Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T

Authors: Marcus Carlsson, Johannes Töger, Mikael Kanski, Karin Markenroth Bloch, Freddy Ståhlberg, Einar Heiberg, Håkan Arheden

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2011

Login to get access

Abstract

Background

Three-dimensional time-resolved (4D) phase-contrast (PC) CMR can visualize and quantify cardiovascular flow but is hampered by long acquisition times. Acceleration with SENSE or k-t BLAST are two possibilities but results on validation are lacking, especially at 3 T. The aim of this study was therefore to validate quantitative in vivo cardiac 4D-acquisitions accelerated with parallel imaging and k-t BLAST at 1.5 T and 3 T with 2D-flow as the reference and to investigate if field strengths and type of acceleration have major effects on intracardiac flow visualization.

Methods

The local ethical committee approved the study. 13 healthy volunteers were scanned at both 1.5 T and 3 T in random order with 2D-flow of the aorta and main pulmonary artery and two 4D-flow sequences of the heart accelerated with SENSE and k-t BLAST respectively. 2D-image planes were reconstructed at the aortic and pulmonary outflow. Flow curves were calculated and peak flows and stroke volumes (SV) compared to the results from 2D-flow acquisitions. Intra-cardiac flow was visualized using particle tracing and image quality based on the flow patterns of the particles was graded using a four-point scale.

Results

Good accuracy of SV quantification was found using 3 T 4D-SENSE (r2 = 0.86, -0.7 ± 7.6%) and although a larger bias was found on 1.5 T (r2 = 0.71, -3.6 ± 14.8%), the difference was not significant (p = 0.46). Accuracy of 4D k-t BLAST for SV was lower (p < 0.01) on 1.5 T (r2 = 0.65, -15.6 ± 13.7%) compared to 3 T (r2 = 0.64, -4.6 ± 10.0%). Peak flow was lower with 4D-SENSE at both 3 T and 1.5 T compared to 2D-flow (p < 0.01) and even lower with 4D k-t BLAST at both scanners (p < 0.01). Intracardiac flow visualization did not differ between 1.5 T and 3 T (p = 0.09) or between 4D-SENSE or 4D k-t BLAST (p = 0.85).

