Open Access 01-12-2009 | Research article

Reproducibility of wrist home blood pressure measurement with position sensor and automatic data storage

Authors: Sakir Uen, Rolf Fimmers, Miriam Brieger, Georg Nickenig, Thomas Mengden

Published in: BMC Cardiovascular Disorders | Issue 1/2009

Login to get access

Abstract

Background

Wrist blood pressure (BP) devices have physiological limits with regards to accuracy, therefore they were not preferred for home BP monitoring. However some wrist devices have been successfully validated using etablished validation protocols. Therefore this study assessed the reproducibility of wrist home BP measurement with position sensor and automatic data storage.

Methods

To compare the reproducibility of three different(BP) measurement methods: 1) office BP, 2) home BP (Omron wrist device HEM- 637 IT with position sensor), 3) 24-hour ambulatory BP(24-h ABPM) (ABPM-04, Meditech, Hun)conventional sphygmomanometric office BP was measured on study days 1 and 7, 24-h ABPM on study days 7 and 14 and home BP between study days 1 and 7 and between study days 8 and 14 in 69 hypertensive and 28 normotensive subjects. The correlation coeffcient of each BP measurement method with echocardiographic left ventricular mass index was analyzed. The schedule of home readings was performed according to recently published European Society of Hypertension (ESH)- guidelines.

Results

The reproducibility of home BP measurement analyzed by the standard deviation as well as the squared differeces of mean individual differences between the respective BP measurements was significantly higher than the reproducibility of office BP (p < 0.001 for systolic and diastolic BP) and the reproducibility of 24-h ABPM (p < 0.001 systolic BP, p = 0.127 diastolic BP). The reproducibility of systolic and diastolic office versus 24-h ABPM was not significantly different (p = 0.80 systolic BP, p = 0.1 diastolic BP). The correlation coefficient of 24-h ABMP (r = 0.52) with left ventricular mass index was significantly higher than with office BP (r = 0.31). The difference between 24-h ABPM and home BP (r = 0.46) was not significant.

