Skip to main content
Top
Published in: Patient Safety in Surgery 1/2016

Open Access 01-12-2016 | Research

Periprosthetic joint infections in modular endoprostheses of the lower extremities: a retrospective observational study in 101 patients

Authors: Dirk Zajonz, Almut Zieme, Torsten Prietzel, Michael Moche, Solveig Tiepoldt, Andreas Roth, Christoph Josten, Georg Freiherr von Salis-Soglio, Christoph- E. Heyde, Mohamed Ghanem

Published in: Patient Safety in Surgery | Issue 1/2016

Login to get access

Abstract

Background

Modular mega-endoprosthesis systems are used to bridge very large bone defects and have become a widespread method in orthopaedic surgery for the treatment of tumours and revision arthroplasty. However, the indications for the use of modular mega-endoprostheses must be carefully considered. Implanting modular endoprostheses requires major, complication-prone surgery in which the limited salvage procedures should always be borne in mind. The management of periprosthetic infection is particularly difficult and beset with problems.
Given this, the present study was designed to gauge the significance of periprosthetic infections in connection with modular mega-implants in the lower extremities among our own patients.

Methods

Patients who had been fitted with modular endoprosthesis on a lower extremity at our department between September 1994 and December 2011 were examined retrospectively. A total of 101 patients with 114 modular prostheses were identified. Comprising 30 men (29.7 %) and 71 women (70.3 %), their average age at the time of surgery was 67 years (18–92 years).

Results

The average follow-up period was 27 months (5 months and 2 weeks to 14 years and 11 months) and the drop-out rate was about 8.8 %. Altogether, there were 19 (17.7 %) endoprosthesis infections: 3 early infections and 16 late or delayed infections. The pathogen spectrum was dominated by coagulase-negative staphylococci (36 %) and Staphylococcus aureus (16 %), including 26 % multi-resistant pathogens. Reinfection occurred in 37 % of cases of infection. Tumours were followed by significantly fewer infections than the other indications. Infections were twice as likely to occur after previous surgery.

