Preclinical evaluation of posterior spine stabilization devices: can the current standards represent basic everyday life activities?

Authors: Luigi La Barbera, Fabio Galbusera, Hans-Joachim Wilke, Tomaso Villa

Published in: European Spine Journal | Issue 9/2016

Login to get access

Abstract

Purpose

To discuss whether the available standard methods for preclinical evaluation of posterior spine stabilization devices can represent basic everyday life activities and how to compare the results obtained with different procedures.

Methods

A comparative finite element study compared ASTM F1717 and ISO 12189 standards to validated instrumented L2–L4 segments undergoing standing, upper body flexion and extension. The internal loads on the spinal rod and the maximum stress on the implant are analysed.

Results

ISO recommended anterior support stiffness and force allow for reproducing bending moments measured in vivo on an instrumented physiological segment during upper body flexion. Despite the significance of ASTM model from an engineering point of view, the overly conservative vertebrectomy model represents an unrealistic worst case scenario. A method is proposed to determine the load to apply on assemblies with different anterior support stiffnesses to guarantee a comparable bending moment and reproduce specific everyday life activities.

Conclusions

The study increases our awareness on the use of the current standards to achieve meaningful results easy to compare and interpret.
Literature
2.
3.
go back to reference ISO Standard (2008) 12189: Implants for surgery—mechanical testing of implantable spinal devices—fatigue test method for spinal implant assemblies using an anterior support ISO Standard (2008) 12189: Implants for surgery—mechanical testing of implantable spinal devices—fatigue test method for spinal implant assemblies using an anterior support
4.
go back to reference Mosnier T (2008) Contribution à l’analyse bioméchanique et à l’évaluation des implants rachidiens. PhD Thesis, École Nationale Supérieure d’Arts et Métiers, FR Mosnier T (2008) Contribution à l’analyse bioméchanique et à l’évaluation des implants rachidiens. PhD Thesis, École Nationale Supérieure d’Arts et Métiers, FR
5.
go back to reference Carson W, Asher M, Boachie-Adjei O, Akbarnia B, Dzioba R, Lebwohl N (2003) History of isola-vsp fatigue testing results with correlation to clinical implant failures. In: Melkerson M, Griffith S, Kirkpatrick J (eds) Spinal implants: are we evaluating them appropriately? STP 1431. ASTM International, pp 3–16 Carson W, Asher M, Boachie-Adjei O, Akbarnia B, Dzioba R, Lebwohl N (2003) History of isola-vsp fatigue testing results with correlation to clinical implant failures. In: Melkerson M, Griffith S, Kirkpatrick J (eds) Spinal implants: are we evaluating them appropriately? STP 1431. ASTM International, pp 3–16
6.
go back to reference Cunningham BW, Sefter JC, Shono Y, McAfee PC (1993) Static and cyclical biomechanical analysis of pedicle screw spinal constructs. Spine 18(12):1677–1688CrossRefPubMed Cunningham BW, Sefter JC, Shono Y, McAfee PC (1993) Static and cyclical biomechanical analysis of pedicle screw spinal constructs. Spine 18(12):1677–1688CrossRefPubMed
8.
go back to reference Graham J, Anderson P, Speciner D (2014) Letter to the editor in response to Villa T, La Barbera L, Galbusera F, Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems. Spine J 14(12):3067–3068. doi:10.1016/j.spinee.2014.07.026 CrossRefPubMed Graham J, Anderson P, Speciner D (2014) Letter to the editor in response to Villa T, La Barbera L, Galbusera F, Comparative analysis of international standards for the fatigue testing of posterior spinal fixation systems. Spine J 14(12):3067–3068. doi:10.​1016/​j.​spinee.​2014.​07.​026 CrossRefPubMed