Conclusion

The short-term reproducibility of home BP measurement with the Omron HEM-637 IT wrist device was superior to the reproducibility of office BP and 24- h ABPM measurement. Furthermore, home BP with the wrist device showed similar correlations to targed organ damage as recently reported for upper arm devices. Although wrist devices have to be used cautious and with defined limitations, the use of validated devices with position sensor according to recently recommended measurement schedules might have the potential to be used for therapy monitoring.
Appendix
Available only for authorised users
Literature
1.
go back to reference Coats A, Radaelli A, Clark S, Conway J, Sleight P: The influence of ambulatory blood pressure monitoring on the design and interpretation of trials in hypertension. J Hypertens. 1992, 10: 385-391. 10.1097/00004872-199204000-00011.CrossRefPubMed Coats A, Radaelli A, Clark S, Conway J, Sleight P: The influence of ambulatory blood pressure monitoring on the design and interpretation of trials in hypertension. J Hypertens. 1992, 10: 385-391. 10.1097/00004872-199204000-00011.CrossRefPubMed
2.
go back to reference James GD, Pickering TG, Yee LS, Harshfield GA, Riva S, Laragh JH: The reproducibility of average ambulatory, home, and clinic pressures. Hypertension. 1988, 1: 545-549.CrossRef James GD, Pickering TG, Yee LS, Harshfield GA, Riva S, Laragh JH: The reproducibility of average ambulatory, home, and clinic pressures. Hypertension. 1988, 1: 545-549.CrossRef
3.
go back to reference Verdecchia P, Clement D, Fagard R, Palatini P, Parati G: Blood pressure monitoring. Task force III: Target-organ damage, morbidity and mortalità. Blood Press Monit. 1999, 4 (6): 303-17. 10.1097/00126097-199912000-00004.CrossRefPubMed Verdecchia P, Clement D, Fagard R, Palatini P, Parati G: Blood pressure monitoring. Task force III: Target-organ damage, morbidity and mortalità. Blood Press Monit. 1999, 4 (6): 303-17. 10.1097/00126097-199912000-00004.CrossRefPubMed
4.
go back to reference Mengden T, Battig B, Vetter W: Self-measurement of blood pressure improves the accuracy and reduces the number of subjects in clinical trials. J Hypertens. 1991, 9 (suppl 6): 336-337. Mengden T, Battig B, Vetter W: Self-measurement of blood pressure improves the accuracy and reduces the number of subjects in clinical trials. J Hypertens. 1991, 9 (suppl 6): 336-337.
5.
go back to reference Stergiou G, Baibas N, Gantzarou A, Skeva I, Kalkana C, Roussias L, Mountokalakis T: Reproducibility of home, ambulatory, and clinic blood opressure: Implications for the design of trials for the assessment of antihypertensive drug efficacy. Am J Hypertens. 2002, 15: 101-104. 10.1016/S0895-7061(01)02324-X.CrossRefPubMed Stergiou G, Baibas N, Gantzarou A, Skeva I, Kalkana C, Roussias L, Mountokalakis T: Reproducibility of home, ambulatory, and clinic blood opressure: Implications for the design of trials for the assessment of antihypertensive drug efficacy. Am J Hypertens. 2002, 15: 101-104. 10.1016/S0895-7061(01)02324-X.CrossRefPubMed
6.
go back to reference Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al: Recommendations for blood pressure measurement in humans and experimental animals. Part 1: Blood pressure measurement in humans. Circulation. 2005, 111 (5): 697-716. 10.1161/01.CIR.0000154900.76284.F6.CrossRefPubMed Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al: Recommendations for blood pressure measurement in humans and experimental animals. Part 1: Blood pressure measurement in humans. Circulation. 2005, 111 (5): 697-716. 10.1161/01.CIR.0000154900.76284.F6.CrossRefPubMed
7.
go back to reference Kikuya M, Chonan K, Imai Y, Goto E, Ishii M: Accuracy and reliability of wrist-cuff devices for selfmeasurement. J Hypertens. 2002, 20: 629-38. 10.1097/00004872-200204000-00019.CrossRefPubMed Kikuya M, Chonan K, Imai Y, Goto E, Ishii M: Accuracy and reliability of wrist-cuff devices for selfmeasurement. J Hypertens. 2002, 20: 629-38. 10.1097/00004872-200204000-00019.CrossRefPubMed
