Published in: European Radiology 6/2017

Open Access 01-06-2017 | Chest

Low-tube-voltage selection for triple-rule-out CTA: relation to patient size

Authors: Richard A. P. Takx, Radko Krissak, Christian Fink, Valentin Bachmann, Thomas Henzler, Mathias Meyer, John W. Nance, Stefan O. Schoenberg, Paul Apfaltrer

Published in: European Radiology | Issue 6/2017

Login to get access

Abstract

Objectives

To investigate the relationship between image quality and patient size at 100 kilovoltage (kV) compared to 120 kV ECG-gated Triple-Rule-Out CT angiography (TRO-CTA).

Methods

We retrospectively included 73 patients (age 64 ± 14 years) who underwent retrospective ECG-gated chest CTA. 40 patients were scanned with 100 kV while 33 patients with 120 kV. Body mass index (BMI), patients’ chest circumference (PC) and thoracic surface area (TSA) were recorded. Quantitative image quality was assessed as vascular attenuation in the ascending aorta (AA), pulmonary trunk (PA) and left coronary artery (LCA) and the signal-to-noise ratio (SNR) in the AA.

Results

There was no significant difference in BMI (26.0 ± 4.6 vs. 28.0 ± 6.7 kg/m2), PC (103 ± 7 vs. 104 ± 10 cm2) and TSA (92 ± 15 vs. 91 ± 19 cm2) between 100 kV and 120 kV group. Mean vascular attenuation was significantly higher in the 100 kV compared to the 120 kV group (AA 438 vs. 354 HU, PA 460 vs. 349 HU, LCA 370 vs. 299 HU all p < 0.001). SNR was not significantly different, even after adjusting for patient size. Radiation dose was significantly lower in the 100 kV group (10.7 ± 4.1 vs. 20.7 ± 10.7 mSv; p < 0.001).

Conclusions

100 kV TRO-CTA is feasible in normal-to-overweight patients while maintaining image quality and achieving substantial dose reduction.

