Skip to main content
Top
Published in: Journal of Cardiothoracic Surgery 1/2021

Open Access 01-12-2021 | ECMO | Research article

Design optimization of bidirectional arterial perfusion cannula

Authors: Saad Abdel-Sayed, Enrico Ferrari, Philippe Abdel-Sayed, Markus Wilhelm, Ludwig-Karl von Segesser, Denis Berdajs

Published in: Journal of Cardiothoracic Surgery | Issue 1/2021

Login to get access

Abstract

Objectives

Determine if shortening the covered section of a self-expanding bidirectional arterial cannula, can enhance retrograde flow and thus reduce the risk of lower limb ischemia.

Methods

Outlet pressure vs flow rate was determined for three cannulas types: a 15F self-expanding bidirectional cannula having a covered section of 90 mm, the same cannula but with a shorter covered section of 60 mm, and a Biomedicus cannula as control. The performances of all the cannulas were compared using a computerized flow-bench with calibrated sensors and a centrifugal pump. Water retrograde flow was determined using a tank timer technique. Anterograde and retrograde flow rate versus outlet pressure were determined at six different pump speed.

Results

For each of the six pump speed, both bidirectional cannulas, 60-mm covered and 90-mm covered respectively, showed higher performance than Biomedicus cannula control, as demonstrated by higher flow rate and lower pressure. We also observed that for the bidirectional cannula with shorter covered section, i.e. 60 mm coverage, provides enhanced performance as compared to a 90-mm coverage. Finally, the flow rate and the corresponding pressure can be consistently measured by our experimental set-up with low variability.

Conclusions

The new configuration of a shorter covered section in a bidirectional self-expanding cannula design, may present an opportunity to overcome lower leg ischemia during extra-corporal life support with long term peripheral cannulation.
Literature
6.
go back to reference Yoshimura N, Ataka K, Nakagiri K, Azami T, Yoshida M, Yamashita C, et al. A simple technique for the prevention of lower limb ischemia during femoral veno-arterial cardiopulmonary support. J Cardiovasc Surg. 1996;37(6):557–9. Yoshimura N, Ataka K, Nakagiri K, Azami T, Yoshida M, Yamashita C, et al. A simple technique for the prevention of lower limb ischemia during femoral veno-arterial cardiopulmonary support. J Cardiovasc Surg. 1996;37(6):557–9.
7.
go back to reference von Segesser L, Marinakis S, Berdajs D, Ferrari E, Wilhelm M, Maisano F. Prevention and therapy of leg ischaemia in extracorporeal life support and extracorporeal membrane oxygenation with peripheral cannulation. Swiss Med Wkly. 2016;146:w14304. von Segesser L, Marinakis S, Berdajs D, Ferrari E, Wilhelm M, Maisano F. Prevention and therapy of leg ischaemia in extracorporeal life support and extracorporeal membrane oxygenation with peripheral cannulation. Swiss Med Wkly. 2016;146:w14304.
13.
go back to reference Mikael BL, Prahl WL, Jeker WC, et al. Pressure and flow properties of cannulae for extracorporeal membrane Oxygenation II: drainage (venous) Cannulae. Perfusion. 2019;34(IS):65–73. Mikael BL, Prahl WL, Jeker WC, et al. Pressure and flow properties of cannulae for extracorporeal membrane Oxygenation II: drainage (venous) Cannulae. Perfusion. 2019;34(IS):65–73.
17.
go back to reference Mikael BL, Prahl WL, Jeker WC, et al. Pressure and flow properties of cannulae for extracorporeal membrane Oxygenation I: return (arterial) Cannulae. Perfusion. 2019;34(IS):58–64. Mikael BL, Prahl WL, Jeker WC, et al. Pressure and flow properties of cannulae for extracorporeal membrane Oxygenation I: return (arterial) Cannulae. Perfusion. 2019;34(IS):58–64.
Metadata
Title
Design optimization of bidirectional arterial perfusion cannula
Authors
Saad Abdel-Sayed
Enrico Ferrari
Philippe Abdel-Sayed
Markus Wilhelm
Ludwig-Karl von Segesser
Denis Berdajs
Publication date
01-12-2021
Publisher
BioMed Central
Keywords
ECMO
ECMO
Published in
Journal of Cardiothoracic Surgery / Issue 1/2021
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-021-01500-3

Other articles of this Issue 1/2021

Journal of Cardiothoracic Surgery 1/2021 Go to the issue