8.
go back to reference O'Brien E, Asmar R, Beilin L, Imai Y, Mancia G, Mengden T, Myers M, Padfield P, Palatini P, Parati G, Pickering T, Redon J, Staessen J, Stergiou G, Verdecchia P, On behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring: Practice Guidelines of the European Society of Hypertension for Clinic, Ambulatory, and Self Blood Pressure Measurement. J Hypertens. 2005, 23 (4): 697-701.CrossRefPubMed O'Brien E, Asmar R, Beilin L, Imai Y, Mancia G, Mengden T, Myers M, Padfield P, Palatini P, Parati G, Pickering T, Redon J, Staessen J, Stergiou G, Verdecchia P, On behalf of the European Society of Hypertension Working Group on Blood Pressure Monitoring: Practice Guidelines of the European Society of Hypertension for Clinic, Ambulatory, and Self Blood Pressure Measurement. J Hypertens. 2005, 23 (4): 697-701.CrossRefPubMed
9.
go back to reference Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali LN, Asmar RG: Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM-637-IT). Blood Press Monit. 2006, 11: 165-171. 10.1097/01.mbp.0000209078.17246.34.CrossRefPubMed Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali LN, Asmar RG: Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM-637-IT). Blood Press Monit. 2006, 11: 165-171. 10.1097/01.mbp.0000209078.17246.34.CrossRefPubMed
10.
go back to reference Barna I, Keszei A, Dunai A: Evaluation of Meditech ABPM-04 ambulatory BP measuring device according to the British Hypertension Society protocol. Blood Press Monit. 1998, 3: 363-368.PubMed Barna I, Keszei A, Dunai A: Evaluation of Meditech ABPM-04 ambulatory BP measuring device according to the British Hypertension Society protocol. Blood Press Monit. 1998, 3: 363-368.PubMed
11.
go back to reference Trazzi S, Mutti E, Frattola A, Imholz B, Parati G, Mancia G: Reproducibility of non-invasive and intra-arterial blood pressure monitoring: implications for studies on antihypertensive drugs. J Hypertens. 1991, 9: 115-119. 10.1097/00004872-199102000-00003.CrossRefPubMed Trazzi S, Mutti E, Frattola A, Imholz B, Parati G, Mancia G: Reproducibility of non-invasive and intra-arterial blood pressure monitoring: implications for studies on antihypertensive drugs. J Hypertens. 1991, 9: 115-119. 10.1097/00004872-199102000-00003.CrossRefPubMed
12.
go back to reference Chatellier G, Day M, Bodrie G, Menard J: Feasibility study of n-of-1 trials with blood pressure self-monitoring in hypertension. Hypertension. 1995, 25: 294-301.CrossRefPubMed Chatellier G, Day M, Bodrie G, Menard J: Feasibility study of n-of-1 trials with blood pressure self-monitoring in hypertension. Hypertension. 1995, 25: 294-301.CrossRefPubMed
13.
go back to reference Mengden T, Binswanger B, Weisser B, Vetter W: An evaluation of self-measured blood pressure in a study with calcium-channel antagonist versus β-blocker. Am J Hypertens. 1992, 5: 154-160.CrossRefPubMed Mengden T, Binswanger B, Weisser B, Vetter W: An evaluation of self-measured blood pressure in a study with calcium-channel antagonist versus β-blocker. Am J Hypertens. 1992, 5: 154-160.CrossRefPubMed
14.
go back to reference Ragot S, Genès N, Vaur L, Herpin D: Comparison of three blood pressure measurement methods for the evaluation of two antihypertensive drugs: Feasibility, agreement, and reproducibility of blood pressure response. Am J Hypertens. 2000, 13: 632-639. 10.1016/S0895-7061(99)00258-7.CrossRefPubMed Ragot S, Genès N, Vaur L, Herpin D: Comparison of three blood pressure measurement methods for the evaluation of two antihypertensive drugs: Feasibility, agreement, and reproducibility of blood pressure response. Am J Hypertens. 2000, 13: 632-639. 10.1016/S0895-7061(99)00258-7.CrossRefPubMed
15.
go back to reference Baguet JP, Mallion JM: Self-monitoring of blood pressure should be used in clinical trials. Blood Press Monit. 2002, 7: 55-59. 10.1097/00126097-200202000-00011.CrossRefPubMed Baguet JP, Mallion JM: Self-monitoring of blood pressure should be used in clinical trials. Blood Press Monit. 2002, 7: 55-59. 10.1097/00126097-200202000-00011.CrossRefPubMed
