Published in: Archives of Orthopaedic and Trauma Surgery 6/2008

01-06-2008 | Osteoporotic Fracture Management

Quantification of bone strength by intraoperative torque measurement: a technical note

Authors: Norbert Suhm, Markus Haenni, Ronald Schwyn, Michael Hirschmann, Andreas Marc Müller

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 6/2008

Login to get access

Abstract

Introduction

Bone strength describes the resistance of bone against mechanical failure. Bone strength depends on both the amount of bone and the bone’s quality, and the bone strength may be looked upon as a relevant parameter to judge an osteosynthesis’ stability. Information about bone strength was barely available intraoperatively in the past. The previous work of our group reported on development and laboratory evaluation of mechanical torque measurement as a method for the intraoperative quantification of bone strength. With the clinical series presented here we intend to verify that the im gesamten Text DensiProbe™ instrumentation for intraoperative torque measurement and the related measurement method are eligible for intraoperative use based on the following criteria: application of the method may not create complications, the measurement can be performed by the surgeon himself and may only cause a limited increase in the procedure time.

Patients and methods

From December 2006 until May 2007 ten patients with a pertrochanteric femoral fracture or a lateral femoral neck fracture eligible for stabilization with DHS® were included in the study after having received informed consent. Any medication and comorbidity that might have influenced bone quality or bone mineral density (BMD) in these patients was documented. Bone strength was intraoperatively measured with DensiProbe®. Complications that were obviously related with torque measurement were documented as well as any deviation from the suggested procedure; 6 and 12 weeks postoperative follow-up included clinical and radiological examination. The time required for torque measurement, the overall operating time and the number of persons present in the operating room were protocolled. BMD values of the contralateral femoral neck were postoperatively assessed by dual energy X-ray absorptiometry (DEXA) and compared to intraoperative peak torque values measured by DensiProbe®.

Results

No major complication was observed during intraoperative application of DensiProbe®by trained surgeons. The unintended extraction of the guide wire together with the torque measurement probe was reported only once and is looked upon as a minor complication. Fracture healing was uneventful in all patients. The mean time for torque measurement was 2.35 ± 0.9 min accounting for 2.2 ± 1.1% of total surgery time. The presence of an additional person was not required to perform torque measurement but to protocol the data. There was a tendency towards correlation between BMD values of the femoral neck and intraoperative peak torque values.

