Published in: Radiation Oncology 1/2018

Open Access 01-12-2018 | Research

Significance of intra-fractional motion for pancreatic patients treated with charged particles

Authors: Vania Batista, Daniel Richter, Naved Chaudhri, Patrick Naumann, Klaus Herfarth, Oliver Jäkel

Published in: Radiation Oncology | Issue 1/2018

Login to get access

Abstract

Background

Uncertainties associated with the delivery of treatment to moving organs might compromise the accuracy of treatment. This study explores the impact of intra-fractional anatomical changes in pancreatic patients treated with charged particles delivered using a scanning beam. The aim of this paper is to define the potential source of uncertainties, quantify their effect, and to define clinically feasible strategies to reduce them.

Methods

The study included 14 patients treated at our facility with charged particles (protons or 12C) using intensity modulated particle therapy (IMPT). Treatment plans were optimized using the Treatment Planning System (TPS) Syngo® RT Planning. The pre-treatment dose distribution under motion (4D) was simulated using the TPS TRiP4D and the dose delivered for some of the treatment fractions was reconstructed. The volume receiving at least 95% of the prescribed dose (V95CTV) and the target dose homogeneity were evaluated. The results from the 4D dose calculations were compared with dose distributions in the static case and its variation correlated with the internal motion amplitude and plan modulation, through the Pearson correlation coefficient, as well the significant p-value. The concept of the modulation index (MI) was introduced to assess the degree of modulation of IMPT plans, through the quantification of intensity gradients between neighboring pencil beams.

Results

The induced breathing motion together with dynamic beam delivery results in an interplay effect, which affects the homogeneity and target coverage of the dose distribution. This effect is stronger (∆V95CTV > 10%) for patients with tumor motion amplitude above 5 mm and a highly modulated dose distribution between and within fields. The MI combined with the internal motion amplitude is shown to correlate with the target dose degradation and a lack of plan robustness against range and positioning uncertainties.

