01-07-2003 | Chest

Radiation exposure of multi-row detector spiral computed tomography of the pulmonary arteries: comparison with digital subtraction pulmonary angiography

Authors: Jan W. Kuiper, Jacob Geleijns, Niels A. A. Matheijssen, Wouter Teeuwisse, Peter M. T. Pattynama

Published in: European Radiology | Issue 7/2003

Login to get access

Abstract.

The purpose of this study was to determine and compare the effective dose of multidetector computed tomographic angiography (MDCT) and digital subtraction angiography (DSA) studies for diagnosing a pulmonary embolus (PE). Radiation exposure was measured as computed tomography dose index (MDCT) or as dose-area product (DSA) and was subsequently expressed in the quantity effective dose. Effective doses were obtained in 27 patients who underwent MDCT and in 12 patients who underwent DSA for suspected PE. The MDCT angiography was performed on a Siemens Volume Zoom CT scanner and DSA on a Philips Integris V-3000 system according to standardized protocols. Average effective dose for MDCT angiography of the pulmonary arteries (27 patients) was 4.2 mSv (range 2.2–6.0 mSv). Pulmonary DSA gained an average effective dose (12 patients) of 7.1 mSv (range 3.3–17.3 mSv). Our results show that the effective doses in MDCT angiography studies for PE are moderate and even slightly lower in comparison with pulmonary DSA in a comparable patient group. Variations in patient dose are smaller for MDCT, probably because this procedure can be more strictly protocolized. Patient dose should not be restrictive in the discussion of CTA replacing DSA for diagnosing PE.
Literature
1.
go back to reference Van Rossum AB, Pattynama PMT, Tjin A, Ton ER, Treurniet F, Arndt JW, Van Eck B, Kieft GJ (1996) Spiral CT angiography for pulmonary embolism: validation in 149 patients. Radiology 201:467–470 Van Rossum AB, Pattynama PMT, Tjin A, Ton ER, Treurniet F, Arndt JW, Van Eck B, Kieft GJ (1996) Spiral CT angiography for pulmonary embolism: validation in 149 patients. Radiology 201:467–470
2.
go back to reference Remy-Jardin M, Remy J, Deschildre F, Artaud D, Beregi JP, Hossein-Foucher C, Marchandise X, Duhamel A (1996) Diagnosis of pulmonary embolism with spiral CT: comparison with pulmonary angiography and scintigraphy. Radiology 200:699–706 Remy-Jardin M, Remy J, Deschildre F, Artaud D, Beregi JP, Hossein-Foucher C, Marchandise X, Duhamel A (1996) Diagnosis of pulmonary embolism with spiral CT: comparison with pulmonary angiography and scintigraphy. Radiology 200:699–706
3.
go back to reference Rathbun SW, Raskob GE, Whitsett TL (2000) Sensitivity and specificity of helical computed tomography in the diagnosis of pulmonary embolism: a systematical review. Ann Intern Med 132:227–232 Rathbun SW, Raskob GE, Whitsett TL (2000) Sensitivity and specificity of helical computed tomography in the diagnosis of pulmonary embolism: a systematical review. Ann Intern Med 132:227–232
4.
go back to reference Van Rossum AB, Pattynama PMT, Mallens WC, Heyermans HGM, Hermans J (1998) Can helical CT replace scintigraphy in the diagnostic process in suspected pulmonary embolism? A retrospective–projective cohort study focusing on total diagnostic yield. Eur Radiol 8:90–96 Van Rossum AB, Pattynama PMT, Mallens WC, Heyermans HGM, Hermans J (1998) Can helical CT replace scintigraphy in the diagnostic process in suspected pulmonary embolism? A retrospective–projective cohort study focusing on total diagnostic yield. Eur Radiol 8:90–96
5.
go back to reference Remy-Jardin M, Remy J, Baghaie F, Fribourg M, Artaud D, Duhamel A (2000) Clinical value of thin collimation in the diagnostic workup of pulmonary embolism. Am J Roentgenol 75:407A1 1 Remy-Jardin M, Remy J, Baghaie F, Fribourg M, Artaud D, Duhamel A (2000) Clinical value of thin collimation in the diagnostic workup of pulmonary embolism. Am J Roentgenol 75:407A1 1
6.
go back to reference Ghaye B, Szapiro D,Mastora I, Delannoy V, Duhamel A, Remy J, Remy-Jardin M (2001) Peripheral pulmonary arteries: How far in the lung does multi-detector row spiral CT allow analysis? Radiology 219:629–636 Ghaye B, Szapiro D,Mastora I, Delannoy V, Duhamel A, Remy J, Remy-Jardin M (2001) Peripheral pulmonary arteries: How far in the lung does multi-detector row spiral CT allow analysis? Radiology 219:629–636