Discussion

The data presented clearly indicate that the DensiProbe® instrumentation and measurement principle are eligible for routine intraoperative use by trained surgeons. Interpretation of possible correlations between BMD values measured by means of DEXA and the Peak Torque values assessed by DensiProbe® has to be considered very carefully, because BMD and Peak Torque analyse bone at a different scale. Only within the framework of a multicenter study it will be possible to include a sufficient number of patients for calculation of the methods’ predictive value towards implant failure and to verify acceptance of the method by the surgeons.
Literature
1.
go back to reference Augat P, Rapp S, Claes L (2002) A modified hip screw incorporating injected cement for the fixation of osteoporotic trochanteric fractures. J Orthop Trauma 16(5):311–316PubMedCrossRef Augat P, Rapp S, Claes L (2002) A modified hip screw incorporating injected cement for the fixation of osteoporotic trochanteric fractures. J Orthop Trauma 16(5):311–316PubMedCrossRef
2.
go back to reference Bonnaire F, Zenker H, Lill C, Weber AT, Linke B (2005) Treatment strategies for proximal femur fractures in osteoporotic patients. Osteoporos Int 16(Suppl 2):S93–S102PubMedCrossRef Bonnaire F, Zenker H, Lill C, Weber AT, Linke B (2005) Treatment strategies for proximal femur fractures in osteoporotic patients. Osteoporos Int 16(Suppl 2):S93–S102PubMedCrossRef
3.
go back to reference Giannoudis PV, Schneider E (2006) Principles of fixation of osteoporotic fractures. J Bone Joint Surg Br 88(10):1272–1278PubMedCrossRef Giannoudis PV, Schneider E (2006) Principles of fixation of osteoporotic fractures. J Bone Joint Surg Br 88(10):1272–1278PubMedCrossRef
4.
go back to reference Gisep A, Curtis R, Flutsch S, Suhm N (2006) Augmentation of osteoporotic bone: effect of pulsed jet-lavage on injection forces, cement distribution, and push-out strength of implants. J Biomed Mater Res B Appl Biomater 78(1):83–88PubMed Gisep A, Curtis R, Flutsch S, Suhm N (2006) Augmentation of osteoporotic bone: effect of pulsed jet-lavage on injection forces, cement distribution, and push-out strength of implants. J Biomed Mater Res B Appl Biomater 78(1):83–88PubMed
5.
go back to reference Heini PF (2005) The current treatment-a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon’s perspective. Osteoporos Int 16(Suppl 2):S85–S92PubMedCrossRef Heini PF (2005) The current treatment-a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon’s perspective. Osteoporos Int 16(Suppl 2):S85–S92PubMedCrossRef
6.
go back to reference Hertel R, Aebi M, Ganz R (1990) Osteosynthesis in high-grade osteoporosis. Unfallchirurg 93(10):479–484PubMed Hertel R, Aebi M, Ganz R (1990) Osteosynthesis in high-grade osteoporosis. Unfallchirurg 93(10):479–484PubMed
7.
go back to reference Hertel R (2005) Fractures of the proximal humerus in osteoporotic bone. Osteoporos Int 16(Suppl 2):S65–S72PubMedCrossRef Hertel R (2005) Fractures of the proximal humerus in osteoporotic bone. Osteoporos Int 16(Suppl 2):S65–S72PubMedCrossRef
8.
go back to reference Kettunen J, Kroger H (2005) Surgical treatment of ankle and foot fractures in the elderly. Osteoporos Int 16(Suppl 2):S103–S106PubMedCrossRef Kettunen J, Kroger H (2005) Surgical treatment of ankle and foot fractures in the elderly. Osteoporos Int 16(Suppl 2):S103–S106PubMedCrossRef
9.
go back to reference Kramer A, Angst M, Gasser B, Ganz R (2000) [Increasing bone screw anchoring in the femur head by cement administration via the implant-a biomechanical study]. Z Orthop Ihre Grenzgeb 138(5):464–469PubMedCrossRef Kramer A, Angst M, Gasser B, Ganz R (2000) [Increasing bone screw anchoring in the femur head by cement administration via the implant-a biomechanical study]. Z Orthop Ihre Grenzgeb 138(5):464–469PubMedCrossRef
10.
go back to reference Larsson S (2006) Cement augmentation in fracture treatment. Scand J Surg 95(2):111–118PubMed Larsson S (2006) Cement augmentation in fracture treatment. Scand J Surg 95(2):111–118PubMed
12.
go back to reference von der Linden P, Gisep A, Boner V, Windolf M, Appelt A, Suhm N (2006) Biomechanical evaluation of a new augmentation method for enhanced screw fixation in osteoporotic proximal femoral fractures. J Orthop Res 24(12):2230–2237PubMedCrossRef von der Linden P, Gisep A, Boner V, Windolf M, Appelt A, Suhm N (2006) Biomechanical evaluation of a new augmentation method for enhanced screw fixation in osteoporotic proximal femoral fractures. J Orthop Res 24(12):2230–2237PubMedCrossRef
14.
go back to reference Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM (1995) The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am 77(7):1058–1064PubMed Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM (1995) The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am 77(7):1058–1064PubMed
