Open Access 01-05-2013 | Original Article

Patient-prosthesis mismatch in patients with aortic valve replacement

Authors: Yuichiro Kaminishi, Yoshio Misawa, Junjiro Kobayashi, Hiroaki Konishi, Hiroaki Miyata, Noboru Motomura, Shin-ichi Takamoto, Japan Cardiovascular Surgery Database Organization

Published in: General Thoracic and Cardiovascular Surgery | Issue 5/2013

Login to get access

Abstract

Objective

Patient-prosthesis mismatch (PPM) may affect clinical outcomes in patients with aortic valve replacement (AVR). We retrospectively examined the PPM in patients with isolated AVR in the Japan Adult Cardiovascular Surgery Database (JACVSD).

Methods

We examined all patients with isolated AVR between January 1, 2008 and December 31, 2009. The JACVSD data collection form has a total of 255 variables. We defined PPM as an effective orifice area index of ≤0.85 m2/cm2.

Results

PPM was observed in 306 of 3,609 cases analyzed, PPM rate was 8.5 %. Body surface area was larger and body mass index was higher in the PPM group than the non-PPM group (P < 0.001). Patients with PPM were older (P = 0.001) and had a higher prevalence of diabetes (P = 0.004), dyslipidemia (P < 0.001), hypertension (P < 0.001), cerebrovascular disease (P = 0.031), old myocardial infarction (P = 0.006), previous percutaneous coronary artery intervention (P = 0.001), coronary artery disease (P = 0.018), and aortic valve stenosis (P < 0.001). Perioperative blood transfusion (P < 0.001) and dialysis (P = 0.005) were more frequent in the PPM group. Postoperative ventilation (P = 0.004) and intensive care unit stay (P = 0.004) were significantly longer in the PPM group.

