Sustained low efficiency dialysis should not be interrupted for performing transpulmonary thermodilution measurements

Authors: Stefanie Geith, Lynne Stecher, Christian Rabe, Stefan Sack, Florian Eyer

Published in: Annals of Intensive Care | Issue 1/2018

Login to get access

Abstract

Background

Treatment of multiple organ failure frequently requires enhanced hemodynamic monitoring. When renal replacement is indicated, it remains unclear whether transpulmonary thermodilution (TPTD) measurements are influenced by renal replacement therapy (RRT) and whether RRT should be paused for TPTD measurements. Our aim was therefore to investigate the effect of pausing RRT on TPTD results in two dialysis catheter locations.

Materials and methods

In total, 62 TPTD measurements in 24 patients (APACHE: 32 ± 7 [mean ± standard deviation (SD)]) were performed using the PiCCO™ system (Pulsion, Germany). Patients were treated with sustained low efficiency dialysis (SLED; Genius™ system, Fresenius, Germany) as RRT. Measurements were taken during ongoing hemodialysis (HD, HDO), during paused HD (HDP) and immediately after termination of HD and blood restitution (HDT). Dialysis catheters were placed either in the superior vena cava (SVC, 19 times) or in the inferior vena cava (IVC, 5 times). Statistical analysis was performed to assess the effects of the measurement setting, SLED (blood flow rate) and the catheter location, on cardiac index (CI), global end-diastolic volume index (GEDVI) and extravascular lung water index (EVLWI) as measured by TPTD. Multilevel models were used for the analysis due to the triplicate measurements and due to 12 out of 19 SVC and 2 out of 5 IVC patients having more than one TPTD measured.

Results

CI and GEDVI were significantly higher at time point HDP compared to both HDO and HDT. In contrast, values for EVLWI were lower at HDP when compared to HDO and HDT. These findings were independent of the site of dialysis catheter insertion and blood flow rate.

