroplasty. 2011;26:606–614.PubMedCrossRef
15.
go back to reference Jung KA, Lee SC, Hawng SH, Kim SM. Fracture of a second-generation highly cross-linked UHMWPE tibial post in a posterior-stabilized Scorpio knee system. Clin Orthop Relat Res. 2008;31:1137–1140. Jung KA, Lee SC, Hawng SH, Kim SM. Fracture of a second-generation highly cross-linked UHMWPE tibial post in a posterior-stabilized Scorpio knee system. Clin Orthop Relat Res. 2008;31:1137–1140.
16.
go back to reference Kurtz SM, Walker PS. How have new designs and new types of joint replacement influenced wear behaviour? J Am Acad Orthop Surg. 2008;16:107–110. Kurtz SM, Walker PS. How have new designs and new types of joint replacement influenced wear behaviour? J Am Acad Orthop Surg. 2008;16:107–110.
17.
go back to reference Lachiewicz PF. How to treat a tibial post fracture in total knee arthroplasty: a systematic review. Clin Orthop Relat Res. 2010;496:1709–1715. Lachiewicz PF. How to treat a tibial post fracture in total knee arthroplasty: a systematic review. Clin Orthop Relat Res. 2010;496:1709–1715.
18.
go back to reference Lee C, Chen W, Kou H, Lo W, Chen C. Case report early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-flex posterior stabilized knee prosthesis. J Arthroplasty. 2009;24:1295–1299. Lee C, Chen W, Kou H, Lo W, Chen C. Case report early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-flex posterior stabilized knee prosthesis. J Arthroplasty. 2009;24:1295–1299.
19.
go back to reference Lim HC, Bae JH, Hwang JH, Kim SJ, Yoon JY. Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis. Clin Orthop Relat Res. 2009;1:118–121. Lim HC, Bae JH, Hwang JH, Kim SJ, Yoon JY. Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis. Clin Orthop Relat Res. 2009;1:118–121.
20.
go back to reference Lombardi AV, Ellison BS, Berend KR. Polyethylene wear is influenced by manufacturing technique in modular TKA. Clin Orthop Relat Res. 2008;466:2798–2805.PubMedCrossRef Lombardi AV, Ellison BS, Berend KR. Polyethylene wear is influenced by manufacturing technique in modular TKA. Clin Orthop Relat Res. 2008;466:2798–2805.PubMedCrossRef
21.
go back to reference Lu YC, Huang HC, Chang TK, Ho FY, Cheng CK, Huang CH. Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Am. 2010;92:500–507.CrossRef Lu YC, Huang HC, Chang TK, Ho FY, Cheng CK, Huang CH. Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Am. 2010;92:500–507.CrossRef
22.
go back to reference Luo SX, Su W, Zhao JM, Sha K, Wei QJ, Li XF. High-flexion vs conventional prostheses total knee arthroplasty: a meta-analysis. J Arthroplasty. 2011;26:847–854.PubMedCrossRef Luo SX, Su W, Zhao JM, Sha K, Wei QJ, Li XF. High-flexion vs conventional prostheses total knee arthroplasty: a meta-analysis. J Arthroplasty. 2011;26:847–854.PubMedCrossRef
23.
go back to reference Mauerhan DR. Case report fracture of the polyethylene tibial post in a posterior cruciate-substituting total knee arthroplasty mimicking patellar clunk syndrome: a report of 5 cases. J Arthroplasty. 2003;18:942–945.PubMedCrossRef Mauerhan DR. Case report fracture of the polyethylene tibial post in a posterior cruciate-substituting total knee arthroplasty mimicking patellar clunk syndrome: a report of 5 cases. J Arthroplasty. 2003;18:942–945.PubMedCrossRef
24.
go back to reference McCalden RW, MacDonald SJ, Bourne RB, Marr JT. A randomized controlled trial comparing ‘high-flex’ vs ‘standard’ posterior cruciate substituting polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty. 2009;24:33–38.PubMedCrossRef McCalden RW, MacDonald SJ, Bourne RB, Marr JT. A randomized controlled trial comparing ‘high-flex’ vs ‘standard’ posterior cruciate substituting polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty. 2009;24:33–38.PubMedCrossRef
25.
