Published in: Journal of Orthopaedic Surgery and Research 1/2019

Open Access 01-12-2019 | Computed Tomography | Research article

Differences in bone mineral density of trajectory between lumbar cortical and traditional pedicle screws

Authors: Renjie Zhang, Hai Gao, Huimin Li, Tao Xing, Chongyu Jia, Jianxiang Zhang, Fulong Dong, Cailiang Shen

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2019

Login to get access

Abstract

Background

Cortical bone trajectory (CBT) has been well-known in spine surgery for obtaining improved fixation while minimizing soft tissue dissection. This study was designed to compare the bone mineral density (BMD) between the CBT and traditional trajectory (TT) by using Hounsfield unit (HU) values and identify the ideal decades of patients and the suitable lumbar segments using this CBT technology from a radiological standpoint.

Methods

Patients were selected randomly from an institutional database based on age (evenly distributed by a decade of life) and gender. A total of 240 healthy patients had a computed tomography (CT) scan of the chest, abdomen, and pelvis. For each patient, axial slices of every vertebra were cut in two planes: one horizontal to the pedicle representing the plane wherein pedicle screws were inserted using the TT and the other in a caudocranial plane representing the plane wherein pedicle screws were inserted using the CBT. For each trajectory, a region of interest (ROI) was selected within the area wherein the screws were inserted. A CT number (HU values) was then calculated within each ROI to represent bone density.

Results

HU values measured at the ROI of CBT were significantly greater than those of the traditional pedicle screw in all age groups, and the specific value (ratio of the HU values of CBT/the HU values of TT) between CBT and TT was 1.92. A significant difference was observed between male and female. The HU values of CBT and TT of males were generally higher than those of females (males: CBT/TT 1.89 ± 0.45; Females: CBT/TT 1.95 ± 0.47). The specific value in HU values significantly increased with increasing age (p = 0.000) and cauda lumbar level (p = 0.000) in males and females.

