Prosthesis-patient mismatch after mitral valve replacement: a single-centered retrospective analysis in East China

Authors: Armah M Akuffu, Haige Zhao, Junnan Zheng, Yiming Ni

Published in: Journal of Cardiothoracic Surgery | Issue 1/2018

Login to get access

Abstract

Background

Prosthesis–patient mismatch (PPM) may affect the clinical outcomes of patients undergoing mitral valve replacement (MVR) surgery. We aimed to investigate the incidence of PPM of the mitral position in our center and analyze the possible predictors of PPM as well as its effect on short-term outcomes.

Methods

We retrospectively examined all consecutive patients with isolated or concomitant MVR at our center from 2013 to 2015. PPM was defined as an indexed effective orifice area (iEOA) of ≤1.2 cm2/m2. After inclusion and exclusion, a total of 1067 patients were analyzed. The baseline information were collected and compared between the two groups. Multivariate logistic regression analysis was conducted to determine the preoperative predictors of PPM as well as the effect of PPM on early mortality.

Results

A total of 1067 patients were included in the study. PPM was detected in 15.9% of the patients while 12 patients (1.12%) met the criteria for severe PPM. Patients with PPM compared to the non-PPM patients had higher age, larger body surface area and were more likely to be male and obese. Logistic regression analysis showed that higher age, larger BSA, bioprosthesis and smaller left ventricle end-diastolic diameter were predictors of PPM. There were no significant differences between the PPM and non-PPM groups regarding post-operative complications. Logistic regression analysis showed that PPM was not a risk factor of short-term mortality (P = 0.654). Also, there were no significant differences regarding short−/mid-term heart function between the PPM and non PPM groups (P = 0.902).