Key Points

100 kV protocols result in a significantly lower radiation dose.
Mean vascular attenuation is significantly higher using 100 kV.
SNR and CNR are not significantly different between 100 kV and 120 kV.
100 kV CTA is feasible regardless of patient size while maintaining image quality.
Literature
1.
go back to reference Halpern EJ (2009) Triple-rule-out CT angiography for evaluation of acute chest pain and possible acute coronary syndrome. Radiology 252:332–345CrossRefPubMed Halpern EJ (2009) Triple-rule-out CT angiography for evaluation of acute chest pain and possible acute coronary syndrome. Radiology 252:332–345CrossRefPubMed
2.
go back to reference Frauenfelder T, Appenzeller P, Karlo C et al (2009) Triple rule-out CT in the emergency department: protocols and spectrum of imaging findings. Eur Radiol 19:789–799CrossRefPubMed Frauenfelder T, Appenzeller P, Karlo C et al (2009) Triple rule-out CT in the emergency department: protocols and spectrum of imaging findings. Eur Radiol 19:789–799CrossRefPubMed
3.
go back to reference Christiaens L, Duchat F, Boudiaf M et al (2012) Impact of 64-slice coronary CT on the management of patients presenting with acute chest pain: results of a prospective two-centre study. Eur Radiol 22:1050–1058CrossRefPubMed Christiaens L, Duchat F, Boudiaf M et al (2012) Impact of 64-slice coronary CT on the management of patients presenting with acute chest pain: results of a prospective two-centre study. Eur Radiol 22:1050–1058CrossRefPubMed
4.
go back to reference Schertler T, Scheffel H, Frauenfelder T et al (2007) Dual-source computed tomography in patients with acute chest pain: feasibility and image quality. Eur Radiol 17:3179–3188CrossRefPubMedPubMedCentral Schertler T, Scheffel H, Frauenfelder T et al (2007) Dual-source computed tomography in patients with acute chest pain: feasibility and image quality. Eur Radiol 17:3179–3188CrossRefPubMedPubMedCentral
5.
go back to reference Zhang LJ, Lu GM, Meinel FG, McQuiston AD, Ravenel JG, Schoepf UJ (2015) Computed tomography of acute pulmonary embolism: state-of-the-art. Eur Radiol 25:2547–2557CrossRefPubMed Zhang LJ, Lu GM, Meinel FG, McQuiston AD, Ravenel JG, Schoepf UJ (2015) Computed tomography of acute pulmonary embolism: state-of-the-art. Eur Radiol 25:2547–2557CrossRefPubMed
6.
go back to reference Durmus T, Rogalla P, Lembcke A, Muhler MR, Hamm B, Hein PA (2011) Low-dose triple-rule-out using 320-row-detector volume MDCT--less contrast medium and lower radiation exposure. Eur Radiol 21:1416–1423CrossRefPubMed Durmus T, Rogalla P, Lembcke A, Muhler MR, Hamm B, Hein PA (2011) Low-dose triple-rule-out using 320-row-detector volume MDCT--less contrast medium and lower radiation exposure. Eur Radiol 21:1416–1423CrossRefPubMed
7.
go back to reference Rist C, Johnson TR, Becker CR, Reiser MF, Nikolaou K (2007) New applications for noninvasive cardiac imaging: dual-source computed tomography. Eur Radiol 17:F16–F25CrossRefPubMed Rist C, Johnson TR, Becker CR, Reiser MF, Nikolaou K (2007) New applications for noninvasive cardiac imaging: dual-source computed tomography. Eur Radiol 17:F16–F25CrossRefPubMed
8.
go back to reference Wnorowski AM, Halpern EJ (2016) Diagnostic yield of triple-rule-Out CT in an emergency setting. AJR Am J Roentgenol 207:295–301CrossRefPubMed Wnorowski AM, Halpern EJ (2016) Diagnostic yield of triple-rule-Out CT in an emergency setting. AJR Am J Roentgenol 207:295–301CrossRefPubMed
9.
go back to reference Kligerman SJ, White CS (2014) Image quality and feasibility of an ultralow-dose high-pitch helical triple-rule-out computed tomography angiography acquired in the caudocranial direction. J Thorac Imaging 29:50–59CrossRefPubMed Kligerman SJ, White CS (2014) Image quality and feasibility of an ultralow-dose high-pitch helical triple-rule-out computed tomography angiography acquired in the caudocranial direction. J Thorac Imaging 29:50–59CrossRefPubMed
10.
go back to reference Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323CrossRefPubMed Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323CrossRefPubMed
11.
go back to reference Bischoff B, Hein F, Meyer T et al (2009) Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study. JACC Cardiovasc Imaging 2:940–946CrossRefPubMed Bischoff B, Hein F, Meyer T et al (2009) Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study. JACC Cardiovasc Imaging 2:940–946CrossRefPubMed
12.
go back to reference Feuchtner GM, Jodocy D, Klauser A et al (2010) Radiation dose reduction by using 100-kV tube voltage in cardiac 64-slice computed tomography: a comparative study. Eur J Radiol 75:e51–e56CrossRefPubMed Feuchtner GM, Jodocy D, Klauser A et al (2010) Radiation dose reduction by using 100-kV tube voltage in cardiac 64-slice computed tomography: a comparative study. Eur J Radiol 75:e51–e56CrossRefPubMed
13.
