Published in: The Ultrasound Journal 1/2017

Open Access 01-12-2017 | Original article

Correlation of carotid blood flow and corrected carotid flow time with invasive cardiac output measurements

Authors: Irene W. Y. Ma, Joshua D. Caplin, Aftab Azad, Christina Wilson, Michael A. Fifer, Aranya Bagchi, Andrew S. Liteplo, Vicki E. Noble

Published in: The Ultrasound Journal | Issue 1/2017

Login to get access

Abstract

Background

Non-invasive measures that can accurately estimate cardiac output may help identify volume-responsive patients. This study seeks to compare two non-invasive measures (corrected carotid flow time and carotid blood flow) and their correlations with invasive reference measurements of cardiac output. Consenting adult patients (n = 51) at Massachusetts General Hospital cardiac catheterization laboratory undergoing right heart catheterization between February and April 2016 were included. Carotid ultrasound images were obtained concurrently with cardiac output measurements, obtained by the thermodilution method in the absence of severe tricuspid regurgitation and by the Fick oxygen method otherwise. Corrected carotid flow time was calculated as systole time/√cycle time. Carotid blood flow was calculated as π × (carotid diameter)2/4 × velocity time integral × heart rate. Measurements were obtained using a single carotid waveform and an average of three carotid waveforms for both measures.

Results

Single waveform measurements of corrected flow time did not correlate with cardiac output (ρ = 0.25, 95% CI −0.03 to 0.49, p = 0.08), but an average of three waveforms correlated significantly, although weakly (ρ = 0.29, 95% CI 0.02–0.53, p = 0.046). Carotid blood flow measurements correlated moderately with cardiac output regardless of if single waveform or an average of three waveforms were used: ρ = 0.44, 95% CI 0.18–0.63, p = 0.004, and ρ = 0.41, 95% CI 0.16–0.62, p = 0.004, respectively.