go back to reference Pinskerova V, Samuelson KM, Stammers J, Maruthainar K, Sosna A, Freeman M. The knee in full flexion: an anatomical study. J Bone Joint Surg Br. 2009;91:830–834.PubMed Pinskerova V, Samuelson KM, Stammers J, Maruthainar K, Sosna A, Freeman M. The knee in full flexion: an anatomical study. J Bone Joint Surg Br. 2009;91:830–834.PubMed
26.
go back to reference Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001;83:390–397.PubMed Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001;83:390–397.PubMed
27.
go back to reference Silva M, Schmalzried TP. Polyethylene in total knee arthroplasty. In: Callaghan JJ, Rosenberg AG, Rubash HE, Simonian PT, Wickieqicz TL, eds. The Adult Knee. 1st ed. Philadelphia, PA, USA: Lippincott Williams & Wilkins; 2003:279–286. Silva M, Schmalzried TP. Polyethylene in total knee arthroplasty. In: Callaghan JJ, Rosenberg AG, Rubash HE, Simonian PT, Wickieqicz TL, eds. The Adult Knee. 1st ed. Philadelphia, PA, USA: Lippincott Williams & Wilkins; 2003:279–286.
28.
go back to reference Stoller AP, Johnson TS, Popoola OO, Humphrey SM, Blanchard CR. Highly crosslinked polyethylene in posterior-stabilized total knee arthroplasty: in vitro performance evaluation of wear, delamination, and tibial post durability. J Arthroplasty. 2011;26:483–491.PubMedCrossRef Stoller AP, Johnson TS, Popoola OO, Humphrey SM, Blanchard CR. Highly crosslinked polyethylene in posterior-stabilized total knee arthroplasty: in vitro performance evaluation of wear, delamination, and tibial post durability. J Arthroplasty. 2011;26:483–491.PubMedCrossRef
29.
go back to reference Teeter MG, Naudie DDR, McErlain DD, Brandt J-M, Yuan X, MacDonald SJ, Holdsworth DW. In vitro quantification of wear in tibial inserts using microcomputed tomography. Clin Orthop Relat Res. 2011;469:107–112.PubMedCrossRef Teeter MG, Naudie DDR, McErlain DD, Brandt J-M, Yuan X, MacDonald SJ, Holdsworth DW. In vitro quantification of wear in tibial inserts using microcomputed tomography. Clin Orthop Relat Res. 2011;469:107–112.PubMedCrossRef
30.
go back to reference Teeter MG, Naudie DDR, Milner JS, Holdsworth DW. Determination of reference geometry for polyethylene tibial insert wear analysis. J Arthroplasty. 2011;26:497–503.PubMedCrossRef Teeter MG, Naudie DDR, Milner JS, Holdsworth DW. Determination of reference geometry for polyethylene tibial insert wear analysis. J Arthroplasty. 2011;26:497–503.PubMedCrossRef
31.
go back to reference Tsao AK, Jones LC, Lewallen DG. What patient and surgical factors contribute to implant wear and osteolysis in total joint arthroplasty. J Am Acad Orthop Surg. 2008;16:7–13. Tsao AK, Jones LC, Lewallen DG. What patient and surgical factors contribute to implant wear and osteolysis in total joint arthroplasty. J Am Acad Orthop Surg. 2008;16:7–13.
Metadata
Title
The 2012 Mark Coventry Award: A Retrieval Analysis of High Flexion versus Posterior-stabilized Tibial Inserts
Authors
Nicholas R. Paterson, BScH
Matthew G. Teeter, BScH, PhD
Steven J. MacDonald, MD, FRCSC
Richard W. McCalden, MD, MPhil(Edin), FRCSC
Douglas D. R. Naudie, MD, FRCSC
Publication date
01-01-2013
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 1/2013
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2387-3

Other articles of this Issue 1/2013

Clinical Orthopaedics and Related Research® 1/2013 Go to the issue

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Patient-specific Total Knee Arthroplasty Required Frequent Surgeon-directed Changes

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Complications of Total Knee Arthroplasty: Standardized List and Definitions of The Knee Society

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Determinants of Direct Medical Costs in Primary and Revision Total Knee Arthroplasty

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Do Fresh Osteochondral Allografts Successfully Treat Femoral Condyle Lesions?