Conclusions

Age, aortic valve stenosis, dyslipidemia, hypertension, old myocardial infarction, previous percutaneous coronary artery intervention, diabetes mellitus, cerebrovascular disease, and high body mass index were the risk factors for PPM. PPM was not an independent risk factor for short-term mortality.
Literature
1.
go back to reference Ueda Y, Fujii Y, Udagawa H. Thoracic and cardiovascular surgery in Japan during 2006. Gen Thorac Cardiovasc Surg. 2008;56:365–88.PubMedCrossRef Ueda Y, Fujii Y, Udagawa H. Thoracic and cardiovascular surgery in Japan during 2006. Gen Thorac Cardiovasc Surg. 2008;56:365–88.PubMedCrossRef
2.
go back to reference Grover FL, Shroyer AL, Edwards FH, Grover FL, Shroyer ALW, Edwards FH, et al. Data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1996;26:1229–31.CrossRef Grover FL, Shroyer AL, Edwards FH, Grover FL, Shroyer ALW, Edwards FH, et al. Data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1996;26:1229–31.CrossRef
3.
go back to reference Shroyer AL, Edwards FH, Grover FL. Updates to the data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1998;65:1494–7.PubMedCrossRef Shroyer AL, Edwards FH, Grover FL. Updates to the data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1998;65:1494–7.PubMedCrossRef
4.
go back to reference Howell NJ, Keogh BE, Barnet V, Bonser RS, Graham TR, Rooney SJ, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30:10–4.PubMedCrossRef Howell NJ, Keogh BE, Barnet V, Bonser RS, Graham TR, Rooney SJ, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30:10–4.PubMedCrossRef
5.
go back to reference Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36:1131–41.PubMedCrossRef Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36:1131–41.PubMedCrossRef
6.
go back to reference Blais C, Dumensnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108:983–8.PubMedCrossRef Blais C, Dumensnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108:983–8.PubMedCrossRef
7.
go back to reference Hosmer DW, Lemeshow S. Applied logistic regression. New York: Wiley; 1989. p. 82–132. Hosmer DW, Lemeshow S. Applied logistic regression. New York: Wiley; 1989. p. 82–132.
8.
go back to reference Motomura N, Miyata H, Tsukihara H, Takamoto S, Japan Cardiovascular Surgery Database Organization. Risk Model of Valve Surgery in Japan Using the Japan Adult Cardiovascular Surgery Database. J Heart Valve Dis. 2010;19(6):684–691. Motomura N, Miyata H, Tsukihara H, Takamoto S, Japan Cardiovascular Surgery Database Organization. Risk Model of Valve Surgery in Japan Using the Japan Adult Cardiovascular Surgery Database. J Heart Valve Dis. 2010;19(6):684–691.
9.
go back to reference Melina G, Angeloni E, Benedetto U, Refice S, Miceli A, Miele C, et al. Relationship between prosthesis-patient mismatch and pro-brain natriuretic peptides after aortic valve replacement. J Heart Valve Dis. 2010;19:171–6.PubMed Melina G, Angeloni E, Benedetto U, Refice S, Miceli A, Miele C, et al. Relationship between prosthesis-patient mismatch and pro-brain natriuretic peptides after aortic valve replacement. J Heart Valve Dis. 2010;19:171–6.PubMed
10.
go back to reference Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121:2123–9.PubMedCrossRef Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121:2123–9.PubMedCrossRef
11.
12.
go back to reference Tomoeda H, Ueda T, Teshima H, Arinaga K, Tayama K, Fukunaga S, et al. Postoperative left ventricular mass regression after aortic valve replacement for aortic stenosis. Ann Thorac Surg. 2010;89:745–50.PubMedCrossRef Tomoeda H, Ueda T, Teshima H, Arinaga K, Tayama K, Fukunaga S, et al. Postoperative left ventricular mass regression after aortic valve replacement for aortic stenosis. Ann Thorac Surg. 2010;89:745–50.PubMedCrossRef
13.
go back to reference Pisano C, D’Amico T, Palmeri C, Franchino R, Fattouch K, Bianco G, et al. Valve prosthesis-patient mismatch: hemodynamic, echocardiographic and clinical consequences. Interact Cardiovasc Thorac Surg. 2011;13:606–10.PubMedCrossRef Pisano C, D’Amico T, Palmeri C, Franchino R, Fattouch K, Bianco G, et al. Valve prosthesis-patient mismatch: hemodynamic, echocardiographic and clinical consequences. Interact Cardiovasc Thorac Surg. 2011;13:606–10.PubMedCrossRef
14.
