01-12-2017 | Cornea

Ocular anatomic changes for different accommodative demands using swept-source optical coherence tomography: a pilot study

Authors: José J. Esteve-Taboada, Teresa Ferrer-Blasco, Miguel A. Aloy, José E. Adsuara, Pablo Cerdá-Durán, Petar Mimica, Robert Montés-Micó

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 12/2017

Login to get access

Abstract

Purpose

The purpose of our study was to assess the changes in ocular parameters for different accommodative demands using a new optical biometer based on swept-source optical coherence tomography (SS-OCT).

Methods

Seven subjects were included in this pilot study, and only one eye per participant was analyzed. Each eye was measured six times with the optical biometer IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany). As this instrument is not able to change the vergence of the stimulus, to enable measurements at different accommodative states, a tilted first-surface mirror attached to the optical biometer was used to place the fixation stimulus at different vergences. Measurements were taken on the right eye of the subject while the left eye was looking through the mirror. Central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), axial length (AL), white-to-white (WTW) distance and keratometric readings were evaluated for three different accommodative states: 0.0 diopters (D), 1.5 D, and 3.0 D.

Results

No statistically significant differences were found for CCT, AL, WTW, K1 and K2 between the three accommodative states. As expected, changing the accommodative condition did not change CCT, AL, WTW, and keratometric outcomes. Nevertheless, statistically significant differences between the accommodative states were found for ACD and LT measurements. In addition, variations in ACD correlated linearly with variations in LT (R2 ≥ 0.99) when changing the vergence of the optotype.

