Radiation dose optimization for the bolus tracking technique in abdominal computed tomography: usefulness of real-time iterative reconstruction for monitoring scan

Authors: Yuya Ishikawa, Atsushi Urikura, Tsukasa Yoshida, Keisuke Takiguchi, Yoshihiro Nakaya

Published in: Radiological Physics and Technology | Issue 2/2017

Login to get access

Abstract

The purpose of this study was to optimize the monitoring dose, obtained using the conventional filtered back projection (FBP) method and iterative reconstruction algorithms, for the bolus tracking technique. A phantom study was performed to assess the effect of the scan start time in patients grouped according to different body weights. An oval torso phantom was used for simulating the time enhancement curve of the bolus tracking technique. To reproduce image noise levels in the two body weight groups, the phantom diameter was adjusted with a water-equivalent material. The tube currents were 10, 20, 30, and 50 mA. The monitoring scan was performed with the conventional FBP method and real-time adaptive iterative dose reduction by three-dimensional processing (AIDR 3D). The results at different doses were compared with those at 50 mA. The volume computed tomography dose index was 1.31, 2.65, 3.93, and 6.56 mGy at tube currents of 10, 20, 30, and 50 mA, respectively. The scan start time, reconstructed using FBP, was significantly faster at 10 and 20 mA in group A (50–59 kg) and at 20 mA in group B (≥80 kg). The CT values in the region of interest could not be measured at 10 mA because of artifacts. With real-time AIDR 3D, both groups showed no significant differences between the measurements obtained at 30 or 20 mA and those obtained at 50 mA. Our study demonstrated that the real-time AIDR 3D algorithm improved the accuracy of the CT measurements with the bolus tracking technique.
Literature
1.
go back to reference Hein PA, Romano VC, Lembcke A, May J, Rogalla P. Initial experience with a chest pain protocol using 320-slice volume MDCT. Eur Radiol. 2009;19:1148–55.CrossRefPubMed Hein PA, Romano VC, Lembcke A, May J, Rogalla P. Initial experience with a chest pain protocol using 320-slice volume MDCT. Eur Radiol. 2009;19:1148–55.CrossRefPubMed
2.
go back to reference Rogalla P, Kloeters C, Hein PA. CT technology overview: 64-slice and beyond. Radiol Clin North Am. 2009;47:1–11.CrossRefPubMed Rogalla P, Kloeters C, Hein PA. CT technology overview: 64-slice and beyond. Radiol Clin North Am. 2009;47:1–11.CrossRefPubMed
3.
go back to reference Hammerstingl RM, Vogl TJ. Abdominal MDCT: protocols and contrast considerations. Eur Radiol. 2005;15(Suppl 5):E78–90.CrossRefPubMed Hammerstingl RM, Vogl TJ. Abdominal MDCT: protocols and contrast considerations. Eur Radiol. 2005;15(Suppl 5):E78–90.CrossRefPubMed
4.
go back to reference Sandstede JJ, Tschammler A, Beer M, Vogelsang C, Wittenberg G, Hahn D. Optimization of automatic bolus tracking for timing of the arterial phase of helical liver CT. Eur Radiol. 2001;11:1396–400.CrossRefPubMed Sandstede JJ, Tschammler A, Beer M, Vogelsang C, Wittenberg G, Hahn D. Optimization of automatic bolus tracking for timing of the arterial phase of helical liver CT. Eur Radiol. 2001;11:1396–400.CrossRefPubMed
5.
go back to reference Mehnert F, Pereira PL, Trübenbach J, Kopp AF, Claussen CD. Biphasic spiral CT of the liver: automatic bolus tracking or time delay? Eur Radiol. 2001;11:427–31.CrossRefPubMed Mehnert F, Pereira PL, Trübenbach J, Kopp AF, Claussen CD. Biphasic spiral CT of the liver: automatic bolus tracking or time delay? Eur Radiol. 2001;11:427–31.CrossRefPubMed
6.
go back to reference Dinkel HP, Fieger M, Knüpffer J, Moll R, Schindler G. Optimizing liver contrast in helical liver CT: value of a real-time bolus-triggering technique. Eur Radiol. 1998;8:1608–12.CrossRefPubMed Dinkel HP, Fieger M, Knüpffer J, Moll R, Schindler G. Optimizing liver contrast in helical liver CT: value of a real-time bolus-triggering technique. Eur Radiol. 1998;8:1608–12.CrossRefPubMed
7.
go back to reference Sultana S, Awai K, Nakayama Y, Nakaura T, Liu D, Hatemura M, Funama Y, Morishita S, Yamashita Y. Hypervascular hepatocellular carcinomas: bolus tracking with a 40-detector CT scanner to time arterial phase imaging. Radiology. 2007;243:140–7.CrossRefPubMed Sultana S, Awai K, Nakayama Y, Nakaura T, Liu D, Hatemura M, Funama Y, Morishita S, Yamashita Y. Hypervascular hepatocellular carcinomas: bolus tracking with a 40-detector CT scanner to time arterial phase imaging. Radiology. 2007;243:140–7.CrossRefPubMed
8.
go back to reference Tatsugami F, Awai K, Takada H, Yoshikawa S, Takeda Y, Morita H, Narumi Y. Reduction of interpatient variability of arterial enhancement using a new bolus tracking system in 320-detector computed tomographic coronary angiography. J Comput Assist Tomogr. 2013;37:79–83.CrossRefPubMed Tatsugami F, Awai K, Takada H, Yoshikawa S, Takeda Y, Morita H, Narumi Y. Reduction of interpatient variability of arterial enhancement using a new bolus tracking system in 320-detector computed tomographic coronary angiography. J Comput Assist Tomogr. 2013;37:79–83.CrossRefPubMed
