Open Access 01-12-2017 | Research article

Impact of prosthesis–patient mismatch on short-term outcomes after aortic valve replacement: a retrospective analysis in East China

Authors: Lei Guo, Junnan Zheng, Liangwei Chen, Renyuan Li, Liang Ma, Yiming Ni, Haige Zhao

Published in: Journal of Cardiothoracic Surgery | Issue 1/2017

Login to get access

Abstract

Background

Prosthesis–patient mismatch (PPM) may affect the clinical outcomes of patients undergoing aortic valve replacement (AVR). We aimed to determine the incidence of PPM, its effect on short-term mortality, and the factors contributing to PPM in China.

Methods

We retrospectively examined all consecutive patients with isolated or concomitant AVR at our hospital between January 1, 2013 and December 31, 2015. PPM was defined as an effective orifice area index (EOAi) of ≤ 0.85 cm2/m2. The baseline, echocardiographic, operative, and outcome data of all patients were collected from the national database.

Results

A total of 869 patients were included in the study. PPM was detected in 15.9% (138/869) of the patients. Four patients (0.5%) met the criteria for severe PPM. Patients with PPM were older and had a higher prevalence of diabetes, coronary heart disease, aortic stenosis (AS), and preoperative left ventricular dysfunction but a lower incidence of smoking history and aortic regurgitation. Logistic regression analysis showed that female gender (P < 0.001), AS (P = 0.014), higher body mass index (BMI) (P < 0.001), and bioprosthesis (P < 0.001) were independent predictors of PPM. We also found that PPM (P = 0.005) was associated with 30-day all-cause mortality, along with smoking history (P = 0.001) and low preoperative left ventricular ejection fraction (LVEF) (P = 0.004).

