Published in: International Orthopaedics 2/2007

01-04-2007 | Original Paper

Long-term results of uncemented acetabular cups with an ACS polyethylene liner. A 14–16-year follow-up study

Authors: E. Garcia-Rey, E. Garcia-Cimbrelo

Published in: International Orthopaedics | Issue 2/2007

Login to get access

Abstract

We assessed the clinical and radiographic results of 40 porous-coated acetabular cups with an Acetabular Cup System polyethylene liner over a minimum 14-year follow-up. Femoral head penetration was estimated using a software package. Fifteen cups were revised, 11 due to polyethylene liner rupture. All cups but two were radiographically stable, and 11 hips showed acetabular osteolysis. The overall femoral head penetration rate in hips without liner fracture with reference to the early penetration point was 0.1188±0.070 mm per year. Polyethylene liner fractures were associated with higher early femoral head penetration (P<0.0001) and a vertical cup position (P=0.0016). The 14-year survival without cup revision for any reason was 63.9%, 71.8% with no ACS polyethylene liner fracture and 65.3% with no acetabular osteolysis. Most cups showed a good clinical outcome in general, but major Acetabular Cup System liner failure and osteolysis were frequent. Patients with the ACS cups still in place should be monitored closely.
Literature
1.
go back to reference Astion DJ, Saluan P, Stulberg BN, Rimnac CM, Li S (1996) The porous-coated anatomic total hip prosthesis: Failure of the metal-backed acetabular component. J Bone Joint Surg Am 78:755–766PubMed Astion DJ, Saluan P, Stulberg BN, Rimnac CM, Li S (1996) The porous-coated anatomic total hip prosthesis: Failure of the metal-backed acetabular component. J Bone Joint Surg Am 78:755–766PubMed
2.
go back to reference Bartel DL, Bicknell VL, Wright TM (1986) The effect of conformity, thickness and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am 68:1041–1051PubMed Bartel DL, Bicknell VL, Wright TM (1986) The effect of conformity, thickness and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am 68:1041–1051PubMed
3.
go back to reference Berry DJ, Barnes CL, Scott RD, Cabanela ME, Poss R (1994) Catastrophic failure of the polyethyhlene liner of uncemented acetabular components. J Bone Joint Surg Br 76:575–578PubMed Berry DJ, Barnes CL, Scott RD, Cabanela ME, Poss R (1994) Catastrophic failure of the polyethyhlene liner of uncemented acetabular components. J Bone Joint Surg Br 76:575–578PubMed
4.
go back to reference Bono JV, Sanford L, Toussaint JT (1994) Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML Plus hip implants with an ACS polyethylene liner. J Arthroplasty 9:119–125PubMedCrossRef Bono JV, Sanford L, Toussaint JT (1994) Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML Plus hip implants with an ACS polyethylene liner. J Arthroplasty 9:119–125PubMedCrossRef
5.
go back to reference Cruz-Pardos A, García-Cimbrelo E (2001) The Harris-Galante total hip arthroplasty. A minimum 8-year follow-up. J Arthroplasty 16:586–597PubMedCrossRef Cruz-Pardos A, García-Cimbrelo E (2001) The Harris-Galante total hip arthroplasty. A minimum 8-year follow-up. J Arthroplasty 16:586–597PubMedCrossRef
6.
go back to reference DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop 121:20–32PubMed DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop 121:20–32PubMed
7.
go back to reference Devane PA, Bourne RB, Rorabeck CH, Hardie RM, Horne JG (1995) Measurement of polyethylene wear in metal backed cups: Three dimensional technique. Clin Orthop 319:303–316PubMed Devane PA, Bourne RB, Rorabeck CH, Hardie RM, Horne JG (1995) Measurement of polyethylene wear in metal backed cups: Three dimensional technique. Clin Orthop 319:303–316PubMed
8.
go back to reference Devane PA, Horne JG, Martin K, Coldham G, Krause B (1997) Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris. J Arthroplasty 12:256–266PubMedCrossRef Devane PA, Horne JG, Martin K, Coldham G, Krause B (1997) Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris. J Arthroplasty 12:256–266PubMedCrossRef
9.
go back to reference Dowd JE, Sychterz CJ, Young AM, Engh CA (2000) Characterization of long-term femoral-head penetration rates. Association with and prediction of osteolysis. J Bone Joint Surg Am 82:1102–1107PubMed Dowd JE, Sychterz CJ, Young AM, Engh CA (2000) Characterization of long-term femoral-head penetration rates. Association with and prediction of osteolysis. J Bone Joint Surg Am 82:1102–1107PubMed
10.
go back to reference Goetz DD, Smith JJ, Harris WH (1994) The prevalence of femoral osteolysis associated with components inserted with or without cement in total hip replacements. J Bone Joint Surg Am 76:1121–1128PubMed Goetz DD, Smith JJ, Harris WH (1994) The prevalence of femoral osteolysis associated with components inserted with or without cement in total hip replacements. J Bone Joint Surg Am 76:1121–1128PubMed
11.
go back to reference Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components. Clin Orthop 171:17–27 Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components. Clin Orthop 171:17–27
12.
