Published in: Cardiovascular Intervention and Therapeutics 3/2017

Open Access 01-07-2017 | Original Article

Prospective multi-center registry to evaluate efficacy and safety of the newly developed diamond-like carbon-coated cobalt–chromium coronary stent system

Authors: Kenji Ando, Katsuhisa Ishii, Eiji Tada, Kazuaki Kataoka, Atsushi Hirohata, Kenji Goto, Katsuyuki Kobayashi, Hiroshi Tsutsui, Makoto Nakahama, Hitoshi Nakashima, Shinichiroh Uchikawa, Junji Kanda, Satoshi Yasuda, Junji Yajima, Hiroshi Kitabayashi, Shumpei Sakurai, Keita Nakanishi, Naoto Inoue, Hirofumi Noike, Terumitsu Hasebe, Tetsuya Sato, Masao Yamasaki, Takeshi Kimura

Published in: Cardiovascular Intervention and Therapeutics | Issue 3/2017

Login to get access

Abstract

The purpose of this multi-center, non-randomized, and open-label clinical trial was to determine the non-inferiority of diamond-like carbon (DLC)-coated cobalt–chromium coronary stent, the MOMO DLC coronary stent, relative to commercially available bare-metal stents (MULTI-LINK VISION®). Nineteen centers in Japan participated. The study cohort consisted of 99 patients from 19 Japanese centers with single or double native coronary vessel disease with de novo and restenosis lesions who met the study eligibility criteria. This cohort formed the safety analysis set. The efficacy analysis set consisted of 98 patients (one case was excluded for violating the eligibility criteria). The primary endpoint was target vessel failure (TVF) rate at 9 months after stent placement. Of the 98 efficacy analysis set patients, TVF occurred in 11 patients (11.2 %, 95 % confidence interval 5.7–19.2 %) at 9 months after the index stent implantation. The upper 95 % confidence interval for TVF of the study stent was lower than that previously reported for the commercially available MULTI-LINK VISION® (19.6 %), demonstrating non-inferiority of the study stent to MULTI-LINK VISION®. All the TVF cases were related to target vascular revascularization. None of the cases developed in-stent thrombosis or myocardial infarction. The average in-stent late loss and binary restenosis rate at the 6-month follow-up angiography were 0.69 mm and 10.5 %, respectively, which are lower than the reported values for commercially available bare-metal stents. In conclusion, the current pivotal clinical study evaluating the new MOMO DLC-coated coronary stent suggested its low rates of TVF and angiographic binary restenosis, and small in-stent late loss, although the data were considered preliminary considering the small sample size and single arm study design.
Literature
1.
go back to reference Nobuyoshi M, Kimura T, Nosaka H, Mioka S, Ueno K, Yokoi H, et al. Restenosis after successful percutaneous transluminal coronary angioplasty: serial angiographic follow-up of 229 patients. J Am Coll Cardiol. 1988;12(3):616–23.CrossRefPubMed Nobuyoshi M, Kimura T, Nosaka H, Mioka S, Ueno K, Yokoi H, et al. Restenosis after successful percutaneous transluminal coronary angioplasty: serial angiographic follow-up of 229 patients. J Am Coll Cardiol. 1988;12(3):616–23.CrossRefPubMed
2.
go back to reference Fischman DL, Leon MB, Baim DS, Schatz RA, Savage MP, Penn I, et al. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med. 1994;331:496–501.CrossRefPubMed Fischman DL, Leon MB, Baim DS, Schatz RA, Savage MP, Penn I, et al. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med. 1994;331:496–501.CrossRefPubMed
3.
go back to reference Serruys PW, de Jaegere P, Kiemeneij F, Macaya C, Rutsch W, Heyndrickx G, et al. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. N Engl J Med. 1994;331:489–95.CrossRefPubMed Serruys PW, de Jaegere P, Kiemeneij F, Macaya C, Rutsch W, Heyndrickx G, et al. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. N Engl J Med. 1994;331:489–95.CrossRefPubMed
4.
go back to reference Kaiser C, Galatius S, Erne P, Eberli F, Alber H, Rickli H, et al. Drug-eluting versus bare-metal stents in large coronary arteries. N Engl J Med. 2010;363:2310–9.CrossRefPubMed Kaiser C, Galatius S, Erne P, Eberli F, Alber H, Rickli H, et al. Drug-eluting versus bare-metal stents in large coronary arteries. N Engl J Med. 2010;363:2310–9.CrossRefPubMed
5.
go back to reference von Birgelen C, Basalus MWZ, Tandjung K, van Houwelingen KG, Stoel MG, Louwerenburg JHW, et al. A randomized controlled trial in second-generation zotarolimus-eluting resolute stents versus everolimus-eluting Xience V stents in real-world patients: the TWENTE trial. J Am Coll Cardiol. 2012;59:1350–61.CrossRef von Birgelen C, Basalus MWZ, Tandjung K, van Houwelingen KG, Stoel MG, Louwerenburg JHW, et al. A randomized controlled trial in second-generation zotarolimus-eluting resolute stents versus everolimus-eluting Xience V stents in real-world patients: the TWENTE trial. J Am Coll Cardiol. 2012;59:1350–61.CrossRef
