Significance of intraoperative testing in right-sided implantable cardioverter-defibrillators

Authors: Andreas Keyser, Michael K Hilker, Ekrem Ücer, Sigrid Wittmann, Christof Schmid, Claudius Diez

Published in: Journal of Cardiothoracic Surgery | Issue 1/2013

Login to get access

Abstract

Background

Implantation of implantable cardioverter-defibrillators (ICD) from the left pectoral region is the standard therapeutical method. Increasing numbers of system revisions due to lead dysfunction and infections will consecutively increase the numbers of right-sided implantations. The reliability of devices implanted on the right pectoral side remains controversially discussed, and the question of testing these devices remains unanswered.

Methods

In a prospectively designed study all 870 patients (60.0±14 years, 689 male) who were treated with a first ICD from July 2005 until May 2012 and tested intraoperatively according to the testing protocol were analyzed. The indication for implantation was primary prophylactic in 71.5%. Underlying diseases included ischemic cardiomyopathy (50%), dilative cardiomyopathy (37%), and others (13%). Mean ejection faction was 27±12%. Implantation site was right in 4.5% and left in 95.5%.

Results

Five patients supplied with right-sided ICD (13%, p = 0.02 as compared to left-sided) failed initial intraoperative testing with 21 J. 3 patients were male. The age of the patients failing intraoperative testing with right-sided devices appeared higher than of patients with left-sided devices (p = 0.07). The ejection fraction was 28±8%. All patients reached a sufficient DFT ≤ 21 J after corrective procedures.