15.
go back to reference Baumgaertner MR, Solberg BD (1997) Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br 79(6):969–971PubMedCrossRef Baumgaertner MR, Solberg BD (1997) Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br 79(6):969–971PubMedCrossRef
16.
go back to reference Felsenberg D, Boonen S (2005) The bone quality framework: determinants of bone strength and their interrelationships, and implications for osteoporosis management. Clin Ther 27(1):1–11PubMedCrossRef Felsenberg D, Boonen S (2005) The bone quality framework: determinants of bone strength and their interrelationships, and implications for osteoporosis management. Clin Ther 27(1):1–11PubMedCrossRef
17.
go back to reference Ammann P, Rizzoli R (2003) Bone strength and its determinants. Osteoporos Int 14(Suppl 3):S13–S18PubMed Ammann P, Rizzoli R (2003) Bone strength and its determinants. Osteoporos Int 14(Suppl 3):S13–S18PubMed
18.
go back to reference Baruffaldi F, Barbanti BG, Testoni M, et al. (1996) Femoral densitometry as potential preoperative indicator for cementation of hip prosthesis. Radiol Med (Torino) 92(3):193–198 Baruffaldi F, Barbanti BG, Testoni M, et al. (1996) Femoral densitometry as potential preoperative indicator for cementation of hip prosthesis. Radiol Med (Torino) 92(3):193–198
19.
go back to reference Suhm N, Hengg C, Schwyn R, Windolf M, Quarz V, Hanni M (2006) Mechanical torque measurement predicts load to implant cut-out: a biomechanical study investigating DHS(R) anchorage in femoral heads. Arch Orthop Trauma Surg 127:469–474PubMedCrossRef Suhm N, Hengg C, Schwyn R, Windolf M, Quarz V, Hanni M (2006) Mechanical torque measurement predicts load to implant cut-out: a biomechanical study investigating DHS(R) anchorage in femoral heads. Arch Orthop Trauma Surg 127:469–474PubMedCrossRef
20.
go back to reference Muller ME (1980) [Classification and international AO-documentation of femur fractures]. Unfallheilkunde 83(5):251–259PubMed Muller ME (1980) [Classification and international AO-documentation of femur fractures]. Unfallheilkunde 83(5):251–259PubMed
21.
go back to reference Garden RS (1974) Reduction and fixation of subcapital fractures of the femur. Orthop Clin North Am 5(4):683–712PubMed Garden RS (1974) Reduction and fixation of subcapital fractures of the femur. Orthop Clin North Am 5(4):683–712PubMed
22.
go back to reference Yang R, Tsai K, Chieng P, Liu T (1997) Symmetry of bone mineral density at the proximal femur with emphasis on the effect of side dominance. Calcif Tissue Int 61(3):189–191PubMedCrossRef Yang R, Tsai K, Chieng P, Liu T (1997) Symmetry of bone mineral density at the proximal femur with emphasis on the effect of side dominance. Calcif Tissue Int 61(3):189–191PubMedCrossRef
23.
go back to reference Moroni A, Orienti L, Stea S, Visentin M (1996) Improvement of the bone-pin interface with hydroxyapatite coating: an in vivo long-term experimental study. J Orthop Trauma 10(4):236–242PubMedCrossRef Moroni A, Orienti L, Stea S, Visentin M (1996) Improvement of the bone-pin interface with hydroxyapatite coating: an in vivo long-term experimental study. J Orthop Trauma 10(4):236–242PubMedCrossRef
24.
go back to reference Moroni A, Toksvig-Larsen S, Maltarello MC, Orienti L, Stea S, Giannini S (1998) A comparison of hydroxyapatite-coated, titanium-coated, and uncoated tapered external-fixation pins. An in vivo study in sheep. J Bone Joint Surg Am 80(4):547–554PubMed Moroni A, Toksvig-Larsen S, Maltarello MC, Orienti L, Stea S, Giannini S (1998) A comparison of hydroxyapatite-coated, titanium-coated, and uncoated tapered external-fixation pins. An in vivo study in sheep. J Bone Joint Surg Am 80(4):547–554PubMed
25.
go back to reference Moroni A, Faldini C, Pegreffi F, Giannini S (2002) Fixation strength of tapered versus bicylindrical hydroxyapatite-coated external fixation pins: an animal study. J Biomed Mater Res 63(1):61–64PubMedCrossRef Moroni A, Faldini C, Pegreffi F, Giannini S (2002) Fixation strength of tapered versus bicylindrical hydroxyapatite-coated external fixation pins: an animal study. J Biomed Mater Res 63(1):61–64PubMedCrossRef
26.
go back to reference Moroni A, Faldini C, Giannini S, Wippermann B (2003) Plate fixation with hydroxyapatite-coated screws: a comparative loaded study. Clin Orthop Relat Res 408:262–267PubMedCrossRef Moroni A, Faldini C, Giannini S, Wippermann B (2003) Plate fixation with hydroxyapatite-coated screws: a comparative loaded study. Clin Orthop Relat Res 408:262–267PubMedCrossRef
27.