16.
go back to reference Mengden T, Ewald S, Kaufmann S, vor dem Esche J, Uen S, Vetter H: Telemonitoring of blood pressure self measurement in the OLMETEL study. Blood Press Monit (England). 2004, 9 (6): Mengden T, Ewald S, Kaufmann S, vor dem Esche J, Uen S, Vetter H: Telemonitoring of blood pressure self measurement in the OLMETEL study. Blood Press Monit (England). 2004, 9 (6):
17.
go back to reference Altunkan S, Öztas K, Altunkan E: Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the international protocol in adults and obese adults. Blood Press Monit. 2006, 11: 79-85. 10.1097/01.mbp.0000200483.49540.dc.CrossRefPubMed Altunkan S, Öztas K, Altunkan E: Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the international protocol in adults and obese adults. Blood Press Monit. 2006, 11: 79-85. 10.1097/01.mbp.0000200483.49540.dc.CrossRefPubMed
18.
go back to reference Altunkan S, Altunkan E: Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the international protocol in the elderly. Blood Press Monit. 2006, 11: 97-102. 10.1097/01.mbp.0000200486.95281.5c.CrossRefPubMed Altunkan S, Altunkan E: Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the international protocol in the elderly. Blood Press Monit. 2006, 11: 97-102. 10.1097/01.mbp.0000200486.95281.5c.CrossRefPubMed
19.
go back to reference Altunkan S, Genc Y, Altunkan E: A comparative study of an ambulatory blood pressure measuring device and a wrist blood pressure monitor with a position sensor versus a mercury sphygmomanometer. Eur J Intern Med. 2007, 18 (2): 118-23. 10.1016/j.ejim.2006.09.018.CrossRefPubMed Altunkan S, Genc Y, Altunkan E: A comparative study of an ambulatory blood pressure measuring device and a wrist blood pressure monitor with a position sensor versus a mercury sphygmomanometer. Eur J Intern Med. 2007, 18 (2): 118-23. 10.1016/j.ejim.2006.09.018.CrossRefPubMed
20.
go back to reference Yarows S: Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist postion sensor on or disabled. Am J Hypertens. 2004, 17: 54-58. 10.1016/j.amjhyper.2003.09.003.CrossRefPubMed Yarows S: Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist postion sensor on or disabled. Am J Hypertens. 2004, 17: 54-58. 10.1016/j.amjhyper.2003.09.003.CrossRefPubMed
21.
go back to reference World Health Organization, International Society of Hypertension Writing Group: World Health Organization (WHO)/International Society of Hypertension (ISH) statement on management of hypertension. J Hypertens. 2003, 21: 1983-92. 10.1097/00004872-200311000-00002.CrossRef World Health Organization, International Society of Hypertension Writing Group: World Health Organization (WHO)/International Society of Hypertension (ISH) statement on management of hypertension. J Hypertens. 2003, 21: 1983-92. 10.1097/00004872-200311000-00002.CrossRef
22.
go back to reference O'Brien E, Asmar R, Beilin L, Imai Y, Mancia G, Mengden T, Myers M, et al: Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. J Hypertens. 2005, 23 (4): 697-701.CrossRefPubMed O'Brien E, Asmar R, Beilin L, Imai Y, Mancia G, Mengden T, Myers M, et al: Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. J Hypertens. 2005, 23 (4): 697-701.CrossRefPubMed
23.
go back to reference Uen S, Weisser B, Wieneke P, Vetter H, Mengden M: Evaluation of the performance of a wrist blood pressure measuring device with a position sensor compared to ambulatory 24-h blood pressure measurements. Am J Hypertens (United States). 2002, 15 (9): 787-92.CrossRef Uen S, Weisser B, Wieneke P, Vetter H, Mengden M: Evaluation of the performance of a wrist blood pressure measuring device with a position sensor compared to ambulatory 24-h blood pressure measurements. Am J Hypertens (United States). 2002, 15 (9): 787-92.CrossRef
24.