Conclusions

Under internal motion the use of inhomogeneous plans results in a decrease in the dose homogeneity and target coverage of dose distributions in comparison to the static case. Plan robustness can be improved by using multiple beams and avoiding beam entrance directions susceptible to density changes. 4D dose calculations support the selection of the most suitable plan for the specific patient’s anatomy.
Appendix
Available only for authorised users
Literature
1.
2.
go back to reference Ferlay J, Soerjomataram I, Dikshit R, Eser S, Mathers C, Rebelo M, et al. Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012. Int J Cancer Wiley Online Library. 2015;136:E359–86. Ferlay J, Soerjomataram I, Dikshit R, Eser S, Mathers C, Rebelo M, et al. Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012. Int J Cancer Wiley Online Library. 2015;136:E359–86.
3.
go back to reference Combs SE, Habermehl D, Kieser M, Dreher C, Werner J, Haselmann R, et al. Phase I study evaluating the treatment of patients with locally advanced pancreatic cancer with carbon ion radiotherapy: the PHOENIX-01 trial. BMC Cancer. 2013;13:419.CrossRefPubMedPubMedCentral Combs SE, Habermehl D, Kieser M, Dreher C, Werner J, Haselmann R, et al. Phase I study evaluating the treatment of patients with locally advanced pancreatic cancer with carbon ion radiotherapy: the PHOENIX-01 trial. BMC Cancer. 2013;13:419.CrossRefPubMedPubMedCentral
4.
go back to reference Shinoto M, Yamada S, Terashima K, Yasuda S, Shioyama Y, Honda H, et al. Carbon ion radiation therapy with concurrent gemcitabine for patients with locally advanced pancreatic Cancer. Int J Radiat Oncol Elsevier Inc. 2016;95:1–7.CrossRef Shinoto M, Yamada S, Terashima K, Yasuda S, Shioyama Y, Honda H, et al. Carbon ion radiation therapy with concurrent gemcitabine for patients with locally advanced pancreatic Cancer. Int J Radiat Oncol Elsevier Inc. 2016;95:1–7.CrossRef
5.
go back to reference Batista V, Richter D, Combs SE, Jäkel O. Planning strategies for inter-fractional robustness in pancreatic patients treated with scanned carbon therapy. Radiat Oncol. 2017;12:1–9.CrossRef Batista V, Richter D, Combs SE, Jäkel O. Planning strategies for inter-fractional robustness in pancreatic patients treated with scanned carbon therapy. Radiat Oncol. 2017;12:1–9.CrossRef
6.
go back to reference Bert C, Durante M. Motion in radiotherapy: particle therapy. Phys Med Biol IOP Publishing. 2011;56:R113.CrossRef Bert C, Durante M. Motion in radiotherapy: particle therapy. Phys Med Biol IOP Publishing. 2011;56:R113.CrossRef
7.
go back to reference Richter D, Saito N, Chaudhri N, Härtig M, Ellerbrock M, Jäkel O, et al. Four-dimensional patient dose reconstruction for scanned ion beam therapy of moving liver tumors. Int J Radiat Oncol Biol Phys [Internet] Elsevier Inc. 2014;89:175–81.CrossRef Richter D, Saito N, Chaudhri N, Härtig M, Ellerbrock M, Jäkel O, et al. Four-dimensional patient dose reconstruction for scanned ion beam therapy of moving liver tumors. Int J Radiat Oncol Biol Phys [Internet] Elsevier Inc. 2014;89:175–81.CrossRef
8.
go back to reference Lomax a J. Intensity modulated proton therapy and its sensitivity to treatment uncertainties 2: the potential effects of inter-fraction and inter-field motions. Phys. Med. Biol [Internet] 2008 cited. 2014;53:1043–56. Lomax a J. Intensity modulated proton therapy and its sensitivity to treatment uncertainties 2: the potential effects of inter-fraction and inter-field motions. Phys. Med. Biol [Internet] 2008 cited. 2014;53:1043–56.
9.
go back to reference Webb S. Use of a quantitative index of beam modulation to characterize dose conformality: illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy. Phys Med Biol Phys Med Biol. 2003;48135:2051–62.CrossRef Webb S. Use of a quantitative index of beam modulation to characterize dose conformality: illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy. Phys Med Biol Phys Med Biol. 2003;48135:2051–62.CrossRef
10.
go back to reference Richter D, Graeff C, Jäkel O, Combs SE, Durante M, Bert C. Residual motion mitigation in scanned carbon ion beam therapy of liver tumors using enlarged pencil beam overlap. Radiother Oncol Elsevier. 2014;113:290–5.CrossRef Richter D, Graeff C, Jäkel O, Combs SE, Durante M, Bert C. Residual motion mitigation in scanned carbon ion beam therapy of liver tumors using enlarged pencil beam overlap. Radiother Oncol Elsevier. 2014;113:290–5.CrossRef
11.
go back to reference Shackleford JA, Kandasamy N, Sharp GC. On developing B-spline registration algorithms for multi-core processors. Phys Med Biol. 2010;55:6329.CrossRefPubMed Shackleford JA, Kandasamy N, Sharp GC. On developing B-spline registration algorithms for multi-core processors. Phys Med Biol. 2010;55:6329.CrossRefPubMed
12.
go back to reference Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin J-C, Pujol S, et al. 3D slicer as an image computing platform for the quantitative imaging network. Magn Reson Imaging Elsevier. 2012;30:1323–41.CrossRef Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin J-C, Pujol S, et al. 3D slicer as an image computing platform for the quantitative imaging network. Magn Reson Imaging Elsevier. 2012;30:1323–41.CrossRef
13.
go back to reference Anderle K. In silico trial of photons versus carbon ions in single fraction therapy of lung cancer. Technische Universität Darmstadt. 2016;32(9):1118–23. Anderle K. In silico trial of photons versus carbon ions in single fraction therapy of lung cancer. Technische Universität Darmstadt. 2016;32(9):1118–23.
14.
go back to reference Lujan AE, Larsen EW, Balter JM, Haken RKT. A method for incorporating organ motion due to breathing into 3D dose calculations. Med Phys. 1999;26:715–20.CrossRefPubMed Lujan AE, Larsen EW, Balter JM, Haken RKT. A method for incorporating organ motion due to breathing into 3D dose calculations. Med Phys. 1999;26:715–20.CrossRefPubMed