Conclusions

Our results demonstrated that higher age, bioprosthesis, larger BSA and smaller left ventricle size were associated with mitral PPM. However, PPM was not associated with poorer early outcomes after MVR surgery. In eastern of China, the prevalence of mitral valve stenosis is high; therefore, whether the standard PPM criteria are suitable for patients of this district needs to be further verified.
Literature
1.
go back to reference Rahimtoola SH. The problem of valve prosthesis-patient mismatch. Circulation. 1978;58(1):20–4.CrossRef Rahimtoola SH. The problem of valve prosthesis-patient mismatch. Circulation. 1978;58(1):20–4.CrossRef
2.
go back to reference Rahimtoola SH, Murphy E. Valve prosthesis--patient mismatch. A long-term sequela. Br Heart J. 1981;45(3):331–5.CrossRef Rahimtoola SH, Murphy E. Valve prosthesis--patient mismatch. A long-term sequela. Br Heart J. 1981;45(3):331–5.CrossRef
3.
go back to reference Hong S, Yi GJ, Youn YN, Lee S, Yoo KJ, Chang BC. Effect of the prosthesis-patient mismatch on long-term clinical outcomes after isolated aortic valve replacement for aortic stenosis: a prospective observational study. J Thorac Cardiov Sur. 2013;146(5):1098–104.CrossRef Hong S, Yi GJ, Youn YN, Lee S, Yoo KJ, Chang BC. Effect of the prosthesis-patient mismatch on long-term clinical outcomes after isolated aortic valve replacement for aortic stenosis: a prospective observational study. J Thorac Cardiov Sur. 2013;146(5):1098–104.CrossRef
4.
go back to reference Fuster RG, Montero Argudo JA, Albarova OG, et al. Patient-prosthesis mismatch in aortic valve replacement: really tolerable? Eur J Cardiothorac Surg. 2005;27(3):441–9 discussion 9.CrossRef Fuster RG, Montero Argudo JA, Albarova OG, et al. Patient-prosthesis mismatch in aortic valve replacement: really tolerable? Eur J Cardiothorac Surg. 2005;27(3):441–9 discussion 9.CrossRef
5.
go back to reference Guo L, Zheng J, Chen L, et al. Impact of prosthesis-patient mismatch on short-term outcomes after aortic valve replacement: a retrospective analysis in East China. J Cardiothorac Surg. 2017;12(1):42.CrossRef Guo L, Zheng J, Chen L, et al. Impact of prosthesis-patient mismatch on short-term outcomes after aortic valve replacement: a retrospective analysis in East China. J Cardiothorac Surg. 2017;12(1):42.CrossRef
6.
go back to reference Dumesnil JG, Honos GN, Lemieux M, Beauchemin J. Validation and applications of mitral prosthetic valvular areas calculated by Doppler echocardiography. Am J Cardiol. 1990;65(22):1443–8.CrossRef Dumesnil JG, Honos GN, Lemieux M, Beauchemin J. Validation and applications of mitral prosthetic valvular areas calculated by Doppler echocardiography. Am J Cardiol. 1990;65(22):1443–8.CrossRef
7.
go back to reference Dumesnil JG, Yoganathan AP. Valve prosthesis hemodynamics and the problem of high transprosthetic pressure gradients. Eur J Cardiothorac Surg. 1992;6(Suppl 1):S34–7 discussion S8.CrossRef Dumesnil JG, Yoganathan AP. Valve prosthesis hemodynamics and the problem of high transprosthetic pressure gradients. Eur J Cardiothorac Surg. 1992;6(Suppl 1):S34–7 discussion S8.CrossRef
8.
go back to reference Leavitt JI, Coats MH, Falk RH. Effects of exercise on transmitral gradient and pulmonary artery pressure in patients with mitral stenosis or a prosthetic mitral valve: a Doppler echocardiographic study. J Am Coll Cardiol. 1991;17(7):1520–6.CrossRef Leavitt JI, Coats MH, Falk RH. Effects of exercise on transmitral gradient and pulmonary artery pressure in patients with mitral stenosis or a prosthetic mitral valve: a Doppler echocardiographic study. J Am Coll Cardiol. 1991;17(7):1520–6.CrossRef
9.
go back to reference Rosenhek R, Binder T, Maurer G, Baumgartner H. Normal values for Doppler echocardiographic assessment of heart valve prostheses. J Am Soc Echocardiogr. 2003;16(11):1116–27.CrossRef Rosenhek R, Binder T, Maurer G, Baumgartner H. Normal values for Doppler echocardiographic assessment of heart valve prostheses. J Am Soc Echocardiogr. 2003;16(11):1116–27.CrossRef
10.
go back to reference Dumesnil JG, Pibarot P. Prosthesis-patient mismatch: an update. Curr Cardiol Rep. 2011;13(3):250–7.CrossRef Dumesnil JG, Pibarot P. Prosthesis-patient mismatch: an update. Curr Cardiol Rep. 2011;13(3):250–7.CrossRef
11.
go back to reference Lee SH, Chang BC, Youn YN, Joo HC, Yoo KJ, Lee S. Impact of prosthesis-patient mismatch after mitral valve replacement in rheumatic population: does mitral position prosthesis-patient mismatch really exist? J Cardiothorac Surg. 2017;12(1):88.CrossRef Lee SH, Chang BC, Youn YN, Joo HC, Yoo KJ, Lee S. Impact of prosthesis-patient mismatch after mitral valve replacement in rheumatic population: does mitral position prosthesis-patient mismatch really exist? J Cardiothorac Surg. 2017;12(1):88.CrossRef
12.
