Skip to main content
Top
Published in: Journal of Interventional Cardiac Electrophysiology 2/2021

01-08-2021 | Pacemaker

Comparison between leadless and transvenous single-chamber pacemaker therapy in a referral centre for lead extraction

Authors: Giulio Zucchelli, Silvio Tolve, Valentina Barletta, Andrea Di Cori, Matteo Parollo, Raffaele De Lucia, Veronica Della Tommasina, Mario Giannotti Santoro, Stefano Viani, Tea Cellamaro, Luca Segreti, Luca Paperini, Ezio Soldati, Maria Grazia Bongiorni

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2021

Login to get access

Abstract

Purpose

The aim of the study was to compare the long-term clinical and electrical performance of Micra leadless pacemaker with transvenous single-chamber pacemaker (TV-VVI PM) in a high-volume centre for transvenous lead extraction (TLE).

Methods

One-hundred patients (group 1) undergoing Micra implant were matched with 100 patients undergoing TV-VVI PM implant (group 2) by age, sex, left ventricular systolic ejection fraction and previous TLE.

Results

The implant procedure was successful in all patients. In group 1, the procedure duration was lower than in group 2 (43.86 ± 22.38 vs 58.38 ± 17.85 min, p < 0.001), while the fluoroscopy time was longer (12.25 ± 6.84 vs 5.32 ± 4.42 min, p < 0.001). There was no difference about the rate of septal implant at the right ventricle (76% vs 86%, p = 0.10). Patients were followed-up for a median of 12 months. No acute and chronic procedure-related complication was observed in group 1, while we reported acute complications in seven patients (7%, p = 0.02) and long-term complications in three patients (3%, p = 0.24), needing for a system revision in 6 cases (6%, p = 0.038), in group 2. One systemic infection occurred in TV-VVI PM group. Electrical measurements were stable during follow-up in both groups, with a longer estimated battery life in group 1 (mean delivered energy at threshold at discharge: 0.14 ± 0.21 vs 0.26 ± 0.22 μJ, p < 0.001).