Conclusion

Implantation of ICDs on the right side results in significantly higher failure rate of successful termination of intraoperatively induced ventricular fibrillation. The data of our study suggest the necessity of intraoperative ICD testing in right-sided implanted ICDs.
Literature
1.
go back to reference Dickstein K, Cohen-Solal A, Filippatos G, McMurray JJ, Ponikowski P, Pole-Wilson PA: ESC guidelines for the diagnosis and treatment of acute and chronic heart failure2008: the task force for the diagnosis and treatment of acute and chronic heart failure 2008 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association of the ESC (HFA) and endorsed by the European Society of Intensive Care Medicine (ESICM). Eur Heart J. 2008, 29: 2388-2442.CrossRefPubMed Dickstein K, Cohen-Solal A, Filippatos G, McMurray JJ, Ponikowski P, Pole-Wilson PA: ESC guidelines for the diagnosis and treatment of acute and chronic heart failure2008: the task force for the diagnosis and treatment of acute and chronic heart failure 2008 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association of the ESC (HFA) and endorsed by the European Society of Intensive Care Medicine (ESICM). Eur Heart J. 2008, 29: 2388-2442.CrossRefPubMed
2.
go back to reference Epstein AE, Kay GN, Plumb VJ, Voshage-Stahl L, Hull M: Elevated defibrillation threshold when right-sided venous access is used for nonthoracotomy ICD lead implantation. J Cardiovesc Electrophysiol. 1995, 11: 979-CrossRef Epstein AE, Kay GN, Plumb VJ, Voshage-Stahl L, Hull M: Elevated defibrillation threshold when right-sided venous access is used for nonthoracotomy ICD lead implantation. J Cardiovesc Electrophysiol. 1995, 11: 979-CrossRef
3.
go back to reference Natale A, Sra J, Geiger MJ, Newby K, Akhtar M, Pacifico A: Right side implant of the unipolar single lead defibrillation system. Pace. 1997, 20: 1910-1912. 10.1111/j.1540-8159.1997.tb03595.x.CrossRefPubMed Natale A, Sra J, Geiger MJ, Newby K, Akhtar M, Pacifico A: Right side implant of the unipolar single lead defibrillation system. Pace. 1997, 20: 1910-1912. 10.1111/j.1540-8159.1997.tb03595.x.CrossRefPubMed
4.
go back to reference Jensen SM, Pietersen A, Chen X: Implantation of Can Implantable Defibrillators in the right pectoral region. Pace. 1998, 21: 476-477. 10.1111/j.1540-8159.1998.tb00078.x.CrossRefPubMed Jensen SM, Pietersen A, Chen X: Implantation of Can Implantable Defibrillators in the right pectoral region. Pace. 1998, 21: 476-477. 10.1111/j.1540-8159.1998.tb00078.x.CrossRefPubMed
5.
go back to reference Roberts PR, Allen S, Betts T, Urban JF, Whitman T, Euler DE, Kallok MJ, Morgan JM: Increased defibrillation threshold with right-sided active pectoral can. J Intervent Cardiac Electrophysiol. 2000, 2: 245-249.CrossRef Roberts PR, Allen S, Betts T, Urban JF, Whitman T, Euler DE, Kallok MJ, Morgan JM: Increased defibrillation threshold with right-sided active pectoral can. J Intervent Cardiac Electrophysiol. 2000, 2: 245-249.CrossRef
6.
go back to reference Fonteyne W, Rottiers H, Tavernier R: Experience with right sided implantation of transvenous active can cardioverter defibrillators. Europace. 2001, 2 (Suppl 1): A30- Fonteyne W, Rottiers H, Tavernier R: Experience with right sided implantation of transvenous active can cardioverter defibrillators. Europace. 2001, 2 (Suppl 1): A30-
7.
go back to reference Kirk MM, Shorofski SR, Gold MR: Comparison of the effects of active left and right pectoral pulse generators on defibrillation efficacy. Am J Cardiol. 2001, 88: 1308-1311. 10.1016/S0002-9149(01)02096-3.CrossRefPubMed Kirk MM, Shorofski SR, Gold MR: Comparison of the effects of active left and right pectoral pulse generators on defibrillation efficacy. Am J Cardiol. 2001, 88: 1308-1311. 10.1016/S0002-9149(01)02096-3.CrossRefPubMed
8.
go back to reference Gold MR, Shih HT, Herre J, Breiter D, Zhang Y, Schwartz M: Comparison of defibrillation efficacy and survival associated with right versus left pectoral placement for implantable defibrillators. Am J Cardiol. 2007, 100: 243-246. 10.1016/j.amjcard.2007.02.087.CrossRefPubMed Gold MR, Shih HT, Herre J, Breiter D, Zhang Y, Schwartz M: Comparison of defibrillation efficacy and survival associated with right versus left pectoral placement for implantable defibrillators. Am J Cardiol. 2007, 100: 243-246. 10.1016/j.amjcard.2007.02.087.CrossRefPubMed