Conclusions

Carotid blood flow may be a better marker of cardiac output and less subject to measurements issues than corrected carotid flow time.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kelm DJ, Perrin JT, Cartin-Ceba R, Gajic O, Schenck L, Kennedy CC (2015) Fluid overload in patients with severe sepsis and septic shock treated with early-goal directed therapy is associated with increased acute need for fluid-related medical interventions and hospital death. Shock 43:68–73CrossRefPubMedPubMedCentral Kelm DJ, Perrin JT, Cartin-Ceba R, Gajic O, Schenck L, Kennedy CC (2015) Fluid overload in patients with severe sepsis and septic shock treated with early-goal directed therapy is associated with increased acute need for fluid-related medical interventions and hospital death. Shock 43:68–73CrossRefPubMedPubMedCentral
2.
go back to reference Boyd JH, Forbes J, Nakada TA, Walley KR, Russell JA (2011) Fluid resuscitation in septic shock: a positive fluid balance and elevated central venous pressure are associated with increased mortality. Crit Care Med 39:259–265CrossRefPubMed Boyd JH, Forbes J, Nakada TA, Walley KR, Russell JA (2011) Fluid resuscitation in septic shock: a positive fluid balance and elevated central venous pressure are associated with increased mortality. Crit Care Med 39:259–265CrossRefPubMed
4.
go back to reference Marik PE, Cavallazzi R (2013) Does the central venous pressure predict fluid responsiveness? An updated meta-analysis and a plea for some common sense. Crit Care Med 41:1774–1781CrossRefPubMed Marik PE, Cavallazzi R (2013) Does the central venous pressure predict fluid responsiveness? An updated meta-analysis and a plea for some common sense. Crit Care Med 41:1774–1781CrossRefPubMed
5.
go back to reference Bentzer P, Griesdale DE, Boyd J, MacLean K, Sirounis D, Ayas NT (2016) WIll this hemodynamically unstable patient respond to a bolus of intravenous fluids? JAMA 316:1298–1309CrossRefPubMed Bentzer P, Griesdale DE, Boyd J, MacLean K, Sirounis D, Ayas NT (2016) WIll this hemodynamically unstable patient respond to a bolus of intravenous fluids? JAMA 316:1298–1309CrossRefPubMed
6.
go back to reference Bendjelid K, Romand JA (2012) Fluid responsiveness in mechanically ventilated patients: a review of indices used in intensive care. Intensive Care Med 29:352–360CrossRef Bendjelid K, Romand JA (2012) Fluid responsiveness in mechanically ventilated patients: a review of indices used in intensive care. Intensive Care Med 29:352–360CrossRef
7.
go back to reference Cavallaro F, Sandroni C, Marano C, La Torre G, Mannocci A, De Waure C, Bello G, Maviglia R, Antonelli M (2012) Diagnostic accuracy of passive leg raising for prediction of fluid responsiveness in adults: systematic review and meta-analysis of clinical studies. Intensive Care Med 36(9):1476–1483 Cavallaro F, Sandroni C, Marano C, La Torre G, Mannocci A, De Waure C, Bello G, Maviglia R, Antonelli M (2012) Diagnostic accuracy of passive leg raising for prediction of fluid responsiveness in adults: systematic review and meta-analysis of clinical studies. Intensive Care Med 36(9):1476–1483
9.
go back to reference Pugsley J, Lerner AB (2010) Cardiac output monitoring: is there a gold standard and how do the newer technologies compare? Semin Cardiothorac Vasc Anesth 14:274–282CrossRefPubMed Pugsley J, Lerner AB (2010) Cardiac output monitoring: is there a gold standard and how do the newer technologies compare? Semin Cardiothorac Vasc Anesth 14:274–282CrossRefPubMed
10.
go back to reference Dee Boyd K, Thomas SJ, Gold J, Boyd AD (1983) A prospective study of complications of pulmonary artery catheterizations in 500 consecutive patients. Chest 84:245–249CrossRef Dee Boyd K, Thomas SJ, Gold J, Boyd AD (1983) A prospective study of complications of pulmonary artery catheterizations in 500 consecutive patients. Chest 84:245–249CrossRef
11.
go back to reference Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K (2005) Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet 366:472–477CrossRefPubMed Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K (2005) Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet 366:472–477CrossRefPubMed
12.
go back to reference Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M (2003) A Randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 348:5–14CrossRefPubMed Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M (2003) A Randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 348:5–14CrossRefPubMed
13.
go back to reference Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D, Boulain T, Lefort Y, Fartoukh M, Baud F (2003) Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA 290:2713–2720CrossRefPubMed Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D, Boulain T, Lefort Y, Fartoukh M, Baud F (2003) Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA 290:2713–2720CrossRefPubMed
14.
go back to reference Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G, Califf RM (2005) Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA 294:1664–1670CrossRefPubMed Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G, Califf RM (2005) Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA 294:1664–1670CrossRefPubMed
15.
