Biomechanical evaluation of double-strand (looped) and single-strand polyamide multifilament suture: influence of knot and suture size

Authors: David T. Netscher, Justin J. Badal, Jonathan Yang, Yoav Kaufman, Jerry Alexander, Philip Noble

Published in: HAND | Issue 3/2015

Login to get access

Abstract

Background

Flexor tendon repair in zone II remains a vexing problem. Repair techniques have been developed to strengthen and optimize the number of core strands crossing a repair. A polyamide looped suture doubles the number of core strands for every needle path. This simplifies repairs, but the knot remains a potential weakness. The purpose of our study was to create a biomechanical model used to evaluate the bulky knot of a looped suture as it may be weaker, resulting in greater deformation.

Methods

Using machined steel rods to hold our suture constructs, we compared four different knot configurations using looped and non-looped sutures in 3-0 and 4-0 varieties using a four-core strand technique. The constructs were tested under increased cyclic loading recording both forces applied and suture construct lengthening (“clinical gapping”) and ultimate breaking strength.

Results

During continuous periods of cyclic loading, we measured permanent deformation and ultimate breaking strength. Permanent deformation results when there is no recoverable change after force removal defined as a permanent rod separation (or gapping) of 2 mm.
Four-strand 3-0 and 4-0 looped sutures failed at 39.9 and 27.1 N faring worse than a four-strand non-looped suture which reached a rod separation of 2 mm at 60.7 and 41.3 N. The ultimate breaking strength demonstrated absolute failure (construct rupture) with the 3-0 looped suture breaking at the knot at 50.3 N and the non-looped suture at 61.5 N. For the 4-0 suture, these values were 32.4 and 41.76 N.

Conclusion

Within the constraints of this model, a looped suture fared worse than a non-looped suture especially when comparing 4-0 and 3-0 sutures. However, two-knot 3-0 looped suture constructs did resist the force generally accepted as occurring with early non-resistive tendon motion protocols, while two-knot 4-0 looped suture constructs did not.