9.
go back to reference Moss AJ, Zareba WJ, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML: Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002, 346: 877-883. 10.1056/NEJMoa013474.CrossRefPubMed Moss AJ, Zareba WJ, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML: Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002, 346: 877-883. 10.1056/NEJMoa013474.CrossRefPubMed
10.
go back to reference Greenspon AJ, Patel JD, Lau E, Ochoa J, Frisch DR, Ho RT, Pavri BB, Kurtz SM: 16-year tends in the infection burden of pacemakers and implantable cardioveter-defibrillators in the United States. J Am Coll Cardiol. 2011, 58: 1001-1006. 10.1016/j.jacc.2011.04.033.CrossRefPubMed Greenspon AJ, Patel JD, Lau E, Ochoa J, Frisch DR, Ho RT, Pavri BB, Kurtz SM: 16-year tends in the infection burden of pacemakers and implantable cardioveter-defibrillators in the United States. J Am Coll Cardiol. 2011, 58: 1001-1006. 10.1016/j.jacc.2011.04.033.CrossRefPubMed
11.
go back to reference Patton K, Reddy V: Optimizing the defibrillation vector with a right-sided implantable cardioverter-defibrillator. Hear Rhythm. 2004, 6: 756-CrossRef Patton K, Reddy V: Optimizing the defibrillation vector with a right-sided implantable cardioverter-defibrillator. Hear Rhythm. 2004, 6: 756-CrossRef
12.
go back to reference Ideker RE, Epstein AE, Plump VJ: Should shocks still be administered during impantable cardioverter-defibrillator insertion?. Hear Rhythm. 2005, 2: 462-463. 10.1016/j.hrthm.2005.02.004.CrossRef Ideker RE, Epstein AE, Plump VJ: Should shocks still be administered during impantable cardioverter-defibrillator insertion?. Hear Rhythm. 2005, 2: 462-463. 10.1016/j.hrthm.2005.02.004.CrossRef
13.
go back to reference Day JD, Doshi RN, Belott P, Birgersdotter-Green U, Behboodikhah M, Ott P, Glatter KA, Tobias S, Frumin H, Lee BK, Merillat J, Wiener I, Wang S, Grogin H, Chun S, Patrawalla R, Crandall B, Osborn JS, Weiss JP, Lappe DL, Neuman S: Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (arrhythmia single shock defibrillation threshold testing versus upper limit of vulnerability: risk reduction evaluation with implantable cardioverter-defibrillator implantations). Circulation. 2007, 115: 2382-2389. 10.1161/CIRCULATIONAHA.106.663112.CrossRefPubMed Day JD, Doshi RN, Belott P, Birgersdotter-Green U, Behboodikhah M, Ott P, Glatter KA, Tobias S, Frumin H, Lee BK, Merillat J, Wiener I, Wang S, Grogin H, Chun S, Patrawalla R, Crandall B, Osborn JS, Weiss JP, Lappe DL, Neuman S: Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (arrhythmia single shock defibrillation threshold testing versus upper limit of vulnerability: risk reduction evaluation with implantable cardioverter-defibrillator implantations). Circulation. 2007, 115: 2382-2389. 10.1161/CIRCULATIONAHA.106.663112.CrossRefPubMed
14.
go back to reference Russo AM, Sauer W, Gerstenfeld EP, Hsia HH, Lin D, Cooper JM, Dixit S, Verdino RJ, Nayak HM, Callans DJ, Patel V, Marchlinski FE: Defibrillation threshold testing: is it really necessary at the time of implantable cardioverter-defibrillator insertion?. Hear Rhythm. 2005, 2: 456-461. 10.1016/j.hrthm.2005.01.015.CrossRef Russo AM, Sauer W, Gerstenfeld EP, Hsia HH, Lin D, Cooper JM, Dixit S, Verdino RJ, Nayak HM, Callans DJ, Patel V, Marchlinski FE: Defibrillation threshold testing: is it really necessary at the time of implantable cardioverter-defibrillator insertion?. Hear Rhythm. 2005, 2: 456-461. 10.1016/j.hrthm.2005.01.015.CrossRef
15.
go back to reference Blatt JA, Poole JE, Johnson GW, Callans DJ, Raitt MH, Reddy RK, Marchlinski FE, Yee R, Guarnieri T, Talajic M, Wilber DJ, Anderson J, Chung K, Wong WS, Mark DB, Lee KL, Bardy GH: SCD-HeFT Investigators. No benefit from defibrillation threshold testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial). J Am Coll Cardiol. 2008, 52: 551-556. 10.1016/j.jacc.2008.04.051.CrossRefPubMed Blatt JA, Poole JE, Johnson GW, Callans DJ, Raitt MH, Reddy RK, Marchlinski FE, Yee R, Guarnieri T, Talajic M, Wilber DJ, Anderson J, Chung K, Wong WS, Mark DB, Lee KL, Bardy GH: SCD-HeFT Investigators. No benefit from defibrillation threshold testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial). J Am Coll Cardiol. 2008, 52: 551-556. 10.1016/j.jacc.2008.04.051.CrossRefPubMed
