Performance of beta- and high-energy gamma probes for the detection of cancer tissue in experimental surgical resection beds

Authors: Rita Garcia-Parra, Neal Clinthorne, Li Wang, Maria Picchio, Morand Piert

Published in: Annals of Nuclear Medicine | Issue 7/2011

Login to get access

Abstract

Objective

While high-energy gamma and beta probes have gained considerable attention due to their ability to detect cancerous lesions using 18F-FDG in humans intraoperatively, it is unknown whether the sensitivity of such probes would allow the detection of remaining tumor tissue in the resection bed after removal of macroscopically evident disease.

Methods

9L tumors (13 primaries, 17 lymph node metastases) were generated at the upper thigh of Fisher 344 rats. After approximately 2 weeks, microPET was performed to verify increased 18F-FDG tumor uptake. Tumors were surgically exposed and probe readings of tumor and background tissues were performed in triplicate. To evaluate the ability to detect tiny tumor lesions, 12 tumor fragments (range 0.001–0.032 g) were placed into the resection bed and measured to obtain a tumor-to-muscle ratio (TMR). Lesions were classified as positive if count rates were above 2.5 SD of muscle background. All tumor and muscle tissues were weighed and counted to obtain the “true” tumor-to-muscle background ratio (TMRcounter) for reference. The presence or absence of tumor tissue was verified by histology.

Results

In the presence of background gamma radiation, the beta probe detected all tumor lesions with TMR greater than 2.5 SD of muscle background (TMR range 1.24–3.9 for all 41 lesions). Despite suitable shielding, lesion identification by the gamma probe was clearly limited by the presence of background radioactivity. As a result, only 11/13 primary tumors, 6/17 lymph nodes and 1/11 tumor fragments were identified above 2.5 SD of muscle background (TMR range 0.64–3.59 for all lesions).