Clinical Relevance

This paper provides a description of a model to evaluate various suture materials and knot strengths in isolation of the tendon itself. This allowed us to evaluate mechanical differences between looped and non-looped sutures for polyamide, which are commonly used in flexor tendon repair. These differences between sutures may impact choices for a suture type selected for these repairs.
Literature
1.
go back to reference Savage R, Risitano G. Flexor tendon repair using a “six strand” method of repair and early active mobilisation. J Hand Surg Br Eur. 1989;14(4):396–9.CrossRef Savage R, Risitano G. Flexor tendon repair using a “six strand” method of repair and early active mobilisation. J Hand Surg Br Eur. 1989;14(4):396–9.CrossRef
2.
go back to reference Trumble TE. Zone-II flexor tendon repair: a randomized prospective trial of active place-and-hold therapy compared with passive motion therapy. J Bone Joint Surg Am. 2010;92(6):1381.CrossRefPubMed Trumble TE. Zone-II flexor tendon repair: a randomized prospective trial of active place-and-hold therapy compared with passive motion therapy. J Bone Joint Surg Am. 2010;92(6):1381.CrossRefPubMed
3.
go back to reference Yalcin L. Biomechanical assessment of suture techniques used for tendon repair. Acta Orthop Traumatol Turc. 2011;45(6):453–7.CrossRefPubMed Yalcin L. Biomechanical assessment of suture techniques used for tendon repair. Acta Orthop Traumatol Turc. 2011;45(6):453–7.CrossRefPubMed
4.
go back to reference Bhatia D, Tanner KE, Bonfield W, Citron ND. Factors affecting the strength of flexor tendon repair. J Hand Surg Br Eur. 1992;17B(5):550–2.CrossRef Bhatia D, Tanner KE, Bonfield W, Citron ND. Factors affecting the strength of flexor tendon repair. J Hand Surg Br Eur. 1992;17B(5):550–2.CrossRef
5.
go back to reference Dogramaci Y, Kalaci A, Sevinç TT, Esen E, Komurcu M, Yanat AN. Does strand configuration and number of purchase points affect the biomechanical behavior of a tendon repair? A biomechanical evaluation using different Kessler methods of flexor tendon repair. Hand. 2008;3(3):266–70.PubMedCentralCrossRefPubMed Dogramaci Y, Kalaci A, Sevinç TT, Esen E, Komurcu M, Yanat AN. Does strand configuration and number of purchase points affect the biomechanical behavior of a tendon repair? A biomechanical evaluation using different Kessler methods of flexor tendon repair. Hand. 2008;3(3):266–70.PubMedCentralCrossRefPubMed
6.
go back to reference Rees L, Matthews A, Masouros SD, Bull AMJ, Haywood R. Comparison of 1- and 2-knot, 4-strand, double-modified Kessler tendon repairs in a porcine model. J Hand Surg. 2009;34(4):705–9.CrossRef Rees L, Matthews A, Masouros SD, Bull AMJ, Haywood R. Comparison of 1- and 2-knot, 4-strand, double-modified Kessler tendon repairs in a porcine model. J Hand Surg. 2009;34(4):705–9.CrossRef
7.
go back to reference Timothy Thurman R, Trumble TE, Hanel DP, Tencer AF, Kiser PK. Two-, four-, and six-strand zone II flexor tendon repairs: an in situ biomechanical comparison using a cadaver model. J Hand Surg. 1998;23(2):261–5.CrossRef Timothy Thurman R, Trumble TE, Hanel DP, Tencer AF, Kiser PK. Two-, four-, and six-strand zone II flexor tendon repairs: an in situ biomechanical comparison using a cadaver model. J Hand Surg. 1998;23(2):261–5.CrossRef
8.
go back to reference Wagner Jr WF, Carroll IV C, Strickland JW, Heck DA, Toombs JP. A biomechanical comparison of techniques of flexor tendon repair. J Hand Surg. 1994;19(6):979–83.CrossRef Wagner Jr WF, Carroll IV C, Strickland JW, Heck DA, Toombs JP. A biomechanical comparison of techniques of flexor tendon repair. J Hand Surg. 1994;19(6):979–83.CrossRef
9.
go back to reference McClellan WT, Schessler MJ, Ruch DS, Levin LS, Goldner RD. A knotless flexor tendon repair technique using a bidirectional barbed suture: an ex vivo comparison of three methods. Plast Reconstr Surg. 2011;128(4):322e–7.CrossRefPubMed McClellan WT, Schessler MJ, Ruch DS, Levin LS, Goldner RD. A knotless flexor tendon repair technique using a bidirectional barbed suture: an ex vivo comparison of three methods. Plast Reconstr Surg. 2011;128(4):322e–7.CrossRefPubMed
10.
go back to reference Kleinert H, Kutz J, Ashbell T, Martinez E. Primary repair of lacerated flexor tendons in “no man’s land” (abstract). J Bone Joint Surg. 1967;49A:577. Kleinert H, Kutz J, Ashbell T, Martinez E. Primary repair of lacerated flexor tendons in “no man’s land” (abstract). J Bone Joint Surg. 1967;49A:577.
11.
go back to reference Schuind F, Garcia-Elias M, Cooney III WP, An K-N. Flexor tendon forces: in vivo measurements. J Hand Surg. 1992;17(2):291–8.CrossRef Schuind F, Garcia-Elias M, Cooney III WP, An K-N. Flexor tendon forces: in vivo measurements. J Hand Surg. 1992;17(2):291–8.CrossRef
13.
go back to reference Moriya T, Larson MC, Zhao C, An K-N, Amadio PC. The effect of core suture flexor tendon repair techniques on gliding resistance during static cycle motion and load to failure: a human cadaver study. J Hand Surg Eur. 2012;37(4):316–22.CrossRef Moriya T, Larson MC, Zhao C, An K-N, Amadio PC. The effect of core suture flexor tendon repair techniques on gliding resistance during static cycle motion and load to failure: a human cadaver study. J Hand Surg Eur. 2012;37(4):316–22.CrossRef
14.
go back to reference Cao Y, Zhu B, Xie RG, Tang JB. Influence of core suture purchase length on strength of four-strand tendon repairs. J Hand Surg. 2006;31(1):107–12.CrossRef Cao Y, Zhu B, Xie RG, Tang JB. Influence of core suture purchase length on strength of four-strand tendon repairs. J Hand Surg. 2006;31(1):107–12.CrossRef
15.
go back to reference Dinopoulos HT, Boyer MI, Burns ME, Gelberman RH, Silva MJ. The resistance of a four- and eight-strand suture technique to gap formation during tensile testing: an experimental study of repaired canine flexor tendons after 10 days of in vivo healing. J Hand Surg. 2000;25(3):489–98.CrossRef Dinopoulos HT, Boyer MI, Burns ME, Gelberman RH, Silva MJ. The resistance of a four- and eight-strand suture technique to gap formation during tensile testing: an experimental study of repaired canine flexor tendons after 10 days of in vivo healing. J Hand Surg. 2000;25(3):489–98.CrossRef
16.
go back to reference Strickland J. Flexor tendon injuries. I. Foundations of treatment. J Am Acad Orthop Surg. 1995;3:44–54.PubMed Strickland J. Flexor tendon injuries. I. Foundations of treatment. J Am Acad Orthop Surg. 1995;3:44–54.PubMed
17.
go back to reference Williamson D, Richards R. Flexor tendon injuries & reconstruction. In: Mathes S, Hentz V, editors. Plastic surgery, the hand & upper limb, vol. 8. 2nd ed. Philadelphia: Saunders/Elsevier; 2006. p. 351–99. Williamson D, Richards R. Flexor tendon injuries & reconstruction. In: Mathes S, Hentz V, editors. Plastic surgery, the hand & upper limb, vol. 8. 2nd ed. Philadelphia: Saunders/Elsevier; 2006. p. 351–99.
18.
go back to reference Peltz TS, Haddad R, Scougall PJ, Nicklin S, Gianoutsos MP, Walsh WR. Influence of locking stitch size in a four-strand cross-locked cruciate flexor tendon repair. J Hand Surg. 2011;36(3):450–5.CrossRef Peltz TS, Haddad R, Scougall PJ, Nicklin S, Gianoutsos MP, Walsh WR. Influence of locking stitch size in a four-strand cross-locked cruciate flexor tendon repair. J Hand Surg. 2011;36(3):450–5.CrossRef
Metadata
Title
Biomechanical evaluation of double-strand (looped) and single-strand polyamide multifilament suture: influence of knot and suture size
Authors
David T. Netscher
Justin J. Badal
Jonathan Yang
Yoav Kaufman
Jerry Alexander
Philip Noble
Publication date
01-09-2015
Publisher
Springer US
Published in
HAND / Issue 3/2015
Print ISSN: 1558-9447
Electronic ISSN: 1558-9455
DOI
https://doi.org/10.1007/s11552-014-9723-x

Other articles of this Issue 3/2015

HAND 3/2015 Go to the issue