Conclusion

In our findings modular endoprostheses (18 %) are much more susceptible to infection than primary endoprostheses (0.5–2,5 %). Infection is the most common complication alongside the dislocation of proximal femur endoprostheses. Consistent, radical surgery is essential – although even with an adequate treatment strategy, the recurrence rate is very high. Unfortunately, the functional results are frequently unsatisfactory, with amputation often being the last resort. Therefore, the indication for implantation must be carefully considered and discussed in great detail, especially in the case of multimorbid patients with previous joint infections.
Literature
1.
go back to reference Kotz R. Tumorendoprothesen bei malignen Knochentumoren. Orthopade. 1993;22:160–66.PubMed Kotz R. Tumorendoprothesen bei malignen Knochentumoren. Orthopade. 1993;22:160–66.PubMed
3.
go back to reference Moore AT, Bohlmann HR. Metal hip joint: A case report. J Bone Joint Surg Am. 1943;25:688–92. Moore AT, Bohlmann HR. Metal hip joint: A case report. J Bone Joint Surg Am. 1943;25:688–92.
5.
go back to reference Dean BJF, Matthews JJ, Price A, Stubbs D, Whitwell D, Gibbons CMLH. Modular endoprosthetic replacement for failed internal fixation of the proximal femur following trauma. Int Orthop. 2012;36:731–34.PubMedCentralCrossRefPubMed Dean BJF, Matthews JJ, Price A, Stubbs D, Whitwell D, Gibbons CMLH. Modular endoprosthetic replacement for failed internal fixation of the proximal femur following trauma. Int Orthop. 2012;36:731–34.PubMedCentralCrossRefPubMed
6.
go back to reference Fakler JK, Hase F, Böhme J, Josten C. Safety aspects in surgical treatment of pathological fractures of the proximal femur - modular endoprosthetic replacement vs. intramedullary nailing. Patient Saf Surg. 2013;7(1):37. Fakler JK, Hase F, Böhme J, Josten C. Safety aspects in surgical treatment of pathological fractures of the proximal femur - modular endoprosthetic replacement vs. intramedullary nailing. Patient Saf Surg. 2013;7(1):37.
8.
go back to reference Höll S, Schlomberg A, Gosheger G, Dieckmann R, Streitbuerger A, Schulz D, et al. Distal femur and proximal tibia replacement with megaprosthesis in revision knee arthroplasty: a limb-saving procedure. Knee Surg Sports Traumatol Arthrosc. 2012;20(12):2513-8. Höll S, Schlomberg A, Gosheger G, Dieckmann R, Streitbuerger A, Schulz D, et al. Distal femur and proximal tibia replacement with megaprosthesis in revision knee arthroplasty: a limb-saving procedure. Knee Surg Sports Traumatol Arthrosc. 2012;20(12):2513-8.
9.
go back to reference Foo LSS, Hardes J, Henrichs M, Ahrens H, Gosheger G, Streitbürger A. Surgical difficulties encountered with use of modular endoprosthesis for limb preserving salvage of failed allograft reconstruction after malignant tumor resection. J Arthroplasty. 2011;26:744–50.CrossRefPubMed Foo LSS, Hardes J, Henrichs M, Ahrens H, Gosheger G, Streitbürger A. Surgical difficulties encountered with use of modular endoprosthesis for limb preserving salvage of failed allograft reconstruction after malignant tumor resection. J Arthroplasty. 2011;26:744–50.CrossRefPubMed
10.
go back to reference von Salis-Soglio G, Ghanem M, Meinecke I, Ellenrieder M, Klinger HM, Kirchhoff C. The modular endoprosthetic system Munich-Luebeck (MML): potential applications and results in the lower extremities. Orthopade. 2010;39:960–67.CrossRef von Salis-Soglio G, Ghanem M, Meinecke I, Ellenrieder M, Klinger HM, Kirchhoff C. The modular endoprosthetic system Munich-Luebeck (MML): potential applications and results in the lower extremities. Orthopade. 2010;39:960–67.CrossRef
11.