Conclusion

A practical methodology to assess the changes in ocular parameters for different accommodative demands using the IOLMaster 700 based on SS-OCT has been described. Statistically significant changes that have been found that agree well with previous reports.
Appendix
Available only for authorised users
Literature
1.
go back to reference Montés-Micó R, Carones F, Buttacchio A, Ferrer-Blasco T, Madrid-Costa D (2011) Comparison of immersion ultrasound, partial coherence interferometry, and low coherence reflectometry for ocular biometry in cataract patients. J Refract Surg 27:665–671CrossRefPubMed Montés-Micó R, Carones F, Buttacchio A, Ferrer-Blasco T, Madrid-Costa D (2011) Comparison of immersion ultrasound, partial coherence interferometry, and low coherence reflectometry for ocular biometry in cataract patients. J Refract Surg 27:665–671CrossRefPubMed
2.
go back to reference Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S (2015) Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg 41:2224–2232CrossRefPubMed Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S (2015) Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg 41:2224–2232CrossRefPubMed
3.
go back to reference Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schütze J, Büehren T (2016) Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography–based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg 42:76–83CrossRefPubMed Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schütze J, Büehren T (2016) Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography–based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg 42:76–83CrossRefPubMed
4.
go back to reference Akman A, Asena L, Güngör SG (2016) Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol 100:1201–1205 Akman A, Asena L, Güngör SG (2016) Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol 100:1201–1205
5.
go back to reference Lehman BM, Berntsen DA, Bailey MD, Zadnik K (2009) Validation of OCT-based crystalline lens thickness measurements in children. Optometry and vision science: official publication of the American Academy of Optometry 86:181CrossRef Lehman BM, Berntsen DA, Bailey MD, Zadnik K (2009) Validation of OCT-based crystalline lens thickness measurements in children. Optometry and vision science: official publication of the American Academy of Optometry 86:181CrossRef
6.
go back to reference Baikoff G, Lutun E, Ferraz C, Wei J (2004) Static and dynamic analysis of the anterior segment with optical coherence tomography. J Cataract Refract Surg 30:1843–1850CrossRefPubMed Baikoff G, Lutun E, Ferraz C, Wei J (2004) Static and dynamic analysis of the anterior segment with optical coherence tomography. J Cataract Refract Surg 30:1843–1850CrossRefPubMed
7.
go back to reference Tsorbatzoglou A, Nemeth G, Szell N, Biro Z, Berta A (2007) Anterior segment changes with age and during accommodation measured with partial coherence interferometry. J Cataract Refract Surg 33:1597–1601CrossRefPubMed Tsorbatzoglou A, Nemeth G, Szell N, Biro Z, Berta A (2007) Anterior segment changes with age and during accommodation measured with partial coherence interferometry. J Cataract Refract Surg 33:1597–1601CrossRefPubMed
8.
go back to reference Leng L, Yuan Y, Chen Q, Shen M, Ma Q, Lin B et al (2014) Biometry of anterior segment of human eye on both horizontal and vertical meridians during accommodation imaged with extended scan depth optical coherence tomography. PLoS One 9:e104775CrossRefPubMedPubMedCentral Leng L, Yuan Y, Chen Q, Shen M, Ma Q, Lin B et al (2014) Biometry of anterior segment of human eye on both horizontal and vertical meridians during accommodation imaged with extended scan depth optical coherence tomography. PLoS One 9:e104775CrossRefPubMedPubMedCentral
9.
go back to reference Richdale K, Bullimore MA, Zadnik K (2008) Lens thickness with age and accommodation by optical coherence tomography. Ophthalmic Physiol Opt 28:441–447CrossRefPubMedPubMedCentral Richdale K, Bullimore MA, Zadnik K (2008) Lens thickness with age and accommodation by optical coherence tomography. Ophthalmic Physiol Opt 28:441–447CrossRefPubMedPubMedCentral
10.
go back to reference McAlinden C, Khadka J, Pesudovs K (2011) Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthal Physiol Opt 31:330–338CrossRef McAlinden C, Khadka J, Pesudovs K (2011) Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthal Physiol Opt 31:330–338CrossRef
11.
go back to reference Montés-Micó R, Alió J, Muñoz G, Charman W (2004) Temporal changes in optical quality of air-tear film interface at anterior cornea after blink. Invest Ophthalmol Vis Sci 45:1752–1757CrossRefPubMed Montés-Micó R, Alió J, Muñoz G, Charman W (2004) Temporal changes in optical quality of air-tear film interface at anterior cornea after blink. Invest Ophthalmol Vis Sci 45:1752–1757CrossRefPubMed
12.
go back to reference Tan WY (1982) Sampling distributions and robustness of t, F and variance-ratio in two samples and ANOVA models with respect to departure from normality. Communication in statistics – Theory and Methods 11:486–511CrossRef Tan WY (1982) Sampling distributions and robustness of t, F and variance-ratio in two samples and ANOVA models with respect to departure from normality. Communication in statistics – Theory and Methods 11:486–511CrossRef
13.
go back to reference Box GEP (1965) Some theorems on quadratic forms applied in the study of analysis of variance problems. II: effects of inequality of variance and of correlation between errors in the two-way classification. Ann Math Stat 25:484–498CrossRef Box GEP (1965) Some theorems on quadratic forms applied in the study of analysis of variance problems. II: effects of inequality of variance and of correlation between errors in the two-way classification. Ann Math Stat 25:484–498CrossRef
14.
15.
go back to reference Neri A, Ruggeri M, Protti A, Leaci R, Gandolfi SA, Macaluso C (2015) Dynamic imaging of accommodation by swept-source anterior segment optical coherence tomography. J Cataract Refract Surg 41:501–510CrossRefPubMedPubMedCentral Neri A, Ruggeri M, Protti A, Leaci R, Gandolfi SA, Macaluso C (2015) Dynamic imaging of accommodation by swept-source anterior segment optical coherence tomography. J Cataract Refract Surg 41:501–510CrossRefPubMedPubMedCentral
16.