Conclusion

Micra pacemaker implant is a safe and effective procedure, with a lower rate of acute complications and system revisions compared with TV-VVI PM, even in a real-life setting including patients who underwent TLE.
Literature
1.
go back to reference Mulpuru SK, Madhavan M, McLeod CJ, Cha YM, Friedman PA. Cardiac pacemakers: function, troubleshooting, and management: part 1 of a 2-part series. J Am Coll Cardiol. 2017;69(2):189–210.CrossRef Mulpuru SK, Madhavan M, McLeod CJ, Cha YM, Friedman PA. Cardiac pacemakers: function, troubleshooting, and management: part 1 of a 2-part series. J Am Coll Cardiol. 2017;69(2):189–210.CrossRef
2.
go back to reference Udo EO, Zuithoff NP, van Hemel NM, de Cock CC, Hendriks T, Doevendans PA, et al. Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm. 2012;9(5):728–35.CrossRef Udo EO, Zuithoff NP, van Hemel NM, de Cock CC, Hendriks T, Doevendans PA, et al. Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm. 2012;9(5):728–35.CrossRef
3.
go back to reference Kirkfeldt RE, Johansen JB, Nohr EA, Jorgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35(18):1186–94.CrossRef Kirkfeldt RE, Johansen JB, Nohr EA, Jorgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35(18):1186–94.CrossRef
4.
go back to reference Lee JZ, Mulpuru SK, Shen WK. Leadless pacemaker: performance and complications. Trends Cardiovasc Med. 2018;28(2):130–41.CrossRef Lee JZ, Mulpuru SK, Shen WK. Leadless pacemaker: performance and complications. Trends Cardiovasc Med. 2018;28(2):130–41.CrossRef
5.
go back to reference Kancharla K, Deshmukh AJ, Friedman PA. Leadless pacemakers - implant, explant and long-term safety and efficacy data. J Atr Fibrillation. 2017;10(2):1581.CrossRef Kancharla K, Deshmukh AJ, Friedman PA. Leadless pacemakers - implant, explant and long-term safety and efficacy data. J Atr Fibrillation. 2017;10(2):1581.CrossRef
6.
go back to reference Sperzel J, Burri H, Gras D, Tjong FV, Knops RE, Hindricks G, et al. State of the art of leadless pacing. Europace. 2015;17(10):1508–13.CrossRef Sperzel J, Burri H, Gras D, Tjong FV, Knops RE, Hindricks G, et al. State of the art of leadless pacing. Europace. 2015;17(10):1508–13.CrossRef
7.
go back to reference Merkel M, Grotherr P, Radzewitz A, Schmitt C. Leadless pacing: current state and future direction. Cardiol Ther. 2017;6(2):175–81.CrossRef Merkel M, Grotherr P, Radzewitz A, Schmitt C. Leadless pacing: current state and future direction. Cardiol Ther. 2017;6(2):175–81.CrossRef
8.
go back to reference Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016;374(6):533–41.CrossRef Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016;374(6):533–41.CrossRef
9.
go back to reference Duray GZ, Ritter P, El-Chami M, Narasimhan C, Omar R, Tolosana JM, et al. Long-term performance of a transcatheter pacing system: 12-month results from the Micra transcatheter pacing study. Heart Rhythm. 2017;14(5):702–9.CrossRef Duray GZ, Ritter P, El-Chami M, Narasimhan C, Omar R, Tolosana JM, et al. Long-term performance of a transcatheter pacing system: 12-month results from the Micra transcatheter pacing study. Heart Rhythm. 2017;14(5):702–9.CrossRef
10.
go back to reference Roberts PR, Clementy N, Al Samadi F, Garweg C, Martinez-Sande JL, Iacopino S, et al. A leadless pacemaker in the real-world setting: the Micra transcatheter pacing system post-approval registry. Heart Rhythm. 2017;14(9):1375–9.CrossRef Roberts PR, Clementy N, Al Samadi F, Garweg C, Martinez-Sande JL, Iacopino S, et al. A leadless pacemaker in the real-world setting: the Micra transcatheter pacing system post-approval registry. Heart Rhythm. 2017;14(9):1375–9.CrossRef
11.
go back to reference El-Chami MF, Al-Samadi F, Clementy N, Garweg C, Martinez-Sande JL, Piccini JP, et al. Updated performance of the Micra transcatheter pacemaker in the real-world setting: a comparison to the investigational study and a transvenous historical control. Heart Rhythm. 2018;15(12):1800–7.CrossRef El-Chami MF, Al-Samadi F, Clementy N, Garweg C, Martinez-Sande JL, Piccini JP, et al. Updated performance of the Micra transcatheter pacemaker in the real-world setting: a comparison to the investigational study and a transvenous historical control. Heart Rhythm. 2018;15(12):1800–7.CrossRef
12.
go back to reference Bongiorni MG, Della Tommasina V, Barletta V, Di Cori A, Rogani S, Viani S, et al. Feasibility and long-term effectiveness of a non-apical Micra pacemaker implantation in a referral centre for lead extraction. Europace. 2019;21(1):114–20.CrossRef Bongiorni MG, Della Tommasina V, Barletta V, Di Cori A, Rogani S, Viani S, et al. Feasibility and long-term effectiveness of a non-apical Micra pacemaker implantation in a referral centre for lead extraction. Europace. 2019;21(1):114–20.CrossRef
13.
go back to reference El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: results from the Micra postapproval registry. J Cardiovasc Electrophysiol. 2019;30(4):569–74.CrossRef El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: results from the Micra postapproval registry. J Cardiovasc Electrophysiol. 2019;30(4):569–74.CrossRef
14.