go back to reference Ascherl R, Grundei H, Hartung I, Gradinger R. Spezialimplantate – Individualprothesen, Sonderanfertigungen. In: Gradinger R, Gollwitzer H, editors. Ossäre Integration. Berlin, Heidelberg: Springer; 2006. p. 195–206.CrossRef Ascherl R, Grundei H, Hartung I, Gradinger R. Spezialimplantate – Individualprothesen, Sonderanfertigungen. In: Gradinger R, Gollwitzer H, editors. Ossäre Integration. Berlin, Heidelberg: Springer; 2006. p. 195–206.CrossRef
12.
go back to reference Gradinger R, Gollwitzer H. Spezialimplantate – Tumorendoprothetik. In: Gradinger R, Gollwitzer H, editors. Ossäre Integration. Berlin, Heidelberg: Springer; 2006. p. 180–9.CrossRef Gradinger R, Gollwitzer H. Spezialimplantate – Tumorendoprothetik. In: Gradinger R, Gollwitzer H, editors. Ossäre Integration. Berlin, Heidelberg: Springer; 2006. p. 180–9.CrossRef
13.
go back to reference Gosheger G, Gebert C, Ahrens H, Streitbuerger A, Winkelmann W, Hardes J. Endoprosthetic reconstruction in 250 patients with sarcoma. Clin Orthop Relat Res. 2006;450:164–71.CrossRefPubMed Gosheger G, Gebert C, Ahrens H, Streitbuerger A, Winkelmann W, Hardes J. Endoprosthetic reconstruction in 250 patients with sarcoma. Clin Orthop Relat Res. 2006;450:164–71.CrossRefPubMed
15.
go back to reference Donati D, Zavatte M, Gozzi E, Giacomini S, Campanacci L, Mercuri M. Modular prosthetic replacement of the proximal femur after resection of a bone tumour. A long-term follow up. J Bone Joint Surg. 2001;83:1156–60.CrossRef Donati D, Zavatte M, Gozzi E, Giacomini S, Campanacci L, Mercuri M. Modular prosthetic replacement of the proximal femur after resection of a bone tumour. A long-term follow up. J Bone Joint Surg. 2001;83:1156–60.CrossRef
16.
go back to reference Mittermayer F, Windhager R, Dominkus M, Krepler P, Schwameis E, Sluga M, et al. Revision of the Kotz type of tumour endoprosthesis for the lower limb. J Bone Joint Surg Br. 2002;84:401–06.CrossRefPubMed Mittermayer F, Windhager R, Dominkus M, Krepler P, Schwameis E, Sluga M, et al. Revision of the Kotz type of tumour endoprosthesis for the lower limb. J Bone Joint Surg Br. 2002;84:401–06.CrossRefPubMed
17.
18.
go back to reference Henderson ER, Groundland JS, Pala E, Dennis JA, Wooten R, Cheong D, et al. Failure mode classification for tumor endoprostheses: retrospective review of five institutions and a literature review. J Bone Joint Surg Am. 2011;93:418–29.CrossRefPubMed Henderson ER, Groundland JS, Pala E, Dennis JA, Wooten R, Cheong D, et al. Failure mode classification for tumor endoprostheses: retrospective review of five institutions and a literature review. J Bone Joint Surg Am. 2011;93:418–29.CrossRefPubMed
19.
go back to reference Jeys LM, Kulkarni A, Grimer RJ, Carter SR, Tillman RM, Abudu A. Endoprosthetic reconstruction for the treatment of musculoskeletal tumors of the appendicular skeleton and pelvis. J Bone Joint Surg Am. 2008;90:1265–71.CrossRefPubMed Jeys LM, Kulkarni A, Grimer RJ, Carter SR, Tillman RM, Abudu A. Endoprosthetic reconstruction for the treatment of musculoskeletal tumors of the appendicular skeleton and pelvis. J Bone Joint Surg Am. 2008;90:1265–71.CrossRefPubMed
20.
go back to reference Ahrens H, Gosheger G, Streitbürger A, Gebert C, Hardes J. Antimikrobielle Silberbeschichtung von Tumorendoprothesen. Onkologe. 2006;12:145–51.CrossRef Ahrens H, Gosheger G, Streitbürger A, Gebert C, Hardes J. Antimikrobielle Silberbeschichtung von Tumorendoprothesen. Onkologe. 2006;12:145–51.CrossRef
21.
go back to reference Zajonz D, Wuthe L, Tiepolt S, Brandmeier P, Prietzel T, von Salis-Soglio GF, et al. Diagnostic work-up strategy for periprosthetic joint infections after total hip and knee arthroplasty: a 12-year experience on 320 consecutive cases. Patient Saf Surg. 2015;9:20.PubMedCentralCrossRefPubMed Zajonz D, Wuthe L, Tiepolt S, Brandmeier P, Prietzel T, von Salis-Soglio GF, et al. Diagnostic work-up strategy for periprosthetic joint infections after total hip and knee arthroplasty: a 12-year experience on 320 consecutive cases. Patient Saf Surg. 2015;9:20.PubMedCentralCrossRefPubMed