Conclusions

PiCCO™ measurements assessed at paused SLED significantly deviate from ongoing and terminated SLED. Therefore, the dialysis system should not be paused for measurements. TPTD measurements in patients with PiCCO monitoring seem sufficiently reliable during ongoing SLED as well as after its termination. An effect of dialysis catheter location (SVC vs IVC) and blood flow rate on PiCCO™ measurements could not be shown.
Literature
1.
go back to reference Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK. Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg. 2010;110(3):799–811.CrossRef Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK. Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg. 2010;110(3):799–811.CrossRef
2.
go back to reference Marx G, Schuerholz T, Sumpelmann R, Simon T, Leuwer M. Comparison of cardiac output measurements by arterial trans-cardiopulmonary and pulmonary arterial thermodilution with direct Fick in septic shock. Eur J Anaesthesiol. 2005;22(2):129–34.CrossRef Marx G, Schuerholz T, Sumpelmann R, Simon T, Leuwer M. Comparison of cardiac output measurements by arterial trans-cardiopulmonary and pulmonary arterial thermodilution with direct Fick in septic shock. Eur J Anaesthesiol. 2005;22(2):129–34.CrossRef
3.
go back to reference Huber W, Fuchs S, Minning A, Kuchle C, Braun M, Beitz A, et al. Transpulmonary thermodilution (TPTD) before, during and after sustained low efficiency dialysis (SLED). A prospective study on feasibility of TPTD and prediction of successful fluid removal. PLoS One. 2016;11(4):e0153430.CrossRef Huber W, Fuchs S, Minning A, Kuchle C, Braun M, Beitz A, et al. Transpulmonary thermodilution (TPTD) before, during and after sustained low efficiency dialysis (SLED). A prospective study on feasibility of TPTD and prediction of successful fluid removal. PLoS One. 2016;11(4):e0153430.CrossRef
4.
go back to reference Martinez-Simon A, Monedero P, Cacho-Asenjo E. Erroneous measurement of haemodynamic parameters by PiCCO monitor in a critically ill patient with renal replacement therapy: a case report. Crit Care. 2006;10(2):410.CrossRef Martinez-Simon A, Monedero P, Cacho-Asenjo E. Erroneous measurement of haemodynamic parameters by PiCCO monitor in a critically ill patient with renal replacement therapy: a case report. Crit Care. 2006;10(2):410.CrossRef
5.
go back to reference Mason N, Froude A, Holst B, Saayman A, Findlay G. The effect of continuous veno-venous haemofiltration on PiCCO(®)haemodynamic parameters. Critical Care. 2005;9(Suppl 1):P65.CrossRef Mason N, Froude A, Holst B, Saayman A, Findlay G. The effect of continuous veno-venous haemofiltration on PiCCO(®)haemodynamic parameters. Critical Care. 2005;9(Suppl 1):P65.CrossRef
6.
go back to reference Neirynck V, Willems A, Peeters D, Van Regenmortel N, De laet I, Schoonheydt K, et al. Influence of continuous venovenous hemofiltration on transpulmonary thermodilution-derived parameters. Critical Care. 2009;13(1):P274.CrossRef Neirynck V, Willems A, Peeters D, Van Regenmortel N, De laet I, Schoonheydt K, et al. Influence of continuous venovenous hemofiltration on transpulmonary thermodilution-derived parameters. Critical Care. 2009;13(1):P274.CrossRef
7.
go back to reference Pathil A, Stremmel W, Schwenger V, Eisenbach C. The influence of haemodialysis on haemodynamic measurements using transpulmonary thermodilution in patients with septic shock: an observational study. Eur J Anaesthesiol. 2013;30(1):16–20.CrossRef Pathil A, Stremmel W, Schwenger V, Eisenbach C. The influence of haemodialysis on haemodynamic measurements using transpulmonary thermodilution in patients with septic shock: an observational study. Eur J Anaesthesiol. 2013;30(1):16–20.CrossRef
8.
go back to reference van Craenenbroeck A, VanIngelgem A, Palmers PJ, Van Regenmortel N, De Laet I, Schoonheydt K, et al. Influence of continuous venovenous hemofiltration (CVVH) and catheter position on transpulmonary thermodilution derived parameters with PiCCO. Intensive Care Med. 2010;36(Suppl 2):S294. van Craenenbroeck A, VanIngelgem A, Palmers PJ, Van Regenmortel N, De Laet I, Schoonheydt K, et al. Influence of continuous venovenous hemofiltration (CVVH) and catheter position on transpulmonary thermodilution derived parameters with PiCCO. Intensive Care Med. 2010;36(Suppl 2):S294.
9.
go back to reference Huber W, Rockmann F. Invasive und nichtinvasive Möglichkeiten des haemodynamischen Monitorings. Intensivmed. 2008;45:337–59.CrossRef Huber W, Rockmann F. Invasive und nichtinvasive Möglichkeiten des haemodynamischen Monitorings. Intensivmed. 2008;45:337–59.CrossRef
10.
go back to reference Sakka SG, Hanusch T, Thuemer O, Wegscheider K. The influence of venovenous renal replacement therapy on measurements by the transpulmonary thermodilution technique. Anesth Analg. 2007;105(4):1079–82.CrossRef Sakka SG, Hanusch T, Thuemer O, Wegscheider K. The influence of venovenous renal replacement therapy on measurements by the transpulmonary thermodilution technique. Anesth Analg. 2007;105(4):1079–82.CrossRef
11.
go back to reference Dufour N, Delville M, Teboul JL, Camous L, Favier du Noyer A, Richard C, et al. Transpulmonary thermodilution measurements are not affected by continuous veno-venous hemofiltration at high blood pump flow. Intensive Care Med. 2012;38(7):1162–8.CrossRef Dufour N, Delville M, Teboul JL, Camous L, Favier du Noyer A, Richard C, et al. Transpulmonary thermodilution measurements are not affected by continuous veno-venous hemofiltration at high blood pump flow. Intensive Care Med. 2012;38(7):1162–8.CrossRef
12.
go back to reference Heise D, Faulstich M, Morer O, Brauer A, Quintel M. Influence of continuous renal replacement therapy on cardiac output measurement using thermodilution techniques. Minerva Anestesiol. 2012;78(3):315–21.PubMed Heise D, Faulstich M, Morer O, Brauer A, Quintel M. Influence of continuous renal replacement therapy on cardiac output measurement using thermodilution techniques. Minerva Anestesiol. 2012;78(3):315–21.PubMed
13.
go back to reference R Core Team. R: A language and environment for statistical computing. R Foundation for Statistical Computing. Vienna, Austria 2014. R Core Team. R: A language and environment for statistical computing. R Foundation for Statistical Computing. Vienna, Austria 2014.
14.
go back to reference Monnet X, Persichini R, Ktari M, Jozwiak M, Richard C, Teboul J-L. Precision of the transpulmonary thermodilution measurements. Crit Care. 2011;15(4):R204.CrossRef Monnet X, Persichini R, Ktari M, Jozwiak M, Richard C, Teboul J-L. Precision of the transpulmonary thermodilution measurements. Crit Care. 2011;15(4):R204.CrossRef
15.
go back to reference Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109.CrossRef Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109.CrossRef
16.
go back to reference Monnet X, Osman D, Ridel C, Lamia B, Richard C, Teboul JL. Predicting volume responsiveness by using the end-expiratory occlusion in mechanically ventilated intensive care unit patients. Crit Care Med. 2009;37(3):951–6.CrossRef Monnet X, Osman D, Ridel C, Lamia B, Richard C, Teboul JL. Predicting volume responsiveness by using the end-expiratory occlusion in mechanically ventilated intensive care unit patients. Crit Care Med. 2009;37(3):951–6.CrossRef
17.
go back to reference Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35(1):85–90.CrossRef Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35(1):85–90.CrossRef
18.
go back to reference Haller M, Zollner C, Manert W, Briegel J, Kilger E, Polasek J, et al. Thermodilution cardiac output may be incorrect in patients on venovenous extracorporeal lung assist. Am J Respir Crit Care Med. 1995;152(6 Pt 1):1812–7.CrossRef Haller M, Zollner C, Manert W, Briegel J, Kilger E, Polasek J, et al. Thermodilution cardiac output may be incorrect in patients on venovenous extracorporeal lung assist. Am J Respir Crit Care Med. 1995;152(6 Pt 1):1812–7.CrossRef
Metadata
Title
Sustained low efficiency dialysis should not be interrupted for performing transpulmonary thermodilution measurements
Authors
Stefanie Geith
Lynne Stecher
Christian Rabe
Stefan Sack
Florian Eyer
Publication date
01-12-2018
Publisher
Springer International Publishing
Published in
Annals of Intensive Care / Issue 1/2018
Electronic ISSN: 2110-5820
DOI
https://doi.org/10.1186/s13613-018-0455-x

Other articles of this Issue 1/2018

Annals of Intensive Care 1/2018 Go to the issue