Published in: Abdominal Radiology 11/2017

01-11-2017

Quality of routine diagnostic abdominal images generated from a novel detector-based spectral CT scanner: a technical report on a phantom and clinical study

Authors: Mojgan Hojjati, Steven Van Hedent, Negin Rassouli, Curtis Tatsuoka, David Jordan, Amar Dhanantwari, Prabhakar Rajiah

Published in: Abdominal Radiology | Issue 11/2017

Login to get access

Abstract

Purpose

To evaluate the image quality of routine diagnostic images generated from a novel detector-based spectral detector CT (SDCT) and compare it with CT images obtained from a conventional scanner with an energy-integrating detector (Brilliance iCT), Routine diagnostic (conventional/polyenergetic) images are non-material-specific images that resemble single-energy images obtained at the same radiation,

Methods

ACR guideline-based phantom evaluations were performed on both SDCT and iCT for CT adult body protocol. Retrospective analysis was performed on 50 abdominal CT scans from each scanner. Identical ROIs were placed at multiple locations in the abdomen and attenuation, noise, SNR, and CNR were measured. Subjective image quality analysis on a 5-point Likert scale was performed by 2 readers for enhancement, noise, and image quality.

Results

On phantom studies, SDCT images met the ACR requirements for CT number and deviation, CNR and effective radiation dose. In patients, the qualitative scores were significantly higher for the SDCT than the iCT, including enhancement (4.79 ± 0.38 vs. 4.60 ± 0.51, p = 0.005), noise (4.63 ± 0.42 vs. 4.29 ± 0.50, p = 0.000), and quality (4.85 ± 0.32, vs. 4.57 ± 0.50, p = 0.000). The SNR was higher in SDCT than iCT for liver (7.4 ± 4.2 vs. 7.2 ± 5.3, p = 0.662), spleen (8.6 ± 4.1 vs. 7.4 ± 3.5, p = 0.152), kidney (11.1 ± 6.3 vs. 8.7 ± 5.0, p = 0.033), pancreas (6.90 ± 3.45 vs 6.11 ± 2.64, p = 0.303), aorta (14.2 ± 6.2 vs. 11.0 ± 4.9, p = 0.007), but was slightly lower in lumbar-vertebra (7.7 ± 4.2 vs. 7.8 ± 4.5, p = 0.937). The CNR of the SDCT was also higher than iCT for all abdominal organs.