22.
go back to reference Tsukayama DT, Goldberg VM, Kyle R. Diagnosis and management of infection after total knee arthroplasty. J Bone Joint Surg Am. 2003;85-A Suppl 1:75–80. Tsukayama DT, Goldberg VM, Kyle R. Diagnosis and management of infection after total knee arthroplasty. J Bone Joint Surg Am. 2003;85-A Suppl 1:75–80.
23.
go back to reference Gosheger G, Hardes J, Ahrens H, Streitburger A, Buerger H, Erren M, et al. Silvercoated megaendoprostheses in a rabbit model – an analysis of the infection rate and toxicological side effects. Biomaterials. 2004;25:5547–56.CrossRefPubMed Gosheger G, Hardes J, Ahrens H, Streitburger A, Buerger H, Erren M, et al. Silvercoated megaendoprostheses in a rabbit model – an analysis of the infection rate and toxicological side effects. Biomaterials. 2004;25:5547–56.CrossRefPubMed
24.
go back to reference Hardes J, Ahrens H, Gebert C, Streitbuerger A, Buerger H, Erren M, et al. Lack of toxicological side-effects in silver-coated megaprostheses in humans. Biomaterials. 2007;28:2869–75.CrossRefPubMed Hardes J, Ahrens H, Gebert C, Streitbuerger A, Buerger H, Erren M, et al. Lack of toxicological side-effects in silver-coated megaprostheses in humans. Biomaterials. 2007;28:2869–75.CrossRefPubMed
25.
go back to reference Roberts P, Chan D, Grimer RJ, Sneath RS, Scales JT. Prosthetic replacement of the distal femur for primary bone tumours. J Bone Joint Surg Br. 1991;73:762–9.PubMed Roberts P, Chan D, Grimer RJ, Sneath RS, Scales JT. Prosthetic replacement of the distal femur for primary bone tumours. J Bone Joint Surg Br. 1991;73:762–9.PubMed
26.
go back to reference Unwin PS, Cannon SR, Grimer RJ, Kemp HB, Sneath RS, Walker PS. Aseptic loosening in cemented custom-made prosthetic replacements for bone tumours of the lower limb. J Bone Joint Surg Br. 1996;78:5–13.PubMed Unwin PS, Cannon SR, Grimer RJ, Kemp HB, Sneath RS, Walker PS. Aseptic loosening in cemented custom-made prosthetic replacements for bone tumours of the lower limb. J Bone Joint Surg Br. 1996;78:5–13.PubMed
27.
go back to reference Mascard E, Anract P, Touchene A, Pouillart P, Tomeno B. Complications from the hinged GUEPAR prosthesis after resection of knee tumor. 102 cases. Rev Chir Orthop Reparatrice Appar Mot. 1998;84:628–37.PubMed Mascard E, Anract P, Touchene A, Pouillart P, Tomeno B. Complications from the hinged GUEPAR prosthesis after resection of knee tumor. 102 cases. Rev Chir Orthop Reparatrice Appar Mot. 1998;84:628–37.PubMed
29.
go back to reference Griffin AM, Parsons JA, Davis AM, Bell RS, Wunder JS. Uncemented tumor endoprostheses at the knee: root causes of failure. Clin Orthop Relat Res. 2005;438:71–9.CrossRefPubMed Griffin AM, Parsons JA, Davis AM, Bell RS, Wunder JS. Uncemented tumor endoprostheses at the knee: root causes of failure. Clin Orthop Relat Res. 2005;438:71–9.CrossRefPubMed
30.
go back to reference Gerdesmeyer L, Töpfer A, Kircher J, Grundei H, Diehl P. The modular MML revision system in knee revision and tumor arthroplasty. Orthopade. 2006;35:975–81.CrossRefPubMed Gerdesmeyer L, Töpfer A, Kircher J, Grundei H, Diehl P. The modular MML revision system in knee revision and tumor arthroplasty. Orthopade. 2006;35:975–81.CrossRefPubMed
31.
go back to reference Hardes J, Budny T, Hauschild G, Balke M, Streitbürger A, Dieckmann R, et al. Proximal femur replacement in revision arthroplasty. Z Orthop Unfall. 2009;147:694–99.PubMed Hardes J, Budny T, Hauschild G, Balke M, Streitbürger A, Dieckmann R, et al. Proximal femur replacement in revision arthroplasty. Z Orthop Unfall. 2009;147:694–99.PubMed