go back to reference Hamadouche M, Witvoet J, Porcher R, Meunier A, Sedel L, Nizard R (2001) Hydroxyapatite-coated versus grit blasted femoral stems. A prospective, randomised study using EBRA-FCA. J Bone Joint Surg Br 83:979–987PubMedCrossRef Hamadouche M, Witvoet J, Porcher R, Meunier A, Sedel L, Nizard R (2001) Hydroxyapatite-coated versus grit blasted femoral stems. A prospective, randomised study using EBRA-FCA. J Bone Joint Surg Br 83:979–987PubMedCrossRef
13.
go back to reference Harris WH (1995) The problem is osteolysis. Clin Orthop 311:46–53PubMed Harris WH (1995) The problem is osteolysis. Clin Orthop 311:46–53PubMed
14.
go back to reference Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH (2003) Validation of two- and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am 85:505–511PubMed Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH (2003) Validation of two- and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am 85:505–511PubMed
15.
go back to reference Johnston RC, Fitzgerald RH, Harris WH, Poss R, Müller ME, Sledge CB (1990) Clinical and radiographic evaluation of total hip arthroplasty. A standard system of terminology for reporting results. J Bone Joint Surg Am 72:161–168PubMed Johnston RC, Fitzgerald RH, Harris WH, Poss R, Müller ME, Sledge CB (1990) Clinical and radiographic evaluation of total hip arthroplasty. A standard system of terminology for reporting results. J Bone Joint Surg Am 72:161–168PubMed
16.
go back to reference Kaplan EL, Meier P (1958) Nonparametric estimation from incomplete observations. J Am Statistc Assn 53:457–481CrossRef Kaplan EL, Meier P (1958) Nonparametric estimation from incomplete observations. J Am Statistc Assn 53:457–481CrossRef
17.
go back to reference Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH (2005) In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty 20:880–886PubMedCrossRef Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH (2005) In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty 20:880–886PubMedCrossRef
18.
go back to reference Martell JM, Berdia S (1997) Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am 79:1635–1641PubMed Martell JM, Berdia S (1997) Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am 79:1635–1641PubMed
19.
go back to reference Massin P, Schmidt L, Engh CA (1989) Evaluation of cementless acetabular component migration. An experimental study. J Arthroplasty 4:245–251PubMedCrossRef Massin P, Schmidt L, Engh CA (1989) Evaluation of cementless acetabular component migration. An experimental study. J Arthroplasty 4:245–251PubMedCrossRef
20.
go back to reference Merle D’Aubigné R, Postel M (1954) Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am 36:451–475 Merle D’Aubigné R, Postel M (1954) Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am 36:451–475
21.
go back to reference Orishimo KF, Claus AM, Sychterz CJ, Engh CA (2003) Relationship between polyethylene wear and osteolysis in hips with a second-generation porous-costed cementless cup after seven years of follow-up. J Bone Joint Surg Am 85:1095–1099PubMed Orishimo KF, Claus AM, Sychterz CJ, Engh CA (2003) Relationship between polyethylene wear and osteolysis in hips with a second-generation porous-costed cementless cup after seven years of follow-up. J Bone Joint Surg Am 85:1095–1099PubMed
22.
go back to reference Patil S, Bergula A, Chen PC, Colwell CW Jr, D’Lima DD (2003) Polyethylene wear acetabular component orientation. J Bone Joint Surg Am 85(Suppl 4):56–63PubMed Patil S, Bergula A, Chen PC, Colwell CW Jr, D’Lima DD (2003) Polyethylene wear acetabular component orientation. J Bone Joint Surg Am 85(Suppl 4):56–63PubMed
23.
go back to reference Sychterz CJ, Engh CA Jr, Shah N, Engh CA Sr (1997). Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J Bone Joint Surg Am 79A:1040–1046 Sychterz CJ, Engh CA Jr, Shah N, Engh CA Sr (1997). Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J Bone Joint Surg Am 79A:1040–1046
24.
go back to reference Sychterz CJ, Yang AM, McAuley JP, Engh CA (1999) Two-dimensional versus three-dimensional radiographic measurements of polyethylene wear. Clin Orthop 365:117–123PubMedCrossRef Sychterz CJ, Yang AM, McAuley JP, Engh CA (1999) Two-dimensional versus three-dimensional radiographic measurements of polyethylene wear. Clin Orthop 365:117–123PubMedCrossRef
25.
go back to reference Sychterz CJ, Young AM, Engh CA (2001) Effect of radiographic quality on computer-assisted head penetration measurements. Clin Orthop 386:150–158PubMedCrossRef Sychterz CJ, Young AM, Engh CA (2001) Effect of radiographic quality on computer-assisted head penetration measurements. Clin Orthop 386:150–158PubMedCrossRef
26.
go back to reference Thanner J, Kärrholm J, Herberts P, Malchau H (1999) Poor outcome of the PCA and Harris-Galante hip prostheses. Randomized study of 171 arthroplasties with 9-year follow-up. Acta Orthop Scand 70:155–162PubMedCrossRef Thanner J, Kärrholm J, Herberts P, Malchau H (1999) Poor outcome of the PCA and Harris-Galante hip prostheses. Randomized study of 171 arthroplasties with 9-year follow-up. Acta Orthop Scand 70:155–162PubMedCrossRef
Metadata
Title
Long-term results of uncemented acetabular cups with an ACS polyethylene liner. A 14–16-year follow-up study
Authors
E. Garcia-Rey
E. Garcia-Cimbrelo
Publication date
01-04-2007
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 2/2007
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-006-0151-6

Other articles of this Issue 2/2007

International Orthopaedics 2/2007 Go to the issue