go back to reference Zucchelli G, Barletta V, Della Tommasina V, Viani S, Parollo M, Mazzocchetti L, et al. Micra pacemaker implant after cardiac implantable electronic device extraction: feasibility and long-term outcomes. Europace. 2019;21:1229–36.CrossRef Zucchelli G, Barletta V, Della Tommasina V, Viani S, Parollo M, Mazzocchetti L, et al. Micra pacemaker implant after cardiac implantable electronic device extraction: feasibility and long-term outcomes. Europace. 2019;21:1229–36.CrossRef
15.
go back to reference Tjong FV, Reddy VY. Permanent leadless cardiac pacemaker therapy: a comprehensive review. Circulation. 2017;135(15):1458–70.CrossRef Tjong FV, Reddy VY. Permanent leadless cardiac pacemaker therapy: a comprehensive review. Circulation. 2017;135(15):1458–70.CrossRef
16.
go back to reference Vaidya VR, Dai M, Asirvatham SJ, Rea RF, Thome TM, Srivathsan K, et al. Real-world experience with leadless cardiac pacing. Pacing Clin Electrophysiol. 2019;42(3):366–73.CrossRef Vaidya VR, Dai M, Asirvatham SJ, Rea RF, Thome TM, Srivathsan K, et al. Real-world experience with leadless cardiac pacing. Pacing Clin Electrophysiol. 2019;42(3):366–73.CrossRef
17.
go back to reference Bongiorni MG, Soldati E, Zucchelli G, Di Cori A, Segreti L, De Lucia R, et al. Transvenous removal of pacing and implantable cardiac defibrillating leads using single sheath mechanical dilatation and multiple venous approaches: high success rate and safety in more than 2000 leads. Eur Heart J. 2008;29(23):2886–93.CrossRef Bongiorni MG, Soldati E, Zucchelli G, Di Cori A, Segreti L, De Lucia R, et al. Transvenous removal of pacing and implantable cardiac defibrillating leads using single sheath mechanical dilatation and multiple venous approaches: high success rate and safety in more than 2000 leads. Eur Heart J. 2008;29(23):2886–93.CrossRef
18.
go back to reference Zucchelli G, Coluccia G, Bongiorni MG. How to temporarily pace a pacemaker-dependent patient after lead extraction for device infection? Europace. 2014;16(11):1697.CrossRef Zucchelli G, Coluccia G, Bongiorni MG. How to temporarily pace a pacemaker-dependent patient after lead extraction for device infection? Europace. 2014;16(11):1697.CrossRef
19.
go back to reference Piccini JP, Stromberg K, Jackson KP, Laager V, Duray GZ, El-Chami M, et al. Long-term outcomes in leadless Micra transcatheter pacemakers with elevated thresholds at implantation: results from the Micra transcatheter pacing system global clinical trial. Heart Rhythm. 2017;14(5):685–91.CrossRef Piccini JP, Stromberg K, Jackson KP, Laager V, Duray GZ, El-Chami M, et al. Long-term outcomes in leadless Micra transcatheter pacemakers with elevated thresholds at implantation: results from the Micra transcatheter pacing system global clinical trial. Heart Rhythm. 2017;14(5):685–91.CrossRef
20.
go back to reference Zucchelli G, Favilli E, Viani S, Barletta V, Di Cori A, Segreti L, et al. Leadless pacing in a patient with superior vena cava syndrome undergoing lead extraction and percutaneous angioplasty. J Cardiol Cases. 2018;17(6):212–4.CrossRef Zucchelli G, Favilli E, Viani S, Barletta V, Di Cori A, Segreti L, et al. Leadless pacing in a patient with superior vena cava syndrome undergoing lead extraction and percutaneous angioplasty. J Cardiol Cases. 2018;17(6):212–4.CrossRef
21.
go back to reference El-Chami MF, Clementy N, Garweg C, Omar R, Duray GZ, Gornick CC, et al. Leadless pacemaker implantation in hemodialysis patients: experience with the Micra transcatheter pacemaker. JACC Clin Electrophysiol. 2019;5(2):162–70.CrossRef El-Chami MF, Clementy N, Garweg C, Omar R, Duray GZ, Gornick CC, et al. Leadless pacemaker implantation in hemodialysis patients: experience with the Micra transcatheter pacemaker. JACC Clin Electrophysiol. 2019;5(2):162–70.CrossRef
22.
go back to reference Tjong FVY, Knops RE, Udo EO, Brouwer TF, Dukkipati SR, Koruth JS, et al. Leadless pacemaker versus transvenous single-chamber pacemaker therapy: a propensity matched analysis. Heart Rhythm. 2018;15(9):1387–93.CrossRef Tjong FVY, Knops RE, Udo EO, Brouwer TF, Dukkipati SR, Koruth JS, et al. Leadless pacemaker versus transvenous single-chamber pacemaker therapy: a propensity matched analysis. Heart Rhythm. 2018;15(9):1387–93.CrossRef
23.
go back to reference Garweg C, Vandenberk B, Foulon S, Haemers P, Ector J, Willems R. Leadless pacing with Micra TPS: a comparison between right ventricular outflow tract, mid-septal, and apical implant sites. J Cardiovasc Electrophysiol. 2019;30:2002–11.CrossRef Garweg C, Vandenberk B, Foulon S, Haemers P, Ector J, Willems R. Leadless pacing with Micra TPS: a comparison between right ventricular outflow tract, mid-septal, and apical implant sites. J Cardiovasc Electrophysiol. 2019;30:2002–11.CrossRef
Metadata
Title
Comparison between leadless and transvenous single-chamber pacemaker therapy in a referral centre for lead extraction
Authors
Giulio Zucchelli
Silvio Tolve
Valentina Barletta
Andrea Di Cori
Matteo Parollo
Raffaele De Lucia
Veronica Della Tommasina
Mario Giannotti Santoro
Stefano Viani
Tea Cellamaro
Luca Segreti
Luca Paperini
Ezio Soldati
Maria Grazia Bongiorni
Publication date
01-08-2021
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2021
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-020-00832-9

Other articles of this Issue 2/2021

Journal of Interventional Cardiac Electrophysiology 2/2021 Go to the issue