go back to reference Fink C, Krissak R, Henzler T et al (2011) Radiation dose at coronary CT angiography: second-generation dual-source CT versus single-source 64-MDCT and first-generation dual-source CT. AJR Am J Roentgenol 196:W550–W557CrossRefPubMed Fink C, Krissak R, Henzler T et al (2011) Radiation dose at coronary CT angiography: second-generation dual-source CT versus single-source 64-MDCT and first-generation dual-source CT. AJR Am J Roentgenol 196:W550–W557CrossRefPubMed
14.
go back to reference Leschka S, Stolzmann P, Schmid FT et al (2008) Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose. Eur Radiol 18:1809–1817CrossRefPubMed Leschka S, Stolzmann P, Schmid FT et al (2008) Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose. Eur Radiol 18:1809–1817CrossRefPubMed
15.
go back to reference Ghafourian K, Younes D, Simprini LA, Weigold WG, Weissman G, Taylor AJ (2012) Scout view X-ray attenuation versus weight-based selection of reduced peak tube voltage in cardiac CT angiography. JACC Cardiovasc Imaging 5:589–595CrossRefPubMed Ghafourian K, Younes D, Simprini LA, Weigold WG, Weissman G, Taylor AJ (2012) Scout view X-ray attenuation versus weight-based selection of reduced peak tube voltage in cardiac CT angiography. JACC Cardiovasc Imaging 5:589–595CrossRefPubMed
16.
go back to reference Ghoshhajra BB, Engel LC, Major GP et al (2011) Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr 5:240–246CrossRefPubMed Ghoshhajra BB, Engel LC, Major GP et al (2011) Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr 5:240–246CrossRefPubMed
17.
go back to reference Commission E (1999) European guidelines on quality criteria for computed tomography. Brussels, Belgium Report EUR 16262 EN Commission E (1999) European guidelines on quality criteria for computed tomography. Brussels, Belgium Report EUR 16262 EN
18.
go back to reference Kalva SP, Sahani DV, Hahn PF, Saini S (2006) Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study. J Comput Assist Tomogr 30:391–397CrossRefPubMed Kalva SP, Sahani DV, Hahn PF, Saini S (2006) Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study. J Comput Assist Tomogr 30:391–397CrossRefPubMed
19.
go back to reference Kidoh M, Nakaura T, Nakamura S et al (2014) Contrast material and radiation dose reduction strategy for triple-rule-out cardiac CT angiography: feasibility study of non-ECG-gated low kVp scan of the whole chest following coronary CT angiography. Acta Radiol 55:1186–1196CrossRefPubMed Kidoh M, Nakaura T, Nakamura S et al (2014) Contrast material and radiation dose reduction strategy for triple-rule-out cardiac CT angiography: feasibility study of non-ECG-gated low kVp scan of the whole chest following coronary CT angiography. Acta Radiol 55:1186–1196CrossRefPubMed
20.
go back to reference Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435CrossRefPubMed Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435CrossRefPubMed
21.
go back to reference Sigal-Cinqualbre AB, Hennequin R, Abada HT, Chen X, Paul JF (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231:169–174CrossRefPubMed Sigal-Cinqualbre AB, Hennequin R, Abada HT, Chen X, Paul JF (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231:169–174CrossRefPubMed
22.
23.
go back to reference Li JL, Huang MP, Liang CH et al (2012) Individualized radiation dose control in 256-slice CT coronary angiography (CTCA) in retrospective ECG-triggered helical scans: Using a measure of body size to adjust tube current selection. Eur J Radiol. doi:10.1016/j.ejrad.2012.04.017 PubMedCentral Li JL, Huang MP, Liang CH et al (2012) Individualized radiation dose control in 256-slice CT coronary angiography (CTCA) in retrospective ECG-triggered helical scans: Using a measure of body size to adjust tube current selection. Eur J Radiol. doi:10.​1016/​j.​ejrad.​2012.​04.​017 PubMedCentral
24.
go back to reference Moscariello A, Takx RA, Schoepf UJ et al (2011) Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection. Eur Radiol 21:2130–2138CrossRefPubMed Moscariello A, Takx RA, Schoepf UJ et al (2011) Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection. Eur Radiol 21:2130–2138CrossRefPubMed
25.
go back to reference Lu GM, Luo S, Meinel FG et al (2014) High-pitch computed tomography pulmonary angiography with iterative reconstruction at 80 kVp and 20 mL contrast agent volume. Eur Radiol 24:3260–3268CrossRefPubMed Lu GM, Luo S, Meinel FG et al (2014) High-pitch computed tomography pulmonary angiography with iterative reconstruction at 80 kVp and 20 mL contrast agent volume. Eur Radiol 24:3260–3268CrossRefPubMed
Metadata
Title
Low-tube-voltage selection for triple-rule-out CTA: relation to patient size
Authors
Richard A. P. Takx
Radko Krissak
Christian Fink
Valentin Bachmann
Thomas Henzler
Mathias Meyer
John W. Nance
Stefan O. Schoenberg
Paul Apfaltrer
Publication date
01-06-2017
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 6/2017
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-016-4607-9

Other articles of this Issue 6/2017

European Radiology 6/2017 Go to the issue