Conclusions

The present study showed that quantitative 4D flow accelerated with SENSE has good accuracy at 3 T and compares favourably to 1.5 T. 4D flow accelerated with k-t BLAST underestimate flow velocities and thereby yield too high bias for intra-cardiac quantitative in vivo use at the present time. For intra-cardiac 4D-flow visualization, however, 1.5 T and 3 T as well as SENSE or k-t BLAST can be used with similar quality.
Appendix
Available only for authorised users
Literature
1.
go back to reference Thomsen C, Cortsen M, Sondergaard L, Henriksen O, Stahlberg F: A segmented K-space velocity mapping protocol for quantification of renal artery blood flow during breath-holding. J Magn Reson Imaging. 1995, 5: 393-401. 10.1002/jmri.1880050405.CrossRefPubMed Thomsen C, Cortsen M, Sondergaard L, Henriksen O, Stahlberg F: A segmented K-space velocity mapping protocol for quantification of renal artery blood flow during breath-holding. J Magn Reson Imaging. 1995, 5: 393-401. 10.1002/jmri.1880050405.CrossRefPubMed
2.
go back to reference Kramer CM, Barkhausen J, Flamm SD, Kim RJ, Nagel E: Standardized cardiovascular magnetic resonance imaging (CMR) protocols, society for cardiovascular magnetic resonance: board of trustees task force on standardized protocols. J Cardiovasc Magn Reson. 2008, 10: 35-10.1186/1532-429X-10-35.CrossRefPubMedPubMedCentral Kramer CM, Barkhausen J, Flamm SD, Kim RJ, Nagel E: Standardized cardiovascular magnetic resonance imaging (CMR) protocols, society for cardiovascular magnetic resonance: board of trustees task force on standardized protocols. J Cardiovasc Magn Reson. 2008, 10: 35-10.1186/1532-429X-10-35.CrossRefPubMedPubMedCentral
3.
go back to reference Arheden H, Holmqvist C, Thilen U, Hanseus K, Bjorkhem G, Pahlm O, Laurin S, Stahlberg F: Left-to-right cardiac shunts: comparison of measurements obtained with MR velocity mapping and with radionuclide angiography. Radiology. 1999, 211: 453-458.CrossRefPubMed Arheden H, Holmqvist C, Thilen U, Hanseus K, Bjorkhem G, Pahlm O, Laurin S, Stahlberg F: Left-to-right cardiac shunts: comparison of measurements obtained with MR velocity mapping and with radionuclide angiography. Radiology. 1999, 211: 453-458.CrossRefPubMed
4.
go back to reference Carlsson M, Ugander M, Heiberg E, Arheden H: The quantitative relationship between longitudinal and radial function in left, right, and total heart pumping in humans. Am J Physiol Heart Circ Physiol. 2007, 293: H636-644. 10.1152/ajpheart.01376.2006.CrossRefPubMed Carlsson M, Ugander M, Heiberg E, Arheden H: The quantitative relationship between longitudinal and radial function in left, right, and total heart pumping in humans. Am J Physiol Heart Circ Physiol. 2007, 293: H636-644. 10.1152/ajpheart.01376.2006.CrossRefPubMed
5.
go back to reference Gatehouse PD, Rolf MP, Graves MJ, Hofman MB, Totman J, Werner B, Quest RA, Liu Y, von Spiczak J, Dieringer M, et al: Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J Cardiovasc Magn Reson. 2010, 12: 5-10.1186/1532-429X-12-5.CrossRefPubMedPubMedCentral Gatehouse PD, Rolf MP, Graves MJ, Hofman MB, Totman J, Werner B, Quest RA, Liu Y, von Spiczak J, Dieringer M, et al: Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J Cardiovasc Magn Reson. 2010, 12: 5-10.1186/1532-429X-12-5.CrossRefPubMedPubMedCentral
6.
go back to reference Rolf MP, Hofman MB, Gatehouse PD, Markenroth Bloch K, Heymans MW, Ebbers T, Graves MJ, Totman JJ, Werner B, Rossum AC, et al: Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification - A multi-vendor study. J Cardiovasc Magn Reson. 2011, 13: 18-10.1186/1532-429X-13-18.CrossRefPubMedPubMedCentral Rolf MP, Hofman MB, Gatehouse PD, Markenroth Bloch K, Heymans MW, Ebbers T, Graves MJ, Totman JJ, Werner B, Rossum AC, et al: Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification - A multi-vendor study. J Cardiovasc Magn Reson. 2011, 13: 18-10.1186/1532-429X-13-18.CrossRefPubMedPubMedCentral
7.