16.
go back to reference Gold M-R, Olsovsky MR, DeGroot PJ, Cuello C, Shorofsky SR: Optimization of transvenous coil position for active can defibrillation thresholds. J Cardiovasc Electrophysiol. 2000, 11: 25-29. 10.1111/j.1540-8167.2000.tb00731.x.CrossRefPubMed Gold M-R, Olsovsky MR, DeGroot PJ, Cuello C, Shorofsky SR: Optimization of transvenous coil position for active can defibrillation thresholds. J Cardiovasc Electrophysiol. 2000, 11: 25-29. 10.1111/j.1540-8167.2000.tb00731.x.CrossRefPubMed
17.
go back to reference Rashba EJ, Bonner M, Wilson J, Shorofsky SR, Peters RW, Gold MR: Distal right ventricular coil position reduces defibrillation thresholds. J Cardiovasc Electrophysiol. 2003, 14: 1036-1040. 10.1046/j.1540-8167.2003.03205.x.CrossRefPubMed Rashba EJ, Bonner M, Wilson J, Shorofsky SR, Peters RW, Gold MR: Distal right ventricular coil position reduces defibrillation thresholds. J Cardiovasc Electrophysiol. 2003, 14: 1036-1040. 10.1046/j.1540-8167.2003.03205.x.CrossRefPubMed
18.
go back to reference Usui M, Walcott GP, KenKnight BH, Walker RG, Rollins DL, Smith WM, Ideker RE: Influence of malpositioned transvenous leads on defibrillation efficacy with and without a subcutaneous array electrode. Pacing Clin Electrophysiol. 1995, 18: 2008-2016. 10.1111/j.1540-8159.1995.tb03861.x.CrossRefPubMed Usui M, Walcott GP, KenKnight BH, Walker RG, Rollins DL, Smith WM, Ideker RE: Influence of malpositioned transvenous leads on defibrillation efficacy with and without a subcutaneous array electrode. Pacing Clin Electrophysiol. 1995, 18: 2008-2016. 10.1111/j.1540-8159.1995.tb03861.x.CrossRefPubMed
19.
go back to reference Atary JZ, Borleffs CJ, van der Bom JG, Trines SA, Bootsma M, Zeppenfeld K, van Erven L, Schalij MJ: Right ventricular stimulation threshold at ICD implant predicts device therapy in primary prevention patients with ischaemic heart disease. Europace. 2010, 12: 1581-1588. 10.1093/europace/euq266.CrossRefPubMed Atary JZ, Borleffs CJ, van der Bom JG, Trines SA, Bootsma M, Zeppenfeld K, van Erven L, Schalij MJ: Right ventricular stimulation threshold at ICD implant predicts device therapy in primary prevention patients with ischaemic heart disease. Europace. 2010, 12: 1581-1588. 10.1093/europace/euq266.CrossRefPubMed
20.
go back to reference Glikson M, Suleiman M, Luria DM, Martin ML, Hodge DO, Shen WK, Bradley DJ, Munger TM, Rea RF, Hayes DL, Hammill SC, Friedman PA: Do abandoned leads pose risk to implantable cardioverter-defibrillator patients?. Hear Rhythm. 2009, 6: 65-68. 10.1016/j.hrthm.2008.10.012.CrossRef Glikson M, Suleiman M, Luria DM, Martin ML, Hodge DO, Shen WK, Bradley DJ, Munger TM, Rea RF, Hayes DL, Hammill SC, Friedman PA: Do abandoned leads pose risk to implantable cardioverter-defibrillator patients?. Hear Rhythm. 2009, 6: 65-68. 10.1016/j.hrthm.2008.10.012.CrossRef
21.
go back to reference Napp A, Birkenhauer F, Lee KM, Gemein C, Zink M, Knackstedt C, Marx N, Schauerte P, Mischke K: ICD-Test: Is the evaluation of the defibrillationthreshold (DFT) still up to date? The ATLANTISRegister. Eur Heart J. 2010, 31: 928-929. Napp A, Birkenhauer F, Lee KM, Gemein C, Zink M, Knackstedt C, Marx N, Schauerte P, Mischke K: ICD-Test: Is the evaluation of the defibrillationthreshold (DFT) still up to date? The ATLANTISRegister. Eur Heart J. 2010, 31: 928-929.
22.
go back to reference Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, Canby RC, Khalighi K, Machado C, Rubenstein DS, Volosin KJ, PainFREE Rx II: Investigators. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing fast ventricular tachycardia reduces shock therapies (PainFREE Rx II) trial results. Circulation. 2004, 110: 2591-2596. 10.1161/01.CIR.0000145610.64014.E4.CrossRefPubMed Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, Canby RC, Khalighi K, Machado C, Rubenstein DS, Volosin KJ, PainFREE Rx II: Investigators. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing fast ventricular tachycardia reduces shock therapies (PainFREE Rx II) trial results. Circulation. 2004, 110: 2591-2596. 10.1161/01.CIR.0000145610.64014.E4.CrossRefPubMed