Conclusions

PPM is associated with high short-term mortality after AVR in China. Female gender, aortic stenosis, bioprosthesis, and high BMI are risk factors for the incidence of PPM.
Literature
1.
2.
go back to reference Howell NJ, Keogh BE, Ray D, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89(1):60–4.CrossRefPubMed Howell NJ, Keogh BE, Ray D, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89(1):60–4.CrossRefPubMed
3.
go back to reference Howell NJ, Keogh BE, Barnet V, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30(1):10–4.CrossRefPubMed Howell NJ, Keogh BE, Barnet V, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30(1):10–4.CrossRefPubMed
4.
go back to reference Sportelli E, Regesta T, Salsano A, et al. Does patient-prosthesis mismatch after aortic valve replacement affect survival and quality of life in elderly patients? J Cardiovasc Med. 2016;17(2):137–43.CrossRef Sportelli E, Regesta T, Salsano A, et al. Does patient-prosthesis mismatch after aortic valve replacement affect survival and quality of life in elderly patients? J Cardiovasc Med. 2016;17(2):137–43.CrossRef
5.
go back to reference Hong S, Yi GJ, Youn YN, Lee S, Yoo KJ, Chang BC. Effect of the prosthesis-patient mismatch on long-term clinical outcomes after isolated aortic valve replacement for aortic stenosis: A prospective observational study. J Thorac Cardiov Sur. 2013;146(5):1098–104.CrossRef Hong S, Yi GJ, Youn YN, Lee S, Yoo KJ, Chang BC. Effect of the prosthesis-patient mismatch on long-term clinical outcomes after isolated aortic valve replacement for aortic stenosis: A prospective observational study. J Thorac Cardiov Sur. 2013;146(5):1098–104.CrossRef
6.
go back to reference Fuster RG, Montero Argudo JA, Albarova OG, et al. Patient-prosthesis mismatch in aortic valve replacement: really tolerable? Eur J Cardiothorac Surg. 2005;27(3):441–9. discussion 9.CrossRefPubMed Fuster RG, Montero Argudo JA, Albarova OG, et al. Patient-prosthesis mismatch in aortic valve replacement: really tolerable? Eur J Cardiothorac Surg. 2005;27(3):441–9. discussion 9.CrossRefPubMed
7.
go back to reference Zhou XM, Zhuang W, Hu JG, Li JM, Yu JF, Jiang L. Low-dose anticoagulation in Chinese patients with mechanical heart valves. Asian Cardiovasc Thorac Ann. 2005;13(4):341–4.CrossRefPubMed Zhou XM, Zhuang W, Hu JG, Li JM, Yu JF, Jiang L. Low-dose anticoagulation in Chinese patients with mechanical heart valves. Asian Cardiovasc Thorac Ann. 2005;13(4):341–4.CrossRefPubMed
8.
go back to reference Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart. 2006;92(8):1022–9.CrossRefPubMed Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart. 2006;92(8):1022–9.CrossRefPubMed
9.
go back to reference Kaminishi Y, Misawa Y, Kobayashi J, et al. Patient-prosthesis mismatch in patients with aortic valve replacement. Gen Thorac Cardiovasc Surg. 2013;61(5):274–9.CrossRefPubMedPubMedCentral Kaminishi Y, Misawa Y, Kobayashi J, et al. Patient-prosthesis mismatch in patients with aortic valve replacement. Gen Thorac Cardiovasc Surg. 2013;61(5):274–9.CrossRefPubMedPubMedCentral
10.
go back to reference Joshi SS, Ashwini T, George A, Jagadeesh AM. Patient prosthesis mismatch after aortic valve replacement: An Indian perspective. Ann Card Anaesth. 2016;19(1):84–8.CrossRefPubMedPubMedCentral Joshi SS, Ashwini T, George A, Jagadeesh AM. Patient prosthesis mismatch after aortic valve replacement: An Indian perspective. Ann Card Anaesth. 2016;19(1):84–8.CrossRefPubMedPubMedCentral
11.
go back to reference Muneretto C, Bisleri G, Negri A, Manfredi J. The concept of patient-prosthesis mismatch. J Heart Valve Dis. 2004;13 Suppl 1:S59–62.PubMed Muneretto C, Bisleri G, Negri A, Manfredi J. The concept of patient-prosthesis mismatch. J Heart Valve Dis. 2004;13 Suppl 1:S59–62.PubMed
12.
go back to reference Head SJ, Mokhles MM, Osnabrugge RL, et al. The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27 186 patients with 133 141 patient-years. Eur Heart J. 2012;33(12):1518–29.CrossRefPubMed Head SJ, Mokhles MM, Osnabrugge RL, et al. The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27 186 patients with 133 141 patient-years. Eur Heart J. 2012;33(12):1518–29.CrossRefPubMed
13.