15.
go back to reference Steidl P. Gating for scanned ion beam therapy. 2011; Steidl P. Gating for scanned ion beam therapy. 2011;
16.
go back to reference Hild S, Durante M, Bert C. Assessment of uncertainties in treatment planning for scanned ion beam therapy of moving tumors. Int J Radiat Oncol Biol Phys. 2013;85(2):528–35.CrossRefPubMed Hild S, Durante M, Bert C. Assessment of uncertainties in treatment planning for scanned ion beam therapy of moving tumors. Int J Radiat Oncol Biol Phys. 2013;85(2):528–35.CrossRefPubMed
17.
go back to reference Richter D. Treatment planning for tumors with residual motion in scanned ion beam therapy. Thesis 2012; Richter D. Treatment planning for tumors with residual motion in scanned ion beam therapy. Thesis 2012;
18.
go back to reference Eley JG, Newhauser WD, Lüchtenborg R, Graeff C, Bert C. 4D optimization of scanned ion beam tracking therapy for moving tumors. Phys Med Biol. 2014;59:3431.CrossRefPubMedPubMedCentral Eley JG, Newhauser WD, Lüchtenborg R, Graeff C, Bert C. 4D optimization of scanned ion beam tracking therapy for moving tumors. Phys Med Biol. 2014;59:3431.CrossRefPubMedPubMedCentral
20.
go back to reference Solla I, Zucca S, Possanzini M, Piras S, Pusceddu C, Porru S, et al. Free breathing conformal irradiation of pancreatic cancer. J. Appl. Clin. Med. Phys. 2013;14(4):4152.CrossRefPubMed Solla I, Zucca S, Possanzini M, Piras S, Pusceddu C, Porru S, et al. Free ­ breathing conformal irradiation of pancreatic cancer. J. Appl. Clin. Med. Phys. 2013;14(4):4152.CrossRefPubMed
21.
go back to reference Tai A, Liang Z, Erickson B, Li XA. Management of respiration-induced motion with 4-dimensional computed tomography (4DCT) for pancreas irradiation. Int. J. Radiat. Oncol. Biol. Phys. [Internet]. Elsevier Inc.; 2013;86:908–913. Tai A, Liang Z, Erickson B, Li XA. Management of respiration-induced motion with 4-dimensional computed tomography (4DCT) for pancreas irradiation. Int. J. Radiat. Oncol. Biol. Phys. [Internet]. Elsevier Inc.; 2013;86:908–913.
22.
go back to reference Fontana G, Riboldi M, Gianoli C, Chirvase CI, Villa G, Paganelli C, et al. MRI quantification of pancreas motion as a function of patient setup for particle therapy -a preliminary study. J Appl Clin Med Phys. 2016;17:60–75.CrossRefPubMedPubMedCentral Fontana G, Riboldi M, Gianoli C, Chirvase CI, Villa G, Paganelli C, et al. MRI quantification of pancreas motion as a function of patient setup for particle therapy -a preliminary study. J Appl Clin Med Phys. 2016;17:60–75.CrossRefPubMedPubMedCentral
23.
go back to reference Shiomi M, Mori S, Shinoto M, Nakayama Y, Kamada T, Yamada S. Comparison of carbon-ion passive and scanning irradiation for pancreatic cancer. Radiother Oncol. 2016;119:326–30.CrossRefPubMed Shiomi M, Mori S, Shinoto M, Nakayama Y, Kamada T, Yamada S. Comparison of carbon-ion passive and scanning irradiation for pancreatic cancer. Radiother Oncol. 2016;119:326–30.CrossRefPubMed
24.
go back to reference Bert C, Grözinger SO, Rietzel E. Quantification of interplay effects of scanned particle beams and moving targets. Phys Med Biol. 2008;53:2253–65.CrossRefPubMed Bert C, Grözinger SO, Rietzel E. Quantification of interplay effects of scanned particle beams and moving targets. Phys Med Biol. 2008;53:2253–65.CrossRefPubMed
25.
go back to reference Li H, Zhang X, Park P, Liu W, Chang J, Liao Z, et al. Robust optimization in intensity-modulated proton therapy to account for anatomy changes in lung cancer patients. Radiother Oncol. 2015;114(3):367–7. Li H, Zhang X, Park P, Liu W, Chang J, Liao Z, et al. Robust optimization in intensity-modulated proton therapy to account for anatomy changes in lung cancer patients. Radiother Oncol. 2015;114(3):367–7.
26.
go back to reference Mori S, Yanagi T, Hara R, Sharp GC, Asakura H, Kumagai M, et al. Comparison of respiratory-gated and respiratory-ungated planning in scattered carbon ion beam treatment of the pancreas using four-dimensional computed tomography. Int J Radiat Oncol Biol Phys [Internet]. 2010;76:303–12.CrossRef Mori S, Yanagi T, Hara R, Sharp GC, Asakura H, Kumagai M, et al. Comparison of respiratory-gated and respiratory-ungated planning in scattered carbon ion beam treatment of the pancreas using four-dimensional computed tomography. Int J Radiat Oncol Biol Phys [Internet]. 2010;76:303–12.CrossRef
27.
go back to reference Bernatowicz K, Lomax AJ, Knopf A. Comparative study of layered and volumetric rescanning for different scanning speeds of proton beam in liver patients. Phys. Med. Biol. [Internet]. 2013;58:7905–7920. Bernatowicz K, Lomax AJ, Knopf A. Comparative study of layered and volumetric rescanning for different scanning speeds of proton beam in liver patients. Phys. Med. Biol. [Internet]. 2013;58:7905–7920.
28.
go back to reference Cheung J. Feasibility of Online Range Adaptive Spot Scanning Proton Therapy. Med. Physics, AAPM Proc. 2012;39(6):3998. Cheung J. Feasibility of Online Range Adaptive Spot Scanning Proton Therapy. Med. Physics, AAPM Proc. 2012;39(6):3998.
29.
go back to reference Sharp GC, Lu HM, Trofimov A, Tang X, Jiang SB, Turcotte J, et al. Assessing residual motion for gated proton-beam radiotherapy. J Radiat Res. 2007;48 Sharp GC, Lu HM, Trofimov A, Tang X, Jiang SB, Turcotte J, et al. Assessing residual motion for gated proton-beam radiotherapy. J Radiat Res. 2007;48
Metadata
Title
Significance of intra-fractional motion for pancreatic patients treated with charged particles
Authors
Vania Batista
Daniel Richter
Naved Chaudhri
Patrick Naumann
Klaus Herfarth
Oliver Jäkel
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2018
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-018-1060-8

Other articles of this Issue 1/2018

Radiation Oncology 1/2018 Go to the issue