Conclusion

BMD, as measured by HU values for the ROI of the CBT screw, was significantly greater than that of the traditional pedicle screw, especially in old patients and cauda lumbar segments.
Literature
1.
go back to reference Hart RA, Hansen BL, Shea M, Hsu F, Anderson GJ. Pedicle screw placement in the thoracic spine: a comparison of image-guided and manual techniques in cadavers. Spine. 2005;30(12):E326–31.CrossRef Hart RA, Hansen BL, Shea M, Hsu F, Anderson GJ. Pedicle screw placement in the thoracic spine: a comparison of image-guided and manual techniques in cadavers. Spine. 2005;30(12):E326–31.CrossRef
2.
go back to reference Wittenberg RH, Shea M, Swartz DE, Lee KS, White AA, Hayes WC. Importance of bone mineral density in instrumented spine fusions. Spine. 1991;16(6):647–52.CrossRef Wittenberg RH, Shea M, Swartz DE, Lee KS, White AA, Hayes WC. Importance of bone mineral density in instrumented spine fusions. Spine. 1991;16(6):647–52.CrossRef
3.
go back to reference Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4(4):402–8.CrossRef Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4(4):402–8.CrossRef
4.
go back to reference Davne SH, Myers DL. Complications of lumbar spinal fusion with transpedicular instrumentation. Spine. 1992;17(6 Suppl):184–9.CrossRef Davne SH, Myers DL. Complications of lumbar spinal fusion with transpedicular instrumentation. Spine. 1992;17(6 Suppl):184–9.CrossRef
5.
go back to reference Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Influence of bone mineral density on pedicle screw fixation: a study of pedicle screw fixation augmenting posterior lumbar interbody fusion in elderly patients. Spine J. 2001;1(6):402–7.CrossRef Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Influence of bone mineral density on pedicle screw fixation: a study of pedicle screw fixation augmenting posterior lumbar interbody fusion in elderly patients. Spine J. 2001;1(6):402–7.CrossRef
6.
go back to reference Suzuki T, Abe E, Okuyama K, Sato K. Improving the pullout strength of pedicle screws by screw coupling. J Spinal Disord. 2001;14(5):399.CrossRef Suzuki T, Abe E, Okuyama K, Sato K. Improving the pullout strength of pedicle screws by screw coupling. J Spinal Disord. 2001;14(5):399.CrossRef
7.
go back to reference Santoni BG, Hynes RA, McGilvray KC, Rodriguez-Canessa G, Lyons AS, Henson MAW, Womack WJ, Puttlitz CM. Cortical bone trajectory for lumbar pedicle screws. Spine J. 2009;9(5):366–73.CrossRef Santoni BG, Hynes RA, McGilvray KC, Rodriguez-Canessa G, Lyons AS, Henson MAW, Womack WJ, Puttlitz CM. Cortical bone trajectory for lumbar pedicle screws. Spine J. 2009;9(5):366–73.CrossRef
8.
go back to reference Matsukawa K, Yato Y, Kato T, Imabayashi H, Asazuma T, Nemoto K. In vivo analysis of insertional torque during pedicle screwing using cortical bone trajectory technique. Spine. 2014;39(4):E240–5.CrossRef Matsukawa K, Yato Y, Kato T, Imabayashi H, Asazuma T, Nemoto K. In vivo analysis of insertional torque during pedicle screwing using cortical bone trajectory technique. Spine. 2014;39(4):E240–5.CrossRef
9.
go back to reference Baluch DA, Patel AA, Lullo B, Havey RM, Voronov LI, Nguyen N-L, Carandang G, Ghanayem AJ, Patwardhan AG. Effect of physiological loads on cortical and traditional pedicle screw fixation. Spine. 2014;39(22):E1297–302.CrossRef Baluch DA, Patel AA, Lullo B, Havey RM, Voronov LI, Nguyen N-L, Carandang G, Ghanayem AJ, Patwardhan AG. Effect of physiological loads on cortical and traditional pedicle screw fixation. Spine. 2014;39(22):E1297–302.CrossRef
10.
go back to reference Crookshank M, Ploeg HL, Ellis RE, Macintyre NJ. Repeatable calibration of Hounsfield units to mineral density and effect of scanning medium. Advances in Biomechanics and Applications. 2013;1(1):15–22.CrossRef Crookshank M, Ploeg HL, Ellis RE, Macintyre NJ. Repeatable calibration of Hounsfield units to mineral density and effect of scanning medium. Advances in Biomechanics and Applications. 2013;1(1):15–22.CrossRef
11.
go back to reference Eswaran SK, Gupta A, Adams MF, Keaveny TM. Cortical and trabecular load sharing in the human vertebral body. J Bone Miner Res. 2006;21(2):307–14.CrossRef Eswaran SK, Gupta A, Adams MF, Keaveny TM. Cortical and trabecular load sharing in the human vertebral body. J Bone Miner Res. 2006;21(2):307–14.CrossRef
12.
go back to reference Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am. 2011;93(11):1057–63.CrossRef Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am. 2011;93(11):1057–63.CrossRef
13.
go back to reference Ilser T, Lars S, Mcglumphy EA, Tözüm TF. Biomechanical aspects of primary implant stability: a human cadaver study. Clin Implant Dent Relat Res. 2009;11(2):113.CrossRef Ilser T, Lars S, Mcglumphy EA, Tözüm TF. Biomechanical aspects of primary implant stability: a human cadaver study. Clin Implant Dent Relat Res. 2009;11(2):113.CrossRef
14.
go back to reference Turkyilmaz I, Tumer C, Ozbek EN, Tözüm TF. Relations between the bone density values from computerized tomography, and implant stability parameters: a clinical study of 230 regular platform implants. J Clin Periodontol. 2007;34(8):716–22.CrossRef Turkyilmaz I, Tumer C, Ozbek EN, Tözüm TF. Relations between the bone density values from computerized tomography, and implant stability parameters: a clinical study of 230 regular platform implants. J Clin Periodontol. 2007;34(8):716–22.CrossRef
15.
go back to reference Turkyilmaz I, Aksoy U, Mcglumphy EA. Two alternative surgical techniques for enhancing primary implant stability in the posterior maxilla: a clinical study including bone density, insertion torque, and resonance frequency analysis data. Clin Implant Dent Relat Res. 2008;10(4):231.PubMed Turkyilmaz I, Aksoy U, Mcglumphy EA. Two alternative surgical techniques for enhancing primary implant stability in the posterior maxilla: a clinical study including bone density, insertion torque, and resonance frequency analysis data. Clin Implant Dent Relat Res. 2008;10(4):231.PubMed
