Published in: BMC Ophthalmology 1/2015

Open Access 01-12-2015 | Research article

Optical and material analysis of opacified hydrophilic intraocular lenses after explantation: a laboratory study

Authors: Tamer Tandogan, Ramin Khoramnia, Chul Young Choi, Alexander Scheuerle, Martin Wenzel, Philipp Hugger, Gerd U. Auffarth

Published in: BMC Ophthalmology | Issue 1/2015

Login to get access

Abstract

Background

The opacification of hydrophilic intraocular lenses (IOLs) is a very rare complication in terms of absolute numbers. We report on the analyses of opacified Euromaxx ALI313Y and ALI313 IOLs (Argonoptics, Germany) using light and scanning electron microscopy, X-ray spectroscopy and optical bench analysis.

Methods

Opacified Euromaxx ALI313Y and ALI313 IOLs were explanted after patients presented with a decrease in visual acuity. The explants were sent to our laboratory and examined using light and scanning electron microscopy. The composition of the deposits was analysed using X-ray spectroscopy. The optical quality of the intraocular lens (IOL) was assessed using the OptiSpheric IOL PRO optical bench (Trioptics GmbH Wedel, Germany). Modulation transfer function (MTF) was measured at all spatial frequencies and United States Air Force (USAF) 1951 resolution target pictures were documented.

Results

Macroscopically, the entire optic was opacified in all IOLs. Light and scanning electron microscopy revealed numerous fine, granular, crystalline-like deposits, which were always distributed in a line parallel to the anterior and posterior surfaces of the IOLs. X-ray spectroscopy could prove the deposits consisted of Calcium and Phosphate. Measurements in the optical bench showed deterioration of MTF values at all spatial frequencies and the USAF target pictures demonstrated a significant reduction of brightness as well as resolution with the opacified IOLs.