Conclusion

Image quality of routine diagnostic images from the SDCT is comparable to images of a conventional CT scanner with energy-integrating detectors, making it suitable for diagnostic purposes.
Literature
2.
go back to reference Yu L, Leng S, McCollough CH (2012) Dual-energy CT based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15CrossRefPubMed Yu L, Leng S, McCollough CH (2012) Dual-energy CT based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15CrossRefPubMed
3.
go back to reference Henzler T, Fink C, Schoenberg SO, Schopef J (2012) Dual-energy CT: radiation dose aspects. AJR Am J Roentgenol 199:S16–S25CrossRefPubMed Henzler T, Fink C, Schoenberg SO, Schopef J (2012) Dual-energy CT: radiation dose aspects. AJR Am J Roentgenol 199:S16–S25CrossRefPubMed
4.
go back to reference Holmes DR, Fletcher JG, Apel A, Huprich JE, et al. (2008) Evaluation of non-linear blending in dual-energy computed tomography. Eur J Radiol 68:409–413CrossRefPubMedPubMedCentral Holmes DR, Fletcher JG, Apel A, Huprich JE, et al. (2008) Evaluation of non-linear blending in dual-energy computed tomography. Eur J Radiol 68:409–413CrossRefPubMedPubMedCentral
5.
go back to reference Darras KE, McLaughlin PD, Kang H, et al. (2016) Virtual monoenergetic reconstruction of contrast-enhanced dual energy CT at 70 keV maximizes mural enhancement in acute small bowel obstruction. Eur J Radiol 85(5):950–956CrossRefPubMed Darras KE, McLaughlin PD, Kang H, et al. (2016) Virtual monoenergetic reconstruction of contrast-enhanced dual energy CT at 70 keV maximizes mural enhancement in acute small bowel obstruction. Eur J Radiol 85(5):950–956CrossRefPubMed
7.
go back to reference Ananthakrishnan L, Rajiah P, Ahn R, et al. (2017) Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT. Abdom Radiol 42(3):702–709CrossRef Ananthakrishnan L, Rajiah P, Ahn R, et al. (2017) Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT. Abdom Radiol 42(3):702–709CrossRef
8.
go back to reference Brooks RA, Di Chiro G (1978) Split-detector computed tomography: a preliminary report. Radiology 126(1):255–257CrossRefPubMed Brooks RA, Di Chiro G (1978) Split-detector computed tomography: a preliminary report. Radiology 126(1):255–257CrossRefPubMed
9.
go back to reference Can Hamersvelt RW, Schilham AM, Engelke K, et al. (2017) Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study. Eur Radiol . doi:10.1007/s00330-017-4801-4 Can Hamersvelt RW, Schilham AM, Engelke K, et al. (2017) Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study. Eur Radiol . doi:10.​1007/​s00330-017-4801-4
10.
go back to reference Wellenberg RH, Boomsma MF, van Osch JA, et al. (2017) Quantifying metal artifact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses. Eur J Radiol 88:61–70CrossRefPubMed Wellenberg RH, Boomsma MF, van Osch JA, et al. (2017) Quantifying metal artifact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses. Eur J Radiol 88:61–70CrossRefPubMed
11.
go back to reference Hicthethier T, Baebler B, Kroeger JR, et al. (2017) Monoenergetic reconstructions for imaging of coronary artery stents using spectral detector CT: in vitro experience and comparison to conventional images. J Ardiovasc Comput Tomogr 11(1):33–39CrossRef Hicthethier T, Baebler B, Kroeger JR, et al. (2017) Monoenergetic reconstructions for imaging of coronary artery stents using spectral detector CT: in vitro experience and comparison to conventional images. J Ardiovasc Comput Tomogr 11(1):33–39CrossRef
13.
go back to reference Van Hamersvelt RW, Willemink MJ, de Jong PA, et al. (2017) Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study. Eur Radiol . doi:10.1007/s00330=017-4737-8 Van Hamersvelt RW, Willemink MJ, de Jong PA, et al. (2017) Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study. Eur Radiol . doi:10.​1007/​s00330=​017-4737-8
14.
go back to reference Gabbai, M, Leichter I, Zimam R, et al. The Clinical Impact of Retrospective Analysis in Spectral Detector Dual Energy Body CT. Radiological Society of North America 2013 Scientific Assembly and Annual Meeting, December 1–December 6, 2013, Chicago. http://archive.rsna.org/2013/13018312.html Gabbai, M, Leichter I, Zimam R, et al. The Clinical Impact of Retrospective Analysis in Spectral Detector Dual Energy Body CT. Radiological Society of North America 2013 Scientific Assembly and Annual Meeting, December 1–December 6, 2013, Chicago. http://​archive.​rsna.​org/​2013/​13018312.​html
15.