32.
go back to reference Chandrasekar CR, Grimer RJ, Carter SR, Tillman RM, Abudu A, Jeys LM. Unipolar proximal femoral endoprosthetic replacement for tumour: the risk of revision in young patients. J Bone Joint Surg Br. 2009;91:401–04.CrossRefPubMed Chandrasekar CR, Grimer RJ, Carter SR, Tillman RM, Abudu A, Jeys LM. Unipolar proximal femoral endoprosthetic replacement for tumour: the risk of revision in young patients. J Bone Joint Surg Br. 2009;91:401–04.CrossRefPubMed
33.
go back to reference Kinkel S, Lehner B, Kleinhans JA, Jakubowitz E, Ewerbeck V, Heisel C. Medium to long-term results after reconstruction of bone defects at the knee with tumor endoprostheses. J Surg Oncol. 2010;101:166–69.PubMed Kinkel S, Lehner B, Kleinhans JA, Jakubowitz E, Ewerbeck V, Heisel C. Medium to long-term results after reconstruction of bone defects at the knee with tumor endoprostheses. J Surg Oncol. 2010;101:166–69.PubMed
34.
go back to reference Ruggieri P, Bosco G, Pala E, Errani C, Mercuri M. Local recurrence, survival and function after total femur resection and megaprosthetic reconstruction for bone sarcomas. J Bone Joint Surg Br. 2009;91:108–12.CrossRef Ruggieri P, Bosco G, Pala E, Errani C, Mercuri M. Local recurrence, survival and function after total femur resection and megaprosthetic reconstruction for bone sarcomas. J Bone Joint Surg Br. 2009;91:108–12.CrossRef
35.
go back to reference Winkelmann W. Reconstruction of the proximal femur with the MUTARS® system. Orthopade. 2010;39:942–48.CrossRefPubMed Winkelmann W. Reconstruction of the proximal femur with the MUTARS® system. Orthopade. 2010;39:942–48.CrossRefPubMed
36.
go back to reference Pala E, Trovarelli G, Calabrò T, Angelini A, Abati CN, Ruggieri P. Survival of modern knee tumor megaprostheses: failures, functional results, and a comparative statistical analysis. Clin Orthop Relat Res. 2015;473:820–30.CrossRef Pala E, Trovarelli G, Calabrò T, Angelini A, Abati CN, Ruggieri P. Survival of modern knee tumor megaprostheses: failures, functional results, and a comparative statistical analysis. Clin Orthop Relat Res. 2015;473:820–30.CrossRef
37.
go back to reference Capanna R, Scoccianti G, Frenos F, Vilardi A, Beltrami G, Campanacci DA. What was the survival of megaprostheses in lower limb reconstructions after tumor resections? Clin Orthop Relat Res. 2015;473:2079–87.CrossRef Capanna R, Scoccianti G, Frenos F, Vilardi A, Beltrami G, Campanacci DA. What was the survival of megaprostheses in lower limb reconstructions after tumor resections? Clin Orthop Relat Res. 2015;473:2079–87.CrossRef
38.
go back to reference Sevelda F, Schuh R, Hofstaetter JG, Schinhan M, Windhager R, Funovics PT. Total Femur Replacement After Tumor Resection: Limb Salvage Usually Achieved But Complications and Failures are Common. Clin Orthop Relat Res. 2010;468:2860–66.CrossRef Sevelda F, Schuh R, Hofstaetter JG, Schinhan M, Windhager R, Funovics PT. Total Femur Replacement After Tumor Resection: Limb Salvage Usually Achieved But Complications and Failures are Common. Clin Orthop Relat Res. 2010;468:2860–66.CrossRef
Metadata
Title
Periprosthetic joint infections in modular endoprostheses of the lower extremities: a retrospective observational study in 101 patients
Authors
Dirk Zajonz
Almut Zieme
Torsten Prietzel
Michael Moche
Solveig Tiepoldt
Andreas Roth
Christoph Josten
Georg Freiherr von Salis-Soglio
Christoph- E. Heyde
Mohamed Ghanem
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Patient Safety in Surgery / Issue 1/2016
Electronic ISSN: 1754-9493
DOI
https://doi.org/10.1186/s13037-016-0095-8

Other articles of this Issue 1/2016