Conclusions

The calcification of hydrophilic IOLs only occurs rarely. The exact chemical composition of the deposits can be assessed by means of X-ray spectroscopy. Optical quality analysis of the explanted Euromaxx ALI313Y and ALI313 IOLs showed significant reduction of MTF values, which was confirmed by USAF target pictures.
Literature
1.
go back to reference Khoramnia R, Salgado JP, Auffarth GU, Schmidt S, Wegner A, Kobuch KA, et al. [Opacification of a hydrophilic intraocular lens 4 years after cataract surgery. A biomaterial analysis]. Ophthalmologe. 2012;109(5):483–6.CrossRefPubMed Khoramnia R, Salgado JP, Auffarth GU, Schmidt S, Wegner A, Kobuch KA, et al. [Opacification of a hydrophilic intraocular lens 4 years after cataract surgery. A biomaterial analysis]. Ophthalmologe. 2012;109(5):483–6.CrossRefPubMed
2.
go back to reference Fung SS, Sykakis E, Islam NM, Zambarakji HJ, Khoramnia R, Auffarth GU, et al. Intraocular Lens Opacification following Intracameral Injection of Recombinant Tissue Plasminogen Activator to Treat Inflammatory Membranes after Cataract Surgery. J Ophthalmol. 2015;2015:975075.PubMedCentralCrossRefPubMed Fung SS, Sykakis E, Islam NM, Zambarakji HJ, Khoramnia R, Auffarth GU, et al. Intraocular Lens Opacification following Intracameral Injection of Recombinant Tissue Plasminogen Activator to Treat Inflammatory Membranes after Cataract Surgery. J Ophthalmol. 2015;2015:975075.PubMedCentralCrossRefPubMed
3.
go back to reference Schmidinger G, Pemp B, Werner L. [Opacification of an intraocular lens: calcification of hydrophilic intraocular lenses after gas tamponade of the anterior chamber]. Ophthalmologe. 2013;110(11):1066–8.CrossRefPubMed Schmidinger G, Pemp B, Werner L. [Opacification of an intraocular lens: calcification of hydrophilic intraocular lenses after gas tamponade of the anterior chamber]. Ophthalmologe. 2013;110(11):1066–8.CrossRefPubMed
4.
go back to reference Izak AM, Werner L, Pandey SK, Apple DJ. Calcification of modern foldable hydrogel intraocular lens designs. Eye. 2003;17(3):393–406.CrossRefPubMed Izak AM, Werner L, Pandey SK, Apple DJ. Calcification of modern foldable hydrogel intraocular lens designs. Eye. 2003;17(3):393–406.CrossRefPubMed
5.
go back to reference Werner L, Hunter B, Stevens S, Chew JJ, Mamalis N. Role of silicon contamination on calcification of hydrophilic acrylic intraocular lenses. Am J Ophthalmol. 2006;141(1):35–43.CrossRefPubMed Werner L, Hunter B, Stevens S, Chew JJ, Mamalis N. Role of silicon contamination on calcification of hydrophilic acrylic intraocular lenses. Am J Ophthalmol. 2006;141(1):35–43.CrossRefPubMed
6.
go back to reference Dagres E, Khan MA, Kyle GM, Clark D. Perioperative complications of intraocular lens exchange in patients with opacified Aqua-Sense lenses. J Cataract Refract Surg. 2004;30(12):2569–73.CrossRefPubMed Dagres E, Khan MA, Kyle GM, Clark D. Perioperative complications of intraocular lens exchange in patients with opacified Aqua-Sense lenses. J Cataract Refract Surg. 2004;30(12):2569–73.CrossRefPubMed
7.
go back to reference Madrid-Costa D, Ruiz-Alcocer J, Ferrer-Blasco T, Garcia-Lazaro S, Montes-Mico R. Optical quality differences between three multifocal intraocular lenses: bifocal low add, bifocal moderate add, and trifocal. J Refract Surg. 2013;29(11):749–54.CrossRefPubMed Madrid-Costa D, Ruiz-Alcocer J, Ferrer-Blasco T, Garcia-Lazaro S, Montes-Mico R. Optical quality differences between three multifocal intraocular lenses: bifocal low add, bifocal moderate add, and trifocal. J Refract Surg. 2013;29(11):749–54.CrossRefPubMed
8.
go back to reference Gatinel D, Houbrechts Y. Comparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench. J Cataract Refract Surg. 2013;39(7):1093–9.CrossRefPubMed Gatinel D, Houbrechts Y. Comparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench. J Cataract Refract Surg. 2013;39(7):1093–9.CrossRefPubMed
9.
go back to reference Neuhann I, Lege B, Sommer B, Neuhann T. [Conspicuous intraocular lens]. Ophthalmologe. 2008;105(12):1154–6.CrossRefPubMed Neuhann I, Lege B, Sommer B, Neuhann T. [Conspicuous intraocular lens]. Ophthalmologe. 2008;105(12):1154–6.CrossRefPubMed
10.
go back to reference Neuhann IM, Neuhann TF, Kleinmann G, Apple DJ. Clinical presentation of intraocular lens schisis. J Cataract Refract Surg. 2006;32(9):1503–8.CrossRefPubMed Neuhann IM, Neuhann TF, Kleinmann G, Apple DJ. Clinical presentation of intraocular lens schisis. J Cataract Refract Surg. 2006;32(9):1503–8.CrossRefPubMed
11.
go back to reference Werner L. Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg. 2010;36(8):1398–420.CrossRefPubMed Werner L. Glistenings and surface light scattering in intraocular lenses. J Cataract Refract Surg. 2010;36(8):1398–420.CrossRefPubMed
12.
go back to reference Moreno-Montanes J, Alvarez A, Rodriguez-Conde R, Fernandez-Hortelano A. Clinical factors related to the frequency and intensity of glistenings in AcrySof intraocular lenses. J Cataract Refract Surg. 2003;29(10):1980–4.CrossRefPubMed Moreno-Montanes J, Alvarez A, Rodriguez-Conde R, Fernandez-Hortelano A. Clinical factors related to the frequency and intensity of glistenings in AcrySof intraocular lenses. J Cataract Refract Surg. 2003;29(10):1980–4.CrossRefPubMed
13.
go back to reference Dhaliwal DK, Mamalis N, Olson RJ, Crandall AS, Zimmerman P, Alldredge OC, et al. Visual significance of glistenings seen in the AcrySof intraocular lens. J Cataract Refract Surg. 1996;22(4):452–7.CrossRefPubMed Dhaliwal DK, Mamalis N, Olson RJ, Crandall AS, Zimmerman P, Alldredge OC, et al. Visual significance of glistenings seen in the AcrySof intraocular lens. J Cataract Refract Surg. 1996;22(4):452–7.CrossRefPubMed
14.
go back to reference Allers A, Baumeister M, Steinkamp GW, Ohrloff C, Kohnen T. Intra-individual comparison of intraocular lenses of highly refractive silicone (Allergan SI40NB) and hydrophobic acrylate (Alcon Acrysof MA60BM). 1-year follow-up. Ophthalmologe. 2000;97(10):669–75.CrossRefPubMed Allers A, Baumeister M, Steinkamp GW, Ohrloff C, Kohnen T. Intra-individual comparison of intraocular lenses of highly refractive silicone (Allergan SI40NB) and hydrophobic acrylate (Alcon Acrysof MA60BM). 1-year follow-up. Ophthalmologe. 2000;97(10):669–75.CrossRefPubMed
Metadata
Title
Optical and material analysis of opacified hydrophilic intraocular lenses after explantation: a laboratory study
Authors
Tamer Tandogan
Ramin Khoramnia
Chul Young Choi
Alexander Scheuerle
Martin Wenzel
Philipp Hugger
Gerd U. Auffarth
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2015
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-015-0149-1

Other articles of this Issue 1/2015

BMC Ophthalmology 1/2015 Go to the issue