go back to reference Moroni A, Faldini C, Marchetti S, Manca M, Consoli V, Giannini S (2001) Improvement of the bone-pin interface strength in osteoporotic bone with use of hydroxyapatite-coated tapered external-fixation pins. A prospective, randomized clinical study of wrist fractures. J Bone Joint Surg Am 83-A(5):717–721PubMed Moroni A, Faldini C, Marchetti S, Manca M, Consoli V, Giannini S (2001) Improvement of the bone-pin interface strength in osteoporotic bone with use of hydroxyapatite-coated tapered external-fixation pins. A prospective, randomized clinical study of wrist fractures. J Bone Joint Surg Am 83-A(5):717–721PubMed
28.
go back to reference Moroni A, Faldini C, Pegreffi F, Giannini S (2004) HA-coated screws decrease the incidence of fixation failure in osteoporotic trochanteric fractures. Clin Orthop Relat Res 425:87–92PubMedCrossRef Moroni A, Faldini C, Pegreffi F, Giannini S (2004) HA-coated screws decrease the incidence of fixation failure in osteoporotic trochanteric fractures. Clin Orthop Relat Res 425:87–92PubMedCrossRef
29.
go back to reference Moroni A, Faldini C, Pegreffi F, Hoang-Kim A, Vannini F, Giannini S (2005) Dynamic hip screw compared with external fixation for treatment of osteoporotic pertrochanteric fractures. A prospective, randomized study. J Bone Joint Surg Am 87(4):753–759PubMedCrossRef Moroni A, Faldini C, Pegreffi F, Hoang-Kim A, Vannini F, Giannini S (2005) Dynamic hip screw compared with external fixation for treatment of osteoporotic pertrochanteric fractures. A prospective, randomized study. J Bone Joint Surg Am 87(4):753–759PubMedCrossRef
30.
go back to reference Gisep A, Curtis R, Hanni M, Suhm N (2006) Augmentation of implant purchase with bone cements: an in vitro study of injectability and dough distribution. J Biomed Mater Res B Appl Biomater 77(1):114–119PubMed Gisep A, Curtis R, Hanni M, Suhm N (2006) Augmentation of implant purchase with bone cements: an in vitro study of injectability and dough distribution. J Biomed Mater Res B Appl Biomater 77(1):114–119PubMed
31.
go back to reference Krischak GD, Augat P, Wachter NJ, Kinzl L, Claes LE (1999) Predictive value of bone mineral density and Singh index for the in vitro mechanical properties of cancellous bone in the femoral head. Clin Biomech (Bristol, Avon) 14(5):346–351CrossRef Krischak GD, Augat P, Wachter NJ, Kinzl L, Claes LE (1999) Predictive value of bone mineral density and Singh index for the in vitro mechanical properties of cancellous bone in the femoral head. Clin Biomech (Bristol, Avon) 14(5):346–351CrossRef
32.
go back to reference Wachter NJ, Augat P, Hoellen IP et al (2001) Predictive value of Singh index and bone mineral density measured by quantitative computed tomography in determining the local cancellous bone quality of the proximal femur. Clin Biomech (Bristol, Avon) 16(3):257–262CrossRef Wachter NJ, Augat P, Hoellen IP et al (2001) Predictive value of Singh index and bone mineral density measured by quantitative computed tomography in determining the local cancellous bone quality of the proximal femur. Clin Biomech (Bristol, Avon) 16(3):257–262CrossRef
33.
go back to reference Mizuno K, Shinomiya K, Nakai O, Shindo S, Otani K (2005) Intraoperative insertion torque of lumbar pedicle screw and postoperative radiographic evaluation: short-term observation. J Orthop Sci 10(2):137–144PubMedCrossRef Mizuno K, Shinomiya K, Nakai O, Shindo S, Otani K (2005) Intraoperative insertion torque of lumbar pedicle screw and postoperative radiographic evaluation: short-term observation. J Orthop Sci 10(2):137–144PubMedCrossRef
34.
go back to reference Reitman CA, Nguyen L, Fogel GR (2004) Biomechanical evaluation of relationship of screw pullout strength, insertional torque, and bone mineral density in the cervical spine. J Spinal Disord Tech 17(4):306–311PubMedCrossRef Reitman CA, Nguyen L, Fogel GR (2004) Biomechanical evaluation of relationship of screw pullout strength, insertional torque, and bone mineral density in the cervical spine. J Spinal Disord Tech 17(4):306–311PubMedCrossRef
35.
go back to reference Ozawa T, Takahashi K, Yamagata M, et al. (2005) Insertional torque of the lumbar pedicle screw during surgery. J Orthop Sci 10(2):133–136PubMedCrossRef Ozawa T, Takahashi K, Yamagata M, et al. (2005) Insertional torque of the lumbar pedicle screw during surgery. J Orthop Sci 10(2):133–136PubMedCrossRef
Metadata
Title
Quantification of bone strength by intraoperative torque measurement: a technical note
Authors
Norbert Suhm
Markus Haenni
Ronald Schwyn
Michael Hirschmann
Andreas Marc Müller
Publication date
01-06-2008
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 6/2008
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-008-0566-1

Other articles of this Issue 6/2008

Archives of Orthopaedic and Trauma Surgery 6/2008 Go to the issue