23.
go back to reference Viskin S, Rosse R: The top 10 reasons to avoid defibrillation threshold testing during ICD implantation. HeartRhythm. 2008, 5: 391-393. Viskin S, Rosse R: The top 10 reasons to avoid defibrillation threshold testing during ICD implantation. HeartRhythm. 2008, 5: 391-393.
24.
go back to reference Alter P, Waldhans S, Plachta E, Moosdorf R, Grimm W: Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing Clin Electrophysiol. 2005, 28: 926-932. 10.1111/j.1540-8159.2005.00195.x.CrossRefPubMed Alter P, Waldhans S, Plachta E, Moosdorf R, Grimm W: Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing Clin Electrophysiol. 2005, 28: 926-932. 10.1111/j.1540-8159.2005.00195.x.CrossRefPubMed
25.
go back to reference Birnie D, Tung S, Simpson C, Crystal E, Exner D, Ayala Paredes FA, Krahn A, Parkash R, Khaykin Y, Philippon F, Guerra P, Kimber S, Cameron D, Healey JS: Complications associated with defibrillation threshold testing: the Canadian experience. Hear Rhythm. 2008, 5: 391-393. 10.1016/j.hrthm.2008.01.006.CrossRef Birnie D, Tung S, Simpson C, Crystal E, Exner D, Ayala Paredes FA, Krahn A, Parkash R, Khaykin Y, Philippon F, Guerra P, Kimber S, Cameron D, Healey JS: Complications associated with defibrillation threshold testing: the Canadian experience. Hear Rhythm. 2008, 5: 391-393. 10.1016/j.hrthm.2008.01.006.CrossRef
26.
go back to reference Brignole M, Raciti G, Bongiorni MG, De Martino G, Favale S, Gaparini M, Luise R: Defibrillation testing at the time of implantation of cardioverter defibrillation in the clinical practice: a nation-wide survey. Europace. 2007, 9: 540-543. 10.1093/europace/eum083.CrossRefPubMed Brignole M, Raciti G, Bongiorni MG, De Martino G, Favale S, Gaparini M, Luise R: Defibrillation testing at the time of implantation of cardioverter defibrillation in the clinical practice: a nation-wide survey. Europace. 2007, 9: 540-543. 10.1093/europace/eum083.CrossRefPubMed
27.
go back to reference Kolb C, Tzeis S, Zrenner B: Defibrillation threshold testing: tradition or necessity?. Pacing Clin Electrophysiol. 2009, 32: 570-572. 10.1111/j.1540-8159.2009.02328.x.CrossRefPubMed Kolb C, Tzeis S, Zrenner B: Defibrillation threshold testing: tradition or necessity?. Pacing Clin Electrophysiol. 2009, 32: 570-572. 10.1111/j.1540-8159.2009.02328.x.CrossRefPubMed
28.
go back to reference Healey JS, Birnie DH, Lee DS, Krahn AD, Crystal E, Simpson CS, Dorian P, Chen Z, Cameron D, Verma A, Connolly SJ, Gula LJ, Lockwood E, Nair G, Tu JV: Defibrillation testing at the time of ICD insertion: an analysis from the Ontario ICD registry. J Cardiovasc Electrophysiol. 2010, 21: 1344-1348. 10.1111/j.1540-8167.2010.01850.x.CrossRefPubMed Healey JS, Birnie DH, Lee DS, Krahn AD, Crystal E, Simpson CS, Dorian P, Chen Z, Cameron D, Verma A, Connolly SJ, Gula LJ, Lockwood E, Nair G, Tu JV: Defibrillation testing at the time of ICD insertion: an analysis from the Ontario ICD registry. J Cardiovasc Electrophysiol. 2010, 21: 1344-1348. 10.1111/j.1540-8167.2010.01850.x.CrossRefPubMed
29.
go back to reference Strickberger SA, Klein GJ: Is defibrillation testing required for defibrillator implantation?. J Am Coll Cardiol. 2004, 44: 88-91. 10.1016/j.jacc.2003.11.068.CrossRefPubMed Strickberger SA, Klein GJ: Is defibrillation testing required for defibrillator implantation?. J Am Coll Cardiol. 2004, 44: 88-91. 10.1016/j.jacc.2003.11.068.CrossRefPubMed
Metadata
Title
Significance of intraoperative testing in right-sided implantable cardioverter-defibrillators
Authors
Andreas Keyser
Michael K Hilker
Ekrem Ücer
Sigrid Wittmann
Christof Schmid
Claudius Diez
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2013
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-8-77

Other articles of this Issue 1/2013

Journal of Cardiothoracic Surgery 1/2013 Go to the issue