go back to reference Bonderman D, Graf A, Kammerlander AA, et al. Factors determining patient-prosthesis mismatch after aortic valve replacement--a prospective cohort study. PLoS One. 2013;8(12):e81940.CrossRefPubMedPubMedCentral Bonderman D, Graf A, Kammerlander AA, et al. Factors determining patient-prosthesis mismatch after aortic valve replacement--a prospective cohort study. PLoS One. 2013;8(12):e81940.CrossRefPubMedPubMedCentral
14.
go back to reference Blais C, Dumesnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108(8):983–8.CrossRefPubMed Blais C, Dumesnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108(8):983–8.CrossRefPubMed
15.
go back to reference Walther T, Rastan A, Falk V, et al. Patient prosthesis mismatch affects short- and long-term outcomes after aortic valve replacement. Eur J Cardiothorac Surg. 2006;30(1):15–9.CrossRefPubMed Walther T, Rastan A, Falk V, et al. Patient prosthesis mismatch affects short- and long-term outcomes after aortic valve replacement. Eur J Cardiothorac Surg. 2006;30(1):15–9.CrossRefPubMed
16.
go back to reference Hoffmann G, Ogbamicael SA, Jochens A, et al. Impact of patient-prosthesis mismatch following aortic valve replacement on short-term survival: a retrospective single center analysis of 632 consecutive patients with isolated stented biological aortic valve replacement. Thorac Cardiov Surg. 2014;62(6):469–74.CrossRef Hoffmann G, Ogbamicael SA, Jochens A, et al. Impact of patient-prosthesis mismatch following aortic valve replacement on short-term survival: a retrospective single center analysis of 632 consecutive patients with isolated stented biological aortic valve replacement. Thorac Cardiov Surg. 2014;62(6):469–74.CrossRef
17.
go back to reference Sommers KE, David TE. Aortic valve replacement with patch enlargement of the aortic annulus. Ann Thorac Surg. 1997;63(6):1608–12.CrossRefPubMed Sommers KE, David TE. Aortic valve replacement with patch enlargement of the aortic annulus. Ann Thorac Surg. 1997;63(6):1608–12.CrossRefPubMed
18.
go back to reference Carrier M, Pellerin M, Perrault LP, et al. Experience with the 19-mm Carpentier-Edwards pericardial bioprosthesis in the elderly. Ann Thorac Surg. 2001;71(5 Suppl):S249–52.CrossRefPubMed Carrier M, Pellerin M, Perrault LP, et al. Experience with the 19-mm Carpentier-Edwards pericardial bioprosthesis in the elderly. Ann Thorac Surg. 2001;71(5 Suppl):S249–52.CrossRefPubMed
19.
go back to reference Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36(4):1131–41.CrossRefPubMed Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36(4):1131–41.CrossRefPubMed
20.
go back to reference Sakamoto Y, Hashimoto K. Update on aortic valve prosthesis-patient mismatch in Japan. Gen Thorac Cardiovasc Surg. 2013;61(12):669–75.CrossRefPubMed Sakamoto Y, Hashimoto K. Update on aortic valve prosthesis-patient mismatch in Japan. Gen Thorac Cardiovasc Surg. 2013;61(12):669–75.CrossRefPubMed
21.
go back to reference Bach DS, Sakwa MP, Goldbach M, Petracek MR, Emery RW, Mohr FW. Hemodynamics and early clinical performance of the St. Jude Medical Regent mechanical aortic valve. Ann Thorac Surg. 2002;74(6):2003–9. discussion 9.CrossRefPubMed Bach DS, Sakwa MP, Goldbach M, Petracek MR, Emery RW, Mohr FW. Hemodynamics and early clinical performance of the St. Jude Medical Regent mechanical aortic valve. Ann Thorac Surg. 2002;74(6):2003–9. discussion 9.CrossRefPubMed
23.
go back to reference Thomson DJ, Jamieson WR, Dumesnil JG, et al. Medtronic mosaic porcine bioprosthesis satisfactory early clinical performance. Ann Thorac Surg. 1998;66(6 Suppl):S122–5.CrossRefPubMed Thomson DJ, Jamieson WR, Dumesnil JG, et al. Medtronic mosaic porcine bioprosthesis satisfactory early clinical performance. Ann Thorac Surg. 1998;66(6 Suppl):S122–5.CrossRefPubMed
24.
go back to reference Pibarot P, Dumesnil JG. Prosthetic heart valves: selection of the optimal prosthesis and long-term management. Circulation. 2009;119(7):1034–48.CrossRefPubMed Pibarot P, Dumesnil JG. Prosthetic heart valves: selection of the optimal prosthesis and long-term management. Circulation. 2009;119(7):1034–48.CrossRefPubMed
Metadata
Title
Impact of prosthesis–patient mismatch on short-term outcomes after aortic valve replacement: a retrospective analysis in East China
Authors
Lei Guo
Junnan Zheng
Liangwei Chen
Renyuan Li
Liang Ma
Yiming Ni
Haige Zhao
Publication date
01-12-2017
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2017
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-017-0596-2

Other articles of this Issue 1/2017

Journal of Cardiothoracic Surgery 1/2017 Go to the issue