6.
go back to reference Gott VL, Daggett RL, Young WP. Development of a carbon-coated, central-hinging, bileaflet valve. Ann Thorac Surg. 1989;48:S28–30.CrossRefPubMed Gott VL, Daggett RL, Young WP. Development of a carbon-coated, central-hinging, bileaflet valve. Ann Thorac Surg. 1989;48:S28–30.CrossRefPubMed
7.
go back to reference Hauert R. A review of modified DLC coatings for biological applications. Diam Relat Mater. 2003;12:583–9.CrossRef Hauert R. A review of modified DLC coatings for biological applications. Diam Relat Mater. 2003;12:583–9.CrossRef
8.
go back to reference Gutensohn K, Beythien C, Bau J, Fenner T, Grewe P, Koester R, et al. In vitro analyses of diamond-like carbon coated stents: reduction of metal ion release, platelet activation, and thrombogenicity. Thromb Res. 2000;99:577–85.CrossRefPubMed Gutensohn K, Beythien C, Bau J, Fenner T, Grewe P, Koester R, et al. In vitro analyses of diamond-like carbon coated stents: reduction of metal ion release, platelet activation, and thrombogenicity. Thromb Res. 2000;99:577–85.CrossRefPubMed
9.
go back to reference Mongrain R, Rodés CJ. Role of shear stress in atherosclerosis and restenosis after coronary stent implantation. Rev Esp Cardiol. 2006;59(1):1–4.CrossRefPubMed Mongrain R, Rodés CJ. Role of shear stress in atherosclerosis and restenosis after coronary stent implantation. Rev Esp Cardiol. 2006;59(1):1–4.CrossRefPubMed
10.
go back to reference Kastrati A, Mehilli J, Dirschinger J, Dotzer F, Schühlen H, Neumann FJ, et al. Intracoronary stenting and angiographic results:strut thickness effect on restenosis outcome (ISAR-STEREO) trial. Circulation. 2001;103:2816–21.CrossRefPubMed Kastrati A, Mehilli J, Dirschinger J, Dotzer F, Schühlen H, Neumann FJ, et al. Intracoronary stenting and angiographic results:strut thickness effect on restenosis outcome (ISAR-STEREO) trial. Circulation. 2001;103:2816–21.CrossRefPubMed
12.
go back to reference Morice MC, Serruys PW, Sousa JE, Fajadet J, Hayashi EB, Perin M, et al. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002;346(23):1773–80.CrossRefPubMed Morice MC, Serruys PW, Sousa JE, Fajadet J, Hayashi EB, Perin M, et al. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002;346(23):1773–80.CrossRefPubMed
13.
go back to reference Kereiakes DJ, Cox DA, Hermiller JB, Midei MG, Bachinsky WB, Nukta ED, et al. Usefulness of a cobalt chromium coronary stent alloy. Am J Cardiol. 2003;92:463–6.CrossRefPubMed Kereiakes DJ, Cox DA, Hermiller JB, Midei MG, Bachinsky WB, Nukta ED, et al. Usefulness of a cobalt chromium coronary stent alloy. Am J Cardiol. 2003;92:463–6.CrossRefPubMed
14.
go back to reference Kesavan S, Strange JW, Johnson TW, Flohr RS, Baumbach A. First-in-man evaluation of the MOMO cobalt–chromium carbon-coated stent. EuroIntervention. 2013;8(9):1012–8.CrossRefPubMed Kesavan S, Strange JW, Johnson TW, Flohr RS, Baumbach A. First-in-man evaluation of the MOMO cobalt–chromium carbon-coated stent. EuroIntervention. 2013;8(9):1012–8.CrossRefPubMed
15.
go back to reference Iijima R, Ikari Y, Amiya E, Tanimoto S, Nakazawa G, Kyono H, et al. The impact of metallic allergy on stent implantation: metal allergy and recurrence of in-stent restenosis. Int J Cardiol. 2005;104(3):319–25.CrossRefPubMed Iijima R, Ikari Y, Amiya E, Tanimoto S, Nakazawa G, Kyono H, et al. The impact of metallic allergy on stent implantation: metal allergy and recurrence of in-stent restenosis. Int J Cardiol. 2005;104(3):319–25.CrossRefPubMed
16.
go back to reference Inoue K, Abe K, Ando K, Shirai S, Nishiyama K, Nakanishi M, et al. Pathological analyses of long-term intracoronary Palmaz-Schatz stenting; Is its efficacy permanent? Cardiovasc Pathol. 2004;13(2):109–15.CrossRefPubMed Inoue K, Abe K, Ando K, Shirai S, Nishiyama K, Nakanishi M, et al. Pathological analyses of long-term intracoronary Palmaz-Schatz stenting; Is its efficacy permanent? Cardiovasc Pathol. 2004;13(2):109–15.CrossRefPubMed
Metadata
Title
Prospective multi-center registry to evaluate efficacy and safety of the newly developed diamond-like carbon-coated cobalt–chromium coronary stent system
Authors
Kenji Ando
Katsuhisa Ishii
Eiji Tada
Kazuaki Kataoka
Atsushi Hirohata
Kenji Goto
Katsuyuki Kobayashi
Hiroshi Tsutsui
Makoto Nakahama
Hitoshi Nakashima
Shinichiroh Uchikawa
Junji Kanda
Satoshi Yasuda
Junji Yajima
Hiroshi Kitabayashi
Shumpei Sakurai
Keita Nakanishi
Naoto Inoue
Hirofumi Noike
Terumitsu Hasebe
Tetsuya Sato
Masao Yamasaki
Takeshi Kimura
Publication date
01-07-2017
Publisher
Springer Japan
Published in
Cardiovascular Intervention and Therapeutics / Issue 3/2017
Print ISSN: 1868-4300
Electronic ISSN: 1868-4297
DOI
https://doi.org/10.1007/s12928-016-0407-z

Other articles of this Issue 3/2017

Cardiovascular Intervention and Therapeutics 3/2017 Go to the issue