9.
go back to reference Catalano C, Francone M, Ascarelli A, Mangia M, Iacucci I, Passariello R. Optimizing radiation dose and image quality. Eur Radiol. 2007;17(Suppl 6):F26–32.CrossRefPubMed Catalano C, Francone M, Ascarelli A, Mangia M, Iacucci I, Passariello R. Optimizing radiation dose and image quality. Eur Radiol. 2007;17(Suppl 6):F26–32.CrossRefPubMed
10.
go back to reference Matsubara K, Sugai M, Toyoda A, Koshida H, Sakuta K, Takata T, Koshida K, Iida H, Matsui O. Assessment of an organ-based tube current modulation in thoracic computed tomography. J Appl Clin Med Phys. 2012;13:3731.PubMed Matsubara K, Sugai M, Toyoda A, Koshida H, Sakuta K, Takata T, Koshida K, Iida H, Matsui O. Assessment of an organ-based tube current modulation in thoracic computed tomography. J Appl Clin Med Phys. 2012;13:3731.PubMed
11.
go back to reference Singh S, Kalra MK, Hsieh J, Licato PE, Do S, Pien HH, Blake MA. Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology. 2010;257:373–83.CrossRefPubMed Singh S, Kalra MK, Hsieh J, Licato PE, Do S, Pien HH, Blake MA. Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology. 2010;257:373–83.CrossRefPubMed
12.
go back to reference Willemink MJ, de Jong PA, Leiner T, de Heer LM, Nievelstein RA, Budde RP, Schilham AM. Iterative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol. 2013;23:1623–31.CrossRefPubMed Willemink MJ, de Jong PA, Leiner T, de Heer LM, Nievelstein RA, Budde RP, Schilham AM. Iterative reconstruction techniques for computed tomography Part 1: technical principles. Eur Radiol. 2013;23:1623–31.CrossRefPubMed
13.
go back to reference Yamaguchi I, Kidoya E, Suzuki M, Kimura H. Evaluation of required saline volume in dynamic contrast-enhanced computed tomography using saline flush technique. Comput Med Imaging Graph. 2009;33(1):23–8.CrossRefPubMed Yamaguchi I, Kidoya E, Suzuki M, Kimura H. Evaluation of required saline volume in dynamic contrast-enhanced computed tomography using saline flush technique. Comput Med Imaging Graph. 2009;33(1):23–8.CrossRefPubMed
14.
go back to reference Kanda Y. Investigation of the freely available easy-to-use software ‘EZR’ for medical statistics. Bone Marrow Transplant. 2013;48:452–8.CrossRefPubMed Kanda Y. Investigation of the freely available easy-to-use software ‘EZR’ for medical statistics. Bone Marrow Transplant. 2013;48:452–8.CrossRefPubMed
15.
16.
go back to reference Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39:4115–22.CrossRef Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39:4115–22.CrossRef
17.
go back to reference Urikura A, Ichikawa K, Hara T, Nishimaru E, Nakaya Y. Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography. Radiol Phys Technol. 2014;7:358–66.CrossRefPubMed Urikura A, Ichikawa K, Hara T, Nishimaru E, Nakaya Y. Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography. Radiol Phys Technol. 2014;7:358–66.CrossRefPubMed
18.
go back to reference Smith EA, Dillman JR, Goodsitt MM, Christodoulou EG, Keshavarzi N, Strouse PJ. Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology. 2014;270:526–34.CrossRefPubMedPubMedCentral Smith EA, Dillman JR, Goodsitt MM, Christodoulou EG, Keshavarzi N, Strouse PJ. Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology. 2014;270:526–34.CrossRefPubMedPubMedCentral
19.
go back to reference Löve A, Olsson ML, Siemund R, Stålhammar F, Björkman-Burtscher IM, Söderberg M. Six iterative reconstruction algorithms in brain CT: a phantom study on image quality at different radiation dose levels. Br J Radiol. 2013;86:20130388.CrossRefPubMedPubMedCentral Löve A, Olsson ML, Siemund R, Stålhammar F, Björkman-Burtscher IM, Söderberg M. Six iterative reconstruction algorithms in brain CT: a phantom study on image quality at different radiation dose levels. Br J Radiol. 2013;86:20130388.CrossRefPubMedPubMedCentral
20.
go back to reference Schindera ST, Odedra D, Raza SA, Kim TK, Jang HJ, Szucs-Farkas Z, Rogalla P. Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved? Radiology. 2013;269:511–8.CrossRefPubMed Schindera ST, Odedra D, Raza SA, Kim TK, Jang HJ, Szucs-Farkas Z, Rogalla P. Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved? Radiology. 2013;269:511–8.CrossRefPubMed
Metadata
Title
Radiation dose optimization for the bolus tracking technique in abdominal computed tomography: usefulness of real-time iterative reconstruction for monitoring scan
Authors
Yuya Ishikawa
Atsushi Urikura
Tsukasa Yoshida
Keisuke Takiguchi
Yoshihiro Nakaya
Publication date
01-06-2017
Publisher
Springer Singapore
Published in
Radiological Physics and Technology / Issue 2/2017
Print ISSN: 1865-0333
Electronic ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-016-0378-x

Other articles of this Issue 2/2017

Radiological Physics and Technology 2/2017 Go to the issue