Conclusion

Under experimental conditions, the beta probe was capable to detect minute amounts of tumor tissue at the surface of resection beds. While clinical application of the current beta probe design may depend on the particular intraoperative circumstances including time requirements for surface scanning of resection beds, the data indicate clinical potential for novel designs of hand-held beta probes.
Literature
1.
go back to reference Wells BJ, Stotland P, Ko MA, Al-Sukhni W, Wunder J, Ferguson P, et al. Results of an aggressive approach to resection of locally recurrent rectal cancer. Ann Surg Oncol. 2007;14:390–5.PubMedCrossRef Wells BJ, Stotland P, Ko MA, Al-Sukhni W, Wunder J, Ferguson P, et al. Results of an aggressive approach to resection of locally recurrent rectal cancer. Ann Surg Oncol. 2007;14:390–5.PubMedCrossRef
2.
go back to reference Stewart CJ, Brennan BA, Hammond IG, Leung YC, McCartney AJ. Intraoperative assessment of ovarian tumors: a 5-year review with assessment of discrepant diagnostic cases. Int J Gynecol Pathol. 2006;25:216–22.PubMed Stewart CJ, Brennan BA, Hammond IG, Leung YC, McCartney AJ. Intraoperative assessment of ovarian tumors: a 5-year review with assessment of discrepant diagnostic cases. Int J Gynecol Pathol. 2006;25:216–22.PubMed
3.
go back to reference Yamaguchi K, Chijiiwa K, Saiki S, Shimizu S, Tsuneyoshi M, Tanaka M. Reliability of frozen section diagnosis of gallbladder tumor for detecting carcinoma and depth of its invasion. J Surg Oncol. 1997;65:132–6.PubMedCrossRef Yamaguchi K, Chijiiwa K, Saiki S, Shimizu S, Tsuneyoshi M, Tanaka M. Reliability of frozen section diagnosis of gallbladder tumor for detecting carcinoma and depth of its invasion. J Surg Oncol. 1997;65:132–6.PubMedCrossRef
4.
go back to reference Wahl RL, Hutchins GD, Buchsbaum DJ, Liebert M, Grossman HB, Fisher S. 18F-2-deoxy-2-fluoro-d-glucose uptake into human tumor xenografts. Feasibility studies for cancer imaging with positron-emission tomography. Cancer. 1991;67:1544–50.PubMedCrossRef Wahl RL, Hutchins GD, Buchsbaum DJ, Liebert M, Grossman HB, Fisher S. 18F-2-deoxy-2-fluoro-d-glucose uptake into human tumor xenografts. Feasibility studies for cancer imaging with positron-emission tomography. Cancer. 1991;67:1544–50.PubMedCrossRef
5.
go back to reference Bai B, Ruangma A, Laforest R, Tai Y-C, Leahy RM. Positron range modeling for statistical PET image reconstruction. IEEE. 2003;4:2501–5. Bai B, Ruangma A, Laforest R, Tai Y-C, Leahy RM. Positron range modeling for statistical PET image reconstruction. IEEE. 2003;4:2501–5.
6.
go back to reference Fowler JS, Wolf AP. Positron emitter-labeled compounds: priorities and problems. In: Phelps M, Mazziotta J, Schlebert H, editors. Positron emission tomography and autoradiography: principles and applications for the brain and heart. New York: Raven Press; 1986. p. 391–450. Fowler JS, Wolf AP. Positron emitter-labeled compounds: priorities and problems. In: Phelps M, Mazziotta J, Schlebert H, editors. Positron emission tomography and autoradiography: principles and applications for the brain and heart. New York: Raven Press; 1986. p. 391–450.
7.
go back to reference Levin CS, Hoffman EJ. Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution. Phys Med Biol. 1999;44:781–99.PubMedCrossRef Levin CS, Hoffman EJ. Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution. Phys Med Biol. 1999;44:781–99.PubMedCrossRef
8.
go back to reference Yamamoto S, Matsumoto K, Senda M. Optimum threshold setting for a positron-sensitive probe with background rejection capability. Ann Nucl Med. 2004;18:251–6.PubMedCrossRef Yamamoto S, Matsumoto K, Senda M. Optimum threshold setting for a positron-sensitive probe with background rejection capability. Ann Nucl Med. 2004;18:251–6.PubMedCrossRef
9.