16.
go back to reference Kojima K, Asamoto S, Kobayashi Y, Ishikawa M, Fukui Y. Cortical bone trajectory and traditional trajectory--a radiological evaluation of screw-bone contact. Acta Neurochir. 2015;157(7):1173–8.CrossRef Kojima K, Asamoto S, Kobayashi Y, Ishikawa M, Fukui Y. Cortical bone trajectory and traditional trajectory--a radiological evaluation of screw-bone contact. Acta Neurochir. 2015;157(7):1173–8.CrossRef
17.
go back to reference Steven W, Ronnie M, Sasidhar V, Shetye SS, Mcgilvray KC, Puttlitz CM. Pedicle screw placement in the lumbar spine: effect of trajectory and screw design on acute biomechanical purchase. J Neurosurg Spine. 2015;22(5):503–10.CrossRef Steven W, Ronnie M, Sasidhar V, Shetye SS, Mcgilvray KC, Puttlitz CM. Pedicle screw placement in the lumbar spine: effect of trajectory and screw design on acute biomechanical purchase. J Neurosurg Spine. 2015;22(5):503–10.CrossRef
18.
go back to reference Ohkawa T, Iwatsuki K, Ohnishi Y, Ninomiya K, Yoshimine T. Isthmus-guided cortical bone trajectory reduces postoperative increases in serum creatinine phosphokinase concentrations. Orthop Surg. 2015;7(3):232–8.CrossRef Ohkawa T, Iwatsuki K, Ohnishi Y, Ninomiya K, Yoshimine T. Isthmus-guided cortical bone trajectory reduces postoperative increases in serum creatinine phosphokinase concentrations. Orthop Surg. 2015;7(3):232–8.CrossRef
19.
go back to reference Lee JH, Lee JH, Park JW, Shin YH. The insertional torque of a pedicle screw has a positive correlation with bone mineral density in posterior lumbar pedicle screw fixation. J Bone Joint Surg Br. 2012;94(1):93.CrossRef Lee JH, Lee JH, Park JW, Shin YH. The insertional torque of a pedicle screw has a positive correlation with bone mineral density in posterior lumbar pedicle screw fixation. J Bone Joint Surg Br. 2012;94(1):93.CrossRef
20.
go back to reference Pappou IP, Girardi FP, Sandhu HS, Parvataneni HK, Jr CF, Schneider R, Frelinghuysen P, Lane JM. Discordantly high spinal bone mineral density values in patients with adult lumbar scoliosis. Spine. 2006;31(14):1614–20.CrossRef Pappou IP, Girardi FP, Sandhu HS, Parvataneni HK, Jr CF, Schneider R, Frelinghuysen P, Lane JM. Discordantly high spinal bone mineral density values in patients with adult lumbar scoliosis. Spine. 2006;31(14):1614–20.CrossRef
21.
go back to reference Rand T, Seidl G, Kainberger F, Resch A, Hittmair K, Schneider B, Glüer CC, Imhof H. Impact of spinal degenerative changes on the evaluation of bone mineral density with dual energy X-ray absorptiometry (DXA). Calcif Tissue Int. 1997;60(5):430–3.CrossRef Rand T, Seidl G, Kainberger F, Resch A, Hittmair K, Schneider B, Glüer CC, Imhof H. Impact of spinal degenerative changes on the evaluation of bone mineral density with dual energy X-ray absorptiometry (DXA). Calcif Tissue Int. 1997;60(5):430–3.CrossRef
22.
go back to reference Overton TR. Quantitative computed tomography. Acta Orthop Scand. 1989;60(5):629–30.CrossRef Overton TR. Quantitative computed tomography. Acta Orthop Scand. 1989;60(5):629–30.CrossRef
23.
24.
go back to reference Richardson ML, Genant HK, Cann CE, Ettinger B, Gordan GS, Kolb FO, Reiser UJ. Assessment of metabolic bone diseases by quantitative computed tomography. Clin Orthop Relat Res. 1985;(195):224–38. Richardson ML, Genant HK, Cann CE, Ettinger B, Gordan GS, Kolb FO, Reiser UJ. Assessment of metabolic bone diseases by quantitative computed tomography. Clin Orthop Relat Res. 1985;(195):224–38.
25.
go back to reference Mai HT, Mitchell SM, Hashmi SZ, Jenkins TJ, Patel AA, Hsu WK. Differences in bone mineral density of fixation points between lumbar cortical and traditional pedicle screws. Spine J. 2016;16(7):835–41.CrossRef Mai HT, Mitchell SM, Hashmi SZ, Jenkins TJ, Patel AA, Hsu WK. Differences in bone mineral density of fixation points between lumbar cortical and traditional pedicle screws. Spine J. 2016;16(7):835–41.CrossRef
26.
go back to reference Schreiber JJ, Anderson PA, Hsu WK. Use of computed tomography for assessing bone mineral density. Neurosurg Focus. 2014;37(1):E4.CrossRef Schreiber JJ, Anderson PA, Hsu WK. Use of computed tomography for assessing bone mineral density. Neurosurg Focus. 2014;37(1):E4.CrossRef
27.
go back to reference Tortolani PJ, Stroh DA. Cortical bone trajectory technique for posterior spinal instrumentation. J Am Acad Orthop Surg. 2016;24(11):755–61.CrossRef Tortolani PJ, Stroh DA. Cortical bone trajectory technique for posterior spinal instrumentation. J Am Acad Orthop Surg. 2016;24(11):755–61.CrossRef
28.
go back to reference Rodriguez A, Neal MT, Liu A, Somasundaram A, Hsu W, Jr BC. Novel placement of cortical bone trajectory screws in previously instrumented pedicles for adjacent-segment lumbar disease using CT image-guided navigation. Neurosurg Focus. 2014;36(3):E9.CrossRef Rodriguez A, Neal MT, Liu A, Somasundaram A, Hsu W, Jr BC. Novel placement of cortical bone trajectory screws in previously instrumented pedicles for adjacent-segment lumbar disease using CT image-guided navigation. Neurosurg Focus. 2014;36(3):E9.CrossRef
Metadata
Title
Differences in bone mineral density of trajectory between lumbar cortical and traditional pedicle screws
Authors
Renjie Zhang
Hai Gao
Huimin Li
Tao Xing
Chongyu Jia
Jianxiang Zhang
Fulong Dong
Cailiang Shen
Publication date
01-12-2019
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2019
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-019-1169-y

Other articles of this Issue 1/2019

Journal of Orthopaedic Surgery and Research 1/2019 Go to the issue