go back to reference Garatti A, Mori F, Innocente F, Canziani A, Gagliardotto P, Mossuto E, et al. Aortic valve replacement with 17-mm mechanical prostheses: is patient-prosthesis mismatch a relevant phenomenon? Ann Thorac Surg. 2011;91:71–7.PubMedCrossRef Garatti A, Mori F, Innocente F, Canziani A, Gagliardotto P, Mossuto E, et al. Aortic valve replacement with 17-mm mechanical prostheses: is patient-prosthesis mismatch a relevant phenomenon? Ann Thorac Surg. 2011;91:71–7.PubMedCrossRef
15.
go back to reference Brown ML, Schaff HV, Suri RM, Li Z, Sundt TM, Dearani JA, et al. Regression in left ventricular mass after aortic valve replacement for chronic aortic regurgitation is unrelated to prosthetic valve size. J Thorac Cardiovasc Surg. 2011;142:e5–9.PubMedCrossRef Brown ML, Schaff HV, Suri RM, Li Z, Sundt TM, Dearani JA, et al. Regression in left ventricular mass after aortic valve replacement for chronic aortic regurgitation is unrelated to prosthetic valve size. J Thorac Cardiovasc Surg. 2011;142:e5–9.PubMedCrossRef
16.
go back to reference Kobayashi Y, Fukushima Y, Hayase T, Kojima K, Endo G. Clinical outcome of aortic valve replacement with 16-mm ATS-advanced performance valve for small aortic annulus. Ann Thorac Surg. 2010;89:1195–9.PubMedCrossRef Kobayashi Y, Fukushima Y, Hayase T, Kojima K, Endo G. Clinical outcome of aortic valve replacement with 16-mm ATS-advanced performance valve for small aortic annulus. Ann Thorac Surg. 2010;89:1195–9.PubMedCrossRef
17.
go back to reference Cotoni DA, Palac RT, Dacey LJ, O’Rourke DJ. Defining patient-prosthesis mismatch and its effect on survival in patients with impaired ejection fraction. Ann Thorac Surg. 2011;91:692–9.PubMedCrossRef Cotoni DA, Palac RT, Dacey LJ, O’Rourke DJ. Defining patient-prosthesis mismatch and its effect on survival in patients with impaired ejection fraction. Ann Thorac Surg. 2011;91:692–9.PubMedCrossRef
18.
go back to reference Howell NJ, Keogh BE, Ray D, Bonser RS, Graham TR, Mascaro J, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89:60–4.PubMedCrossRef Howell NJ, Keogh BE, Ray D, Bonser RS, Graham TR, Mascaro J, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89:60–4.PubMedCrossRef
19.
go back to reference Jamieson WR, Ye J, Higgins J, Cheung A, Fradet GJ, Skasgard P, et al. Effect of prosthesis-patient mismatch on long-term survival with aortic valve replacement: assessment to 15 years. Ann Thorac Surg. 2010;89:51–8.PubMedCrossRef Jamieson WR, Ye J, Higgins J, Cheung A, Fradet GJ, Skasgard P, et al. Effect of prosthesis-patient mismatch on long-term survival with aortic valve replacement: assessment to 15 years. Ann Thorac Surg. 2010;89:51–8.PubMedCrossRef
20.
go back to reference Okamura H, Yamaguchi A, Noguchi K, Naito K, Yuri K, Adachi H. Hemodynamics and outcomes of aortic valve replacement with a 17- or 19-mm valve. Asian Cardiovasc Thorac Ann. 2010;18:450–5.PubMed Okamura H, Yamaguchi A, Noguchi K, Naito K, Yuri K, Adachi H. Hemodynamics and outcomes of aortic valve replacement with a 17- or 19-mm valve. Asian Cardiovasc Thorac Ann. 2010;18:450–5.PubMed
21.
go back to reference Coutinho GF, Correia PM, Paupério G, de Oliveira F, Antunes MJ. Aortic root enlargement does not increase the surgical risk and short-term patient outcome? Eur J Cardiothorac Surg. 2011;40:441–7.PubMed Coutinho GF, Correia PM, Paupério G, de Oliveira F, Antunes MJ. Aortic root enlargement does not increase the surgical risk and short-term patient outcome? Eur J Cardiothorac Surg. 2011;40:441–7.PubMed
22.
go back to reference Kulik A, Al-Saigh M, Chan V, Masters RG, Bédard P, Lam BK, et al. Enlargement of the small aortic root during aortic valve replacement: is there a benefit? Ann Thorac Surg. 2008;85:94–100.PubMedCrossRef Kulik A, Al-Saigh M, Chan V, Masters RG, Bédard P, Lam BK, et al. Enlargement of the small aortic root during aortic valve replacement: is there a benefit? Ann Thorac Surg. 2008;85:94–100.PubMedCrossRef
Metadata
Title
Patient-prosthesis mismatch in patients with aortic valve replacement
Authors
Yuichiro Kaminishi
Yoshio Misawa
Junjiro Kobayashi
Hiroaki Konishi
Hiroaki Miyata
Noboru Motomura
Shin-ichi Takamoto
Japan Cardiovascular Surgery Database Organization
Publication date
01-05-2013
Publisher
Springer Japan
Published in
General Thoracic and Cardiovascular Surgery / Issue 5/2013
Print ISSN: 1863-6705
Electronic ISSN: 1863-6713
DOI
https://doi.org/10.1007/s11748-013-0216-6

Other articles of this Issue 5/2013

General Thoracic and Cardiovascular Surgery 5/2013 Go to the issue