go back to reference Markl M, Kilner PJ, Ebbers T: Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2011, 13: 7-10.1186/1532-429X-13-7.CrossRefPubMedPubMedCentral Markl M, Kilner PJ, Ebbers T: Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2011, 13: 7-10.1186/1532-429X-13-7.CrossRefPubMedPubMedCentral
8.
go back to reference Tsao J, Boesiger P, Pruessmann KP: k-t BLAST and k-t SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med. 2003, 50: 1031-1042. 10.1002/mrm.10611.CrossRefPubMed Tsao J, Boesiger P, Pruessmann KP: k-t BLAST and k-t SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med. 2003, 50: 1031-1042. 10.1002/mrm.10611.CrossRefPubMed
9.
go back to reference Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P: Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE. Magn Reson Med. 2005, 54: 1430-1438. 10.1002/mrm.20730.CrossRefPubMed Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P: Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE. Magn Reson Med. 2005, 54: 1430-1438. 10.1002/mrm.20730.CrossRefPubMed
10.
go back to reference Stadlbauer A, van der Riet W, Globits S, Crelier G, Salomonowitz E: Accelerated Phase-Contrast MR Imaging: Comparison of k-t BLAST With SENSE and Doppler Ultrasound for Velocity and Flow Measurements in the Aorta. Journal of Magnetic Resonance Imaging. 2009, 29: 817-824. 10.1002/jmri.21706.CrossRefPubMed Stadlbauer A, van der Riet W, Globits S, Crelier G, Salomonowitz E: Accelerated Phase-Contrast MR Imaging: Comparison of k-t BLAST With SENSE and Doppler Ultrasound for Velocity and Flow Measurements in the Aorta. Journal of Magnetic Resonance Imaging. 2009, 29: 817-824. 10.1002/jmri.21706.CrossRefPubMed
11.
go back to reference Marshall I: Feasibility of k-t BLAST technique for measuring "seven-dimensional" fluid flow. J Magn Reson Imaging. 2006, 23: 189-196. 10.1002/jmri.20494.CrossRefPubMed Marshall I: Feasibility of k-t BLAST technique for measuring "seven-dimensional" fluid flow. J Magn Reson Imaging. 2006, 23: 189-196. 10.1002/jmri.20494.CrossRefPubMed
12.
go back to reference Stadlbauer A, van der Riet W, Crelier G, Salomonowitz E: Accelerated time-resolved three-dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and k-t BLAST. European Journal of Radiology. 2010, 75: E15-E21.CrossRefPubMed Stadlbauer A, van der Riet W, Crelier G, Salomonowitz E: Accelerated time-resolved three-dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and k-t BLAST. European Journal of Radiology. 2010, 75: E15-E21.CrossRefPubMed
13.
go back to reference Westenberg JJ, Roes SD, Ajmone Marsan N, Binnendijk NM, Doornbos J, Bax JJ, Reiber JH, de Roos A, van der Geest RJ: Mitral valve and tricuspid valve blood flow: accurate quantification with 3D velocity-encoded MR imaging with retrospective valve tracking. Radiology. 2008, 249: 792-800. 10.1148/radiol.2492080146.CrossRefPubMed Westenberg JJ, Roes SD, Ajmone Marsan N, Binnendijk NM, Doornbos J, Bax JJ, Reiber JH, de Roos A, van der Geest RJ: Mitral valve and tricuspid valve blood flow: accurate quantification with 3D velocity-encoded MR imaging with retrospective valve tracking. Radiology. 2008, 249: 792-800. 10.1148/radiol.2492080146.CrossRefPubMed
14.
go back to reference Lotz J, Doker R, Noeske R, Schuttert M, Felix R, Galanski M, Gutberlet M, Meyer GP: In vitro validation of phase-contrast flow measurements at 3 T in comparison to 1.5 T: precision, accuracy, and signal-to-noise ratios. J Magn Reson Imaging. 2005, 21: 604-610. 10.1002/jmri.20275.CrossRefPubMed Lotz J, Doker R, Noeske R, Schuttert M, Felix R, Galanski M, Gutberlet M, Meyer GP: In vitro validation of phase-contrast flow measurements at 3 T in comparison to 1.5 T: precision, accuracy, and signal-to-noise ratios. J Magn Reson Imaging. 2005, 21: 604-610. 10.1002/jmri.20275.CrossRefPubMed
15.