go back to reference Lam BK, Chan V, Hendry P, et al. The impact of patient-prosthesis mismatch on late outcomes after mitral valve replacement. J Thorac Cardiovasc Surg. 2007;133(6):1464–73.CrossRef Lam BK, Chan V, Hendry P, et al. The impact of patient-prosthesis mismatch on late outcomes after mitral valve replacement. J Thorac Cardiovasc Surg. 2007;133(6):1464–73.CrossRef
13.
go back to reference Magne J, Mathieu P, Dumesnil JG, et al. Impact of prosthesis-patient mismatch on survival after mitral valve replacement. Circulation. 2007;115(11):1417–25.CrossRef Magne J, Mathieu P, Dumesnil JG, et al. Impact of prosthesis-patient mismatch on survival after mitral valve replacement. Circulation. 2007;115(11):1417–25.CrossRef
14.
go back to reference Li M, Dumesnil JG, Mathieu P, Pibarot P. Impact of valve prosthesis-patient mismatch on pulmonary arterial pressure after mitral valve replacement. J Am Coll Cardiol. 2005;45(7):1034–40.CrossRef Li M, Dumesnil JG, Mathieu P, Pibarot P. Impact of valve prosthesis-patient mismatch on pulmonary arterial pressure after mitral valve replacement. J Am Coll Cardiol. 2005;45(7):1034–40.CrossRef
15.
go back to reference Jamieson WR, Germann E, Ye J, et al. Effect of prosthesis-patient mismatch on long-term survival with mitral valve replacement: assessment to 15 years. Ann Thorac Surg. 2009;87(4):1135–41 discussion 42.CrossRef Jamieson WR, Germann E, Ye J, et al. Effect of prosthesis-patient mismatch on long-term survival with mitral valve replacement: assessment to 15 years. Ann Thorac Surg. 2009;87(4):1135–41 discussion 42.CrossRef
16.
go back to reference Aziz A, Lawton JS, Maniar HS, Pasque MK, Damiano RJ Jr, Moon MR. Factors affecting survival after mitral valve replacement in patients with prosthesis-patient mismatch. Ann Thorac Surg. 2010;90(4):1202–10 discussion 10-1.CrossRef Aziz A, Lawton JS, Maniar HS, Pasque MK, Damiano RJ Jr, Moon MR. Factors affecting survival after mitral valve replacement in patients with prosthesis-patient mismatch. Ann Thorac Surg. 2010;90(4):1202–10 discussion 10-1.CrossRef
17.
go back to reference Angeloni E, Melina G, Pibarot P, et al. Impact of prosthesis-patient mismatch on the regression of secondary mitral regurgitation after isolated aortic valve replacement with a bioprosthetic valve in patients with severe aortic stenosis. Circ Cardiovasc Imaging. 2012;5(1):36–42.CrossRef Angeloni E, Melina G, Pibarot P, et al. Impact of prosthesis-patient mismatch on the regression of secondary mitral regurgitation after isolated aortic valve replacement with a bioprosthetic valve in patients with severe aortic stenosis. Circ Cardiovasc Imaging. 2012;5(1):36–42.CrossRef
18.
go back to reference Sakamoto H, Watanabe Y. Does patient-prosthesis mismatch affect long-term results after mitral valve replacement? Ann Thorac Cardiovasc Surg. 2010;16(3):163–7.PubMed Sakamoto H, Watanabe Y. Does patient-prosthesis mismatch affect long-term results after mitral valve replacement? Ann Thorac Cardiovasc Surg. 2010;16(3):163–7.PubMed
19.
go back to reference Shi WY, Yap CH, Hayward PA, et al. Impact of prosthesis--patient mismatch after mitral valve replacement: a multicentre analysis of early outcomes and mid-term survival. Heart. 2011;97(13):1074–81.CrossRef Shi WY, Yap CH, Hayward PA, et al. Impact of prosthesis--patient mismatch after mitral valve replacement: a multicentre analysis of early outcomes and mid-term survival. Heart. 2011;97(13):1074–81.CrossRef
20.
go back to reference Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart. 2006;92(8):1022–9.CrossRef Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart. 2006;92(8):1022–9.CrossRef
21.
go back to reference Bitar JN, Lechin ME, Salazar G, Zoghbi WA. Doppler echocardiographic assessment with the continuity equation of St. Jude Medical mechanical prostheses in the mitral valve position. Am J Cardiol. 1995;76(4):287–93.CrossRef Bitar JN, Lechin ME, Salazar G, Zoghbi WA. Doppler echocardiographic assessment with the continuity equation of St. Jude Medical mechanical prostheses in the mitral valve position. Am J Cardiol. 1995;76(4):287–93.CrossRef
22.
go back to reference Cohn LH, Edmunds LH. Cardiac surgery in the adult. 3rd ed. New York: McGraw-Hill Medical; 2008. Cohn LH, Edmunds LH. Cardiac surgery in the adult. 3rd ed. New York: McGraw-Hill Medical; 2008.
23.
go back to reference Borracci RA, Rubio M, Sestito ML, Ingino CA, Barrero C, Rapallo CA. Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor(R) porcine prosthetic valves. Cardiol J. 2016;23(2):178–83.CrossRef Borracci RA, Rubio M, Sestito ML, Ingino CA, Barrero C, Rapallo CA. Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor(R) porcine prosthetic valves. Cardiol J. 2016;23(2):178–83.CrossRef
Metadata
Title
Prosthesis-patient mismatch after mitral valve replacement: a single-centered retrospective analysis in East China
Authors
Armah M Akuffu
Haige Zhao
Junnan Zheng
Yiming Ni
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2018
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-018-0788-4

Other articles of this Issue 1/2018

Journal of Cardiothoracic Surgery 1/2018 Go to the issue