go back to reference Martinez C, Rong Rong, Gilkeson RC, et al. Incremental benefit and clinical significance of retrospectively obtained spectral data in a novel spectral detector CT technology- Initial experiences and results. Radiological Society of North America 2014 Scientific Assembly and Annual Meeting, December 1 2014, http://archive.rsna.org/2014/14019367.html Martinez C, Rong Rong, Gilkeson RC, et al. Incremental benefit and clinical significance of retrospectively obtained spectral data in a novel spectral detector CT technology- Initial experiences and results. Radiological Society of North America 2014 Scientific Assembly and Annual Meeting, December 1 2014, http://​archive.​rsna.​org/​2014/​14019367.​html
16.
go back to reference McCollough CH, Bruesewitz MR, McNitt-Gray MF, et al. (2004) The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: practical tips, artifact examples, and pitfalls to avoid. Med Phys 31(9):2423–2442CrossRefPubMed McCollough CH, Bruesewitz MR, McNitt-Gray MF, et al. (2004) The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: practical tips, artifact examples, and pitfalls to avoid. Med Phys 31(9):2423–2442CrossRefPubMed
18.
go back to reference Godoy MCB, Heller SL, Naidich DP, et al. (2011) Dual-energy MDCT: comparison of pulmonary artery enhancement on dedicated CT pulmonary angiography, routine and low contrast volume studies. Eur J Radiol 79:311–317CrossRef Godoy MCB, Heller SL, Naidich DP, et al. (2011) Dual-energy MDCT: comparison of pulmonary artery enhancement on dedicated CT pulmonary angiography, routine and low contrast volume studies. Eur J Radiol 79:311–317CrossRef
19.
go back to reference Yu L, Primak AN, Liu X, McCollough CH (2009) Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT. Med Phys 36:1019–1024CrossRefPubMedPubMedCentral Yu L, Primak AN, Liu X, McCollough CH (2009) Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT. Med Phys 36:1019–1024CrossRefPubMedPubMedCentral
20.
go back to reference Megibow AJ, Sahani D (2012) Best Practice: implementation and use of abdominal dual-energy CT in routine patient care. AJR Am J Roentgenol 199:S71–S77CrossRefPubMed Megibow AJ, Sahani D (2012) Best Practice: implementation and use of abdominal dual-energy CT in routine patient care. AJR Am J Roentgenol 199:S71–S77CrossRefPubMed
21.
go back to reference Graser A, Johnson T, Chandarana H, et al. (2009) Dual energy CT: preliminary observations and potential clinical applications in the abdomen. Eur Radiol 19:13–23CrossRefPubMed Graser A, Johnson T, Chandarana H, et al. (2009) Dual energy CT: preliminary observations and potential clinical applications in the abdomen. Eur Radiol 19:13–23CrossRefPubMed
23.
go back to reference Matsumoto K, Jinzaki M, Tanami Y, et al. (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 102 kVp CT. Radiology 259(1):257–262CrossRefPubMed Matsumoto K, Jinzaki M, Tanami Y, et al. (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 102 kVp CT. Radiology 259(1):257–262CrossRefPubMed
24.
go back to reference Patel BN, Thomas JV, Lockaart ME, et al. (2013) Single source dual-energy spectral multidetector CT of pancreatic adenocarcinoma: optimization of energy level viewing significantly increases lesion contrast. Clin Radiol 68(2):148–154CrossRefPubMed Patel BN, Thomas JV, Lockaart ME, et al. (2013) Single source dual-energy spectral multidetector CT of pancreatic adenocarcinoma: optimization of energy level viewing significantly increases lesion contrast. Clin Radiol 68(2):148–154CrossRefPubMed
25.
go back to reference Wang U, Qian B, Li B, et al. (2013) Metal artifacts reduction using monochromatic images from spectral CT: evaluation of pedicle screws in patients with scoliosis. Eur J Radiol 82(8):e320–e326CrossRef Wang U, Qian B, Li B, et al. (2013) Metal artifacts reduction using monochromatic images from spectral CT: evaluation of pedicle screws in patients with scoliosis. Eur J Radiol 82(8):e320–e326CrossRef
Metadata
Title
Quality of routine diagnostic abdominal images generated from a novel detector-based spectral CT scanner: a technical report on a phantom and clinical study
Authors
Mojgan Hojjati
Steven Van Hedent
Negin Rassouli
Curtis Tatsuoka
David Jordan
Amar Dhanantwari
Prabhakar Rajiah
Publication date
01-11-2017
Publisher
Springer US
Published in
Abdominal Radiology / Issue 11/2017
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-017-1170-z

Other articles of this Issue 11/2017

Abdominal Radiology 11/2017 Go to the issue

Classics in Abdominal Imaging

The football sign

Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.