go back to reference Dyverfeldt P, Kvitting JP, Sigfridsson A, Engvall J, Bolger AF, Ebbers T: Assessment of fluctuating velocities in disturbed cardiovascular blood flow: in vivo feasibility of generalized phase-contrast MRI. J Magn Reson Imaging. 2008, 28: 655-663. 10.1002/jmri.21475.CrossRefPubMed Dyverfeldt P, Kvitting JP, Sigfridsson A, Engvall J, Bolger AF, Ebbers T: Assessment of fluctuating velocities in disturbed cardiovascular blood flow: in vivo feasibility of generalized phase-contrast MRI. J Magn Reson Imaging. 2008, 28: 655-663. 10.1002/jmri.21475.CrossRefPubMed
16.
go back to reference Heiberg E, Sjogren J, Ugander M, Carlsson M, Engblom H, Arheden H: Design and validation of Segment--freely available software for cardiovascular image analysis. BMC Med Imaging. 2010, 10: 1-10.1186/1471-2342-10-1.CrossRefPubMedPubMedCentral Heiberg E, Sjogren J, Ugander M, Carlsson M, Engblom H, Arheden H: Design and validation of Segment--freely available software for cardiovascular image analysis. BMC Med Imaging. 2010, 10: 1-10.1186/1471-2342-10-1.CrossRefPubMedPubMedCentral
17.
go back to reference Reeder SB, Wintersperger BJ, Dietrich O, Lanz T, Greiser A, Reiser MF, Glazer GM, Schoenberg SO: Practical approaches to the evaluation of signal-to-noise ratio performance with parallel imaging: application with cardiac imaging and a 32-channel cardiac coil. Magn Reson Med. 2005, 54: 748-754. 10.1002/mrm.20636.CrossRefPubMed Reeder SB, Wintersperger BJ, Dietrich O, Lanz T, Greiser A, Reiser MF, Glazer GM, Schoenberg SO: Practical approaches to the evaluation of signal-to-noise ratio performance with parallel imaging: application with cardiac imaging and a 32-channel cardiac coil. Magn Reson Med. 2005, 54: 748-754. 10.1002/mrm.20636.CrossRefPubMed
18.
go back to reference Eriksson J, Carlhall CJ, Dyverfeldt P, Engvall J, Bolger AF, Ebbers T: Semi-automatic quantification of 4D left ventricular blood flow. J Cardiovasc Magn Reson. 2010, 12: 9-10.1186/1532-429X-12-9.CrossRefPubMedPubMedCentral Eriksson J, Carlhall CJ, Dyverfeldt P, Engvall J, Bolger AF, Ebbers T: Semi-automatic quantification of 4D left ventricular blood flow. J Cardiovasc Magn Reson. 2010, 12: 9-10.1186/1532-429X-12-9.CrossRefPubMedPubMedCentral
19.
go back to reference Markl M, Chan FP, Alley MT, Wedding KL, Draney MT, Elkins CJ, Parker DW, Wicker R, Taylor CA, Herfkens RJ, Pelc NJ: Time-resolved three-dimensional phase-contrast MRI. J Magn Reson Imaging. 2003, 17: 499-506. 10.1002/jmri.10272.CrossRefPubMed Markl M, Chan FP, Alley MT, Wedding KL, Draney MT, Elkins CJ, Parker DW, Wicker R, Taylor CA, Herfkens RJ, Pelc NJ: Time-resolved three-dimensional phase-contrast MRI. J Magn Reson Imaging. 2003, 17: 499-506. 10.1002/jmri.10272.CrossRefPubMed
20.
go back to reference Markl M, Draney MT, Hope MD, Levin JM, Chan FP, Alley MT, Pelc NJ, Herfkens RJ: Time-resolved 3-dimensional velocity mapping in the thoracic aorta: visualization of 3-directional blood flow patterns in healthy volunteers and patients. J Comput Assist Tomogr. 2004, 28: 459-468. 10.1097/00004728-200407000-00005.CrossRefPubMed Markl M, Draney MT, Hope MD, Levin JM, Chan FP, Alley MT, Pelc NJ, Herfkens RJ: Time-resolved 3-dimensional velocity mapping in the thoracic aorta: visualization of 3-directional blood flow patterns in healthy volunteers and patients. J Comput Assist Tomogr. 2004, 28: 459-468. 10.1097/00004728-200407000-00005.CrossRefPubMed
21.
go back to reference Stalder AF, Russe MF, Frydrychowicz A, Bock J, Hennig J, Markl M: Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008, 60: 1218-1231. 10.1002/mrm.21778.CrossRefPubMed Stalder AF, Russe MF, Frydrychowicz A, Bock J, Hennig J, Markl M: Quantitative 2D and 3D phase contrast MRI: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008, 60: 1218-1231. 10.1002/mrm.21778.CrossRefPubMed
22.