11.
go back to reference Sapkas G, Kateros K, Papadakis SA, Brilakis E, Macheras G, Katonis P (2010) Treatment of unstable thoracolumbar burst fractures by indirect reduction and posterior stabilization: short-segment versus long-segment stabilization. Open Orthop J. 15(4):7–13. doi:10.2174/1874325001004010007 CrossRef Sapkas G, Kateros K, Papadakis SA, Brilakis E, Macheras G, Katonis P (2010) Treatment of unstable thoracolumbar burst fractures by indirect reduction and posterior stabilization: short-segment versus long-segment stabilization. Open Orthop J. 15(4):7–13. doi:10.​2174/​1874325001004010​007 CrossRef
12.
go back to reference Hwang JH, Modi HN, Yang JH, Kim SJ, Lee SH (2009) Short segment pedicle screw fixation for unstable T11-L2 fractures: with or without fusion? A three-year follow-up study. Acta Orthop Belg 75(6):822–827PubMed Hwang JH, Modi HN, Yang JH, Kim SJ, Lee SH (2009) Short segment pedicle screw fixation for unstable T11-L2 fractures: with or without fusion? A three-year follow-up study. Acta Orthop Belg 75(6):822–827PubMed
13.
go back to reference Korovessis P, Papazisis Z, Koureas G, Lambiris E (2004) Rigid, semirigid versus dynamic instrumentation for degenerative lumbar spinal stenosis: a correlative radiological and clinical analysis of short-term results. Spine (Phila Pa 1976) 29(7):735–42 Korovessis P, Papazisis Z, Koureas G, Lambiris E (2004) Rigid, semirigid versus dynamic instrumentation for degenerative lumbar spinal stenosis: a correlative radiological and clinical analysis of short-term results. Spine (Phila Pa 1976) 29(7):735–42
14.
go back to reference Chamoli U, Diwan AD (2014) Tsafnat N Pedicle screw-based posterior dynamic stabilizers for degenerative spine: in vitro biomechanical testing and clinical outcomes. J Biomed Mater Res A (United States) 102(9):p3324–p3340CrossRef Chamoli U, Diwan AD (2014) Tsafnat N Pedicle screw-based posterior dynamic stabilizers for degenerative spine: in vitro biomechanical testing and clinical outcomes. J Biomed Mater Res A (United States) 102(9):p3324–p3340CrossRef
15.
go back to reference Smith JS, Shaffrey CI, Ames CP, Demakakos J, Fu KM, Keshavarzi S, Li CM, Deviren V, Schwab FJ, Lafage V, Bess S (2012) Assessment of symptomatic rod fracture after posterior instrumented fusion for adult spinal deformity. Neurosurgery 71(4):862–867CrossRefPubMed Smith JS, Shaffrey CI, Ames CP, Demakakos J, Fu KM, Keshavarzi S, Li CM, Deviren V, Schwab FJ, Lafage V, Bess S (2012) Assessment of symptomatic rod fracture after posterior instrumented fusion for adult spinal deformity. Neurosurgery 71(4):862–867CrossRefPubMed
19.
go back to reference Rohlmann A, Bergmann G, Graichen F, Weber U (1997) Comparison of loads on internal spinal fixation devices measured in vitro and in vivo. Med Eng Phys 19(6):539–546CrossRefPubMed Rohlmann A, Bergmann G, Graichen F, Weber U (1997) Comparison of loads on internal spinal fixation devices measured in vitro and in vivo. Med Eng Phys 19(6):539–546CrossRefPubMed
20.
go back to reference Wilke HJ, Rohlmann A, Neller S, Schultheiss M, Bergmann G, Graichen F, Claes L (2001) Is it possible to simulate physiologic loading conditions by applying pure moments? A comparison of in vivo and in vitro load components in an internal fixator. Spine 26(6):636–642CrossRefPubMed Wilke HJ, Rohlmann A, Neller S, Schultheiss M, Bergmann G, Graichen F, Claes L (2001) Is it possible to simulate physiologic loading conditions by applying pure moments? A comparison of in vivo and in vitro load components in an internal fixator. Spine 26(6):636–642CrossRefPubMed