go back to reference Mengden T, Hernandez Medina RM, Beltran B, Alvarez E, Kraft K, Vetter H: Reliability of reporting self-measured blood pressure values by hypertensive patients. Am J Hypertens. 1998, 11: 1413-1417. 10.1016/S0895-7061(98)00241-6.CrossRefPubMed Mengden T, Hernandez Medina RM, Beltran B, Alvarez E, Kraft K, Vetter H: Reliability of reporting self-measured blood pressure values by hypertensive patients. Am J Hypertens. 1998, 11: 1413-1417. 10.1016/S0895-7061(98)00241-6.CrossRefPubMed
25.
go back to reference Kok RH, Beltman FW, Terpstra WF, Smit AJ, May JF, de Graeff PA, Meyboom-de Jong B: Home blood pressure measurement: reproducibility and relationship with left ventricular mass. Blood Press Monit. 1999, 4 (2): 65-9.CrossRefPubMed Kok RH, Beltman FW, Terpstra WF, Smit AJ, May JF, de Graeff PA, Meyboom-de Jong B: Home blood pressure measurement: reproducibility and relationship with left ventricular mass. Blood Press Monit. 1999, 4 (2): 65-9.CrossRefPubMed
26.
go back to reference Fagard R, Staessen JA, Thijs L: The relationships between left ventricular mass and daytime and night-time blood pressures: a meta-analysis of comparative studies. J Hypertens. 1995, 13: 823-9. 10.1097/00004872-199508000-00002.CrossRefPubMed Fagard R, Staessen JA, Thijs L: The relationships between left ventricular mass and daytime and night-time blood pressures: a meta-analysis of comparative studies. J Hypertens. 1995, 13: 823-9. 10.1097/00004872-199508000-00002.CrossRefPubMed
27.
go back to reference Iula A, Puukka P, Karanko A: Multiple clinic and home blood pressure measurements versus ambulatory blood pressure monitoring. Hypertension. 1999, 34: 261-266.CrossRef Iula A, Puukka P, Karanko A: Multiple clinic and home blood pressure measurements versus ambulatory blood pressure monitoring. Hypertension. 1999, 34: 261-266.CrossRef
28.
go back to reference Nyström F, Malmqvist K, Lind L, Kahan T: Nurse-recorded clinic and ambulatory blood pressures correlate equally well with left ventricular mass and carotid intima-media thickness. J I M. 2005, 257: 514-522. 10.1111/j.1365-2796.2005.01489.x. Nyström F, Malmqvist K, Lind L, Kahan T: Nurse-recorded clinic and ambulatory blood pressures correlate equally well with left ventricular mass and carotid intima-media thickness. J I M. 2005, 257: 514-522. 10.1111/j.1365-2796.2005.01489.x.
29.
go back to reference Ibrahim MM, Tarazi RC, Dustan HP, Gifford RW: Electrocardiogram in evaluation of resistance to antihypertensive therapy. Arch Intern Med. 1977, 137: 15-18. 10.1001/archinte.137.9.1125.CrossRef Ibrahim MM, Tarazi RC, Dustan HP, Gifford RW: Electrocardiogram in evaluation of resistance to antihypertensive therapy. Arch Intern Med. 1977, 137: 15-18. 10.1001/archinte.137.9.1125.CrossRef
30.
go back to reference Kleinert HD, Harshlfield GA, Pickering TG, Devereux RB, Sullivan PA, Marlon RM, et al: What is the value of home blood pressure measurement in patients with mild hypertension?. Hypertension. 1984, 6: 574-578.CrossRefPubMed Kleinert HD, Harshlfield GA, Pickering TG, Devereux RB, Sullivan PA, Marlon RM, et al: What is the value of home blood pressure measurement in patients with mild hypertension?. Hypertension. 1984, 6: 574-578.CrossRefPubMed
31.
go back to reference Abé H, Yokouchi M, Nagata S, Ashida T, Yoshimi H, Kawano Y, Kojima S, Matsuoka S, Kuramochi M, Omae T: Relation of office and home blood pressure to left ventricular hypertrophy and performance in patients with hypertension. High Blood Press. 1992, 1: 279-285. Abé H, Yokouchi M, Nagata S, Ashida T, Yoshimi H, Kawano Y, Kojima S, Matsuoka S, Kuramochi M, Omae T: Relation of office and home blood pressure to left ventricular hypertrophy and performance in patients with hypertension. High Blood Press. 1992, 1: 279-285.
Metadata
Title
Reproducibility of wrist home blood pressure measurement with position sensor and automatic data storage
Authors
Sakir Uen
Rolf Fimmers
Miriam Brieger
Georg Nickenig
Thomas Mengden
Publication date
01-12-2009
Publisher
BioMed Central
Published in
BMC Cardiovascular Disorders / Issue 1/2009
Electronic ISSN: 1471-2261
DOI
https://doi.org/10.1186/1471-2261-9-20

Other articles of this Issue 1/2009

BMC Cardiovascular Disorders 1/2009 Go to the issue