go back to reference van der Hulst AE, Westenberg JJ, Kroft LJ, Bax JJ, Blom NA, de Roos A, Roest AA: Tetralogy of fallot: 3D velocity-encoded MR imaging for evaluation of right ventricular valve flow and diastolic function in patients after correction. Radiology. 2010, 256: 724-734. 10.1148/radiol.10092269.CrossRefPubMed van der Hulst AE, Westenberg JJ, Kroft LJ, Bax JJ, Blom NA, de Roos A, Roest AA: Tetralogy of fallot: 3D velocity-encoded MR imaging for evaluation of right ventricular valve flow and diastolic function in patients after correction. Radiology. 2010, 256: 724-734. 10.1148/radiol.10092269.CrossRefPubMed
23.
go back to reference Brix L, Ringgaard S, Rasmusson A, Sorensen TS, Kim WY: Three dimensional three component whole heart cardiovascular magnetic resonance velocity mapping: comparison of flow measurements from 3D and 2D acquisitions. J Cardiovasc Magn Reson. 2009, 11: 3-10.1186/1532-429X-11-3.CrossRefPubMedPubMedCentral Brix L, Ringgaard S, Rasmusson A, Sorensen TS, Kim WY: Three dimensional three component whole heart cardiovascular magnetic resonance velocity mapping: comparison of flow measurements from 3D and 2D acquisitions. J Cardiovasc Magn Reson. 2009, 11: 3-10.1186/1532-429X-11-3.CrossRefPubMedPubMedCentral
24.
go back to reference Johnson KM, Lum DP, Turski PA, Block WF, Mistretta CA, Wieben O: Improved 3D phase contrast MRI with off-resonance corrected dual echo VIPR. Magn Reson Med. 2008, 60: 1329-1336. 10.1002/mrm.21763.CrossRefPubMedPubMedCentral Johnson KM, Lum DP, Turski PA, Block WF, Mistretta CA, Wieben O: Improved 3D phase contrast MRI with off-resonance corrected dual echo VIPR. Magn Reson Med. 2008, 60: 1329-1336. 10.1002/mrm.21763.CrossRefPubMedPubMedCentral
25.
go back to reference Francois CJ, Lum DP, Johnson KM, Landgraf BR, Bley TA, Reeder SB, Schiebler ML, Grist TM, Wieben O: Renal arteries: isotropic, high-spatial-resolution, unenhanced MR angiography with three-dimensional radial phase contrast. Radiology. 258: 254-260. Francois CJ, Lum DP, Johnson KM, Landgraf BR, Bley TA, Reeder SB, Schiebler ML, Grist TM, Wieben O: Renal arteries: isotropic, high-spatial-resolution, unenhanced MR angiography with three-dimensional radial phase contrast. Radiology. 258: 254-260.
26.
go back to reference Chang W, Landgraf B, Johnson KM, Kecskemeti S, Wu Y, Velikina J, Rowley H, Wieben O, Mistretta C, Turski P: Velocity measurements in the middle cerebral arteries of healthy volunteers using 3D radial phase-contrast HYPRFlow: comparison with transcranial Doppler sonography and 2D phase-contrast MR imaging. AJNR Am J Neuroradiol. 2011, 32: 54-59. 10.3174/ajnr.A2457.CrossRefPubMedPubMedCentral Chang W, Landgraf B, Johnson KM, Kecskemeti S, Wu Y, Velikina J, Rowley H, Wieben O, Mistretta C, Turski P: Velocity measurements in the middle cerebral arteries of healthy volunteers using 3D radial phase-contrast HYPRFlow: comparison with transcranial Doppler sonography and 2D phase-contrast MR imaging. AJNR Am J Neuroradiol. 2011, 32: 54-59. 10.3174/ajnr.A2457.CrossRefPubMedPubMedCentral
27.
go back to reference Hope MD, Hope TA, Meadows AK, Ordovas KG, Urbania TH, Alley MT, Higgins CB: Bicuspid aortic valve: four-dimensional MR evaluation of ascending aortic systolic flow patterns. Radiology. 2010, 255: 53-61. 10.1148/radiol.09091437.CrossRefPubMed Hope MD, Hope TA, Meadows AK, Ordovas KG, Urbania TH, Alley MT, Higgins CB: Bicuspid aortic valve: four-dimensional MR evaluation of ascending aortic systolic flow patterns. Radiology. 2010, 255: 53-61. 10.1148/radiol.09091437.CrossRefPubMed
28.
go back to reference Harloff A, Simon J, Brendecke S, Assefa D, Helbing T, Frydrychowicz A, Weber J, Olschewski M, Strecker C, Hennig J, et al: Complex plaques in the proximal descending aorta: an underestimated embolic source of stroke. Stroke. 2010, 41: 1145-1150. 10.1161/STROKEAHA.109.577775.CrossRefPubMed Harloff A, Simon J, Brendecke S, Assefa D, Helbing T, Frydrychowicz A, Weber J, Olschewski M, Strecker C, Hennig J, et al: Complex plaques in the proximal descending aorta: an underestimated embolic source of stroke. Stroke. 2010, 41: 1145-1150. 10.1161/STROKEAHA.109.577775.CrossRefPubMed
29.