go back to reference Raylman RR, Fisher SJ, Brown RS, Ethier SP, Wahl RL. Fluorine-18-fluorodeoxyglucose-guided breast cancer surgery with a positron-sensitive probe: validation in preclinical studies. J Nucl Med. 1995;36:1869–74.PubMed Raylman RR, Fisher SJ, Brown RS, Ethier SP, Wahl RL. Fluorine-18-fluorodeoxyglucose-guided breast cancer surgery with a positron-sensitive probe: validation in preclinical studies. J Nucl Med. 1995;36:1869–74.PubMed
10.
go back to reference Essner R, Hsueh EC, Haigh PI, Glass EC, Huynh Y, Daghighian F. Application of an [(18)F]fluorodeoxyglucose-sensitive probe for the intraoperative detection of malignancy. J Surg Res. 2001;96:120–6.PubMedCrossRef Essner R, Hsueh EC, Haigh PI, Glass EC, Huynh Y, Daghighian F. Application of an [(18)F]fluorodeoxyglucose-sensitive probe for the intraoperative detection of malignancy. J Surg Res. 2001;96:120–6.PubMedCrossRef
11.
go back to reference Essner R, Daghighian F, Giuliano AE. Advances in FDG PET probes in surgical oncology. Cancer J. 2002;8:100–8.PubMedCrossRef Essner R, Daghighian F, Giuliano AE. Advances in FDG PET probes in surgical oncology. Cancer J. 2002;8:100–8.PubMedCrossRef
12.
go back to reference Hoffman EJ, Tornai MP, Janecek M, Patt BE, Iwanczyk JS. Intraoperative probes and imaging probes (see comment). Eur J Nucl Med. 1999;26:913–35.PubMedCrossRef Hoffman EJ, Tornai MP, Janecek M, Patt BE, Iwanczyk JS. Intraoperative probes and imaging probes (see comment). Eur J Nucl Med. 1999;26:913–35.PubMedCrossRef
13.
go back to reference Gamelin J, Yang Y, Biswal N, Chen Y, Yan S, Zhang X, et al. A prototype hybrid intraoperative probe for ovarian cancer detection. Opt Express. 2009;17:7245–58.PubMedCrossRef Gamelin J, Yang Y, Biswal N, Chen Y, Yan S, Zhang X, et al. A prototype hybrid intraoperative probe for ovarian cancer detection. Opt Express. 2009;17:7245–58.PubMedCrossRef
14.
go back to reference Raylman RR. Performance of a dual, solid-state intraoperative probe system with 18F, 99mTc, and (111)In. J Nucl Med. 2001;42:352–60.PubMed Raylman RR. Performance of a dual, solid-state intraoperative probe system with 18F, 99mTc, and (111)In. J Nucl Med. 2001;42:352–60.PubMed
15.
go back to reference Zervos EE, Desai DC, DePalatis LR, Soble D, Martin EW. 18F-labeled fluorodeoxyglucose positron emission tomography-guided surgery for recurrent colorectal cancer: a feasibility study. J Surg Res. 2001;97:9–13.PubMedCrossRef Zervos EE, Desai DC, DePalatis LR, Soble D, Martin EW. 18F-labeled fluorodeoxyglucose positron emission tomography-guided surgery for recurrent colorectal cancer: a feasibility study. J Surg Res. 2001;97:9–13.PubMedCrossRef
16.
go back to reference Nwogu C, Fischer G, Tan D, Glinianski M, Lamonica D, Demmy T. Radioguided detection of lymph node metastasis in non-small cell lung cancer. Ann Thorac Surg. 2006;82:1815–20.PubMedCrossRef Nwogu C, Fischer G, Tan D, Glinianski M, Lamonica D, Demmy T. Radioguided detection of lymph node metastasis in non-small cell lung cancer. Ann Thorac Surg. 2006;82:1815–20.PubMedCrossRef
17.
go back to reference Piert M, Burian M, Meisetschlager G, Stein HJ, Ziegler S, Nahrig J, et al. Positron detection for the intraoperative localisation of cancer deposits. Eur J Nucl Med Mol Imaging. 2007;34:1534–44.PubMedCrossRef Piert M, Burian M, Meisetschlager G, Stein HJ, Ziegler S, Nahrig J, et al. Positron detection for the intraoperative localisation of cancer deposits. Eur J Nucl Med Mol Imaging. 2007;34:1534–44.PubMedCrossRef
18.
go back to reference Bogalhas F, Charon Y, Duval MA, Lefebvre F, Palfi S, Pinot L, et al. Development of a positron probe for localization and excision of brain tumours during surgery. Phys Med Biol. 2009;54:4439–53.PubMedCrossRef Bogalhas F, Charon Y, Duval MA, Lefebvre F, Palfi S, Pinot L, et al. Development of a positron probe for localization and excision of brain tumours during surgery. Phys Med Biol. 2009;54:4439–53.PubMedCrossRef