go back to reference Wheeler A, Bernard G, Thompson B, Schoenfeld D, Wiedemann H, DeBoisblanc B, Connors A Jr, Hite R, Harabin A (2006) Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury. N Engl J Med 354:2213–2224CrossRefPubMed Wheeler A, Bernard G, Thompson B, Schoenfeld D, Wiedemann H, DeBoisblanc B, Connors A Jr, Hite R, Harabin A (2006) Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury. N Engl J Med 354:2213–2224CrossRefPubMed
16.
go back to reference Benomar B, Ouattara A, Estagnasie P, Brusset A, Squara P (2010) Fluid responsiveness predicted by noninvasive bioreactance-based passive leg raise test. Intensive Care Med 36:1875–1881CrossRefPubMed Benomar B, Ouattara A, Estagnasie P, Brusset A, Squara P (2010) Fluid responsiveness predicted by noninvasive bioreactance-based passive leg raise test. Intensive Care Med 36:1875–1881CrossRefPubMed
17.
18.
go back to reference Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D (2008) Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput 22:113–119CrossRefPubMed Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D (2008) Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput 22:113–119CrossRefPubMed
19.
go back to reference Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C (2007) Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med 33:1191–1194CrossRefPubMed Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C (2007) Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med 33:1191–1194CrossRefPubMed
20.
go back to reference Kupersztych-Hagege E, Teboul JL, Artigas A, Talbot A, Sabatier C, Richard C, Monnet X (2013) Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients. Br J Anaesth 111:961–966CrossRefPubMed Kupersztych-Hagege E, Teboul JL, Artigas A, Talbot A, Sabatier C, Richard C, Monnet X (2013) Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients. Br J Anaesth 111:961–966CrossRefPubMed
21.
go back to reference Gunst M, Ghaemmaghami V, Sperry J, Robinson M, O’Keeffe T, Friese R, Frankel H (2008) Accuracy of cardiac function and volume status estimates using the bedside echocardiographic assessment in trauma/critical care. J Trauma 65:509–516CrossRefPubMed Gunst M, Ghaemmaghami V, Sperry J, Robinson M, O’Keeffe T, Friese R, Frankel H (2008) Accuracy of cardiac function and volume status estimates using the bedside echocardiographic assessment in trauma/critical care. J Trauma 65:509–516CrossRefPubMed
22.
go back to reference Hossein-Nejad H, Mohammadinejad P, Lessan-Pezeshki M, Davarani SS, Banaie M (2015) Carotid artery corrected flow time measurement via bedside ultrasonography in monitoring volume status. J Crit Care 30:1199–1203CrossRefPubMed Hossein-Nejad H, Mohammadinejad P, Lessan-Pezeshki M, Davarani SS, Banaie M (2015) Carotid artery corrected flow time measurement via bedside ultrasonography in monitoring volume status. J Crit Care 30:1199–1203CrossRefPubMed
23.
go back to reference Stolz LA, Mosier JM, Gross AM, Douglas MJ, Blaivas M, Adhikari S (2015) Can emergency physicians perform common carotid Doppler flow measurements to assess volume responsiveness? W J Emerg Med 16:255–259CrossRef Stolz LA, Mosier JM, Gross AM, Douglas MJ, Blaivas M, Adhikari S (2015) Can emergency physicians perform common carotid Doppler flow measurements to assess volume responsiveness? W J Emerg Med 16:255–259CrossRef
24.
go back to reference Adedipe AA, Fly DL, Schwitz SD, Jorgenson DB, Duric H, Sayre MR, Nichol G (2015) Carotid Doppler blood flow measurement during cardiopulmonary resuscitation is feasible: a first in man study. Resuscitation 96:121–125CrossRefPubMed Adedipe AA, Fly DL, Schwitz SD, Jorgenson DB, Duric H, Sayre MR, Nichol G (2015) Carotid Doppler blood flow measurement during cardiopulmonary resuscitation is feasible: a first in man study. Resuscitation 96:121–125CrossRefPubMed
25.
go back to reference Blehar DJ, Glazier S, Gaspari RJ (2014) Correlation of corrected flow time in the carotid artery with changes in intravascular volume status. J Crit Care 29:486–488CrossRefPubMed Blehar DJ, Glazier S, Gaspari RJ (2014) Correlation of corrected flow time in the carotid artery with changes in intravascular volume status. J Crit Care 29:486–488CrossRefPubMed
26.
go back to reference Shokoohi H, Berry GW, Shahkolahi M, King J, Salimian M, Poshtmashad A, Pourmand A (2016) The diagnostic utility of sonographic carotid flow time in determining volume responsiveness. J Crit Care 38:231–235CrossRefPubMed Shokoohi H, Berry GW, Shahkolahi M, King J, Salimian M, Poshtmashad A, Pourmand A (2016) The diagnostic utility of sonographic carotid flow time in determining volume responsiveness. J Crit Care 38:231–235CrossRefPubMed
27.
go back to reference Mackenzie DC, Khan NA, Blehar D, Glazier S, Chang Y, Stowell CP, Noble VE, Liteplo AS (2015) Carotid flow time changes with volume status in acute blood loss. Ann Emerg Med 66:277–282CrossRefPubMed Mackenzie DC, Khan NA, Blehar D, Glazier S, Chang Y, Stowell CP, Noble VE, Liteplo AS (2015) Carotid flow time changes with volume status in acute blood loss. Ann Emerg Med 66:277–282CrossRefPubMed
28.
go back to reference Marik PE, Levitov A, Young A, Andrews L (2013) The use of bioreactance and carotid doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest 143:364–370CrossRefPubMed Marik PE, Levitov A, Young A, Andrews L (2013) The use of bioreactance and carotid doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest 143:364–370CrossRefPubMed
29.