go back to reference Brandts A, Bertini M, van Dijk EJ, Delgado V, Marsan NA, van der Geest RJ, Siebelink HM, de Roos A, Bax JJ, Westenberg JJ: Left ventricular diastolic function assessment from three-dimensional three-directional velocity-encoded MRI with retrospective valve tracking. J Magn Reson Imaging. 2011, 33: 312-319. 10.1002/jmri.22424.CrossRefPubMed Brandts A, Bertini M, van Dijk EJ, Delgado V, Marsan NA, van der Geest RJ, Siebelink HM, de Roos A, Bax JJ, Westenberg JJ: Left ventricular diastolic function assessment from three-dimensional three-directional velocity-encoded MRI with retrospective valve tracking. J Magn Reson Imaging. 2011, 33: 312-319. 10.1002/jmri.22424.CrossRefPubMed
30.
go back to reference Bolger AF, Heiberg E, Karlsson M, Wigstrom L, Engvall J, Sigfridsson A, Ebbers T, Kvitting JP, Carlhall CJ, Wranne B: Transit of blood flow through the human left ventricle mapped by cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2007, 9: 741-747. 10.1080/10976640701544530.CrossRefPubMed Bolger AF, Heiberg E, Karlsson M, Wigstrom L, Engvall J, Sigfridsson A, Ebbers T, Kvitting JP, Carlhall CJ, Wranne B: Transit of blood flow through the human left ventricle mapped by cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2007, 9: 741-747. 10.1080/10976640701544530.CrossRefPubMed
31.
go back to reference Eriksson J, Dyverfeldt P, Engvall J, Bolger AF, Ebbers T, Carlhall CJ: Quantification of presystolic blood flow organization and energetics in the human left ventricle. Am J Physiol Heart Circ Physiol. 2011, 300: H2135-2141. 10.1152/ajpheart.00993.2010.CrossRefPubMed Eriksson J, Dyverfeldt P, Engvall J, Bolger AF, Ebbers T, Carlhall CJ: Quantification of presystolic blood flow organization and energetics in the human left ventricle. Am J Physiol Heart Circ Physiol. 2011, 300: H2135-2141. 10.1152/ajpheart.00993.2010.CrossRefPubMed
32.
go back to reference Kellman P, McVeigh ER: Image reconstruction in SNR units: a general method for SNR measurement. Magn Reson Med. 2005, 54: 1439-1447. 10.1002/mrm.20713.CrossRefPubMedPubMedCentral Kellman P, McVeigh ER: Image reconstruction in SNR units: a general method for SNR measurement. Magn Reson Med. 2005, 54: 1439-1447. 10.1002/mrm.20713.CrossRefPubMedPubMedCentral
33.
go back to reference Plein S, Ryf S, Schwitter J, Radjenovic A, Boesiger P, Kozerke S: Dynamic contrast-enhanced myocardial perfusion MRI accelerated with k-t sense. Magn Reson Med. 2007, 58: 777-785. 10.1002/mrm.21381.CrossRefPubMed Plein S, Ryf S, Schwitter J, Radjenovic A, Boesiger P, Kozerke S: Dynamic contrast-enhanced myocardial perfusion MRI accelerated with k-t sense. Magn Reson Med. 2007, 58: 777-785. 10.1002/mrm.21381.CrossRefPubMed
34.
go back to reference Hansen MS, Baltes C, Tsao J, Kozerke S, Pruessmann KP, Eggers H: k-t BLAST reconstruction from non-Cartesian k-t space sampling. Magn Reson Med. 2006, 55: 85-91. 10.1002/mrm.20734.CrossRefPubMed Hansen MS, Baltes C, Tsao J, Kozerke S, Pruessmann KP, Eggers H: k-t BLAST reconstruction from non-Cartesian k-t space sampling. Magn Reson Med. 2006, 55: 85-91. 10.1002/mrm.20734.CrossRefPubMed
35.
go back to reference Madore B, Hoge WS, Chao TC, Zientara GP, Chu R: Retrospectively gated cardiac cine imaging with temporal and spatial acceleration. Magn Reson Imaging. 2011, 29: 457-469. 10.1016/j.mri.2011.01.003.CrossRefPubMedPubMedCentral Madore B, Hoge WS, Chao TC, Zientara GP, Chu R: Retrospectively gated cardiac cine imaging with temporal and spatial acceleration. Magn Reson Imaging. 2011, 29: 457-469. 10.1016/j.mri.2011.01.003.CrossRefPubMedPubMedCentral
Metadata
Title
Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
Authors
Marcus Carlsson
Johannes Töger
Mikael Kanski
Karin Markenroth Bloch
Freddy Ståhlberg
Einar Heiberg
Håkan Arheden
Publication date
01-12-2011
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2011
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-13-55

Other articles of this Issue 1/2011

Journal of Cardiovascular Magnetic Resonance 1/2011 Go to the issue