19.
go back to reference Cohn DE, Hall NC, Povoski SP, Seamon LG, Farrar WB, Martin EW Jr. Novel perioperative imaging with 18F-FDG PET/CT and intraoperative 18F-FDG detection using a handheld gamma probe in recurrent ovarian cancer. Gynecol Oncol. 2008;110:152–7.PubMedCrossRef Cohn DE, Hall NC, Povoski SP, Seamon LG, Farrar WB, Martin EW Jr. Novel perioperative imaging with 18F-FDG PET/CT and intraoperative 18F-FDG detection using a handheld gamma probe in recurrent ovarian cancer. Gynecol Oncol. 2008;110:152–7.PubMedCrossRef
20.
go back to reference Povoski SP, Hall NC, Martin EW Jr, Walker MJ. Multimodality approach of perioperative 18F-FDG PET/CT imaging, intraoperative 18F-FDG handheld gamma probe detection, and intraoperative ultrasound for tumor localization and verification of resection of all sites of hypermetabolic activity in a case of occult recurrent metastatic melanoma. World J Surg Oncol. 2008;6:1. Povoski SP, Hall NC, Martin EW Jr, Walker MJ. Multimodality approach of perioperative 18F-FDG PET/CT imaging, intraoperative 18F-FDG handheld gamma probe detection, and intraoperative ultrasound for tumor localization and verification of resection of all sites of hypermetabolic activity in a case of occult recurrent metastatic melanoma. World J Surg Oncol. 2008;6:1.
21.
go back to reference Strong VE, Galanis CJ, Riedl CC, Longo VA, Daghighian F, Humm JL, et al. Portable PET probes are a novel tool for intraoperative localization of tumor deposits. Ann Surg Innov Res. 2009;3:2. Strong VE, Galanis CJ, Riedl CC, Longo VA, Daghighian F, Humm JL, et al. Portable PET probes are a novel tool for intraoperative localization of tumor deposits. Ann Surg Innov Res. 2009;3:2.
23.
go back to reference Raylman RR, Wahl RL. Evaluation of ion-implanted-silicon detectors for use in intraoperative positron-sensitive probes. Med Phys. 1996;23:1889–95.PubMedCrossRef Raylman RR, Wahl RL. Evaluation of ion-implanted-silicon detectors for use in intraoperative positron-sensitive probes. Med Phys. 1996;23:1889–95.PubMedCrossRef
24.
go back to reference Piert M, Carey J, Clinthorne N. Probe-guided localization of cancer deposits using [(18)F]fluorodeoxyglucose. Q J Nucl Med Mol Imaging. 2008;52:37–49.PubMed Piert M, Carey J, Clinthorne N. Probe-guided localization of cancer deposits using [(18)F]fluorodeoxyglucose. Q J Nucl Med Mol Imaging. 2008;52:37–49.PubMed
25.
go back to reference Lederman RJ, Raylman RR, Fisher SJ, Kison PV, San H, Nabel EG, et al. Detection of atherosclerosis using a novel positron-sensitive probe and 18-fluorodeoxyglucose (FDG). Nucl Med Commun. 2001;22:747–53.PubMedCrossRef Lederman RJ, Raylman RR, Fisher SJ, Kison PV, San H, Nabel EG, et al. Detection of atherosclerosis using a novel positron-sensitive probe and 18-fluorodeoxyglucose (FDG). Nucl Med Commun. 2001;22:747–53.PubMedCrossRef
26.
go back to reference Gonzalez SJ, Gonzalez L, Wong J, Brader P, Zakowski M, Gonen M, et al. An analysis of the utility of handheld PET probes for the intraoperative localization of malignant tissue. J Gastrointest Surg. 2010;15:358–66.PubMedCrossRef Gonzalez SJ, Gonzalez L, Wong J, Brader P, Zakowski M, Gonen M, et al. An analysis of the utility of handheld PET probes for the intraoperative localization of malignant tissue. J Gastrointest Surg. 2010;15:358–66.PubMedCrossRef
27.
go back to reference Bai B, Li Q, Holdsworth CH, Asma E, Tai YC, Chatziioannou A, et al. Model-based normalization for iterative 3D PET image reconstruction. Phys Med Biol. 2002;47:2773–84.PubMedCrossRef Bai B, Li Q, Holdsworth CH, Asma E, Tai YC, Chatziioannou A, et al. Model-based normalization for iterative 3D PET image reconstruction. Phys Med Biol. 2002;47:2773–84.PubMedCrossRef
28.