7.
go back to reference Baile EM, King GG, Muller NL, D'Yachkova Y, Coche EE, Pare PD, Mayo JR (2000) Spiral computed tomography is comparable to angiography for the diagnosis of pulmonary embolism. Am J Respir Crit Care Med 161:1010–1015 Baile EM, King GG, Muller NL, D'Yachkova Y, Coche EE, Pare PD, Mayo JR (2000) Spiral computed tomography is comparable to angiography for the diagnosis of pulmonary embolism. Am J Respir Crit Care Med 161:1010–1015
8.
go back to reference ICRP (1991) 1990 recommendations of the International Commission on Radiological Protection. Publication 60. Pergamon, Oxford ICRP (1991) 1990 recommendations of the International Commission on Radiological Protection. Publication 60. Pergamon, Oxford
9.
go back to reference Shope TB, Gagne RM, Johnson GC (1981) A method for describing the doses delivered by transmission X-ray computed tomography. Med Phys 8:488–495 Shope TB, Gagne RM, Johnson GC (1981) A method for describing the doses delivered by transmission X-ray computed tomography. Med Phys 8:488–495
10.
go back to reference Shrimpton PC, Wall BF (1982) An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinations. Phys Med Biol 27:871–878 Shrimpton PC, Wall BF (1982) An evaluation of the Diamentor transmission ionisation chamber in indicating exposure-area product (R cm2) during diagnostic radiological examinations. Phys Med Biol 27:871–878
11.
go back to reference Jones DG, Shrimpton PC (1991) Normalised organ doses for X-ray computed tomography calculated using Monte Carlo techniques. NRPB-SR250, Chilton Jones DG, Shrimpton PC (1991) Normalised organ doses for X-ray computed tomography calculated using Monte Carlo techniques. NRPB-SR250, Chilton
12.
go back to reference St. George's Hospital (2001) ImPACT CT Patient Dosimetry Calculator, version 0.99e, ImPACT, London St. George's Hospital (2001) ImPACT CT Patient Dosimetry Calculator, version 0.99e, ImPACT, London
13.
go back to reference Servomaa A, Tapiovaara M (1998) Organ dose calculation in medical X-ray examinations by the program PCXMC. Radiat Prot Dosim 80:213–219 Servomaa A, Tapiovaara M (1998) Organ dose calculation in medical X-ray examinations by the program PCXMC. Radiat Prot Dosim 80:213–219
14.
go back to reference ICRP (1993) Radiological protection in biomedical research, a report by the International Commission on Radiological Protection. Publication 62. Pergamon, Oxford ICRP (1993) Radiological protection in biomedical research, a report by the International Commission on Radiological Protection. Publication 62. Pergamon, Oxford
15.
go back to reference UNSCEAR (2000) 2000 report, sources and effects of ionizing radiation, vol I: sources. United Nations Scientific Committee on the Effects of Atomic Radiation UNSCEAR (2000) 2000 report, sources and effects of ionizing radiation, vol I: sources. United Nations Scientific Committee on the Effects of Atomic Radiation
16.
go back to reference Wilting JE, Zwartkruis A, van Leeuwen MS, Timmer J, Kamphuis AG, Feldberg M (2001) A rational approach to dose reduction in CT: individualized scan protocols. Eur Radiol 11:2627–2632 Wilting JE, Zwartkruis A, van Leeuwen MS, Timmer J, Kamphuis AG, Feldberg M (2001) A rational approach to dose reduction in CT: individualized scan protocols. Eur Radiol 11:2627–2632
17.
go back to reference Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as function of patient weight at abdominal CT. Radiology 217:430–435 Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as function of patient weight at abdominal CT. Radiology 217:430–435
Metadata
Title
Radiation exposure of multi-row detector spiral computed tomography of the pulmonary arteries: comparison with digital subtraction pulmonary angiography
Authors
Jan W. Kuiper
Jacob Geleijns
Niels A. A. Matheijssen
Wouter Teeuwisse
Peter M. T. Pattynama
Publication date
01-07-2003
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 7/2003
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-002-1753-z

Other articles of this Issue 7/2003

European Radiology 7/2003 Go to the issue