go back to reference Monnet X, Rienzo M, Osman D, Anguel N, Richard C, Pinsky MR, Teboul JL (2006) Passive leg raising predicts fluid responsiveness in the critically ill. Crit Care Med 34:1402–1407CrossRefPubMed Monnet X, Rienzo M, Osman D, Anguel N, Richard C, Pinsky MR, Teboul JL (2006) Passive leg raising predicts fluid responsiveness in the critically ill. Crit Care Med 34:1402–1407CrossRefPubMed
31.
go back to reference Tola M, Yurdakul M (2006) Effect of Doppler angle in diagnosis of internal carotid artery stenosis. J Ultrasound Med 25:1187–1192CrossRefPubMed Tola M, Yurdakul M (2006) Effect of Doppler angle in diagnosis of internal carotid artery stenosis. J Ultrasound Med 25:1187–1192CrossRefPubMed
32.
go back to reference Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ (1971) A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396CrossRefPubMed Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ (1971) A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396CrossRefPubMed
33.
go back to reference Callan P, Clark AL (2016) Right heart catheterisation: indications and interpretation. Heart 102:147–157CrossRefPubMed Callan P, Clark AL (2016) Right heart catheterisation: indications and interpretation. Heart 102:147–157CrossRefPubMed
36.
go back to reference Cohen J (1988) Statistical power analysis for the behavioral sciences. Academic Press, New York Cohen J (1988) Statistical power analysis for the behavioral sciences. Academic Press, New York
37.
go back to reference Benjamini Y, Hochberg Y (1995) Controlling the false discovery rate: a practical and powerful approach to multiple testing. J R Stat Soc B Methodol 57:289–300 Benjamini Y, Hochberg Y (1995) Controlling the false discovery rate: a practical and powerful approach to multiple testing. J R Stat Soc B Methodol 57:289–300
38.
go back to reference Garcia X, Mateu L, Maynar J, Mercadal J, Ochagavía A, Ferrandiz A (2011) Estimating cardiac output. Utility in the clinical practice. Available invasive and non-invasive monitoring. Med Intensiva 35:552–561CrossRefPubMed Garcia X, Mateu L, Maynar J, Mercadal J, Ochagavía A, Ferrandiz A (2011) Estimating cardiac output. Utility in the clinical practice. Available invasive and non-invasive monitoring. Med Intensiva 35:552–561CrossRefPubMed
39.
go back to reference Nicolosi GL, Pungercic E, Cervesato E, Pavan D, Modena L, Moro E, Dall’Aglio V, Zanuttini D (1988) Feasibility and variability of six methods for the echocardiographic and Doppler determination of cardiac output. Br Heart J 59:299–303CrossRefPubMedPubMedCentral Nicolosi GL, Pungercic E, Cervesato E, Pavan D, Modena L, Moro E, Dall’Aglio V, Zanuttini D (1988) Feasibility and variability of six methods for the echocardiographic and Doppler determination of cardiac output. Br Heart J 59:299–303CrossRefPubMedPubMedCentral
40.
go back to reference Madhwal S, Yesenko S, Kim ESH, Park M, Begelman SM, Gornik HL (2014) Manifestations of cardiac disease in carotid Duplex ultrasound examination. JACC 7:200–203PubMed Madhwal S, Yesenko S, Kim ESH, Park M, Begelman SM, Gornik HL (2014) Manifestations of cardiac disease in carotid Duplex ultrasound examination. JACC 7:200–203PubMed
41.
go back to reference O’Boyle MK, Vibhakar NI, Chung J, Keen WD, Gosink BB (1996) Duplex sonography of the carotid arteries in patients with isolated aortic stenosis: imaging findings and relation to severity of stenosis. AJR Am J Roentgenol 166:197–202CrossRefPubMed O’Boyle MK, Vibhakar NI, Chung J, Keen WD, Gosink BB (1996) Duplex sonography of the carotid arteries in patients with isolated aortic stenosis: imaging findings and relation to severity of stenosis. AJR Am J Roentgenol 166:197–202CrossRefPubMed
42.
go back to reference Kallman C, Gosink B, Gardner D (1991) Carotid duplex sonography: bisferious pulse contour in patients with aortic valvular disease. AJR Am J Roentgenol 157:403–407CrossRefPubMed Kallman C, Gosink B, Gardner D (1991) Carotid duplex sonography: bisferious pulse contour in patients with aortic valvular disease. AJR Am J Roentgenol 157:403–407CrossRefPubMed
43.
go back to reference Kelsey JL, Whittemore AS, Evans AS, Thompson WD (1996) Methods in observational epidemiology. Oxford University Press Inc, New York Kelsey JL, Whittemore AS, Evans AS, Thompson WD (1996) Methods in observational epidemiology. Oxford University Press Inc, New York
44.
go back to reference Lee AJ, Cohn JH (2011) Ranasinghe JS (2011) Cardiac output assessed by invasive and minimally invasive techniques. Anesthesiol Res Pract 17:475151 Lee AJ, Cohn JH (2011) Ranasinghe JS (2011) Cardiac output assessed by invasive and minimally invasive techniques. Anesthesiol Res Pract 17:475151
45.
go back to reference Marino PL (2014) The ICU Bok, 4th edn. Lippincott Williams & Wilkins, Philadelphia Marino PL (2014) The ICU Bok, 4th edn. Lippincott Williams & Wilkins, Philadelphia
Metadata
Title
Correlation of carotid blood flow and corrected carotid flow time with invasive cardiac output measurements
Authors
Irene W. Y. Ma
Joshua D. Caplin
Aftab Azad
Christina Wilson
Michael A. Fifer
Aranya Bagchi
Andrew S. Liteplo
Vicki E. Noble
Publication date
01-12-2017
Publisher
Springer Milan
Published in
The Ultrasound Journal / Issue 1/2017
Electronic ISSN: 2524-8987
DOI
https://doi.org/10.1186/s13089-017-0065-0

Other articles of this Issue 1/2017

The Ultrasound Journal 1/2017 Go to the issue