go back to reference Franc BL, Mari C, Johnson D, Leong SP. The role of a positron- and high-energy gamma photon probe in intraoperative localization of recurrent melanoma. Clin Nucl Med. 2005;30:787–91.PubMedCrossRef Franc BL, Mari C, Johnson D, Leong SP. The role of a positron- and high-energy gamma photon probe in intraoperative localization of recurrent melanoma. Clin Nucl Med. 2005;30:787–91.PubMedCrossRef
29.
go back to reference Gulec SA, Daghighian F, Essner R. PET-probe: evaluation of technical performance and clinical utility of a handheld high-energy gamma probe in oncologic surgery. Ann Surg Oncol. 2006;Jul 24. Gulec SA, Daghighian F, Essner R. PET-probe: evaluation of technical performance and clinical utility of a handheld high-energy gamma probe in oncologic surgery. Ann Surg Oncol. 2006;Jul 24.
30.
go back to reference Daghighian F, Mazziotta JC, Hoffman EJ, Shenderov P, Eshaghian B, Siegel S, et al. Intraoperative beta probe: a device for detecting tissue labeled with positron or electron emitting isotopes during surgery. Med Phys. 1994;21:153–7.PubMedCrossRef Daghighian F, Mazziotta JC, Hoffman EJ, Shenderov P, Eshaghian B, Siegel S, et al. Intraoperative beta probe: a device for detecting tissue labeled with positron or electron emitting isotopes during surgery. Med Phys. 1994;21:153–7.PubMedCrossRef
31.
go back to reference Virnig BA, Tuttle TM, Shamliyan T, Kane RL. Ductal carcinoma in situ of the breast: a systematic review of incidence, treatment, and outcomes. J Natl Cancer Inst. 2010;102:170–8.PubMedCrossRef Virnig BA, Tuttle TM, Shamliyan T, Kane RL. Ductal carcinoma in situ of the breast: a systematic review of incidence, treatment, and outcomes. J Natl Cancer Inst. 2010;102:170–8.PubMedCrossRef
32.
go back to reference Yamamoto S, Matsumoto K, Senda M. Development of a GSO positron/single-photon imaging detector. Phys Med Biol. 2006;51:457–69.PubMedCrossRef Yamamoto S, Matsumoto K, Senda M. Development of a GSO positron/single-photon imaging detector. Phys Med Biol. 2006;51:457–69.PubMedCrossRef
33.
go back to reference Yamamoto S, Higashi T, Matsumoto K, Senda M. Development of a positron-imaging detector with background rejection capability. Ann Nucl Med. 2006;20:655–62.PubMedCrossRef Yamamoto S, Higashi T, Matsumoto K, Senda M. Development of a positron-imaging detector with background rejection capability. Ann Nucl Med. 2006;20:655–62.PubMedCrossRef
34.
go back to reference Weinberg I, Majewski S, Weisenberger A, Markowitz A, Aloj L, Majewski L, et al. Preliminary results for positron emission mammography: real-time functional breast imaging in a conventional mammography gantry. Eur J Nucl Med. 1996;23:804–6.PubMedCrossRef Weinberg I, Majewski S, Weisenberger A, Markowitz A, Aloj L, Majewski L, et al. Preliminary results for positron emission mammography: real-time functional breast imaging in a conventional mammography gantry. Eur J Nucl Med. 1996;23:804–6.PubMedCrossRef
35.
go back to reference Wendler T, Feuerstein M, Traub J, Lasser T, Vogel J, Daghighian F, et al. Real-time fusion of ultrasound and gamma probe for navigated localization of liver metastases. Med Image Comput Comput Assist Interv. 2007;10:252–60.PubMed Wendler T, Feuerstein M, Traub J, Lasser T, Vogel J, Daghighian F, et al. Real-time fusion of ultrasound and gamma probe for navigated localization of liver metastases. Med Image Comput Comput Assist Interv. 2007;10:252–60.PubMed
Metadata
Title
Performance of beta- and high-energy gamma probes for the detection of cancer tissue in experimental surgical resection beds
Authors
Rita Garcia-Parra
Neal Clinthorne
Li Wang
Maria Picchio
Morand Piert
Publication date
01-08-2011
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 7/2011
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-011-0492-0

Other articles of this Issue 7/2011

Annals of Nuclear Medicine 7/2011 Go to the issue