go back to reference Piñero DP (2015) Technologies for anatomical and geometric characterization of the corneal structure and anterior segment: a review. Semin Ophthalmol 30:161–170CrossRefPubMed Piñero DP (2015) Technologies for anatomical and geometric characterization of the corneal structure and anterior segment: a review. Semin Ophthalmol 30:161–170CrossRefPubMed
17.
go back to reference Ramos JL, Li Y, Huang D (2009) Clinical and research applications of anterior segment optical coherence tomography—a review. Clin Experiment Ophthalmol 37:81–89CrossRefPubMed Ramos JL, Li Y, Huang D (2009) Clinical and research applications of anterior segment optical coherence tomography—a review. Clin Experiment Ophthalmol 37:81–89CrossRefPubMed
18.
go back to reference Bizheva K, Haines L, Mason E, MacLellan B, Tan B, Hileeto D, Sorbara L (2016) In vivo imaging and morphometry of the human pre-Descemet's layer and endothelium with ultrahigh-resolution optical coherence tomography. Invest Ophthalmol Vis Sci 57(1):2782–2787CrossRefPubMed Bizheva K, Haines L, Mason E, MacLellan B, Tan B, Hileeto D, Sorbara L (2016) In vivo imaging and morphometry of the human pre-Descemet's layer and endothelium with ultrahigh-resolution optical coherence tomography. Invest Ophthalmol Vis Sci 57(1):2782–2787CrossRefPubMed
19.
go back to reference Sisó-Fuertes I, Domínguez-Vicent A, del Águila-Carrasco A, Ferrer-Blasco T, Montés-Micó R (2015) Corneal changes with accommodation using dual Scheimpflug photography. J Cataract Refract Surg 41:981–989CrossRefPubMed Sisó-Fuertes I, Domínguez-Vicent A, del Águila-Carrasco A, Ferrer-Blasco T, Montés-Micó R (2015) Corneal changes with accommodation using dual Scheimpflug photography. J Cataract Refract Surg 41:981–989CrossRefPubMed
20.
go back to reference Bayramlar H, Sadigov F, Yildirim A (2013) Effect of accommodation on corneal topography. Cornea 32:1251–1254CrossRefPubMed Bayramlar H, Sadigov F, Yildirim A (2013) Effect of accommodation on corneal topography. Cornea 32:1251–1254CrossRefPubMed
21.
go back to reference Woodman EC, Read SA, Collins MJ (2012) Axial length and choroidal thickness changes accompanying prolonged accommodation in myopes and emmetropes. Vis Res 72(1):34–41CrossRefPubMed Woodman EC, Read SA, Collins MJ (2012) Axial length and choroidal thickness changes accompanying prolonged accommodation in myopes and emmetropes. Vis Res 72(1):34–41CrossRefPubMed
22.
go back to reference Zhong J, Tao A, Xu Z, Jiang H, Shao Y, Zhang H, Liu C, Wang J (2014) Whole eye axial biometry during accommodation using ultra-long scan depth optical coherence tomography. Am J Ophthalmol 157:1064–1069CrossRefPubMedPubMedCentral Zhong J, Tao A, Xu Z, Jiang H, Shao Y, Zhang H, Liu C, Wang J (2014) Whole eye axial biometry during accommodation using ultra-long scan depth optical coherence tomography. Am J Ophthalmol 157:1064–1069CrossRefPubMedPubMedCentral
23.
go back to reference Ostrin L, Kasthurirangan S, Win-Hall D, Glasser A (2006) Simultaneous measurements of refraction and A-scan biometry during accommodation in humans. Optom Vis Sci 83:657–665CrossRefPubMed Ostrin L, Kasthurirangan S, Win-Hall D, Glasser A (2006) Simultaneous measurements of refraction and A-scan biometry during accommodation in humans. Optom Vis Sci 83:657–665CrossRefPubMed
24.
go back to reference Bolz M, Prinz A, Drexler W, Findl O (2007) Linear relationship of refractive and biometric lenticular changes during accommodation in emmetropic and myopic eyes. Br J Ophthalmol 91:360–365CrossRefPubMed Bolz M, Prinz A, Drexler W, Findl O (2007) Linear relationship of refractive and biometric lenticular changes during accommodation in emmetropic and myopic eyes. Br J Ophthalmol 91:360–365CrossRefPubMed
25.
go back to reference Laughton DS, Sheppard AL, Davies LN (2016) A longitudinal study of accommodative changes in biometry during incipient presbyopia. Ophthalmic Physiol Opt 36:33–42CrossRefPubMed Laughton DS, Sheppard AL, Davies LN (2016) A longitudinal study of accommodative changes in biometry during incipient presbyopia. Ophthalmic Physiol Opt 36:33–42CrossRefPubMed
26.
go back to reference Ramasubramanian V, Glasser A (2015) Objective measurement of accommodative biometric changes using ultrasound biomicroscopy. J Cataract Refract Surg 41:511–526CrossRefPubMedPubMedCentral Ramasubramanian V, Glasser A (2015) Objective measurement of accommodative biometric changes using ultrasound biomicroscopy. J Cataract Refract Surg 41:511–526CrossRefPubMedPubMedCentral
27.
go back to reference Ramasubramanian V, Glasser A (2015) Prediction of accommodative optical response in prepresbyopic subjects using ultrasound biomicroscopy. J Cataract Refract Surg 41:964–980CrossRefPubMedPubMedCentral Ramasubramanian V, Glasser A (2015) Prediction of accommodative optical response in prepresbyopic subjects using ultrasound biomicroscopy. J Cataract Refract Surg 41:964–980CrossRefPubMedPubMedCentral
28.
go back to reference Farouk MM, Naito T, Shinomiya K, Eguchi H, Sayed KM, Nagasawa T, Katome T, Mitamura Y (2015) Optical coherence tomography reveals new insights into the accommodation mechanism. J Ophthalmol 2015:510459CrossRefPubMedPubMedCentral Farouk MM, Naito T, Shinomiya K, Eguchi H, Sayed KM, Nagasawa T, Katome T, Mitamura Y (2015) Optical coherence tomography reveals new insights into the accommodation mechanism. J Ophthalmol 2015:510459CrossRefPubMedPubMedCentral
29.
go back to reference Aloy MA, Adsuara JE, Cerdá-Durán P, Obergaulinger M, Esteve-Taboada JJ, Ferrer-Blasco T, Montés-Micó R (2017) PLOS One (accepted for publication) Aloy MA, Adsuara JE, Cerdá-Durán P, Obergaulinger M, Esteve-Taboada JJ, Ferrer-Blasco T, Montés-Micó R (2017) PLOS One (accepted for publication)
Metadata
Title
Ocular anatomic changes for different accommodative demands using swept-source optical coherence tomography: a pilot study
Authors
José J. Esteve-Taboada
Teresa Ferrer-Blasco
Miguel A. Aloy
José E. Adsuara
Pablo Cerdá-Durán
Petar Mimica
Robert Montés-Micó
Publication date
01-12-2017
Publisher
Springer Berlin Heidelberg
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 12/2017
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-017-3801-z

Other articles of this Issue 12/2017

Graefe's Archive for Clinical and Experimental Ophthalmology 12/2017 Go to the issue