Skip to main content
Top
Published in: HAND 3/2015

01-09-2015

Routine use of locking shaft screws is not necessary in volar plate fixation of distal radius fractures

Authors: Kevin Lutsky, C. Edward Hoffler, Nayoung Kim, Jonas L. Matzon

Published in: HAND | Issue 3/2015

Login to get access

Abstract

Hypothesis

Volar locking plate fixation is a common treatment method for distal radius fractures. Recently, implants have been designed with an option to use locking screws in the shaft portion of the plate. While there is a high incidence of low bone mineral density in patients who sustain fragility fractures of the distal radius, the need for locking shaft screws is not well defined. Our hypothesis is that the routine use of locking screws in the shaft portion of volar plates is not required to maintain reduction or to prevent hardware failure.

Methods

A retrospective review was performed in all patients over age 50 years who underwent volar plate fixation using an implant with non-locking shaft screws for a distal radius fracture during a 2-year period. Patients were permitted to perform early range of motion exercises. Radiographs were examined and measurements were obtained to assess maintenance of reduction and incidence of hardware failure. Patients were followed at least until fracture healing. Patients were excluded from analysis if locking shaft screws were utilized or if follow-up was inadequate.

Results

Forty-one patients met the inclusion criteria. The average age was 62 years (range 50–79). There were 12 men and 29 women. The implant used incorporated 3.5-mm shaft screws in 26 patients and 2.4-mm shaft screws in 15 patients. All patients healed within acceptable radiographic parameters (mean volar tilt = 4.9°, mean radial inclination = 21.7°, mean radial height = 11.6 mm). There were not any instances of hardware failure.

Discussion

Distal radius fractures frequently occur in patients with low bone mineral density. Non-locking, bicortically placed shaft screws provide adequate stability to allow for early range of motion without loss of reduction or hardware failure. The routine use of locking screws in the shaft portion of volar plates does not appear justified.
Literature
1.
go back to reference Anglen J, Kyle RF, Marsh JL, Virkus WW, Watters 3rd WC, Keith MW, et al. Locking plates for extremity fractures. J Am Acad Orthop Surg. 2009;17:465–72.PubMed Anglen J, Kyle RF, Marsh JL, Virkus WW, Watters 3rd WC, Keith MW, et al. Locking plates for extremity fractures. J Am Acad Orthop Surg. 2009;17:465–72.PubMed
2.
go back to reference Berglund LM, Messer TM. Complications of volar plate fixation for managing distal radius fractures. J Am Acad Orthop Surg. 2009;17:369–77.PubMed Berglund LM, Messer TM. Complications of volar plate fixation for managing distal radius fractures. J Am Acad Orthop Surg. 2009;17:369–77.PubMed
3.
go back to reference Haidukewych GJ. Innovations in locking plate technology. J Am Acad Orthop Surg. 2004;12:205–12.PubMed Haidukewych GJ. Innovations in locking plate technology. J Am Acad Orthop Surg. 2004;12:205–12.PubMed
4.
go back to reference Harness NG, Funahashi T, Dell R, Adams AL, Burchette R, Chen X, et al. Distal radius fracture risk reduction with a comprehensive osteoporosis management program. J Hand Surg [Am]. 2012;37:1543–9.CrossRef Harness NG, Funahashi T, Dell R, Adams AL, Burchette R, Chen X, et al. Distal radius fracture risk reduction with a comprehensive osteoporosis management program. J Hand Surg [Am]. 2012;37:1543–9.CrossRef
5.
go back to reference Koval KJ, Harrast JJ, Anglen JO, Weinstein JN. Fractures of the distal part of the radius. The evolution of practice over time. Where’s the evidence? J Bone Joint Surg Am. 2008;90:1855–61.CrossRefPubMed Koval KJ, Harrast JJ, Anglen JO, Weinstein JN. Fractures of the distal part of the radius. The evolution of practice over time. Where’s the evidence? J Bone Joint Surg Am. 2008;90:1855–61.CrossRefPubMed
6.
go back to reference Kreder HJ, Hanel DP, McKee M, Jupiter J, McGillivary G, Swiontkowski MF. Consistency of AO fracture classification for the distal radius. J Bone Joint Surg (Br). 1996;78:726–31. Kreder HJ, Hanel DP, McKee M, Jupiter J, McGillivary G, Swiontkowski MF. Consistency of AO fracture classification for the distal radius. J Bone Joint Surg (Br). 1996;78:726–31.
7.
go back to reference Kubiak EN, Fulkerson E, Strauss E, Egol KA. The evolution of locked plates. J Bone Joint Surg Am. 2006;88 Suppl 4:189–200.CrossRefPubMed Kubiak EN, Fulkerson E, Strauss E, Egol KA. The evolution of locked plates. J Bone Joint Surg Am. 2006;88 Suppl 4:189–200.CrossRefPubMed
8.
go back to reference Rozental TD, Makhni EC, Day CS, Bouxsein ML. Improving evaluation and treatment for osteoporosis following distal radial fractures. A prospective randomized intervention. J Bone Joint Surg Am. 2008;90:953–61.CrossRefPubMed Rozental TD, Makhni EC, Day CS, Bouxsein ML. Improving evaluation and treatment for osteoporosis following distal radial fractures. A prospective randomized intervention. J Bone Joint Surg Am. 2008;90:953–61.CrossRefPubMed
9.
go back to reference Wall LB, Brodt MD, Silva MJ, Boyer MI, Calfee RP. The effects of screw length on stability of simulated osteoporotic distal radius fractures fixed with volar locking plates. J Hand Surg [Am]. 2012;37:446–53.CrossRef Wall LB, Brodt MD, Silva MJ, Boyer MI, Calfee RP. The effects of screw length on stability of simulated osteoporotic distal radius fractures fixed with volar locking plates. J Hand Surg [Am]. 2012;37:446–53.CrossRef
Metadata
Title
Routine use of locking shaft screws is not necessary in volar plate fixation of distal radius fractures
Authors
Kevin Lutsky
C. Edward Hoffler
Nayoung Kim
Jonas L. Matzon
Publication date
01-09-2015
Publisher
Springer US
Published in
HAND / Issue 3/2015
Print ISSN: 1558-9447
Electronic ISSN: 1558-9455
DOI
https://doi.org/10.1007/s11552-014-9722-y

Other articles of this Issue 3/2015

HAND 3/2015 Go to the issue