Open Access 01-12-2019 | Original research

Head-to-head comparison of image quality between brain 18F-FDG images recorded with a fully digital versus a last-generation analog PET camera

Authors: Julien Salvadori, Laetitia Imbert, Mathieu Perrin, Gilles Karcher, Zohra Lamiral, Pierre-Yves Marie, Antoine Verger

Published in: EJNMMI Research | Issue 1/2019

Login to get access

Abstract

Background

The quality of phantom images was previously shown to be higher on digital (Vereos Philips®) compared to analog PET (Ingenuity Philips®) cameras. This study aimed to determine the extent to which this difference still remains significant on normal brain 18F-FDG PET images.

Methods

Relative noise and contrast as well as border sharpness (a spatial resolution index) of central (striata) and peripheral (occiput) gray-matter structures were compared between 10 sets of normal brain 18F-FDG PET images recorded and reconstructed on digital and analog last-generation PET cameras, together with a subjective visual analysis of image quality provided by experienced physicians.

Results

Compared with analog PET, digital PET provided marked improvements in image quality parameters. The median relative noise was decreased (− 22%), while gray/white-matter contrast was increased (+ 27%/+ 41% for central/peripheral gray-matter structures), with these results being consistent with visual analysis. In addition, a clear enhancement in image sharpness was further documented for digital PET owing to the possible use of a 1-mm3 voxel size (+ 24%/+ 21%).

Conclusions

On normal brain 18F-FDG images and compared with a last-generation analog PET, the fully digital PET camera offers marked improvements in image noise and contrast, as well as significant potential for further enhancing spatial resolution.
Appendix
Available only for authorised users
Literature
2.
go back to reference Budinger TF. Time-of-flight positron emission tomography: status relative to conventional PET. J Nucl Med. 1983;24:73–8.PubMed Budinger TF. Time-of-flight positron emission tomography: status relative to conventional PET. J Nucl Med. 1983;24:73–8.PubMed
3.
go back to reference NationalElectricalManufacturersAssociation. NEMA NU-2 2012: performance measurement of positron emission tomography. 2013. NationalElectricalManufacturersAssociation. NEMA NU-2 2012: performance measurement of positron emission tomography. 2013.
4.
go back to reference Rausch I, Ruiz A, Valverde-Pascual I, Cal-Gonzalez J, Beyer T, Ignasi C. Performance evaluation of the Philips Vereos PET/CT system according to the NEMA NU2–2012 standards. J Nucl Med. 2018;60:561–7.CrossRef Rausch I, Ruiz A, Valverde-Pascual I, Cal-Gonzalez J, Beyer T, Ignasi C. Performance evaluation of the Philips Vereos PET/CT system according to the NEMA NU2–2012 standards. J Nucl Med. 2018;60:561–7.CrossRef
15.
go back to reference Salvadori J, Verger A, Chevalier E, Perrin M, Tissot H, Claudin M, et al. Head-to-head comparisons of the PET performance measured using the NEMA standards between digital and analogic last-generation cameras. J Nucl Med. 2018;59(Suppl 1):574. Abstract. Salvadori J, Verger A, Chevalier E, Perrin M, Tissot H, Claudin M, et al. Head-to-head comparisons of the PET performance measured using the NEMA standards between digital and analogic last-generation cameras. J Nucl Med. 2018;59(Suppl 1):574. Abstract.
Metadata
Title
Head-to-head comparison of image quality between brain 18F-FDG images recorded with a fully digital versus a last-generation analog PET camera
Authors
Julien Salvadori
Laetitia Imbert
Mathieu Perrin
Gilles Karcher
Zohra Lamiral
Pierre-Yves Marie
Antoine Verger
Publication date
01-12-2019
Publisher
Springer Berlin Heidelberg
Published in
EJNMMI Research / Issue 1/2019
Electronic ISSN: 2191-219X
DOI
https://doi.org/10.1186/s13550-019-0526-5

Other articles of this Issue 1/2019

EJNMMI Research 1/2019 Go to the issue