23.
25.
26.
27.
go back to reference ISO Standard (2010) 10243: tools for pressing—compression springs with rectangular section—housing dimensions and colour coding ISO Standard (2010) 10243: tools for pressing—compression springs with rectangular section—housing dimensions and colour coding
28.
go back to reference Rohlmann A, Bergmann G, Graichen F (1999) Loads on internal spinal fixators measured in different body positions. Europ Spine J 8(5):354–359CrossRef Rohlmann A, Bergmann G, Graichen F (1999) Loads on internal spinal fixators measured in different body positions. Europ Spine J 8(5):354–359CrossRef
29.
go back to reference Rohlmann A, Claes L, Graichen F, Neef P, Wilke HJ (1997) Loads on an internal spinal fixation device during walking. J Biomech 30(1):41–47CrossRefPubMed Rohlmann A, Claes L, Graichen F, Neef P, Wilke HJ (1997) Loads on an internal spinal fixation device during walking. J Biomech 30(1):41–47CrossRefPubMed
30.
go back to reference Rohlmann A, Graichen F, Bergmann G (2000) Changes in the loads on an internal spinal fixator after iliac-crest autograft. J Bone Joint Surg Br 82(3):445–449CrossRefPubMed Rohlmann A, Graichen F, Bergmann G (2000) Changes in the loads on an internal spinal fixator after iliac-crest autograft. J Bone Joint Surg Br 82(3):445–449CrossRefPubMed
31.
go back to reference Rohlmann A, Claes L, Graichen F, Neef P, Wilke H (2001) Comparison of intradiscal pressures and spinal fixator loads for different body positions and exercises. Ergonomics 44(8):781–794CrossRef Rohlmann A, Claes L, Graichen F, Neef P, Wilke H (2001) Comparison of intradiscal pressures and spinal fixator loads for different body positions and exercises. Ergonomics 44(8):781–794CrossRef
32.
go back to reference Skalli W, Robin S, Lavaste F, Dubousset J (1993) A biomechanical analysis of short segment spinal fixation using a three-dimensional geometric and mechanical model. Spine 18(5):536–545CrossRefPubMed Skalli W, Robin S, Lavaste F, Dubousset J (1993) A biomechanical analysis of short segment spinal fixation using a three-dimensional geometric and mechanical model. Spine 18(5):536–545CrossRefPubMed
33.
34.
go back to reference Rohlmann A, Graichen F, Bergmann G (2002) Loads on an internal spinal fixation device during physical therapy. Phys Ther 82(1):44–52PubMed Rohlmann A, Graichen F, Bergmann G (2002) Loads on an internal spinal fixation device during physical therapy. Phys Ther 82(1):44–52PubMed
35.
go back to reference Rohlmann A, Consmuller T, Dreischarf M, Bashkuev M, Disch A, Pries E, Duda GN, Schmidt H (2014) Measurement of the number of lumbar spinal movements in the sagittal plane in a 24-hour period. Eur Spine J (Germany) 23(11):2375–2384. doi:10.1007/s00586-014-3588-0 CrossRef Rohlmann A, Consmuller T, Dreischarf M, Bashkuev M, Disch A, Pries E, Duda GN, Schmidt H (2014) Measurement of the number of lumbar spinal movements in the sagittal plane in a 24-hour period. Eur Spine J (Germany) 23(11):2375–2384. doi:10.​1007/​s00586-014-3588-0 CrossRef
Metadata
Title
Preclinical evaluation of posterior spine stabilization devices: can the current standards represent basic everyday life activities?
Authors
Luigi La Barbera
Fabio Galbusera
Hans-Joachim Wilke
Tomaso Villa
Publication date
01-09-2016
Publisher
Springer Berlin Heidelberg
Published in
European Spine Journal / Issue 9/2016
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-016-4622-1

Other articles of this Issue 9/2016

European Spine Journal 9/2016 Go to the issue